The MiVi 6F Guide Catheter is indicated for use in facilitating the insertion and guidance of microcathers into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter.
Device Story
Single lumen, braided, variable stiffness catheter; used to facilitate insertion/guidance of microcatheters into peripheral, coronary, or neurovascular vessels under fluoroscopy; includes radiopaque marker for visualization; hydrophilic coating reduces friction; proximal luer hub allows flushing, aspiration, and device insertion; used in conjunction with rotating hemostatic valve; operated by physicians in clinical settings; enables delivery of contrast agents or therapeutic devices; benefits patient by providing endovascular access for diagnostic/therapeutic procedures.
Clinical Evidence
No clinical data was needed; substantial equivalence demonstrated via bench testing and biocompatibility studies. Bench testing included push/track, stent crossing, tip stiffness, torque response, kink resistance, tensile strength, luer leakage, and coating performance. Biocompatibility testing included cytotoxicity, irritation, hemocompatibility (hemolysis, thrombosis, complement activation), pyrogenicity, sensitization, and systemic toxicity.
Technological Characteristics
Single lumen, braided, variable stiffness shaft; PTFE-lined PEBAX construction with stainless steel braid; hydrophilic coating; 6F diameter, 110 cm length; straight distal tip; radiopaque marker; proximal female luer hub; ethylene oxide sterilization; compliant with ISO 10555-1, ISO 10555-3, ISO 594-1/2, and AAMI TIR42.
Indications for Use
Indicated for patients requiring insertion and guidance of microcatheters into peripheral, coronary, or neurovascular blood vessels, or for diagnostic angiography.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Stryker Concentric Medical, Inc. Modified HD Guide Catheter (K110483)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them, possibly representing water or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2014
MiVi Neuroscience, LLC Ms. Michelle Straight VP of Quality & Regulatory 10900 73td Avenue North, Suite 150 Maple Grove, MN 55369
Re: K140557 Trade/Device Name: MiVi 6F Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: August 14, 2014 Received: August 15, 2014
Dear Ms. Straight,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140557
Device Name MiVi 6F Guide Catheter
Indications for Use (Describe)
The MiVi 6F Guide Catheter is indicated for use in facilitating the insertion and guidance of microcathers into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Carlos L. Pena -S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for MIVI Neuroscience. The logo features the word "MIVI" in large, bold, teal letters, with the word "NEUROSCIENCE" in smaller letters underneath. To the right of the text is a teal silhouette of a human head, with the brain and blood vessels visible. The entire logo is enclosed in a rounded teal rectangle.
10900 73rd Ave. North, Suite 150 Maple Grove, MN 55369
# 510(k) SUMMARY: K140557
## Date Prepared: 3-Mar-2014
### Submitter's Name / Contact Person
#### Manufacturer
MiVi Neuroscience, LLC 10900 73rd Ave N, Suite 150 Maple Grove, MN 55369
### Contact Person
Michelle Straight Vice President of RA/QA Phone: 612-940-7240 Fax: 763-577-1044 Email: mstraight@mivi.neuro.com
#### General Information
| Trade Name | MiVi 6F Guide Catheter |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common / Usual Name | Guide Catheter |
| Classification Name | Percutaneous Catheter (21 CFR 870.1250, Product Code DQY) |
| Regulatory Classification | |
| Class | II |
| Panel | Neurodiagnostic and Neurotherapeutic Device (NNDB) Division of Neurological and Physical Medicine Devices (DNPMD) |
| Predicate Device(s) | Stryker Concentric <sup>®</sup> Medical, Inc. K110483 Modified HD Guide Catheter Stryker Concentric <sup>®</sup> Distal Access Catheters Models 90120, 90121, 90130, 90131, 90160, 90161 and 90162 |
#### Device Description
The MiVi 6F Guide Catheter consists of a single lumen, braided, variable stiffness shaft designed for use in facilitating the insertion and guidance microcatheters into a selected blood vessel in the peripheral, coronary or neurovascular system. A radiopaque marker is included on the distal end for angiographic visualization. The catheter shaft has a hydrophilic coating to reduce friction during use. A luer hub on the proximal end allows attachments for flushing, insertion of catheters and aspiration. It is used in junction with a rotating hemostatic valve with side-arm adapter for flushing, catheter insertion and aspiration. The MiVi 6F Guide Catheter has a straight distal tip and is available in a 110 cm length and 6F diameter.
{5}------------------------------------------------
#### Intended Use / Indications
The MiVi 6F Guide Catheter is indicated for: use in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems. It may also be used as a diagnostic angiographic catheter.
#### Principle of Operation
The device will be utilized to endovascularly insert and guide microcatheters under fluoroscopy so that the microcatheters can deliver contrast agents or other devices during diagnostic and/or therapeutic procedures for patients with arterial disease or damage.
#### Accessories
There are no accessories to the MiVi 6F Guide Catheters.
| | Subject Device | Predicate Device | Predicate Device |
|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | MiVi 6F Guide<br>Catheter | Stryker Concentric®<br>Medical, Inc. K110483<br>Modified HD Guide<br>Catheter<br>Stryker Concentric<br>REF 90170 and 90171 | Stryker Concentric®<br>Distal Access Catheters<br>REF 90120, 90121,<br>90130, 90131, 90160,<br>90161 and 90162 |
| Classification | II, DQY | II, DQY | II, DQY |
| Indications for<br>Use | The MiVi 6F Guide<br>Catheter is indicated<br>for use in facilitating<br>the insertion and<br>guidance of<br>microcatheters into a<br>selected blood vessel<br>in the peripheral,<br>coronary and neuro<br>vascular systems. | The Modified HD Guide<br>Catheter is indicated for<br>facilitating the insertion<br>and guidance of an<br>occlusion catheter,<br>infusion catheter or<br>other appropriate<br>microcatheter into a<br>selected blood vessel in<br>the peripheral, coronary<br>and neuro vascular<br>systems. It may also be<br>used as a diagnostic<br>angiographic catheter. | The Distal Access<br>Catheter is indicated for<br>use in facilitating the<br>insertion and guidance of<br>an occlusion catheter,<br>infusion catheter or other<br>appropriate<br>microcatheter into a<br>selected blood vessel in<br>the peripheral, coronary<br>and neuro vascular<br>systems. It may also be<br>used as a diagnostic<br>angiographic catheter. |
| Shaft<br>Materials<br>Materials, Packaging and Sterilization | PTFE lined PEBAX<br>with stainless steel<br>braid | PTFE lined polymer<br>with stainless steel braid | PTFE lined polymer<br>with stainless steel braid |
| Proximal End | Hub with female luer | Hub with female luer | Hub with female luer |
| Configuration | conical fitting | conical fitting | conical fitting |
| Packaging | Catheter attached to an<br>SBS packaging card<br>inside PET/PE/Tyvek<br>pouch, inside SBS box | Catheter attached to SBS<br>packaging card inside<br>carton | Catheter attached to SBS<br>packaging card inside<br>carton |
| Sterilization | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
#### Comparison to Predicate Devices:
{6}------------------------------------------------
| | Subject Device | Predicate Device | Predicate Device |
|---------------------|----------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Feature | MiVi 6F Guide<br>Catheter | Stryker Concentric®<br>Medical, Inc. K110483<br>Modified HD Guide<br>Catheter<br>Stryker Concentric<br>REF 90170 and 90171 | Stryker Concentric®<br>Distal Access Catheters<br>REF 90120, 90121,<br>90130, 90131, 90160,<br>90161 and 90162 |
| Dimensions | | | |
| Effective<br>Length | 107.3 cm | 105 cm<br>120 cm | 115 cm to 136 cm |
| Outer Diameter | 6F | 6.3F | 3.9F to 5.2F |
| Inner Diameter | Distal: .054"<br>Proximal: .064" | .070" | .038" to .057" |
| Tip Shape | Straight | Straight | Straight |
#### Design verification testing of the MiVi 6F Guide Catheter consisted of:
| Performance Test | Result |
|----------------------|--------------------------|
| Push/Track | Met established criteria |
| Stent Crossing | Met established criteria |
| Tip Stiffness | Met established criteria |
| Torque Response | Met established criteria |
| Kink Resistance | Met established criteria |
| Tensile | Met established criteria |
| Luer Leakage | Met established criteria |
| Coating Adhesion | Met established criteria |
| Coating Uniformity | Met established criteria |
| Coating Thickness | Met established criteria |
| Surface Integrity | Met established criteria |
| Radiopacity | Met established criteria |
| System Introduction | Met established criteria |
| Device Compatibility | Met established criteria |
Testing was conducted in accordance with ISO 10555-1:2013, ISO 10555-3:2013, AAMI TIR42:2010, ISO 594-1 and ISO 594-2 standards and FDA guidance: FDA Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters September 8, 2010.
Shelf Life Testing (Product and Packaging) and Distribution Shipping Challenge Conditioning and Testing were performed and the devices met established criteria.
Packaging Verification Testing complies with EN ISO 11607-1 and assessed the ability of finished packages to withstand the effects of anticipated hazards of the distribution environment on essential packaging characteristics and the ability of packaging to protect the device and to maintain sterility (sterile barrier testing).
Biocompatibility testing was conducted in accordance with EN ISO 10993-1. MiVi 6F Guide Catheters have been classified according to EN ISO 10993-1:2009, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing as follows:
{7}------------------------------------------------
Category: Externally Communicating Contact type: Circulating blood Contact Duration: Limited exposure (≤24 hours)
Based on this classification, tests relevant to the device described within this premarket notification were selected and conducted in accordance with EN ISO 10993-1 and its applicable sub-parts.
The MiVi 6F Guide Catheter devices meet ethylene Oxide (EO) and ethylene chlorohydrin (ECH) residuals specified in EN ISO 10993-7:2008, Biological Evaluation of Medical Devices -Part 7: Ethylene Oxide Sterilization Residuals. A sterility assurance level (SAL) of 10-6 will be demonstrated.
Testing showed the device, including its packaging, to be biocompatible for its intended use as an externally communicating, circulating blood contacting device as classified under EN ISO 10993-1:2009. The MiVi 6F Guide Catheter successfully passed all of the following biocompatibility tests and was found to be safe and effective and substantially equivalent to the predicate device from a biocompatibility perspective.
| Test | Results | Conclusions |
|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity - ISO MEM Elution<br>Assay | The test article scored '0' at 24, 48, and 72 hours and is<br>considered non-cytotoxic | Not cytotoxic |
| Irritation or Intracutaneous<br>Reactivity - ISO Intracutaneous<br>Reactivity Test | Animals in the study showed no<br>abnormal clinical signs, no<br>significant dermal reactions at the<br>injection and control sites at the<br>24, 48, and 72 hour observation<br>period. | Non-irritating |
| Hemocompatibility<br>Hemolysis - ASTM Hemolysis<br>Direct Contact and Extract<br>Methods | The test article had a blank<br>corrected hemolytic index of<br>0.1% above the control in the<br>direct contact test. The test<br>article had a blank corrected<br>hemolytic index of 0.0 % above<br>the control in the extract test. | Non-hemolytic |
| Hemocompatibility<br>Thrombosis - Four Hour<br>Thromboresistance Evaluation in<br>Dogs | The combined analysis of APTT,<br>platelet counts, device weights,<br>and clinical observations<br>indicated that the animal's<br>clotting abilities were not<br>compromised after implantation<br>of the devices. The test article<br>had similar thromboresistance<br>results to the predicate device so<br>was found to be substantially<br>equivalent to it. | Similar<br>thromboresistance<br>characteristics as control<br>Stryker Concentric<br>Guide Catheter<br>Predicate Device. |
{8}------------------------------------------------
| Test | Results | Conclusions |
|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Hemocompatibility<br>Complement Activation -<br>Complement Activation C3a and<br>SC5b-9) | The test article concentrations of<br>C3a was 1.0 and the predicate<br>device was 3.2 and SC5b-9 was<br>0.0 and the predicate device was<br>1.1 so the test article was not<br>found to be an activator of<br>complements. The test article<br>results were also comparable to<br>the predicate device so was found<br>to be substantially equivalent to<br>it. | Comparable to Stryker<br>Concentric Guide<br>Catheter predicate<br>device for complement<br>activation. Not an<br>activator |
| Pyrogenicity - Limulus<br>Amebocyte Lysate (LAL) Limit<br>Test | Both test articles contained<br><0.005 EU/mL and <0.200<br>EU/device | Non-pyrogenic |
| Sensitization - ISO Guinea Pig<br>Maximization Test | None of the animals challenged<br>with the test article extracts were<br>observed with a sensitization<br>response greater than '0'. | Non-sensitizing |
| Systemic Toxicity (Acute) - ISO<br>Acute Systemic Injection Test | No clinical signs consistent with<br>toxicity were observed and body<br>weight changes were within<br>acceptable parameters over the<br>course of the study. | Non-toxic |
| Ethylene Oxide Residuals -<br>Ethylene Oxide and Ethylene<br>Chlorohydrin | The 6F MiVi Guide Catheter had<br>.094 mg EO/device and .015 mg<br>ECH/device so met both the EO<br>and ECH requirements | Met the ISO 10993-<br>7:2008/ requirements<br>for EO residuals. |
## Summary of Clinical Performance Data
Not applicable as no clinical data was needed to demonstrate substantial equivalence to the predicate devices.
## Conclusion
MiVi Neurosciences has determined its 6F Guide Catheter to be substantially equivalent to the current legally marketed predicate devices because of the following:
- 1. Intended use and indications for use are the same as for the predicate devices.
- 2. There is no difference in the fundamental scientific technology or the devices.
- 3. The risk assessments and successful verification testing including testing to EN ISO 10555-1 and EN ISO 10555-3, raise no new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.