K161152 · Micro Therapeutics, Inc. d/b/a ev3 Neurovascular · DQY · Oct 12, 2016 · Cardiovascular
Device Facts
Record ID
K161152
Device Name
Navien Intracranial Support Catheter
Applicant
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular
Product Code
DQY · Cardiovascular
Decision Date
Oct 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Navien™ Intracranial Support Catheter is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
Device Story
Navien Intracranial Support Catheter is a single-lumen, flexible, variable-stiffness composite catheter; features luer hub for accessory attachment and liquid infusion; includes hydrophilic coating to reduce friction; distal tip marker visible under fluoroscopy. Used in neurovascular and peripheral procedures to facilitate placement of interventional devices (e.g., guidewires up to 0.038 inches). Operated by physicians in clinical settings. Device provides a pathway for interventional tools; aids in navigating tortuous anatomy. Benefits include atraumatic access to target vasculature.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing and design validation comparing the new coating to the predicate device. Bench testing included dimensional inspection, torque strength, coating lubricity/durability, leakage, dynamic burst, tip buckling, and particulate testing per USP <788>.
Technological Characteristics
Single-lumen, variable-stiffness composite catheter; PTFE-lined polymer shaft; Nitinol support; platinum marker band; hydrophilic coating. Dimensions: 90-130 cm length; 0.046"-0.072" ID; 0.058"-0.084" OD. Sterilized via Ethylene Oxide per ISO 11135-1:2007. Biocompatibility per ISO 10993-1. Luer hub per ISO 594-1.
Indications for Use
Indicated for introduction of interventional devices into peripheral and neurovasculature. No specific patient population, age, or gender restrictions provided.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with three figures in profile, representing the department's focus on people and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 12, 2016
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Ms. Jennifer Correa Sr. Product Specialist/Regulatory Affairs 9775 Toledo Way Irvine, California 92618
Re: K161152
Trade/Device Name: Navien Intracranial Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: September 6, 2016 Received: September 8, 2016
Dear Ms. Correa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -SA
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161152
#### Device Name
Navien™ Intracranial Support Catheter
Indications for Use (Describe)
The Navien™ Intracranial Support Catheter is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(k) Owner: | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular<br>9775 Toledo Way<br>Irvine, CA 92618<br>Establishment Registration No. 2029214 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jennifer Correa<br>Senior Regulatory Affairs Specialist<br>Telephone: (949) 297-9563<br>E-mail: jennifer.l.correa@medtronic.com |
| Date Summary Prepared: | September 6, 2016 |
| Trade Name of Device: | Navien™ Intracranial Support Catheter |
| Common Name of Device: | Catheter, Percutaneous |
| Classification of Device: | 21 CFR 870.1250 – Class II |
| Product Code: | DQY |
| Predicate Device: | Navien™ Intracranial Support Catheter, 510(k)#: K110055<br>Originally cleared under trade name: ReFlex™ Guide Catheter |
| Performance Data: | The following testing was performed to support the coating change to the Navien™ Intracranial Support Catheter with new coating to establish substantial equivalence to the Navien™ Intracranial Support Catheter with old coating:<br>Visual/Dimensional Inspection Torque Strength Coating Lubricity Catheter Leakage Test (Liquid) Dynamic Pressure Test Distal Tip Buckling Particulate Testing Physician Usability Testing The following testing was leveraged from the predicate device due to sufficiency of the existing testing record and justification for adoption: System Dimensional- Length Kink Resistance Flow Rate During Continuous Flush (for information only) Leakage (Static Burst Test) |
{4}------------------------------------------------
- Aspiration ●
- Labeling ●
- Catheter Hub (ISO 594-1) ●
- Tensile Srength
- Corrosion Resistance ●
- Tip Configuration ●
- Radiopacity ●
A Design Validation study was performed on a bench model to assess the substantial equivalence and usability of Navien™ Intracranial Support Catheter with new coating compared to the previously cleared Navien™ Intracranial Support Catheter with old coating. Biocompatibility testing, sterilization, and a 2-year accelerated aging study were also performed. No clinical studies were performed as there is no change to the indications for use or the fundamental scientific technology of the device.
- The Navien™ Intracranial Support Catheter with new coating is Conclusion: substantially equivalent to the currently cleared Navien™ Intracranial Support Catheter with old coating based on the successful completion of non-clinical bench and design validation testing as well as identical principles of design, operation and indications for use.
#### Device Description:
The Navien™ Intracranial Support Catheter is a single lumen, flexible, variable stiffness composite catheter which has a luer hub on the proximal end. The catheter shaft has a hydrophilic coating to reduce friction during use. The Navien™ Intracranial Support Catheter has a marker on the distal tip that is visible under fluoroscopy. The dimensions are included in the Navien™ Intracranial Support Catheter inner lumen can accommodate guidewires up to 0.038 inches in diameter to aid in placement of the catheter system. The catheter has both straight and pre-shaped tips.
The proximal end of the Navier™ Intracranial Support Catheter has a Luer fitting to allow attachment of accessories and infusion of liquids through the system. The Catheter is offered in various sizes to accommodate physician preferences and anatomical variations. The catheter is provided sterile, nonpyrogenic, and is intended for single use only.
#### Indications for Use:
The Navien™ Intracranial Support Catheter is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
#### Device Comparison
The table below provides a comparison of the technological characteristics of the Navien™ Intracranial Support Catheter with old coating and the currently cleared Navien™ Intracranial Support Catheter.
{5}------------------------------------------------
| | Navien™ Intracranial Support<br>Catheter with old coating<br>(K110055) | Navien™ Intracranial Support<br>Catheter with new coating |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Indication for<br>Use | The Navien™ Intracranial Support<br>Catheter is indicated for the<br>introduction of interventional devices<br>into the peripheral and<br>neurovasculature. | Same |
| Catheter Shaft<br>Materials | PTFE lined polymeric catheter, with<br>hydrophilic coating | Same |
| Catheter Shaft<br>Support | Nitinol | Same |
| Marker band | Platinum | Same |
| Usable Length | 90 - 130 cm | Same |
| Distal ID | 0.046" – 0.072" | Same |
| Distal OD | 0.058"–0.084" max | Same |
| Proximal ID | 0.046" – 0.072" | Same |
| Proximal OD | 0.058"–0.084" max | Same |
| Tip<br>Configuration | Single, straight flexible tip | Same |
| Sterilization<br>Method | Ethylene Oxide | Same |
| Packaging | Catheter in polyethylene hoop<br>attached to packaging card inside<br>PET/PE/Tyvek pouch inside SBS<br>carton | Same |
#### Sterilization and Shelf Life
The packaged Navien™ Intracranial Support Catheter with new coating is sterilized using a validated ethylene oxide (EO) sterilization cycle at the Sterigenics US; LLC facility located at 4900 Gifford Avenue, Los Angeles, CA. The sterilization cycle has been validated to ensure a sterility assurance level (SAL) of 10° in accordance with ISO 11135-1:2007, Sterilization of health care products - Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
Aging studies for the Navien™ Intracranial Support Catheter with new coating device has established the product and packaging remain functional and maintain sterility for up to 24 months. Aging studies for packaging integrity, seal strength, and device functionality were performed and met all acceptance criteria.
#### Biocompatibility
Biocompatibility testing was performed in compliance with the FDA consensus standard, recognition number 2-156, AAMI/ANSI/ISO 10993-1: 2009, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process and U.S. Food and Drug Administration (FDA)
{6}------------------------------------------------
Blue Book Memorandum G95-1 (1995) guidelines. All studies were conducted in compliance with U.S. Food and Drug Administration Good Laboratory Practice (GLP) regulations set forth in 21 CFR Part 58.
| Test | Result | Conclusion |
|-----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Plastics (USP) | Meets USP Physicochemical<br>extraction parameters. | Passes physical chemical<br>characteristics. |
| L929 MEM Elution Test - ISO | The test article scored "0" at 24, 48<br>and 72 ± 4 hours and is considered<br>non-cytotoxic under the conditions<br>of this test. | Non-cytotoxic |
| Klingman Maximization Test –<br>ISO (Guinea Pig Sensitization) | Under the conditions of this<br>protocol, the test article did not<br>elicit a sensitization response. | Non-sensitizer |
| Intracutaneous Injection Test -<br>ISO | The differences in the mean test<br>and control scores of the extract<br>dermal observations were less than<br>1.0, indicating that the requirements<br>of the ISO Intracutaneous<br>Reactivity Test have been met by<br>the test article. | Non-irritant |
| Acute Systemic Injection Test –<br>ISO | None of the test article extract<br>treated animals were observed with<br>clinical signs consistent with<br>toxicity at any of the observation<br>periods. | Non-cytotoxic |
| Materials Mediated Rabbit<br>Pyrogen - ISO | This response did not exceed the<br>USP limit and meets the<br>requirements for this test. Therefore<br>these results indicate that the test<br>article was determined to be non-<br>pyrogenic. | Non-pyrogenic |
| Hemolysis: Direct Contact /<br>Indirect Extract | There were no significant<br>differences between the test article<br>extract and negative control article<br>results. The test article is<br>considered non-hemolytic | Non-hemolytic |
| Complement activation C3a and<br>SC5b-9 | The levels of C3a and SC5b-9 of<br>the NavienTM catheter with new<br>coating are comparable to the<br>NavienTM with old coating and less<br>than that of the positive control. | Levels of the compliments C3a<br>and SC5b complements were<br>similar for Navien and control<br>device |
| Thrombosis ( <i>in vivo</i> ) – Canine<br>(Navien) | The thromboresistance properties<br>of the NavienTM Intracranial<br>Support Catheters are acceptable in<br>clinical use. | Acceptable, expected to be<br>equivalent to Navien in clinical<br>use |
| Test | Result | Conclusion |
| <i>in vitro</i> Hemocompatibility<br>Assay | The Navien Guide Catheter did not<br>result in a decrease in any blood<br>component as compared to the<br>reference material. These results<br>indicate that the cause of thrombi is<br>not related to the materials exposed<br>to human blood during use. | No adverse effect on platelet<br>and leukocyte counts |
| Partial Thromboplastin Time | Clotting times for the Navien<br>(predicate device) test arms were<br>similar to the negative control and<br>the reference material (HDPE),<br>indicating that the device materials<br>are not an activator of the intrinsic<br>coagulation pathway. | No adverse effect on<br>prothrombin coagulation time<br>of human plasma. |
| Ames bacterial Mutagenicity 4<br>salmonella+1e. coli | Based on the criteria and conditions<br>of the study protocol, the test article<br>is considered non-mutagenic. | Non-mutagenic |
| <i>in vitro</i> Mouse Lymphoma Assay<br>with Extended Treatment | The test article is considered to be<br>non-mutagenic (non-genotoxic and<br>non-clastogenic) in this test system. | Non-mutagenic |
| <i>in vivo</i> Mouse Micronucleus<br>Assay | Based on the criteria of the assay,<br>the test article is considered non-<br>mutagenic in this test system. | Non-mutagenic |
{7}------------------------------------------------
### Performance Testing – Bench
A summary of the pre-clinical bench testing performed for the Navien™ Intracranial Support Catheter is presented in the table below.
{8}------------------------------------------------
| Test | Method | Acceptance Criteria | Conclusions |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Dimensional Inspection | The usable length,<br>proximal and distal inner<br>and outer diameters were<br>measured and recorded. | Usable length, proximal<br>and distal inner and outer<br>diameters must meet all<br>inspection criteria | All device met<br>acceptance criteria. |
| Torque to Failure | Device was tested for full-<br>length torque strength to<br>determine number of<br>rotations to failure. | The catheter must<br>withstand multiple<br>revolutions without failure | Torque to failure testing<br>met acceptance criteria. |
| Coating<br>Lubricity/Durability | Device coating was<br>evaluated for average<br>frictional force and<br>durability. | Average Friction Force<br>must be comparable to the<br>predicate device | Coating lubricity and<br>durability testing met<br>acceptance criteria. |
| Liquid Leakage | Device was tested for ISO<br>10555-1, Annex C liquid<br>leakage testing. | Catheter shall not leak per<br>ISO 10555-1 | Liquid leakage met<br>acceptance criteria. |
| Dynamic Burst | Device was tested under<br>full-length static<br>conditions to burst and at<br>pressures experienced<br>during worst-case dynamic<br>injections. | Catheter shall not burst or<br>leak during a dynamic<br>pressure test | Dynamic burst testing<br>met acceptance criteria. |
| Tip Buckling | Repeated distal tip<br>buckling force under<br>compressive load was<br>evaluated for stiffness. | Compressive load shall not<br>be statistically<br>significantly different to<br>comparable and/or the<br>predicate device | Tip buckling force met<br>acceptance criteria. |
| Particulate Testing | Device was evaluated for<br>particulate generation<br>under simulated use in a<br>representative tortuous<br>anatomical model per<br>USP<788> | Particulate must be within<br>limits per USP<788> | Number of particulates<br>generated met<br>acceptance criteria. |
| Physician Usability<br>Testing | The device was navigated<br>through a tortuous<br>benchtop model to assess<br>accessibility and the tip<br>atraumatic profile. | Assess navigation of<br>Navien™ with new<br>coating to the ICA and<br>comparing Navien™ with<br>new to Navien™ with old<br>coating with regards to tip<br>atraumaticity. | All test results met the<br>acceptance criteria. |
#### Performance Testing - Animal
No animal study was performed as there is no change to the indications for use or the fundamental scientific technology for the new device. Substantial equivalence of the Navien™ Intracranial Support Catheter with new coating has been established to the predicate device through the results of bench testing.
## Performance Testing – Clinical
{9}------------------------------------------------
No clinical study was performed as there is no change to the indications for use or the fundamental scientific technology for the new devices. Substantial equivalence of the Navien™ Intracranial Support Catheter with new coating has been established to the predicate device through the results of bench and design validation testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.