The GRAEL SYSTEM is intended for use in the recording, displaying, analysis, printing and storage of human biological parameters such as heart and muscle activity, eye movement, breathing and body movements to assist in the diagnosis of various sleep disorders or neurological disorders. The Grael is designed for use in a hospital or other clinical environment. The Grael is only to be used under the direction of a physician.
Device Story
Grael System (Grael/Grael EEG models) records, displays, analyzes, and stores human physiological signals including EEG, EOG, EMG, ECG, respiratory, and body movement. System comprises standalone hardware unit connected to PC running Profusion PSG or Profusion EEG software. Operates via Power over Ethernet (PoE). Used in hospitals/clinics by trained personnel under physician direction. Inputs include sensor-derived biological signals; software performs automated analysis to assist clinicians in diagnosing sleep and neurological disorders. Output provided as visual data for review and printed reports. Benefits include diagnostic aid for sleep/neurological conditions.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on technological similarities and shared components with previously cleared Compumedics devices.
Technological Characteristics
Standalone hardware unit; PoE powered. Supports EEG, EOG, EMG, ECG, respiratory, and body movement sensors. High-frequency channels (EEG/EMG/EOG) feature increased sampling rates compared to predicate. Connectivity via PC interface. Software-based analysis (Profusion PSG/EEG).
Indications for Use
Indicated for patients requiring diagnostic assessment of sleep or neurological disorders via recording of physiological parameters (EEG, EOG, EMG, ECG, respiratory, body movement). Intended for use in hospital or clinical settings under physician direction.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
# K093222
. Hela 1 & Co.,
## 5.510(k) Summary
In accordance with 21 CFR 807.92, the following information constitutes the Compumedics Limited summary for the GRAEL SYSTEM.
JAN 1 5 2010
| SUBMITTER'S NAME: | Compumedics Limited |
|---------------------|--------------------------------------------------------------------|
| ADDRESS: | 30-40 Flockhart Street,<br>Abbotsford, Victoria, 3067<br>Australia |
| CONTACT PERSON: | Gordon Ferguson |
| TELEPHONE NUMBER: | +61 3 8420 7300 |
| FAX NUMBER: | +61 3 8420 7399 |
| DATE OF SUBMISSION: | 6 October 2009 |
### Identification of device
| Proprietary Name: | Compumedics Grael System |
|------------------------|---------------------------------------------|
| Common Name: | GRAEL SYSTEM |
| Classification Status: | Class II per regulations 882.1400, 868.2375 |
| Product Codes: | GWQ, MNR |
#### Predicate devices
| Compumedics Limited believes the GRAEL SYSTEM is substantially equivalent to: | |
|-------------------------------------------------------------------------------|---------------------------|
| Device name: | Compumedics Siesta System |
| Supplier: | Compumedics Limited |
| 510(k) number: | K003175 |
Other predicate devices have been used for comparison where features or technology in the Grael are not present in the Compumedics Siesta System. Other predicate devices compared are:
- で Compumedics Summit IP, 510(k) number K040194
- . Compumedics Somté PSG, 510(k) number K072201
- Compumedics E-Series System, 510(k) No .: K000068
- 0 Compumedics P-Series System, 510(k) No .: K955841
- Neuroscan Labs Scan LT40, 510(k) No .: K001564 .
- . Compumedics Neuvo System, 510(k) No .: K081151
{1}------------------------------------------------
#### Description of the System
S
ಿ
Two different Grael models are available:
- . Grael, is a full channel device suitable for conducting either - PSG (polysomnographic - sleep), or EEG (electroencephalographic neurological) studies;
- Grael EEG, is a reduced channel device suitable for EEG studies only. o
NOTE - Unless otherwise noted the information in this submission applies to both devices and the name Grael is used to refer to either device type. Where necessary the name Grael EEG is used to explicitly indicate functionality that is either not present in or unique to the Grael EEG.
NOTE - "Grail" is the project name that was assigned to the Grael hardware development project.
Grael and Grael EEG are standalone units, which together with sensors and a PC running Compumedics PSG Online and Profusion PSG software for sleep studies and/or Profusion EEG for EEG studies, can record, review and analyze a number of physiological parameters, including EEG, EOG, EOG and respiratory signals. These are then used as an aid in the diagnosis of respiratory and/or neurological related sleep disorders or other neurological conditions by qualified physicians. The system is comprised of hardware and software which, depending on the model type, provides up to 60 separate parameters for recording, review and analysis.
Power is supplied to the Grael via the network cable using either a Power over Ethernet (PoE) switch or power injector.
#### Intended use
The GRAEL SYSTEM is intended for use in the recording, displaying, analysis, printing and storage of human biological parameters such as heart and muscle activity, eye movement, breathing and body movements to assist in the diagnosis of various sleep disorders or neurological disorders. The Grael is designed for use in a hospital or other clinical environment. The Grael is only to be used under the direction of a physician.
#### Shared Technologies
The Grael System incorporates technology shared with other approved devices:
- Pulse Oximetry Vampire module as used in Somté PSG (K072201) ●
- . Inductive plethysmography - Technology as used in Summit IP (K040194)
- High frequency channels (EEG, EMG, and EOG) are identical to the . technology used in the Siesta system (K003175), with the exception that these are sampled at a higher rate than the Siesta.
- Recording of the ECG channel is identical to the Somté PSG (K072201) .
- Pressure monitoring is identical to the Somté PSG (K072201) .
{2}------------------------------------------------
- Airflow monitoring via thermal oro-nasal sensor is identical to the Somté . PSG (K072201)
- Automatic analysis via the Profusion PSG software is identical to the . Somté PSG System (K072201)
- . Profusion EEG (4.0) software is identical to that in the Neuvo System (K081151)
- Photic Stimulator Identical to that included with the Scan LT40 (K001564) .
Compumedics Grael System 510(k) Submission
i
-T
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are written vertically along the left side of the symbol. The symbol is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
### JAN 1 5 2010
Stuart King Product Manager Compumedics Limited 30-40 Flockhart Street Abbotsford Vic 3067 Australia
Re: K093223
Trade/Device Name: Compumedics GRAEL SYSTEM (Models: Grael and Grael EEG), Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: GWQ, OLV, and OLZ Dated: December 10, 2009 Received: December 11, 2009
Dear Mr. King:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act. or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1.15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Kesia Alexander for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use Statement
510(k) Number K095223
Device Name: GRAEL SYSTEM
Indications for Use:
- The GRAEL SYSTEM is intended for use in the recording, displaying, analysis, printing and storage of human biological parameters such as heart and muscle activity, eye movement, breathing and body movements to assist in the diagnosis of various sleep disorders or neurological disorders. The Grael is designed for use in a hospital or other clinical environment. The Grael is only to be used under the direction of a physician.
Prescription Use OR (Per Part 21 CFR 801 Subpart D)
Over the Counter Use
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kristen Bowsher
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number_KQ93223
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.