K171332 · Penumbra, Inc. · GWG · Aug 14, 2017 · Neurology
Device Facts
Record ID
K171332
Device Name
Artemis Neuro Evacuation Device
Applicant
Penumbra, Inc.
Product Code
GWG · Neurology
Decision Date
Aug 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1480
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Artemis™ Neuro Evacuation Device is used for the controlled aspiration of tissue and/or fluid during surgery of the Ventricular System or Cerebrum in conjunction with a Penumbra Aspiration Pump.
Device Story
Artemis Neuro Evacuation Device is a surgical instrument for image-guided neurosurgery; aids removal of tissue/fluid from Ventricular System or Cerebrum. Device consists of a wand with a rigid hypotube containing an integrated wire. Operates via vacuum aspiration generated by an external Penumbra Aspiration Pump; rotational energy is provided by an internal 1.5V DC motor powered by a lithium-ion battery in the handle. Rotational energy facilitates movement of material to prevent lumen clogging. Used by trained physicians in clinical settings. Output is the physical removal of tissue/fluid; healthcare provider visualizes procedure via image guidance. Benefits include efficient evacuation of intracranial material during surgery.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included biocompatibility (ISO 10993), design verification (packaging, dimensional, dynamic vacuum, simulated use, tensile/torque), and electrical safety (IEC 60601-1 series). All tests passed.
Technological Characteristics
Materials: Polymer and metal (304 stainless steel hypotube). Energy: Vacuum aspiration (external pump) and rotational energy (internal 1.5V DC motor, Li-ion battery). Dimensions: 26.1-27.0 cm length. Connectivity: Standalone wand connected to external pump via tubing. Sterilization: Gamma irradiation (ISO 11137, ISO 11737). Electrical safety: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366.
Indications for Use
Indicated for controlled aspiration of tissue and/or fluid during surgery of the Ventricular System or Cerebrum in patients undergoing image-guided neurosurgery.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 14. 2017
Penumbra, Inc. Mary Rose Director, Regulatory Affairs One Penumbra Place Alameda, California 94502
Re: K171332
Trade/Device Name: Artemis Neuro Evacuation Device Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG Dated: July 14, 2017 Received: July 17, 2017
Dear Ms. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171332
Device Name ArtemisTM Neuro Evacuation Device
Indications for Use (Describe)
The Artemis™ Neuro Evacuation Device is used for the controlled aspiration of tissue and/or fluid during surgery of the Ventricular System or Cerebrum in conjunction with a Penumbra Aspiration Pump.
Penumbra Aspiration Pump:
The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Penumbra" in a bold, red font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern. The word "Penumbra" is the main focus of the image.
#### 510(k) SUMMARY 1
(as required by 21 CFR 807.92)
#### Sponsor/Applicant Name and Address 1.1
Penumbra Inc. One Penumbra Place Alameda, CA 94502, USA
#### Sponsor Contact Information 1.2
Mary Rose Director, Regulatory Affairs Phone: 510-748-3346 FAX: 510-217-6414 email: mary.rose@penumbrainc.com
### 1.3 Date of Preparation of 510(k) Summary
July 14, 2017
#### Device Trade or Proprietary Name 1.4
Artemis™ Neuro Evacuation Device
#### Device Common/Usual or Classification Name 1.5
Endoscope, neurological (Product Code: GWG)
#### Primary Device Classification 1.6
| Regulatory Class: | II |
|-----------------------|-------------------------|
| Classification Panel: | Neurology |
| Classification Name: | Endoscope, neurological |
| Regulation Number: | 21 CFR 882.1480 |
| Product Code: | GWG |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "Penumbra" in a red, sans-serif font. To the right of the word is a circular logo, also in red, with a white "P" shape inside. The logo is simple and modern, complementing the clean typography of the wordmark.
#### Predicate Devices 1.7
| 510(k)<br>Number | Clearance<br>Date | Name of Predicate Device | Name of<br>Manufacturer |
|-------------------|-------------------|------------------------------------------------------------------------|-------------------------|
| K132931 | 01/17/14 | Apollo System | Penumbra, Inc. |
| K152699 | 03/17/16 | Apollo System | Penumbra, Inc. |
| Reference Devices | | | |
| K160533 | 05/24/16 | Penumbra Embolectomy Aspiration System<br>(INDIGOTM Aspiration System) | Penumbra, Inc. |
| K160449 | 05/25/16 | Penumbra System® and Penumbra Pump MAX | Penumbra, Inc. |
#### 1.8 Predicate Comparison
| | Apollo™ System | Subject Device:<br>Artemis™ Neuro Evacuation<br>Device |
|-----------------------------------------|--------------------------------------------------|---------------------------------------------------------------------------------------|
| 510(k) No. | K132931, K152699 | K171332 |
| Classification | Class II, GWG | Class II, GWG |
| Intended Use | Used for the controlled | SAME |
| | aspiration of tissue and/or | |
| | fluid during surgery of the | |
| | Ventricular System or | |
| | Cerebrum. | |
| Aspiration Control | Fingertip controlled via | SAME |
| | vacuum regulator hole on | |
| | Wand handle. | |
| Mechanism of Action | Vacuum Aspiration aided by<br>Vibrational Energy | Vacuum Aspiration aided by<br>Rotational Energy |
| Power Source for<br>Mechanism of Action | Apollo Generator | Primary Cell Lithium Ion<br>Battery, 1.5 Volt DC Motor<br>contained within the handle |
| Aspiration Source | Apollo Pump | Penumbra Aspiration Pump |
| Wand Hypotube | 304 Stainless Steel | SAME |
| Wand Materials | Polymer, Metal | SAME |
| Wand Dimensions | L: 27.0 cm – 28.0 cm | L: 26.1 cm - 27.0 cm |
| | ID: 0.060″ — 0.088″ | ID: 00.048″ — 0.100″ |
| | OD: 0.072″ — 0.101″ | OD: 00.058″ — 0.109″ |
| Aspiration Tubing | Polymer | SAME |
| Material | | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo features the word "Penumbra" in a bold, red font. To the right of the name is a circular graphic, also in red, with a white stylized "P" inside.
| | Apollo™ System | Subject Device:<br>Artemis™ Neuro Evacuation<br>Device |
|---------------------------------|--------------------------|--------------------------------------------------------|
| Aspiration Tubing<br>Dimensions | OD: 0.188"<br>ID: 0.088" | OD: 0.263"<br>ID: 0.163" |
| Single Use or<br>Reusable | Single Use | SAME |
| Sterilization | EO | Gamma |
| Shelf-Life | 12-Months - Wand | SAME |
| Aspiration Pump | Apollo System Pump | Penumbra Aspiration Pump |
| IEC 60601-1<br>Compliance | Yes | SAME |
| IEC 60601-1-2<br>Compliance | Yes | SAME |
| Voltage | 100-115 VAC/230 VAC | SAME |
| Frequency | 50 Hz/60 Hz | SAME |
| Maximum Vacuum | 29 inHg | SAME |
| Maximum Flow rate | 0.8 SCFM | SAME |
#### 1.9 Device Description
The Artemis™ Neuro Evacuation Device, is a surgical instrument designed to aid a physician in the removal of tissue and/or fluid during image-guided neurosurgery. The Artemis Wand has two functions. These functions are control and transfer of aspiration and generation of rotational energy. Aspiration is generated by a Penumbra Aspiration Pump, which the Artemis Wand connects to through its flexible tubing. The Artemis Wand has a rigid hypotube containing a wire to facilitate removing tissue and/or fluid with the assistance of rotational energy and aspiration.
The Artemis Wand is designed to be image-guided, allowing visualization of the procedure. The method of removal is vacuum aspiration, which draws the tissue and/or fluid into the lumen of the Wand hypotube. The integrated wire is fully contained within the lumen of the Wand hypotube, and has rotational capability facilitating movement of any tissue and/or fluid that may otherwise clog the hypotube lumen.
Intended users for this device are physicians who have received appropriate training in image-guided neurosurgery.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Penumbra" in a red font. To the right of the word is a red circle with a white "P" in the center. The logo is simple and modern. The red color is bold and eye-catching.
### 1.10 Indications for Use
The Artemis™ Neuro Evacuation Device is used for the controlled aspiration of tissue and/or fluid during surgery of the Ventricular System or Cerebrum in conjunction with a Penumbra Aspiration Pump.
### Penumbra Aspiration Pump:
The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.
### 1.11 Summary of Non-clinical Data
Included in this section is a description of the testing, which substantiates the safe and effective performance of the subject Artemis™ Neuro Evacuation Device as well as its substantial equivalence to the predicate device:
- Biocompatibility .
- Design Verification (Bench-Top Testing) ●
- EMC ●
- Sterilization
- . Shelf-life
### 1.11.1 Biocompatibility
Biocompatibility testing was performed on the subject Artemis™ Neuro Evacuation Device. Tests were selected in accordance with EN ISO 10993-1 guidelines (Biological Evaluation of Medical Devices) and FDA Guidance. All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices.
| Test | Method | Conclusions |
|----------------------------------------------|------------------------------------------------|-----------------|
| In vitro Cytotoxicity | MEM Elution Test | Non-toxic |
| Sensitization (Magnusson-<br>Kligman Method) | ISO Guinea Pig Maximization Sensitization Test | Non-sensitizing |
The following tests were performed based contact type and duration:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Penumbra" in red font, followed by a red circle with a white "P" inside. The word "Penumbra" is written in a sans-serif font. The red circle is slightly to the right of the word "Penumbra". The logo is simple and modern.
| Test | Method | Conclusions |
|-------------------------------------------------|-------------------------------------------------|------------------|
| Irritation (Intracutaneous<br>Reactivity (ISO)) | ISO Tests for Irritation and Skin Sensitization | Non-irritant |
| Systemic Toxicity | | |
| Systemic Injection (ISO) | ISO Acute Systemic Injection Test | Non-toxic |
| Material Mediated Pyrogen | ISO Materials-Mediated Rabbit Pyrogen Test | Non-pyrogenic |
| Hemocompatibility | | |
| Thrombosis (Dog<br>Thrombogenicity) | Thrombogenicity Study in Dogs - ISO | Non-Thrombogenic |
| Coagulation (PT) | Prothrombin Time (PT) Test | Non-hemolytic |
| Coagulation (PTT) | Partial Thromboplastin Time (PTT) Test | Non-hemolytic |
| Hematology (Hemolysis) -<br>Direct Contact | ASTM Hemolysis (Direct Contact Method) | Non-hemolytic |
| Hematology (Hemolysis) -<br>Indirect Contact | ASTM Hemolysis (Extract Method) | Non-hemolytic |
| Genotoxicity | | |
| Ames Mutagenicity | Ames Test | Non-mutagenic |
| In vivo Mouse micronucleus | ISO In Vivo Mouse Micronucleus Assay | Non-mutagenic |
The non-clinical testing found the Penumbra Artemis™ Neuro Evacuation Device to be biocompatible according to EN ISO 10993 requirements. Therefore, Penumbra Artemis™ Neuro Evacuation Device is substantial equivalent to the predicate device.
# 1.11.2 Bench-Top Testing
Design Verification testing was conducted to evaluate the physical and mechanical properties of the Artemis™ Neuro Evacuation Device and demonstrate substantial equivalence to predicate. The following tests were successfully performed:
| Test | Test Method Summary | Results |
|------------------------------------|-----------------------------------------------------------------------------------------------------|--------------|
| Packaging Inspection | Confirm the packaging outputs meet all product<br>specifications. | 100%<br>Pass |
| Dimensional / Visual<br>Inspection | Confirm the dimensions / visual outputs meet all<br>product specifications. | 100%<br>Pass |
| Dynamic Vacuum<br>Testing | Wand can sustain maximum vacuum and transfer<br>vacuum to distal tip of wand. | 100%<br>Pass |
| Simulated Use | Evaluate the effectiveness of the device to generate<br>rotational energy to remove simulated clot. | 100%<br>Pass |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern, with the red color giving it a sense of energy.
| Tensile/Torque | All components are tested to ensure connections/joints meet all product specifications. | 100% |
|-------------------------|-------------------------------------------------------------------------------------------------|------|
| | | Pass |
| Electric Safety Testing | Compliant to requirements of IEC 60601-1 (3rd Ed.),<br>IEC 60601-1-2, 60601-1-6, and IEC 62366. | 100% |
| | | Pass |
Design Verification (Bench-Top Testing) found the Penumbra Artemis™ Neuro Evacuation Device to meet all design inputs. Therefore, Penumbra Artemis™ Neuro Evacuation Device is substantially equivalent to the predicate device.
### 1.11.3 Sterilization
The Artemis™ Neuro Evacuation Device was validated for gamma irradiation sterilization to a 6-log sterility assurance utilizing the VDmax25 sterilization validation in accordance with ISO 11137 and ISO 11737.
### 1.11.4 Shelf Life
The Artemis™ Neuro Evacuation Device has a 12-month Shelf-life
### 1.12 Summary of Substantial Equivalence
The subject Artemis™ Neuro Evacuation Device is equivalent to the predicate Apollo System™. The subject device has an identical intended use as the predicate device. The subject device and predicate device differ slightly in regards to technological and material variations. However, these differences do not raise different questions of safety and effectiveness. The device testing described in Section 1.11 demonstrate the subject device is equivalent to the predicate device in regards to operating principle fundamental technology, materials, and device performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.