The EagleRay Long Sheath and EagleRay Access Catheter are indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Device Story
EagleRay Long Sheath and Access Catheter are single-lumen, guidewire-compatible catheters (0.014"-0.038") used to access peripheral, coronary, and neuro vasculature for interventional device delivery. Constructed of metal coils/braids and medical-grade polymers with hydrophilic coating and angled distal tip for tracking. Radiopaque marker at distal tip allows fluoroscopic visualization. Operated by physicians in clinical settings. Used with guidewires, support/diagnostic catheters, and introducer sheaths. Includes rotating hemostasis valve (RHV) to maintain hemostasis. Facilitates device delivery to target anatomy, potentially improving procedural outcomes.
Clinical Evidence
No human clinical data. Evidence based on bench testing (ISO 10555-1) and subacute/chronic porcine animal studies. Animal studies compared EagleRay to predicate Penumbra Neuron Max System; results showed no angiographic vessel injury, no significant histological findings, and comparable vascular response.
Technological Characteristics
Single-lumen catheter; metal coils/braids and medical-grade polymers; hydrophilic coating (Surmodics); radiopaque marker; 80-158cm lengths; 0.035"-0.088" ID. Sterilization: EtO (ISO 11135). Luer fittings per ISO 80369-7. No software/algorithm.
Indications for Use
Indicated for introduction of interventional devices into peripheral, coronary, and neuro vasculature in patients requiring endovascular procedures.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 8, 2018
Imperative Care Inc. Jake Wolenberg Ouality Assurance and Regulatory Affairs Manager 1221 Innsbruck Drive Sunnyvale, California 94089
Re: K180169
Trade/Device Name: EagleRay Long Sheath, EagleRay Access Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: Julv 3. 2018 Received: July 9, 2018
Dear Jake Wolenberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Xiaolin Zheng -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180169
Device Name
EagleRay Long Sheath, EagleRay Access Catheter
#### Indications for Use (Describe)
The EagleRay Long Sheath and EagleRay Access Catheter are indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CFR 801 Subpart D)</b></span> |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Imperative Care. The logo features an abstract human head made of orange triangles on the left. To the right of the head is the company name, "Imperative," in a bold, sans-serif font. Below "Imperative" is the word "CARE" in a smaller, lighter font.
# 510(k) Summary EagleRay Long Sheath and EagleRay Access Catheter
#### A. Submitter Information
Submitter's Name: Address:
- Telephone: Email: Contact Person: Date of Preparation:
#### B. Subject Device
Proprietary Name: 510(k) #: Common/Usual Name: Classification Name: Product Code:
### C. Primary Predicate Device
- Primary Proprietary Name: Manufacturer: 510(k) #'s: Common/Usual Name: Classification Name: Product Code:
### D. Additional Predicate Devices
Proprietary Name: Manufacturer: 510(k) #: Common/Usual Name: Classification Name: Product Code:
Proprietary Name: Manufacturer: 510(k) #: Common/Usual Name: Classification Name: Product Code:
Imperative Care Inc. 1359 Dell Avenue Campbell, CA 95008 248-496-0198 JWolenberg@ImperativeCare.com Jake Wolenberg July 3, 2018
EagleRay Long Sheath, EagleRay Access Catheter K180169 Guide Sheath & Guide Catheter Catheter, Percutaneous DQY per 21 C.F.R. 870.1250
Neuron Max System Penumbra Inc. K111380 Guide Sheath Catheter, Percutaneous DQY per 21 C.F.R. 870.1250
Arc & Arc Mini Intracranial Support Catheter Micro Therapeutics, Inc. Dba Ev3 Neurovascular K150107 Guide Catheter Catheter, Percutaneous DQY per 21 C.F.R 870.1250
Navien Intracranial Support Catheter Micro Therapeutics, Inc. Dba Ev3 Neurovascular K161152 Guide Catheter Catheter, Percutaneous DQY per 21 C.F.R 870.1250
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Imperative Care. The logo features an abstract human head made of orange triangles on the left. To the right of the head is the text "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font below it.
#### Device Description: E.
The EagleRay Long Sheath and EagleRay Access Catheter, hereinafter referred to as EagleRay Catheters, are 0.014"-0.038" guidewire compatible single lumen catheters that provide access to peripheral, coronary and neuro vasculature which allows insertion of other interventional devices.
The catheters are comprised of a hollow cvlindrical tube which is bonded to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers. The distal section of each catheter has a hydrophilic coating to enhance tracking through tortuous vasculature. Additionally, an angled distal tip facilitates smoother tracking past vessel branches. The distal tip also has a radiopaque marker to provide the user with clear visual confirmation of the distal tip location under fluoroscopy.
The EagleRay Catheters will be offered with various working lengths and nominal inner diameters (ID) and outer diameters (OD) as shown in Table 1 below.
| Commercial | Model | Distal Diameter | | Proximal Diameter | | Nominal Usable |
|--------------------------------|------------|-----------------|--------|-------------------|--------|-----------------|
| Name | Number | Inner | Outer | Inner | Outer | Catheter Length |
| EagleRay<br>Long Sheath | ICLS088080 | 0.088" | 0.106" | 0.088" | 0.108" | 80cm |
| | ICLS088090 | | | | | 90cm |
| | ICLS088100 | | | | | 100 cm |
| | ICLS088110 | | | | | 110 cm |
| EagleRay<br>Access<br>Catheter | ICAC071137 | 0.071" | 0.083" | 0.071" | 0.083" | 137cm |
| | ICAC055137 | 0.055" | 0.069" | 0.067" | 0.080" | 137cm |
| | ICAC045144 | 0.045" | 0.060" | 0.064" | 0.080" | 144cm |
| | ICAC035158 | 0.035" | 0.051" | 0.047" | 0.060" | 158cm |
#### Table 1: EagleRay Catheter Device Sizes
All EagleRay Catheters are packaged with an accessory Rotating Hemostasis Valve (RHV). The RHV is designed to be attached to the proximal luer of the catheter and helps the user maintain hemostasis.
The 0.088" ID EagleRay Long Sheath is packaged with an additional accessory, a short, retractable Catheter Introducer. This accessory is intended to prevent damage to the EagleRay Long Sheath when advancing it through the hemostasis valve on the access site introducer sheath.
In addition to the supplied accessories discussed above, the adjunctive devices and supplies listed below are intended to be used with the EagleRay Catheters. All of these accessories are also commonly used with the predicate devices.
- Guidewires
- Support/Diagnostic Catheters ●
- Introducer Sheaths ●
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Imperative Care. The logo features a stylized head made of orange and yellow triangles on the left. To the right of the head is the text "Imperative" in a bold, dark gray font, with the word "CARE" in a smaller, lighter orange font underneath.
#### F. Indications for Use:
The EagleRay Long Sheath and EagleRay Access Catheter are indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
### G. Predicate Comparison:
The primary predicate device for the EagleRay Catheters is the Neuron Max System (Penumbra, Inc) cleared under K111380. The Neuron Max System was selected as the primary predicate for the EagleRay Catheters as the Neuron Max System has similar dimensions to the largest device sizes for the EagleRay Catheters and has the same intended use.
In addition to the primary predicate device, the devices in the Micro Therapeutics Intracranial Support Catheter product line, which include the Arc and Arc Mini catheters cleared under K150107 and the Navien catheters cleared under K161152, were used as predicate devices for the smaller EagleRay Catheter device sizes as they have similar dimensions and the same intended use.
Table 2 presented below provides a comparison of the similarities and differences between the proposed EagleRay Catheters, the predicate Neuron Max System, and the predicate Micro Therapeutics Intracranial Support Catheter product line. Bench and lab testing was conducted as appropriate to evaluate any differences that were identified.
The comparison between the subject and predicate devices found that the EagleRay Catheters are substantially equivalent to the predicate Neuron Max System and Micro Therapeutics Intracranial Support Catheters, and that there are no new safety or efficacy concerns. This conclusion is based on all devices sharing the same intended use, basic technology characteristics, and performance characteristics, as demonstrated through well-designed bench and lab testing.
| Device Attribute | EagleRay Catheters<br>(subject device) | Penumbra<br>Neuron Max System | Micro Therapeutics<br>Intracranial Support<br>Catheters |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| FDA Product<br>Classification | Class II, DQY, 21 CFR<br>870.1250 | Same | Same |
| Intended Use | Intended for use in<br>supporting the introduction<br>of interventional devices into<br>the vasculature. | Same | Same |
| Condition<br>Supplied | Sterile and Single Use | Same | Same |
| Sterilization<br>Method | Ethylene Oxide (EtO), SAL<br>10-6 | Same | Same |
| Device Attribute | EagleRay Catheters<br>(subject device) | Penumbra<br>Neuron Max System | Micro Therapeutics<br>Intracranial Support<br>Catheters |
| Inner Diameter<br>(Distal) | 0.035" – 0.088" | 0.088" | 0.035" – 0.072" |
| Outer Diameter<br>(Distal) | 0.051" – 0.106" | 0.105" | 0.044" – 0.084" |
| Inner Diameter<br>(Proximal) | 0.047" – 0.088" | 0.088" | 0.044" – 0.072" |
| Outer Diameter<br>(Proximal) | 0.060" – 0.110" | 0.110" | 0.060" – 0.084" |
| Effective Length | 80 – 158cm | 80 – 90cm | 90 – 163cm |
| Tip Design | Beveled edge, soft, flexible, and atraumatic | Square edge, soft, flexible, and atraumatic | Square edge, soft, flexible, and atraumatic |
| Coating | Surmodics Hydrophilic coating | Surmodics Hydrophilic coating | Hydrophilic coating |
| Materials | Commonly used medical grade plastics & metals with hydrophilic coating. | Commonly used medical grade plastics & metals with hydrophilic coating. | PTFE lined polymeric catheter shaft with hydrophilic coating, nitinol support wire, and platinum marker bands. |
| Packaged<br>Accessories | Catheter Introducer<br>Rotating Hemostasis Valve<br>(RHV) | Dilator<br>Rotating Hemostasis<br>Valve (RHV)<br>Hemostasis Valve | Catheter Introducer |
| Packaging<br>Configuration | The catheters are placed in a protective polyethylene tube and then mounted, along with the accessories, onto a polyethylene packaging card.<br>The packaging card is inserted into a Tyvek® pouch which is then sealed.<br>The sealed pouch and IFU | Same | Same |
#### Table 2: EagleRav Catheters, Neuron Max System, and Micro Therapeutics Intracranial Support Catheter Comparison
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Imperative Care. The logo features a stylized human head made up of orange triangles on the left. To the right of the head is the word "Imperative" in black, with the word "CARE" in orange underneath.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Imperative Care. The logo features a geometric head made of orange triangles on the left. To the right of the head is the text "Imperative" in a bold, dark gray font, with the word "CARE" in a smaller, light orange font underneath.
#### Performance Data Supporting Substantial Equivalence: H.
Bench and Lab (in-vitro) testing were conducted in order to evaluate the similarities and differences between the largest diameter EagleRay Catheters and the predicate Neuron Max System. Comparative testing was also performed to evaluate the similarities and differences between the smallest EagleRay Catheters and the predicate Micro Therapeutics Intracranial Support Catheters.
The test results were reviewed and found to demonstrate that any differences between the proposed and predicate devices do not significantly impact any catheter performance parameters that would affect the safety or efficacy of the subject EagleRay Catheters.
A summary of the tests and performance specifications that were evaluated is presented in Table 3. These tests were performed per company approved protocols and test methods based primarily on catheter performance standard ISO 10555-1.
| Test Attribute | Specification |
|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Delivery, Compatibility,<br>and Retraction<br>(Trackability) | The catheter shall be able to be delivered, deployed, and retracted<br>per the IFU within a simulated neurological model without<br>incurring any damage to the catheter. |
| Flexibility and Kink<br>Resistance | There shall be no kinking of shaft (permanent deformation) after<br>simulated use. |
| Compatibility with other<br>devices (external) | The catheters shall be able to be delivered through the minimum<br>introducer sheath or guide catheter size indicated in the product<br>labeling. |
| Guidewire compatibility | The catheters shall be able to be delivered over the maximum size<br>guidewire indicated in the product labeling. |
| Interventional device<br>compatibility (internal) | The catheters shall be able to accommodate other interventional<br>devices (e.g., support catheter, diagnostic catheter) up to the<br>maximum size indicated in the product labeling. |
| Luer compatibility | Devices and accessories shall be compatible with standard syringe<br>luer fittings per ISO 80369-7. |
| Accessory compatibility | Devices shall be compatible with catheter introducer and RHV. |
| Catheter Bond Strength | The catheter shall have sufficient bond strengths to remain intact<br>throughout a procedure. |
| | | Table 3: Tests and Performance Specifications |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| | | |
| Test Attribute | Specification | |
| Flowrate - positive<br>(forward) pressure | The catheter lumen shall allow for a minimum flowrate<br>comparable to competitive products. | |
| Freedom from Leakage<br>– positive pressure | No liquid leakage from the hub or catheter shaft at 46psi for 30<br>seconds. | |
| Freedom from Leakage<br>– negative pressure | No air leakage into a 20cc syringe when vacuum pulled for 15<br>seconds. | |
| Burst Pressure | Catheter does not burst under pressures that could be seen when<br>performing contrast injections with a standard 10cc syringe. | |
| Catheter Torque<br>Strength | No separation of any portion of the catheter when rotated at least<br>two (2) full rotations (360 degrees). | |
| Kink Resistance | There shall be no kinking of the catheter shaft (permanent<br>deformation) after wrapping around anatomically relevant bend<br>radii. | |
| Flexibility | The flexibility of the catheter tip shall be comparable to<br>competitive products and allow for easily tracking the device to<br>the desired target anatomy. | |
| Pushability | The proximal shaft of the catheters shall have sufficient stiffness<br>that the user can easily push the catheter to the target anatomy<br>without buckling. | |
| Coating - Particulate | The amount of particulate matter that comes off the hydrophilic-<br>coated shaft during simulated use testing shall be determined and<br>compared to competitive products and techniques. | |
| Coating - lubricity | Coating must be lubricious and maintain a minimum lubricity over<br>15 test cycles. | |
| Corrosion Resistance | No visible corrosion present on devices after saline immersion<br>followed by 30 minutes in boiling water followed by 48 hours in<br>37°C water bath. | |
| Radiopacity | The radiopaque marker on the catheter can be seen under<br>fluoroscopy during use. | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Imperative Care. The logo features a stylized head made of orange triangles on the left. To the right of the head is the word "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font underneath.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Imperative Care. The logo features a stylized human head made up of orange triangles on the left side. To the right of the head is the word "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font underneath.
#### I. Biocompatibility Testing:
Biocompatibility testing was performed on all direct patient contacting components of the EagleRay Catheters. This testing was conducted to ensure that the components and raw materials used in the proposed EagleRay Catheters, as well as the manufacturing processes and sterilization processes result in a biocompatible product. All biocompatibility tests were conducted pursuant to 21 CFR Part 58, Good Laboratory Practices, ISO 10993-1, and per the recommendations in FDA guidance document titled: Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
As the EagleRay Catheters are in direct contact with circulating blood and the average length of an endovascular procedure is typically less than 2 hours, the devices were evaluated for the tests recommended for the patient contact category: External communicating device, blood path direct, limited contact duration.
A summary of the selected biocompatibility tests performed for the EagleRay Catheters is presented for reference below and on the following pages in Table 4. All test results passed, indicating that the EagleRay Catheters are biocompatible for their intended use.
| Test | Test Method | Extraction<br>Methods/Conditions | Acceptance Criteria | Results |
|---------------------------------------------------------------------|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity: ISO<br>MEM* Elution | ISO 10993-5 | Test device extracted in<br>MEM with 5% serum at 37 ±<br>1°C for 24 ± 2 hours | Sample extracts must<br>yield cell lysis grade 2 or<br>lower. | Pass, Non-<br>cytotoxic |
| Cytotoxicity: ISO<br>MTT* Assay | ISO 10993-5 | Test device extracted in<br>MTT* Assay Media at 37 ±<br>1°C for 24 ± 2 hours | The percentage of cells<br>exhibiting lysis should be<br>similar for all test devices. | Pass, No<br>Significant<br>Differences |
| Sensitization:<br>Magnusson-<br>Kligman Method, 2<br>extracts | ISO 10993-10 | Guinea pigs exposed to test<br>device extracts. Challenged<br>sites observed for skin<br>sensitization 24 ± 2 and 48 ±<br>2 hours after removal of<br>extracts.<br><br>Extracts were prepared at 50<br>± 2°C for 72 ± 2 hours. | Test Group shall yield<br>Grade < 1 score on<br>Magnusson and Kligman<br>scale (provided control<br>Grade < 1). | Pass, Non-<br>Sensitizing |
| Test | Test Method | Extraction<br>Methods/Conditions | Acceptance Criteria | Results |
| Irritation: ISO<br>Intracutaneous<br>Irritation Test | ISO 10993-10 | Rabbits are injected with<br>extracts of test device.<br>Injection sites<br>examined/scored at 24 ± 2,<br>48 ± 2 and 72 ± 2 hours after<br>injections. | The difference in the<br>mean test article and<br>mean control score must<br>be grade 1.0 or lower. | Pass, Non-Irritant |
| | | Extracts were prepared at 50<br>± 2°C for 72 ± 2 hours. | | |
| Systemic Toxicity:<br>ISO Materials<br>Mediated Rabbit<br>Pyrogen | ISO 10993-<br>11:2006 | Test device extracted in<br>0.9% NS* and injected in the<br>marginal ear vein. Rectal<br>temperatures recorded prior<br>to injection and every 30 min<br>until 1-3 hours post injection.<br>Extracts were prepared at 50<br>± 2°C for 72 ± 2 hours | Sample Extracts must not<br>cause a total rise in body<br>temperature of ≥0.5°C. | Pass, Non-<br>pyrogenic |
| Systemic Toxicity:<br>ISO Acute<br>Systemic Injection<br>Test | ISO 10993-11 | Mice are injected with<br>extracts of test device.<br>Animals are observed for<br>signs of toxicity immediately<br>after injection, 4 ± 0.75, 24 ±<br>2, 48 ± 2 and 72 ± 2 hours<br>after injections. Body<br>weights are measured prior<br>to injection and on all 3 days.<br>Extracts were prepared in<br>0.9% NS* and Sesame Oil at<br>50 ± 2°C for 72 ± 2 hours. | Sample extracts must not<br>cause the following:<br>• > 10% weight loss in 3<br>or more test animals<br>• Mortality of 2 or more<br>test animals<br>• Toxic signs such as<br>convulsions and<br>prostration in 2 or more<br>test animals | Pass, Non-Toxic |
| Hemocompatibility:<br>Complement<br>Activation | ISO 10993-4<br>ASTM F756 | Preformed in vitro using a<br>prepared Enzyme<br>Immunoassay (EIA) kit.<br>This kit will detect the presence of<br>specific complement<br>enzymes.<br>Test device incubated in<br>serum at 37± 2°C and tested<br>for C3a and SC5b at 0.5, 1,<br>and 1.5 hours. | The concentrations of C3a<br>and SC5b-9 in the test<br>devices are statistically<br>similar to the predicate<br>device (Penumbra Neuron<br>Max 6F Catheter) and<br>statistically lower than the<br>positive control for all<br>exposure times. | Pass, No<br>Significant<br>Differences<br>The test device had<br>statistically similar<br>or lower<br>concentrations than<br>the predicate and<br>negative controls. |
| Test | Test Method | Extraction<br>Methods/Conditions | Acceptance Criteria | Results |
| Hemocompatibility:<br>Hemolysis (Direct<br>Contact Method) | ASTM F 756<br>ISO 10993-4 | Test device extracted in<br>PBS* and diluted blood at 37<br>$\pm$ 2°C for 3 $\pm$ 5 minutes.<br>Test device exposed to blood<br>cell suspension.<br>% hemolysis is measured. | Sample extracts must be<br>nonhemolytic ( $\leq$ 2%<br>hemolytic index). | Pass, Non-<br>hemolytic |
| Hemocompatibility:<br>Hemolysis (Extract<br>Method) | ASTM F 756<br>ISO 10993-4 | Test device extracted in PBS<br>at 50 $\pm$ 2°C for 72 $\pm$ 2 hours.<br>Extract exposed to blood cell<br>suspension. % hemolysis is<br>measured. | Sample extracts must be<br>nonhemolytic ( $\leq$ 2%<br>hemolytic index). | Pass, Non-<br>hemolytic |
| Standard In vivo<br>Thrombogenicity,<br>Ovine Jugular<br>NAVI (ISO) | ISO 10993-<br>4:2002 | Performed in duplicate. Test<br>and control device are placed<br>in alternate sides of the<br>juggler vein. The devices are<br>implanted and remains in<br>place for 4 $\pm$ 0.5 hours. | The device must have<br>similar or lesser<br>thrombogenic potential<br>after 4 hours in vivo when<br>compared to a control<br>device (Penumbra Neuron<br>Max 6F Catheter). | Pass, Minimal<br>Thrombogenic<br>Potential<br>The test device had<br>less thrombogenic<br>potential than the<br>control device. |
## Table 4: Biocompatibility Test Summary
{10}------------------------------------------------
EagleRay Long Sheath and
EagleRay Access Catheter
Traditional 510(k) K180169
Image /page/10/Picture/1 description: The image shows the logo for Imperative Care. The logo features a geometric human head in shades of orange and yellow on the left. To the right of the head is the word "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font underneath.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Imperative Care. On the left side of the logo is a geometric design of a human head in orange tones. To the right of the head is the text "Imperative" in a bold, sans-serif font. Below "Imperative" is the word "CARE" in a smaller, sans-serif font, also in orange.
*MEM=Minimal Essential Media,
MTT Assay = (3-(4,5-Dimethylthiazol-2-yl)-2,5-Diphenyltetrazolium Bromide)
NS = Normal Saline
PBS=Phosphate Buffered Saline
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for Imperative Care. On the left side of the logo is an orange geometric design that resembles a human head. To the right of the head is the text "Imperative" in a bold, dark gray font. Below "Imperative" is the word "CARE" in a smaller, orange font.
#### J. Sterilization:
The EagleRay Catheters are sterilized using a validated EtO process with a sterility assurance level of 1x10-6. The sterilization process was validated per the overkill method described recognized consensus standard ISO 11135, "Sterilization Of Health-Care Products - Ethylene Oxide - Requirements For The Development, Validation And Routine Control Of A Sterilization Process For Medical Devices." A summary of the completed testing is presented below in Table 5.
## Table 5: Sterilization Validation Summary
| Requirement/Acceptance Criteria | Results | Summary |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Positive BI controls must be positive in<br>fractional, half and full cycles. | PASS | All positive controls in each cycle read<br>positive. |
| Fractional cycle requirements:<br>1) IPCDs must be equal or less difficult to<br>sterilize than the EPCDs<br>2) The IPCDs and EPCDs may show growth<br>after 7 days with the EPCD required to be<br>equal or more EtO resistant.<br>3) Product Sterility testing must show no<br>growth in all samples.<br>4) Bacteriostasis/Fungistasis testing must<br>show that the product is not inhibitory for<br>growth. | PASS | 1) IPCDs showed growth after 7 days.<br>2) EPCDs were equal or more EtO<br>resistant than the IPCDs.<br>3) The devices did not show any<br>growth when tested per the methods<br>in ISO 11737-2.<br>4) Bacteriostasis/Fungistasis Testing<br>showed product is not inhibitory to<br>growth when tested per the methods<br>in ISO 11737-2. |
| Half cycle requirements:<br>1) The IPCDs must show no growth after 7<br>days<br>2) EPCD are expected to show no growth<br>after 7 days. A few positives are allowed<br>by ISO 11135. | PASS | 1) IPCDs showed no growth after 7<br>days.<br>2) EPCDs showed no growth after 7<br>days. |
| Full cycle requirements:<br>1) The IPCDs must show no growth after 7<br>days<br>2) Selected EPCD must show no growth after<br>7 days.<br>3) Samples exposed to 2X sterilization cycle<br>must pass EtO residual testing. | PASS | 1) IPCDs showed no growth after 7<br>days.<br>2) EPCDs showed no growth after 7<br>days.<br>3) Samples passed EtO residual testing<br>at 0 hour aeration time point. |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image contains the logo for Imperative Care. On the left side of the logo is a graphic of a head made up of orange triangles. To the right of the graphic is the text "Imperative" in a bold, dark gray font. Below the word "Imperative" is the word "CARE" in a smaller, light orange font.
## K. Shelf Life and Packaging:
Accelerated aging testing based on ASTM F1980 was conducted to verify device, accessory, and packaging performance. A real time aging equivalent of 13 months was used for this testing and will allow for labeling of product with a 1-year shelf life.
Device and accessory performance was verified by repeating the functional tests previously presented in Section H of this 510(k) summary.
Packaging and sterile barrier integrity through transportation has been verified for both the packaging configurations used for the EagleRay Catheters, a straight packaging configuration and a circular packaging configuration. Aging testing has also been performed that supports the sterile barrier integrity following 13 months of accelerated aging. A summary of the completed packaging test is presented below in Table 6.
| Test | Test Method | Straight Packaging<br>Results<br>(Pass/Fail) | Circular Packaging<br>Results<br>(Pass/Fail) |
|------------------------------------------------|------------------------------------------|----------------------------------------------|----------------------------------------------|
| Packaging Visual<br>Inspection | ASTM F1886<br>Imperative Care<br>TM00071 | Pass | Pass |
| Pouch Integrity Test -<br>Gross Leak Detection | ASTM F2096 | Pass | Pass |
| Pouch Seal Strength<br>– Peel Strength | ASTM F88 | Pass | Pass |
| Label Integrity | Imperative Care<br>TM00071 | Pass | Pass |
#### Table 6: Packaging Validation Summary
### L. Animal Studies
A series of subacute and chronic animal studies were conducted to evaluate the safe use of the EagleRay Catheter design in a porcine model. 510(k) cleared neurovascular catheters, including the predicate Penumbra Neuron Max System, were used as control devices. The studies concluded that:
- . There was no angiographic evidence of vessel injury in any of the treated vessel segments (test or control).
- . No significant histological findings were found in relation to use of the test or control devices.
- Use of the test devices resulted in no significant vascular response during these studies and was found to be comparable to the control devices by the study pathologist and the study director.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image contains the logo for Imperative Care. On the left side of the logo is a geometric design of a human head, with the face made up of many small triangles in shades of orange and yellow. To the right of the head is the word "Imperative" in a bold, sans-serif font, with the word "CARE" in a smaller font underneath.
EagleRay Long Sheath and EagleRay Access Catheter Traditional 510(k) K180169
### M. Conclusions:
Where differences were identified between the subject EagleRay Catheters, the predicate Neuron Max System, and the predicate Micro Therapeutics Intracranial Support Catheters, a risk assessment was completed to determine if the difference would result in new safety or efficacy concerns regarding the use of the device. As appropriate, bench, lab, and animal testing was conducted to support this assessment.
Based on the results of the conducted risk assessments and associated bench, lab, and animal testing, it is concluded that EagleRay Catheters are substantially equivalent to the predicate Neuron Max System and predicate Micro Therapeutics Intracranial Support Catheters and that there are no new safety or efficacy concerns associated with the identified differences. This conclusion was determined based on the fact that the subject device and multiple predicate devices share the same intended use, basic technology characteristics, and performance characteristics, as confirmed through well-designed bench, lab, and animal testing.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.