CLARUS NEUROPEN ENDOSCOPIC PLACEMENT KIT MODEL 2125-XXX
K960915 · Clarus Medical Systems, Inc. · GWG · Dec 24, 1996 · Neurology
Device Facts
Record ID
K960915
Device Name
CLARUS NEUROPEN ENDOSCOPIC PLACEMENT KIT MODEL 2125-XXX
Applicant
Clarus Medical Systems, Inc.
Product Code
GWG · Neurology
Decision Date
Dec 24, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1480
Device Class
Class 2
Indications for Use
The Clarus Model 2121-XXX NeuroPEN Endoscopic Shunt Placement Kit is intended for endoscopic intraoperative intracranial neurosurgical shunt placement procedures where cerebrospinal fluid (CSF) may be contacted. The endoscope and the shunt are supplied sterile and are intended for single use.
Device Story
Device is a sterile, single-use endoscopic shunt placement kit; includes fiber optic endoscope and FDA-cleared shunt. Endoscope features fiber optic image bundle, distal lens, fiber light guide, and fluid irrigation channel with Luer-Lok connector. Used by neurosurgeons in OR for intracranial shunt placement. Physician inserts endoscope to visualize surgical site; irrigation channel allows fluid management. Provides direct visualization to assist in accurate shunt positioning. Benefits include minimally invasive access to intracranial space for CSF management.
Clinical Evidence
Bench testing only. Biocompatibility testing performed on all construction materials. Physical testing included dimensional inspection, visual workmanship, bond strength, optical clarity, light transmittance, distal tip temperature study, and fluid flow. All tests yielded acceptable results.
Technological Characteristics
Medical grade biocompatible materials; fiber optic image bundle; fiber light guide; fluid irrigation channel with Luer-Lok connector. Tubular form factor. Sterile, single-use. No electronic or software components.
Indications for Use
Indicated for patients undergoing endoscopic intraoperative intracranial neurosurgical shunt placement procedures requiring contact with cerebrospinal fluid (CSF).
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
K170462 — KARL STORZ ShuntScope · KARL STORZ Endoscopy-America, Inc. · May 16, 2017
K020507 — OLYMPUS NEUROSCOPE, MODELS S-1349/1, S-1349/2, S-1349/3 AND S-1349/4 · Olympus Winter & Ibe GmbH · Oct 8, 2002
K962761 — REIGEL TUNNELING DEVICE · Cook, Inc. · Oct 11, 1996
K970162 — NEUROLOGICAL ENDOSCOPE · Richard Wolf Medical Instruments Corp. · Oct 29, 1997
K990333 — MEDTRONIC PS MEDICAL INTRODUCER SHEATH 12F,14F,15F, MODELS 3805-012, 3805-014, 3805-015 · Medtronic PS Medical · Apr 28, 1999
Submission Summary (Full Text)
{0}
DEC 24 1996
K960915
# 510(k) Summary
## Model 2121-XXX NeuroPen Endoscopic Shunt Placement Kit
### General Information
**Classifications** Class II
**Trade Name** Clarus Model 2121-XXX NeuroPEN™ Endoscopic Shunt Placement Kit
**Submitter** Clarus Medical Systems, Inc.
1000 Boone Avenue North #100
Minneapolis, Minnesota 55427
**Contact** John Vanden Hoek
Director - Research and Development
(612) 525-8400
### Predicate Devices
Model 2120 NeuroPen Endoscope by Clarus Medical
Model 2125 MurphyScope by Clarus Medical
Catalog No. 82-6570 NEUROGUIDE Endoscopic Shunt Placement Kit by J & J
### Device Description
The Model 2121-XXX NeuroPEN Endoscopic Shunt Placement Kit contains an endoscope designed and manufactured by Clarus Medical and a FDA cleared shunt.
The Model 2121-XXX NeuroPEN endoscope is manufactured using medical grade biocompatible materials. The design of and materials used are the same as those used in other Clarus Series 2100 and Series 2200 Endoscopes. Further, the sizes and configurations of the endoscope are equivalent as well. The Model 2121-XXX endoscope is a tubular device with molded plastic connectors. The optical element of the endoscope consists of a fiber optic image bundle with a distal lens and a fiber light guide. The endoscope also has a separate fluid irrigation channel. The light and image guides are terminated with standard connectors designed to interface with light cables and video cameras. The fluid irrigation channel is terminated with a standard Luer-Lok™ connector.
{1}
# Intended Use
The design, materials used and intended use of the Clarus 2121-XXX Endoscope Shunt Placement Kit is substantially equivalent to the Johnson & Johnson #B2-6570 Endoscopic Shunt Placement Kit.
The Clarus Model 2121-XXX NeuroPEN Endoscopic Shunt Placement Kit is intended for endoscopic intraoperative intracranial neurosurgical shunt placement procedures where cerebrospinal fluid (CSF) may be contacted. The endoscope and the shunt are supplied sterile and are intended for single use.
# Testing
Biocompatibility testing was performed on the materials used in the construction of the endoscopes. All materials passed biocompatibility testing and are suitable for this application.
Physical testing of the endoscope included: dimensional inspection, visual examination for workmanship, bond strength testing, optical clarity, light transmittance, distal tip temperature study and fluid flow. All testing of the product yielded acceptable results.
# Summary of Substantial Equivalence
The Clarus Model 2121-XXX NeuroPEN Kit endoscopes are constructed of the same materials as other Series 2100 and Series 2200 endoscopes and Models 1100 and 1102 Laser Endoscopic Disc Decompression Kits as well as other Clarus products. The sizes and configurations available along with the packaging and sterilization methods are also equivalent.
The clinical indications for use of the Model 2121-XXX NeuroPEN Kit are equivalent to those of the NEUROGUIDE Endoscopic Shunt Placement Kit marketed by Johnson & Johnson. The sizes and configurations available along with the packaging and sterilization methods are also equivalent.
Therefore, due to the similarity of materials to other predicate devices, the test results and the equivalent indications for use to other predicate devices, Clarus believes these products do not raise any new safety or effectiveness issues.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.