K191492 · Neurovirtual USA, Inc. · GWQ · Apr 24, 2020 · Neurology
Device Facts
Record ID
K191492
Device Name
Maxxi Position Sensor
Applicant
Neurovirtual USA, Inc.
Product Code
GWQ · Neurology
Decision Date
Apr 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The Maxxi Position Sensor is intended for use with the BWMini polysomnograph system, to acquire the body position of adult patients during sleep studies. The Sensor produces signals for five positions: supine, prone, left and right. It is intended for use in research, home sleep studies, ambulatory, and clinical environments.
Device Story
Maxxi Position Sensor is a wearable device for sleep studies; attaches to chest or abdominal belts via velcro. Uses electronic tilt switches to detect body position (supine, prone, left, right, upright). Produces 0-1 VDC output signals transmitted via 7ft cable to a polysomnograph (PSG) recorder. Used in clinical, home, or research environments; operated by healthcare professionals. Provides real-time body position data to the PSG system; assists clinicians in correlating sleep disorders with patient posture. Benefits patients by enabling accurate monitoring of position-dependent sleep events.
Clinical Evidence
No clinical trial was performed. Evidence consists of bench testing and a simulated use study with N=20 volunteers. The simulated use study compared the subject device to predicates, reporting 100% accuracy and a 1-second latency difference in detecting body position.
Indicated for adult patients undergoing sleep studies to acquire body position (supine, prone, left, right, upright) for research, home, ambulatory, or clinical environments.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Neurovirtual USA, Inc. Eduardo Faria CEO 3303 W Commercial Blvd. Suite #100 Fort Lauderdale, Florida 33309
## Re: K191492
Trade/Device Name: Maxxi Position Sensor Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ, OLV Dated: January 28, 2020 Received: February 3, 2020
Dear Eduardo Faria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191492
Device Name Maxxi Position Sensor
Indications for Use (Describe)
The Maxxi Position Sensor is intended for use with the BWMini polysomnograph system, to acquire the body position of adult patients during sleep studies. The Sensor produces signals for five positions: supine, prone, left and right. It is intended for use in research, home sleep studies, ambulatory, and clinical environments.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10pt"> <span style="font-family: Wingdings">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt"> <span style="font-family: Wingdings">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform graphic in shades of orange and yellow, followed by the word "NEUROVIRTUAL" in blue. The "O" in "NEUROVIRTUAL" is colored orange, matching the waveform graphic.
## 510(k) SUMMARY
- A) Submitter's Name: Neurovirtual USA, Inc.
#### Owner / Operator Registration Number: 9091724 Manufacture Registration Number: 3006125362
- B) Address: 3303 W Commercial Blvd #100 Fort Lauderdale, FL - 33309
- C) Phone and Fax Numbers Phone: (786) 693-8200 Fax: (305) 393-8429
- D) Contact Person: Eduardo J. Faria
- E) Revision Date: March 24, 2020
- F) Classification Name: Common / Usual Name: Body Position Sensor Proprietary Name: Maxxi Position Sensor Product code: GWQ, OLV Class: Class II Regulation: 21 CFR 882.1400
#### G) Device Description
The Maxxi Position Sensor is a sensor that attaches to either around chest or abdominal belts using velcro tapes- so no additional belts or attachment systems are needed. The sensor features a fully encapsulated active element for troublefree cleaning.
The Maxxi Position Sensor produces a clear and reliable signal for five positions: Upright, supine, prone, left and right.
It comes with 7ft long cable for a convenient connection with the PSG device recorder.
#### H) Substantial Equivalence:
The Maxxi Position Sensor is equivalent with the following products:
| 510(k) Number | Model | Company |
|---------------|-----------------------------|-----------------------|
| K131335 | BWMINI | NEUROVIRTUAL USA INC. |
| K981969 | ULTIMA BODY POSITION SENSOR | BRAEBON MEDICAL CORP. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform graphic in shades of orange, followed by the word "NEUROVIRTUAL" in blue, sans-serif font. The "O" in "NEUROVIRTUAL" is replaced with a solid orange circle.
#### 1. Indications for Use:
| Indications for Use Comparison | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Neurovirtual<br>Maxxi Position Sensor | BRAEBON MEDICAL CORP.<br>Ultima Body Position Sensor | Neurovirtual<br>BWMini |
| The Maxxi Position<br>Sensor is intended for<br>use with the BWMini<br>polysomnograph system,<br>to acquire the body<br>position of adult patients<br>during sleep studies. The<br>Sensor produces signals<br>for five positions:<br>standing/sitting, supine,<br>prone, left and right. It is<br>intended for use in<br>research, home sleep<br>studies, ambulatory, and<br>clinical environments. | The BRAEBON MEDICAL<br>CORPORATION Ultima Body<br>Position Sensor™ is intended for<br>use during sleep disorder studies<br>as an indicator of body position.<br>The sensor uses a three-volt<br>lithium battery and plugs directly<br>into either a DC amplifier or<br>multiplexer.<br><br>The target population of the<br>Ultima Body Position Sensor is all<br>children and adult patients who<br>are screened during sleep<br>disorder studies. The majority of<br>the screenings occur at a sleep<br>laboratory although the sensor<br>can also be used in home<br>studies.<br><br>The Ultima Body Position Sensor<br>is intended for use only by or on<br>the order of a physician. | BWMini is an electroencephalograph,<br>which is a device used to measure and<br>record the electrical activity of the<br>patient's brain obtained by placing two<br>or more electrodes on the head.<br><br>BWMini is multi-channel (up to 32<br>channels) system designed for<br>Electroencephalograph (EEG),<br>Polysomnography (PSG) and Home<br>Sleep Testing (HST) recording<br>application, in research, home sleep<br>studies, ambulatory and clinical<br>environments.<br><br>The BWMini does not make any<br>judgment of normality or abnormality<br>of the displayed signals or the results<br>of an analysis. In no way are any of<br>the functions represented as being in<br>and of themselves diagnostic. |
#### 2. Technological Characteristics Comparison:
The predicate devices used to establish substantial equivalence for the Maxxi Position Sensor is outlined below. This section of this submission will provide a comparison of design, materials, and technical specifications of the Maxxi Position Sensor to each of the predicate devices stratified by functional modality.
| Specifications Comparison | | | | |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Device Brand<br>and Common<br>Name | Neurovirtual<br>Maxxi Position<br>Sensor | Neurovirtual<br>BWMini | BRAEBON<br>MEDICAL CORP.<br>Ultima Body<br>Position Sensor | Comments |
| 510(k)<br>Number | K191492 | K131335 | K923033 | NA |
| Classification | GWQ, OLV | GWQ, OLV | LEL | The predicate BWMini<br>is classified as GWQ |
| Regulation # | 21 CFR 882.1400 | 21 CFR 882.1400 | 21 CFR 882.5050 | is classified as GWQ |
| Classification<br>Name | Electroencephalogr<br>aph used for<br>polysomnography<br>or sleep studies | Electroencephalogr<br>aph used for<br>polysomnography<br>or sleep studies | Device, Sleep<br>Assessment | because it is a full<br>polysomnograph<br>recording device that<br>includes an internal<br>body position sensor<br>with same |
| | | | | technological<br>characteristics as the<br>subject device. This<br>information may not<br>publicly available. |
| Patient<br>Population | Adult patient | Adult patient | Adult patient | Identical |
| Indented<br>Environment | The intended<br>environments are<br>research, home<br>sleep studies,<br>ambulatory, and<br>clinical<br>environments. It<br>is intended for<br>adult patients. | The intended<br>environments are<br>research, home<br>sleep studies,<br>ambulatory, and<br>clinical<br>environments. It is<br>intended for adult<br>patients. | The intended<br>environments are<br>research, home<br>sleep studies,<br>ambulatory, and<br>clinical<br>environments. It is<br>intended for adult<br>patients. | Identical |
| Prescription<br>Use | YES | YES | YES | Identical |
| Sensor<br>Material | Plastic ABS<br>enclosure<br>Electronic<br>components<br>Printed Circuit<br>board | Plastic ABS<br>enclosure<br>Electronic<br>components<br>Printed Circuit<br>board | Plastic ABS<br>enclosure<br>Electronic<br>components<br>Printed Circuit<br>board | Equivalent |
| Cables<br>Material | ABS - plastic<br>injection | ABS - plastic<br>injection | ABS - wire jacket | Equivalent |
| Interface<br>with Patient | Around the<br>abdomen or<br>thorax belts<br>using velcro<br>tapes | Around the<br>abdomen or thorax<br>belts using velcro<br>tapes | Around the<br>abdomen or thorax<br>belts using velcro<br>tapes | Identical |
| Type of<br>Equipment to<br>Be Connected<br>to | Polysomnography<br>recorder | NA | Polysomnography<br>recorder | Identical |
| Connector<br>Type | Monopolar DIN<br>42-802 touch<br>proof | NA | Monopolar DIN 42-<br>802 touch proof | Identical |
| Signals<br>output | Body position:<br>upright, supine,<br>prone, left and<br>right<br>0 to 1 VDC<br>output | Body position:<br>upright, supine,<br>prone, left and<br>right<br>0 to 1 VDC output | Body position:<br>upright, supine,<br>prone, left and<br>right<br>0 to 1 VDC output | Identical |
| Position<br>Detection<br>Technology | Combination of<br>electronic tilt<br>switches<br>generates<br>different types of<br>signals<br>depending on the<br>sensor position. | Combination of<br>electronic tilt<br>switches generates<br>different types of<br>signals depending<br>on the sensor<br>position. | Combination of<br>electronic tilt<br>switches generates<br>different types of<br>signals depending<br>on the sensor<br>position. | Identical |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Neurovirtual. The logo features a stylized waveform graphic in shades of orange on the left. To the right of the waveform is the company name "NEUROVIRTUAL" in blue, with the "O" in Neurovirtual being a solid orange circle.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform in shades of yellow and orange, followed by the word "NEUROVIRTUAL" in blue. The "O" in "NEUROVIRTUAL" is colored orange, matching the waveform.
| Power Source | 3VDC coin<br>battery | AA batteries | 3VDC coin battery | Similar |
|-------------------------|------------------------------------------------|------------------------------|--------------------------------|------------|
| Applicable<br>Standards | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 | Identical |
| Packing | Plastic bag | Cartoon box | Plastic bag | Equivalent |
| Sensor<br>Dimensions | 39x31x15 mm | NA | 40x29x14 mm | Equivalent |
| Device<br>Picture | Image: Device with black strap and white wires | Image: BW device | Image: Device with black strap | Equivalent |
Substantial Equivalency Discussion: The Maxxi Position and the predicate devices BWMini and Ultima Body Position Sensor are substantially equivalent in technology, function, and intended use: the devices are intended to provide body position signals for sleep disordered studies; the devices use the same method of action; the devices use equivalent material and also provide identical output signals.
### I) Performance Testing Summary:
### 1. Electrical Safety and Effectiveness Testing
The Maxxi Position was submitted to standard IEC 60601-1 test which resulted in full compliance as reported in the test reports.
#### 2. EMC Testing
The Maxxi Position was submitted to electromagnetic compatibility test, standard IEC 60601-1-2 which resulted in full compliance as reported in the test reports.
#### 3. Signal Quality and Comparison Testing:
Signal integrity tests were conducted for the Maxxi Position Sensor with focus on signal to noise ratio, signal range, bandwidth and linearity and the test results compared to the signal integrity test conducted for the predicate Ultima Body Position Sensor.
#### 4. Performance Testing:
The performance test was conducted in order to verify technological and functional performance characteristics when compared with the predicate. The summary of the results is listed below.
| Performance<br>Test | Description | Acceptance Criteria<br>(Predicate Specs) | Maxxi Position<br>Results |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|-------------------------------------------------------|
| Sensor<br>Dimensions | Verify if the dimensions of the<br>sensor are within the acceptable<br>range in accordance with the<br>product specifications. | 40x29x14 mm<br>Allowance: +-10% | Measured Value:<br>39x31x15 mm<br>Pass ( x ) Fail ( ) |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform graphic in shades of orange, followed by the word "NEUROVIRTUAL" in blue. The "O" in Neurovirtual is replaced with an orange circle. The logo is simple and modern, and it is likely used to represent a company or organization that is involved in neuroscience or virtual reality.
| Cable<br>Dimensions | Verify if the cable length is within<br>the acceptable range. | Cable length: 7ft<br>Allowance: +-5% | Measured Value: 7ft<br>Pass ( x ) Fail ( ) |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Cable<br>Connectivity | Using a multimeter in continuity<br>scale, check if the internal belt<br>wiring has connectivity. | Not allowed false<br>contact, or no<br>connectivity. | Pass ( x ) Fail ( ) |
| Sensor<br>Visual<br>Aspects | Verify the visual aspects of the<br>product including, labeling,<br>mechanism of action, method of<br>connection, application,<br>interaction with the user and<br>cable specifications. | The visual aspects of<br>the Maxxi Position<br>must be<br>similar/equivalent to<br>the predicate device<br>Ultima Body Position<br>Sensor. | Pass ( x ) Fail ( ) |
| Sensor<br>Signal<br>Aspect | Signal integrity tests were<br>conducted for the Maxxi Position<br>Sensor with focus on signal to<br>noise ratio, signal range,<br>bandwidth and linearity and the<br>test results compared to the<br>signal integrity test conducted<br>for the predicate Ultima Body<br>Position Sensor. | Qualitative signal<br>analysis must be<br>similar in signal type,<br>amplitude and<br>linearity when<br>compared with the<br>predicate device<br>Ultima Body Position<br>Sensor. | Pass ( x ) Fail ( ) |
| Sensor<br>Functional<br>Aspect | Verify if the PSG system<br>identifies the sensor's position<br>Supine, Prone, Left, Right and<br>Stand accordingly with the<br>respective sensor output signals. | All positions signal<br>shown in the user<br>manual must follow<br>the patient body<br>position based on the<br>sensor axis position | Pass ( x ) Fail ( ) |
#### 5. Performance Testing in Simulated Use:
A comparative test in simulated use was performed in N=20 volunteers in a simulated use study between the subject device and predicates in order to compare the accuracy and latency to detect the patient's body position in a PSG system.
The results showed the average time to detect the patient body position was 1 second of latency difference between the devices and 100% of accuracy between them. This result demonstrated substantially equivalence between the subject device and the predicates.
Based on the substantial equivalence comparison, and nonclinical Conclusion: performance tests applied to the Maxxi Position Sensor and the predicates, we conclude that the devices have similar indications for use, technological characteristics and method of action, additionally the nonclinical performance testing demonstrate that the subject device is safe and effectively as well substantially equivalency to the predicates.
## J) Clinical Testing:
No clinical trial was performed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.