The Nemus 2 System is intended to acquire the bioelectric signals produced by the patient's central and peripheral nervous system and muscles. The acquired signals are transmitted to a PC during recording of neurophysiology examinations. The Nemus 2 System consists of a physiological signal amplifier, an evoked response auditory stimulator, an evoked response electrical stimulator, and an evoked response photic stimulator. The user needs to assure compatibility of the Nemus 2 System to the other components of the user's own system.
Device Story
Nemus 2 is a physiological signal amplifier/stimulator system for neurophysiology examinations. It captures bioelectric signals from the central/peripheral nervous system and muscles via electrodes/sensors. The device amplifies, filters, and performs analog-to-digital conversion of these signals before transmitting them to a host PC for analysis. It includes integrated electrical, auditory, and photic stimulators for evoked response analysis. Used in hospitals, laboratories, and institutions by clinicians. The system consists of a base unit (amplifier), AC/DC converter, junction box, and optional peripherals (keyboard, LED flash stimulator). The host PC runs application software to process and display data; the device itself does not perform data management. The output allows clinicians to record and analyze neurophysiological activity to aid in diagnosis and prognosis. Benefits include enhanced diagnostic capability through combined EMG and EEG/EP channel acquisition.
Clinical Evidence
No clinical tests were performed. Substantial equivalence is supported by bench testing, including electrical, electromagnetic, and mechanical safety evaluations (IEC 60601-1, 60601-1-1, 60601-1-2, 60601-1-4, 60601-2-26, 60601-2-40) and verification of technical specifications such as saturation, polarization, offset, gain, noise, CMRR, IMRR, and input impedance.
Technological Characteristics
Physiological signal amplifier with 2 EMG and 20 EEG/EP channels. Materials: Polycarbonate (Macrolon 6557 UL VO). Connectivity: LAN Ethernet IEEE 802.3. Power: External IEC 60601-1 mains adapter. Isolation: BF Type (EMG/Acoustic), CF Type (EEG/EP). Sampling: 16384 Hz (EEG/EP), 32768 Hz (EMG). A/D conversion: 24-bit (EMG), 16-bit (EEG/EP). Includes constant current electrical stimulator, audio stimulator, and optional LED flash stimulator.
Indications for Use
Indicated for pediatric through adult patients requiring acquisition of bioelectric signals from the central and peripheral nervous system and muscles for neurophysiology examinations. Contraindicated for use in the presence of flammable anesthetics or in conjunction with defibrillation equipment.
Regulatory Classification
Identification
A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).
{1}------------------------------------------------
K133517
FEB 26 2014
# Traditional 510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92.
807.92(a)(1)
| Submitter Information | |
|----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EB Neuro, S.p.A.<br>Via Pietro Fanfani 97/a<br>Florence, Italy 50127 | |
| Contact Person: | Allison Scott<br>317.228.8719<br>Allison.Scott@Navigant.com |
| Date: | January 17, 2014 |
| 807.92(a)(2) | |
| | Devices |
| Common Name: | Physiological Signal Amplifier |
| Trade Name: | Nemus 2 |
| Classification Name(s): | Physiological Signal Amplifier |
| Classification Number: | 21 CFR 882.1835 Physiological signal amplifier, Product Code GWL<br>21 CFR 882.1870 Evoked response electrical stimulator, Product Code<br>GWF<br>21 CFR 882.1890 Evoked response photic stimulator, Product Code GWE<br>21 CFR 882.1900 Evoked response auditory stimulator, Product Code<br>GWJ |
| Regulatory Class: | Class II |
| 807.92(a)(2) | |
| Predicate Device(s) | | |
|---------------------|----------|---------|
| Device | Owner | 510(k) |
| Nemus 1 | EB Neuro | K073415 |
| BE Plus LTM | EB Neuro | K121996 |
{2}------------------------------------------------
807.92(a)(4)
### Device Description
The Nemus 2 System is made up of the Nemus 2 Amplifier/Stimulator, an AC/DC converter to power the amplifier, a "junction box" to connect the system to the Host PC, and (optionally) by a dedicate keyboard and a Flash LED stimulator.
The Nemus 2 System is intended to acquire bioelectric signals produced by the patient's central and peripheral nervous system and muscles. The acquired signals are transmitted to a PC during recording of neurophysiology examinations. The device may use electrical stimulus, sound stimulus, or visual stimulus for use in evoked response analysis.
The Nemus 2 System is an amplifier/stimulator device especially developed to amplify biological signals. The amplifier captures the biological signal from the human body through specialized sensors or electrodes, amplifies the very low electrical signal, and filters it to accomplish an antialiasing in order to make an optimal ANALOG to DIGITAL conversion. The data, once converted in numerical form, are "passed" to a host computer, which at this point is free to elaborate the data following the logic of the application software running on the host. The Nemus 2 System is not involved in the data management performed by the host.
807.92(a)(5)
### Intended Use
The Nemus 2 System is intended to acquire the bioelectric signals produced by the patient's central and peripheral nervous system and muscles. The acquired signals are transmitted to a PC during recording of neurophysiology examinations. The Nemus 2 System consists of a physiological signal amplifier, an evoked response auditory stimulator, an evoked response electrical stimulator, and an evoked response photic stimulator. The user needs to assure compatibility of the Nemus 2 System to the other components of the user's own system.
{3}------------------------------------------------
807.92(a)(6)
### Technological Characteristics
The Nemus 2 is very similar to the Nemus 1 both from the hardware and the software point of view. Nemus 2 may be considered an "enhanced" version of Nemus 1. Practically the Nemus 1 with an added internal board deputed to the Acquisition of the 20 EEG/EP channels. The case is practically identical with the obvious exception of the upper panel in which, in the Nemus 2 have been allocated the input sockets for the added channels.
In particular in relation to the internal stimulators:
- -The audio stimulator is absolutely identical
- । The electric stimulator of the Nemus 2 is the same of the Nemus 1, but with the difference that in the Nemus 2 the electric pulse may be delivered to 5 different output sockets. Under control of the firmware the stimulus may be delivered on one of this 5 output, but only one may be active at the same time, so also the Nemus 2 simulator has "only" one output. Apart from this aspect the two stimulators are identical.
The two devices use the same AC/DC mains adapter.
Parts of the system external the "Base unit" (the amplifier), i.e. the connection box, the link cable, the stimulation probe, the optional dedicated myographic keyboard, are the same for both devices.
The firmware is substantially the same with the only obvious exception of the added 20 EEG/EP channels. Apart from this, the host communication protocol is the same for both devices.
Differently from the Nemus 2 can handle, as an option, the EBNeuro Flash LED stimulator, useful during EEG or Visual Evoked Potential exams.
The Nemus 2 20 EEG/EP channels use the same technology of the 64 channels of the BE Plus LTM.
The part of Nemus 2 host communication protocol related to the transmission of the EEG data is substantially the same to the BE Plus LTM due to the fact that the processing architecture is very similar.
The Nemus 2 and BE Plus LTM devices can handle the same (optional) EBNeuro Flash LED stimulator.
The Nemus 2 and BE Plus LTM devices use the same AC/DC mains adapter. However the BE Plus LTM may be powered also from internal batteries.
An obvious difference is that the BE Plus LTM is not capable to hannels due to absence of internal stimulator and due to the technical characteristic of its acquisition channels (band, sampling rate, etc)
{4}------------------------------------------------
## Substantial Equivalence Comparison Table
| Product<br>Characteristic | Nemus2 OEM System<br>(submission device) | Nemus1 System<br>(Predicate Device) | BE Plus LTM<br>(Predicate Device) |
|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory | | | |
| Manufacturer | EBNeuro S.p.A. | EBNeuro S.p.A. | EBNeuro S.p.A. |
| 510(k) number | TBA | K073415 | K121996 |
| Device class | Class II | Class II | Class II |
| Product code | GWL, GWF,GWJ, GWE | IKN, GWF,GWJ,JXE | GWL |
| Device type | Physiological signal<br>Amplifier, Electrical<br>stimulator, Audio, photic<br>stimulator | Electromyograph | Physiological signal<br>Amplifier, |
| Regulation Number | 882.1835 | 890.1375 | 882.1835 |
| Product | Nemus2 Amplifier<br>(submission device) | Nemus1 System<br>(Predicate Device) | BE Plus LTM<br>(Predicate Device) |
| Characteristic | | | |
| Labeling | | | |
| Intended use | The Nemus 2 System is<br>intended to acquire the<br>bioelectric signals<br>produced by the patient's<br>central and peripheral<br>nervous system and<br>muscles. The acquired<br>signals are transmitted to a<br>PC during recording of<br>neurophysiology<br>examinations.<br>The Nemus 2 System<br>consists of a physiological<br>signal amplifier, an<br>evoked response auditory<br>stimulator, an evoked<br>response electrical<br>stimulator, and an evoked<br>response photic stimulator.<br>The user needs to assure<br>compatibility of the<br>Nemus 2 System to the<br>other components of the<br>user's own system. | The Nemus1 system is<br>intended to monitor,<br>record and display the<br>bioelectric signals<br>produced by muscles, to<br>stimulate peripheral<br>nerves and to monitor<br>record and display the<br>electrical activity<br>produced by nerves to aid<br>the clinician in the<br>diagnosis and prognosis of<br>neuromuscular disease<br>(EMG).<br>The device may use<br>electrical stimulus or<br>sound stimulus for use in<br>evoked response<br>measurements (EP). | Acquisition of EEG,<br>polygraphy and<br>polysomnography<br>signals and<br>transmission of these to<br>a PC during recording<br>of neurophysiology<br>examinations |
| Warnings | Items related to off-<br>label use | same | same |
| Contraindication | Items related to design<br>and indicated use<br>limitations, such as not<br>for use in the presence<br>of flammable<br>anesthetics or in<br>conjunction with<br>defibrillation equipment | same | same |
| Target population | Pediatric through adult | same | same |
| Environment of use | Hospitals, laboratory,<br>institutions, or other test<br>environments<br>The device may be placed<br>in Intensive Care Unit or | Hospitals, institutions,<br>or other test<br>environments. | Hospitals,<br>institutions, or other<br>test environments |
| Product Characteristic | Nemus2 Amplifier<br>(submission device) | Nemus1 System<br>(Predicate Device) | BE Plus LTM<br>(Predicate Device) |
| Prescription status | Operating Room for continuous recording.<br>Available only on the order of a physician | same | same |
| User Service | No user service allowed | same | same |
| Design - General | | | |
| The device acquire<br>bioelectric signals<br>produced by<br>muscles and nerves<br>(EMG) | YES | YES | NO |
| Device acquire the<br>electrical activity of<br>the brain (EEG/EP) | YES | NO | YES |
| Patient inputs | 2 EMG channels<br>20 EEG/EP channels | 2 EMG channels | 64 EEG channels |
| Signal acquisition | Analog-to-digital<br>conversion at variable<br>sampling rate | same | same |
| Safety Standards | IEC 60601-1<br>IEC 60601-1-1<br>IEC 60601-1-2<br>IEC 60601-1-4<br>IEC 60601-2-26<br>IEC 60601-2-40 | IEC 60601-1<br>IEC 60601-1-1<br>IEC 60601-1-2<br>IEC 60601-1-4<br>IEC 60601-2-40 | IEC 60601-1<br>IEC 60601-1-1<br>IEC 60601-1-2<br>IEC 60601-1-4<br>IEC 60601-1-6<br>IEC 60601-2-26 |
| Patient circuitry<br>isolation | Optic<br>EMG channels inputs,<br>electrical stimulator<br>outputs: BF Type<br><br>Acoustic stimulator<br>headphones: BF Type<br><br>External serial ports and<br>I/O: B Type<br><br>EP\EEG channels inputs:<br>CF Type | Optic<br>EMG channels inputs,<br>electrical stimulator<br>outputs: BF Type<br><br>Acoustic stimulator<br>headphones: BF Type<br><br>External serial ports and<br>I/O: B Type | Optic<br>Patient isolation BF<br>type<br><br>Auxiliary I/O<br>components ports:<br>B type |
| Product<br>Characteristic | Nemus2 Amplifier<br>(submission device) | Nemus1 System<br>(Predicate Device) | BE Plus LTM<br>(Predicate Device) |
| System Components | Nemus 2 Base Unit<br>(Amplifier)<br>AC/DC Adapter<br>Dedicated Keyboard<br>(optional)<br>LED Flash Stimulator<br>(optional) | Nemus 1 Base Unit<br>(Amplifier)<br>AC/DC Adapter<br>Dedicated Keyboard<br>(optional) | Amplifier Unit<br>AC/DC Adapter<br>Removable 64/32 ch.<br>Patient Box<br>Rechargeable battery<br>pack<br>Battery pack charger<br>Mixed power/Ethernet<br>link cabling<br>Expansion cable<br>LED Flash Stimulator<br>(optional) |
| Amplifier-Computer<br>interface | LAN Ethernet IEEE 802.3 | same | "wired LAN "<br>ETHERNET/IEEE<br>802.3<br>or<br>"wireless LAN "<br>WLAN/IEEE 802.11 |
| Amplifier Power<br>Supply | External IEC 60601-1<br>mains adapter | same | External IEC 60601-1<br>mains adapter<br>or<br>Internal battery pack |
| Size (H/W/D) mm | Nemus 2 Amplifier<br>170 (L) x 125 (W) x 35<br>(H) (mm) | Nemus 1 Amplifier<br>170 (L) x 125 (W) x 35<br>(H) (mm) | 188 (L) x 148 (W) x 40<br>(H) (mm) |
| Weight | Nemus 2 Amplifier<br>0.6 Kg | Nemus 1 Amplifier<br>0.45 Kg | 0.5 Kg |
| Case material | Nemus 2 Amplifier<br>Policarbonate<br>Macrolon 6557 UL VO | Nemus 1 Amplifier<br>Policarbonate<br>Macrolon 6557 UL VO | ABS Resin -<br>(AF312B). |
| Product | Nemus2 System | Nemus1 System | BE Plus LTM |
| Characteristic | (submission device) | (Predicate Device) | (Predicate Device) |
| Design - Acquisition | | | |
| Number of channels | 2 EMG<br>20 EEG/EP | 2 EMG | 64 EEG |
| CMRR | >100 dB | >100 dB | >98 dB |
| IMRR | >100 dB | >100 dB | >98 dB |
| Input impedance | EMG Channels: same as<br>Nemus 1 | EMG Channels :<br>>100 MΩ | EEG Channels: |
| | EEG/EP channels:<br>same as BE Plus LTM | | Comm. Mode : 3.3 MΩ<br>Differential : 6.6 MΩ |
| Noise | EMG Channels : < 0.3<br>µVrms (0.1 – 100 Hz)<br>EEG/EP channels: <20<br>nV/√Hz (10-10KHz) | EMG Channels : < 0.3<br>µVrms (0.1 - 100 Hz) | EEG Channels:<br><0.3µVrms (@256 Hz<br>sampling rate ) |
| A/D conversion | 24 bit (2 EMG bipolar<br>channels )<br>16bit (20 EEG/EP<br>unipolar channels) | 24 bit (2 EMG bipolar<br>channel) | 16 bit (64 EEG unipolar<br>channels) |
| Sampling rate | 16384 Hz (EEG/EP ch)<br>32768 Hz (EMG ch)<br>Oversampling = Fs<br>x 128 | 32768 Hz (EMG ch)<br>Oversampling = Fs x<br>128 | 8192 Hz/Channel |
| Antialiasing Low pass<br>filter | EMG ch. : 20 KHz<br>EEG/EP ch, 2 KHz | EMG ch. : 20 KHz | 2KHz 3rd order filter |
| High-pass Filter | 1.6s (0.1 Hz) or<br>16ms(10Hz)(EMG<br>channels)<br>1.6/160 ms (0.1/1 Hz)<br>(EEG/EP channels | 1.6s (0.1 Hz) or<br>16ms(10Hz)(EMG<br>channels) | selectable 0.1/10 Hz<br>1st order filter |
| Trigger mode | Free, Auto, Internal,<br>External | same | na |
| Ohmmeter | 0-100 kOhm (auto full<br>scale) | same | same |
| Product<br>Characteristic | Nemus System<br>(submission device) | Nemus1 System<br>(Predicate Device) | BE Plus LTM<br>(Predicate Device) |
| Design - Stimulators | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.