K991054 · Nicolet Biomedical, Inc. · GWQ · Nov 3, 1999 · Neurology
Device Facts
Record ID
K991054
Device Name
BRAVO MULTI-MODALITY SYSTEM
Applicant
Nicolet Biomedical, Inc.
Product Code
GWQ · Neurology
Decision Date
Nov 3, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The Bravo Multi-Modality System is intended to record and display EEG, EP, EMG and TCD data in the clinic and hospital (including the hospital room, operating room, emergency room, intensive care unit, neuro intensive care unit, critical care unit, etc.), and to import and display data from third-party monitoring devices such as vital signs monitors. It is intended to aid the diagnosis and monitoring of potential disorders of the central and peripheral nervous system and muscles.
Device Story
Bravo Multi-Modality System is a PC-based digital recorder for neurological and neurovascular data. Inputs: neuroelectric/neuromuscular signals via skin/muscle electrodes (EEG, EMG, EP); ultrasound signals via transducer (TCD); vital signs data from third-party monitors. Operation: passive recording for EEG/EMG; active stimulation (light, sound, electrical) for EP; ultrasound Doppler shift detection for TCD. System integrates multiple modalities on one platform. Used in clinics/hospitals by clinicians to aid diagnosis/monitoring of nervous system/muscle disorders. Output: displayed, analyzed, or stored physiological data. Clinical decision-making relies on human interpretation of system output.
Clinical Evidence
No clinical data provided. Device validation relies on nonclinical testing of program functions, user inputs, and recording of test signals to verify performance against expected outcomes.
Technological Characteristics
PC-based digital data recorder. Modalities: EEG, EMG, EP, TCD. Inputs: conductive electrodes, ultrasound transducers. Energy: electrical stimulation, ultrasound. Connectivity: imports data from third-party monitors. Software: multi-application concurrent execution on a single platform. Design and materials are consistent with predicate devices.
Indications for Use
Indicated for patients requiring diagnostic monitoring of central/peripheral nervous system and muscle disorders. Used in clinical and hospital settings (OR, ICU, etc.) to record EEG, EP, EMG, and TCD data.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Nov - 3 1999
K 991054
## 510(k) SUMMARY BRAVO MULTI-MODALITY SYSTEM
## (a) INFORMATION REQUIRED FOR ALL SUMMARIES
| | (1) Submitter's Name: | Nicolet Biomedical Inc. |
|--|--------------------------------|-------------------------------------------------------------------------------------------------------------|
| | Submitter's Address: | 5225 Verona Road, Bldg. 2<br>Madison, WI 53711 U.S.A. |
| | Submitter's Telephone Number: | (608) 273-5000 |
| | Contact Name: | Douglas E. Pfrang |
| | Date summary was prepared: | March 26, 1999 |
| | (2) Trade or proprietary name: | Bravo Multi-Modality System |
| | Common or usual name: | Electroencephalograph (EEG)<br>Evoked Potential (EP)<br>Electromyograph (EMG)<br>Transcranial Doppler (TCD) |
| | Classification name: | Electroencephalograph (84GWQ) |
| | | Evoked Potential Electrical Stimulator (84GWF) |
| | | Evoked Potential Photic Stimulator (84GWE) |
| | | Evoked Potential Auditory Stimulator (84GWJ) |
| | | Electromyograph (89GWP) |
| | | Nonfetal Ultrasonic Monitor (90JAF) IKN & OLT |
(3) Identification of the legally marketed devices to which equivalence is claimed.
- 1. Nicolet Spirit EMG/EP (FDA Log No. K905632)
- 2. Nicolet Voyageur EEG (FDA Log No. K921927)
- 3. Nicolet/EME Trans-Scan TCD (FDA Log No. K874685)
(4) Description of the device, including an explanation of how the device functions, the scientific (4) Dosenpton of the significant performance characteristics such as device design, material used, and physical properties.
The Bravo Multi-Modality System is a personal computer-based digital data recorder for continuously monitoring various types of neurological information, including: (i) neuroelectric and neuromuscular data pertaining to the patient's central and peripheral nervous system and muscles, and (ii) neurovascular data pertaining to blood flow in the patient's brain and related blood vessels.
The intended use of this device is to record and display EEG, EMG, EP and TCD signals; and to import and display data from third-party monitoring devices, such as vital signs monitors. EEG signals are passively recorded using electrically-conductive electrodes that are placed in electrical contact with the patient's skin or nervous system. EMG and EP signals are recorded using electrically-conductive electrodes that are placed in electrical contact with the patient's skin, nervous system or muscles. EMG signals are passively recorded, while EP
{1}------------------------------------------------
signals are evoked using a light source, a sound source, or an electrical stimulator. TCD signals are actively recorded using a non-invasive ultrasound transducer that emits and records ultrasound energy. The ultrasound energy is applied externally to the skin, passes through the skin and body tissues, reflects off blood molecules moving in the blood stream, passes back through the body tissues and skin, and returns to the transducer. Movement of the blood molecules causes a frequency or "Doppler" shift in the returned ultrasound energy, which is detected and converted into a signal representing the velocity of the blood from which the ultrasound energy was reflected. Third-party monitoring devices, such as vital signs monitors, acquire and display a variety of physiological data from the patient. Importing and displaying such data is done by taking an output signal directly from such monitoring devices without any additional connections to the patient. Data obtained from the monitoring device is then displayed, analyzed or stored by the Bravo Multi-Modality System independently of how the third-party device handles the data.
(5) Statement of the intended use, including a description of the patient populations, and diseases or conditions, that the device is intended to diagnose, treat, prevent, cure, or mitigate. If different from the predicate device, an explanation of why the differences are not critical to the intended use when the device is used as labeled.
The Bravo Multi-Modality System is intended to record and display EEG, EP, EMG and TCD data in the clinic and hospital (including the hospital room, operating room, emergency room, intensive care unit, neuro intensive care unit, critical care unit, etc.), and to import and display data from third-party monitoring devices. It is intended to aid the diagnosis and monitoring of potential disorders of the central and peripheral nervous system and muscles. It differs from the predicate devices in that it contains multiple modalities in a single device, whereas each predicate device has a very limited number of modalities. Nevertheless, like the predicate devices, the Bravo Multi-Modality System is intended to involve competent human intervention before any impact on human health occurs (i.e., clinical judgment and experience must be used to check and interpret the system's output).
(6) Comparison between the technological characteristics of the new and predicate devices, such as design, material, chemical composition, energy source, etc.
The technological characteristics of the new and predicate devices are substantially equivalent because the hardware and software in the Bravo Multi-Modality System is largely based on the predicate devices. The design, material, chemical composition, energy source, etc., are essentially unchanged. The main difference is that advances in computer technology (i.e., hardware and software) have enabled multiple software applications to run concurrently on a single computer platform, thereby making possible a single integrated system with multimodality capabilities. Nevertheless, each modality, when compared to a predicate device of the same modality, has similar performance specifications, a similar set of user-adjustable parameters, and is designed to comply with substantially the same performance standards.
{2}------------------------------------------------
## (b) INFORMATION REQUIRED IF EQUIVALENCE IS BASED ON PERFORMANCE
DATA
(1) Brief discussion of nonclinical tests.
Nonclinical tests consist of various tests to verify program function, such as testing user
uts and recording various test signals. Tests are ongoing, but in line to for a included tests collais of various tests to verify program function, such as testing used
inputs and recording various tests signals. Tests are ongoing, but indicate that the performing as expected. Test results are summarized below in indicate that the
(2) Brief discussion of clinical tests including, if applicable, a descriptions of the subjects, a
discussion of safety or effectiveness data obtained, a discussion of consis person of safety or effectiveness data obtained, a discussion of the subjects,
discussion of safety or effectiveness data obtained, a discussion of any adverse effects of complications, and any other relevant information.
Clinical tests will be used to validate the device's performance under simulated or actual
le stations in use conditions. Because each of the individual modalities was derived from an existing from an existing the decided data will be a primary validation tool. To provided for an existing
the device will be a primary validation tool. To provided tool and easting
and to vill be ev the device will be evaluated at different health care sites to solicit can annound with the mainting valiation, which in the studion, which is and and to was over of ovaluated at different health care sites to solicit common
(3) Conclusions drawn from nonclinical and clinical tests that demonstrate the device is as safe
(c){ca (1) as safe
Clinical and nonclinical results of testing the Bravo Multi-Modality System are showing
it it performs substantially as expected: i.e. that it is substantially as expected that it performs substantially as expected; i.e., that it is substantially equivalent to the predicate devices of safety as expected; 1.e., that it is
predicate devices in terms of safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol is a stylized eagle or bird-like figure, with three lines forming the body and wings. The logo is simple and recognizable, representing the department's role in health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Douglas E. Pfrang Director, Regulatory and Legal Affairs Nicolet Biomedical, Inc. 5225 Verona Road, Building 2 Madison, Wisconsin 53711-4495
APR - 9 2012
Re: K991054
Trade/Device Name: Bravo Multi-Modality System Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked response electrical stimulator Regulatory Class: II Product Code: GWF, GWQ, GWE, GWJ, OLT, JAF, IKN Dated (Date on orig SE ltr): August 4, 1999 Received (Date on orig SE Itr): August 5, 1999
Dear Mr. Pfrang:
This letter corrects our substantially equivalent letter of November 3, 1999.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Image /page/4/Picture/7 description: The image shows a handwritten word "for" in cursive. The letter 'f' is elongated and loops downward, forming the main body of the word. The letters 'o' and 'r' are connected to the 'f' and are smaller in size. The writing style appears fluid and elegant.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K991654 510(k) Number (if known):
Device Name: _Bravo Multi-Modality System
Indications For Use:
The Bravo Multi-Modality System is intended to record and display EEG, EP, EMG and TCD data in the clinic and hospital (including the hospital room, operating room, emergency room, intensive care unit, neuro intensive care unit, critical care unit, etc.), and to import and display data from third-party monitoring devices such as vital signs monitors. It is intended to aid the diagnosis and monitoring of potential disorders of the central and peripheral nervous system and muscles.
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Milken
for (Division Sign-Off)
Division of General Restorative Devices
510k) Number. K991054
**Prescription Use**
Per 21 CFR 801.109
OR
Over-The-Counter Use
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.