The Nurochek System is intended for prescription use in healthcare facilities or clinical research environments for subjects ages 14 years and older. The Nurochek System is indicated for the generation of visual evoked potentials (VEPs) and to acquire, transmit, display and store electroencephalograms (EEGs) during the generation of VEPs. The Nurochek System only acquires and displays physiological signals: no claims are being made for use as a diagnostic criterion or for the analysis of the acquired signals with respect to the accuracy, precision and reliability.
Device Story
Nurochek System is a headset-based device for generating visual evoked potentials (VEPs) and acquiring EEG signals. Input: short-duration white light flashes from LEDs in the headset front directed at the subject's eyes. Output: EEG signals acquired from the rear of the head via 5 electrode interfaces (hydrophilic foam cylinders with saline). The headset transmits data via Bluetooth to a software application for display and storage. Used in healthcare facilities or clinical research environments by licensed healthcare professionals. The system provides a GUI for study initiation, electrode contact quality monitoring, and signal recording. Acquired signals are intended for physician analysis. Benefits include standardized VEP generation and EEG acquisition for clinical assessment.
Clinical Evidence
Clinical study with 20 participants compared the Nurochek System to the Compumedics Grael EEG (reference device) to evaluate the ability to detect steady-state visual-evoked potentials (SSVEPs). Both systems used a common visual stimulus. Results showed both systems functioned identically in their ability to detect SSVEPs. No adverse events reported.
Technological Characteristics
Headset with LED visual stimulator and 5 EEG electrode interfaces (hydrophilic polyurethane foam on gold-plated copper). Connectivity: Bluetooth wireless. Power: 3.7V, 800 mAH Li-Ion battery. Standards: IEC 60601-1, IEC 60601-2-40, IEC 60601-1-2, ISO 15004-2, ANSI Z80.36, ISO 10993. Software: Firmware and application software for signal acquisition and display. Cleaning: Validated for disinfection via wipes per AAMI TIR12/TIR30.
Indications for Use
Indicated for subjects ages 14 years and older in healthcare or clinical research settings for the generation of visual evoked potentials (VEPs) and acquisition, transmission, display, and storage of EEGs during VEP generation. No diagnostic claims.
Regulatory Classification
Identification
An evoked response photic stimulator is a device used to generate and display a shifting pattern or to apply a brief light stimulus to a patient's eye for use in evoked response measurements or for electroencephalogram (EEG) activation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
Cryptych Pty Ltd % Rafael Aguila Responsible Third-Party Official Accelerated Device Approval Services, LLC 6800 S.W. 40th Street, Ste. 444 Ludlum, Florida 33155-3708
Re: K200705
Trade/Device Name: Nurochek System Regulation Number: 21 CFR 882.1890 Regulation Name: Evoked response photic stimulator Regulatory Class: Class II Product Code: GWE, OMC Dated: April 21, 2020 Received: April 22, 2020
Dear Rafael Aguila:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
#### K200705
Device Name
Nurochek System Indications for Use (Describe)
The Nurochek System is intended for prescription use in healthcare facilities or clinical research environments for subjects ages 14 years and older. The Nurochek System is indicated for the generation of visual evoked potentials (VEPs) and to acquire, transmit, display and store electroencephalograms (EEGs) during the generation of VEPs. The Nurochek System only acquires and displays physiological signals: no claims are being made for use as a diagnostic criterion or for the analysis of the acquired signals with respect to the accuracy, precision and reliability.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a 3D rendering of a hollow square or rectangular frame. The frame is composed of four segments, with two segments appearing in a darker shade of gray or black and the other two in a lighter shade of gray. The segments are joined at the corners to form the square shape, and there is a small notch or cutout visible on the left side of the frame.
# 510(k) SUMMARY
#### for
## Nurochek System
| 510(k) Owner:<br>Address:<br>Phone and fax numbers: | Cryptych Pty Ltd<br>Level 15, Suite 1502<br>275 Alfred Street<br>North Sydney NSW 2060<br>+61 299 595 820 |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Name of contact person: | Angela Roche |
| Trade Name:<br>Common Name:<br>Classification name: | Nurochek System<br>Visual Evoked Photic Stimulator<br>Stimulator, Photic, Evoked Response |
| Regulation number: | 882.1890 |
| Product Code:<br>Device Class:<br>Manufacturer:<br>Submitter: | GWE, OMC<br>II<br>Cryptych Pty Ltd<br>Cryptych Pty Ltd |
| Preparation Date:<br>Predicate Devices: | March 05, 2020<br>SightSaver Visual Stimulator by Anschel Technology INC.(K113785)<br>X-Series System by Advanced Brain Monitoring (K131383) |
### DEVICE DESCRIPTION
The Nurochek System is indicated for the generation of visual evoked potentials (VEPs) and to acquire, transmit, display and store electroencephalograms (EEGs) during the generation of VEPs. The Nurochek System only acquires and displays physiological signals: no claims are being made for use as a diagnostic criterion or for the analysis of the acquired signals with respect to the accuracy, precision and reliability.
The Nurochek System combines hardware, firmware and software to generate and acquire physiological signals, specifically, VEPs. These VEPs are generated by a visual stimulus delivered through the Nurochek headset worn by the subject. This visual stimulus is a short-duration flash of white light. The Nurochek headset acquires the VEPs from the rear of the head and transmits the resulting EEG to the Nurochek software application to be displayed to the user and stored. These acquired signals are intended to be analyzed by a Physician.
The Nurochek System operates on the principles of generating VEPs via photic simulation and acquiring the VEPs via EEG. Photic stimulation is provided through short-duration flashes of white light from multiple LEDs located in the front of the headset to direct the stimulus into the subject's eyes. The VEPs are acquired by an EEG comprising of a total of 5 electrode interfaces with hydrophilic foam cylinders saturated with saline solution to provide electrical contact to the subject's scalp. A Bluetooth receiver and
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a 3D rendering of a hollow square or cube shape. The cube is constructed from four separate rectangular pieces that are joined at the corners. The color of the cube is a gradient, transitioning from a darker shade on the left to a lighter shade on the right, giving it a sense of depth and dimension. The cube appears to be floating or suspended in space, as there is no visible surface or background.
transmitter located within the Nurochek headset allows it to communicate with and be controlled by the Nurochek software application.
The Nurochek software application provides a graphical user interface which allows:
- Collection of the subject details and consent,
- Initiation of a study and tracking of patient information,
- Acquisition and transmission of signals wirelessly to and from the headset.
- Display of the contact quality of electrodes to the subject's scalp,
- Recording, processing and display of EEG signals received from the headset, and ●
- Manage previous EEG recordings of VEPs. ●
### INTENDED USE
The Nurochek System is intended for prescription use in healthcare facilities or clinical research environments for subjects ages 14 years and older. The Nurochek System is indicated for the generation of visual evoked potentials (VEPs) and to acquire, transmit, display and store electroencephalograms (EEGs) during the generation of VEPs. The Nurochek System only acquires and displays physiological signals: no claims are being made for use as a diagnostic criterion or for the analysis of the acquired signals with respect to the accuracy, precision and reliability.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICES
The Nurochek System uses the same fundamental technology as its predicates for the generation of visualevoked potentials (VEPs) via photic stimulation, the capture of electroencephalograph (EEG) signals, wireless acquisition and software. The technologies used in the Nurochek System are used in the same manner as the predicate devices and do not raise new questions related to safety and effectiveness.
The Nurochek System is substantially equivalent to the legally marketed evoked-response photic stimulator and electroencephalograph (EEG) devices:
(1) Evoked-response photic stimulator technology: Device name: SightSaver Visual Stimulator 510(k) number: K113785 510(k) submitter: Anschel Technology Inc. Classification regulation: 882.1890 Product code: GWE
- (2) Electroencephalograph technology: Device name: X-Series System 510(k) number: K131383 Model number: X10 / X24 510(k) submitter: Advanced Brain Monitoring Classification regulation: 882.1400 Product code: GWO, OMC
The Nurochek headset is worn by a subject and combines hardware, firmware and software to generate and acquire physiological signals, specifically, visual-evoked potentials (VEPs). The Nurochek operates on the principles of generating VEPs via photic activation and acquiring the VEPs via EEG.
Photic activation is achieved through short-duration flashes of white light from multiple LEDs located in the front of the headset (LED visual stimulator goggle) to direct the subject's eyes. This is technologically equivalent to the SightSaver Visual Stimulator (K1113785).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a 3D rendering of a hollow cube-like structure. The structure is composed of four rectangular segments that are joined at the corners to form an open, square frame. The segments are shaded in varying tones of gray, suggesting a light source from above. There is a small notch or cutout visible on the left side of the structure.
The VEP is acquired by an EEG located on the rear part of the headset that contacts the subject's scalp (the sensor). The resulting EEG signal is transmitted via Bluetooth to the Nurochek software application to be displayed and stored. These acquired signals are intended to be analyzed by a Physician. This is technologically equivalent to X-Series System (K131383).
The Nurochek System uses software that is technologically equivalent to the X-Series System to control the headset and display the acquired physiological signals. Like the X-Series System, the Nurochek System only acquires and displays the physiological signals and makes no claims in relation to diagnoses.
Table 1 provides a summary of the comparison of the key features between the Nurochek System and its predicates: SightSaver Visual Stimulator and the X-Series System.
{6}------------------------------------------------
| | | | | Table 1: Comparison table of Nurochek with predicates devices | |
|--|--|--|--|---------------------------------------------------------------|--|
| | | | | | |
| Device Name | Subject device: Nurochek System | Primary Predicate SightSaver Visual<br>Stimulator | Predicate X-Series System | Discussion |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Cryptych Pty Ltd | Anschel Technology Inc | Advanced Brain Monitoring | N/A |
| K Number | K | K113785 | K131383 | N/A |
| Indications for<br>Use | The Nurochek System is intended for<br>prescription use in healthcare facilities<br>or clinical research environments for<br>subjects ages 14 years and older. The<br>Nurochek System is indicated for the<br>generation of visual evoked potentials<br>(VEPs) and to acquire, transmit, display<br>and store electroencephalograms (EEGs)<br>during the generation of VEPs. The<br>Nurochek System only acquires and<br>displays physiological signals: no claims<br>are being made for use as a diagnostic<br>criterion or for the analysis of the<br>acquired signals with respect to the<br>accuracy, precision and reliability. | The SightSaver is an evoked response<br>photic stimulator that is used to apply a<br>visible light stimulus to a patient's eyes<br>for use in evoked response measurements<br>or for electroencephalogram (EEG)<br>activation.<br>The SightSaver Visual Stimulator is<br>designed to be used in hospital and<br>clinical settings by trained medical<br>personnel and is for prescription use only. | The X-Series System is intended for<br>prescription use in the home, healthcare<br>facility, or clinical research<br>environment to acquire, transmit,<br>display and store physiological signals<br>from patients ages 6 and older. The X-<br>Series system requires operation by a<br>trained technician. The X-Series<br>System acquires, transmits, displays<br>and stores electroencephalogram<br>(EEG), electrooculogram (EOG),<br>electrocardiogram (ECG), and/or<br>electromyogram (EMG), and<br>accelerometer signals. The X-Series<br>System only acquires and displays<br>physiological signals, no claims are<br>being made for analysis of the acquired<br>signals with respect to the accuracy,<br>precision and reliability. | Equivalent<br>The predicate devices generate a<br>visual stimulus and acquire,<br>transmit, displays and store<br>electroencephalograms (EEGs) |
| Patient population | Ages 14 and older | Children and adults | Ages 6 or older | Equivalent<br>Patient population of the Nurochek<br>System device is within the patient<br>population of the predicates. |
| Users | Licensed healthcare professionals | Trained medical personnel | Trained Technician | Equivalent<br>All devices require professional<br>operators and are prescription<br>only. |
| Anatomical sites<br>for EEG readings | Scalp | N/A (does not measure patient) | Scalp and Chest | Equivalent<br>Nurochek System's EEG sites are<br>the same (the scalp) as the X-<br>Series System. The chest site is not<br>relevant as the Nurochek System<br>does not utilize ECG technology. |
| Device Name | Subject device: Nurochek System | Primary Predicate SightSaver Visual<br>Stimulator | Predicate X-Series System | Discussion |
| Anatomical sites<br>for visual stimulus | Periocular region of the patient's face | Periocular region of the patient's face | N/A (is not a visual stimulator) | Equivalent<br>Both the Nurochek and SightSaver<br>contact the periocular region of<br>patient's face to deliver the visual<br>stimulus. |
| Prescription use<br>or OTC | Prescription use only | Prescription use only | Prescription use only | Identical |
| Environment of<br>Use | Healthcare facility<br>Clinical Research Environment | Hospital<br>Clinical settings | Home<br>Healthcare facility<br>Clinical Research Environment | Equivalent<br>Nurochek System and both<br>predicates are used in a healthcare<br>facility. |
| Cleaning | Cleaned and disinfected by rubbing with<br>disinfectant wipes | N/A (the device is disposable) | Cleaned and disinfected by rubbing<br>isopropyl alcohol | Equivalent<br>The Nurochek headset is cleaned<br>and disinfected to a higher<br>standard. A higher standard is used<br>due to the design of Nurochek's<br>headset. The Nurochek headset<br>contains an anterior visor while the<br>X-series System does not. As tests<br>were performed to demonstrated<br>that cleaning methods are<br>appropriate for ensuring the<br>Nurochek headset is clean between<br>uses they are considered<br>equivalent. |
| | | Technical specifications | | |
| User Interface | User control, visual indicators | N/A (does not contain any controls) | User control, visual indicators | Equivalent<br>Nurochek System has same user<br>interface features as X-Series<br>System. |
| Wireless data<br>transfer | Bluetooth | N/A (does not transfer data) | Bluetooth | Equivalent<br>Nurochek uses the same<br>technology as X-Series System. |
| Signals Acquired | Scalp EEG | N/A (does not acquire signals) | Scalp EEG<br>3-D actigraphy | Equivalent<br>Nurochek System only utilizes<br>scalp EEG technology and does<br>not utilize 3-D actigraphy nor the<br>optional channels of the X-Series |
| Device Name | Subject device: Nurochek System | Primary Predicate SightSaver Visual<br>Stimulator | Predicate X-Series System | Discussion |
| | | | Optional channels<br>ECG/EEG/EOG/EMG | System. Therefore, the other<br>signals are not relevant for the<br>Nurochek System. |
| Power Supply | 3.7V, 800 mAH Li-Ion battery | Externally powered by trigger device. | 2 to 4 240 mAH 3.7V, 240 mAH Li-Ion<br>batteries | No Significant Differences<br>Both Nurochek System and X-<br>series System use rechargeable<br>3.7V lithium-ion batteries. The<br>difference in charge capacity and<br>number of batteries, which only<br>dictates the length of use.<br>The SightSaver is to be connected<br>to a separate control device which<br>supplies the power. |
| Operating time | After full charge, unit is capable of at<br>least 30 complete tests (30 minutes) | N/A (controlled by external trigger) | Monitoring days after charge hours of<br>use:<br>0-4 days: 16 to 17 hours<br>5-10 days: 14 to 15 hours. | No significant differences<br>The Nurochek System is indicated<br>for short-term use i.e. 2 sets of 30<br>second tests completed<br>consecutively.<br>The X-Series System is indicated<br>for long-term monitoring over a<br>large period of time. |
| Battery Charging | Via USB cable from a wall charger | N/A (does not contain battery) | Via JED Connector connected to the<br>USB port or USB wall charger | Equivalent<br>The charging mechanisms between<br>the Nurochek System and X-Series<br>System are the same, differing<br>only in the connector style which<br>does not affect safety of Nurochek<br>System. |
| Typical Charging<br>Time | Typical: 0.5- 3 hours | N/A (does not contain battery) | Typical: 0.5-5 hours. | No significant difference<br>The charging time for a full charge<br>for the Nurochek device is less<br>than the X-Series System, allowing<br>it to be used more frequently. |
| Photic Stimulation Technology |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.