K112052 · Carefusion 209, Inc. · GWF · Mar 15, 2012 · Neurology
Device Facts
Record ID
K112052
Device Name
CAREFUSION NICOLET EDX
Applicant
Carefusion 209, Inc.
Product Code
GWF · Neurology
Decision Date
Mar 15, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1870
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CareFusion Nicolet EDX is intended for the acquisition, display, storage, reporting, and management of electrophysiological information from the human nervous and muscular systems including Nerve Conduction (NCS). Electromyography (EMG). Evoked Potentials (EP). Autonomic Responses and Intra-Operative including Electroencephalography (EEG). Evoked Potential (EP) includes Visual Evoked Potentials (VEP). Auditory Evoked Potentials (AEP), Somatosensory Evoked Potentials (SEP), Electroretinography (ERG), Electrooculography (EOG), P300, Motor Evoked Potentials (MEP). The Nicolet EDX with Viking Software may be used to determine autonomic responses to physiologic stimuli by measuring the change in electrical resistance between two electrodes (Galvanic Skin Response and Sympathetic Skin Response). Autonomic testing also includes assessment of RR Interval variability. The Nicolet EDX with Viking Software is used to detect the physiologic function of the nervous system, for the location of neural structures during surgery, and to support the diagnosis of neuromuscular disease or condition. The listed modalities do include overlap in functionality. In general, Nerve Conduction Studies measure the electrical responses of the nerve; Electromyography measures the electrical activity of the muscle and Evoked Potentials measure electrical activity from the Central Nervous System. The Nicolet EDX with Viking Software is intended to be used by a qualified healthcare provider.
Device Story
The CareFusion Nicolet EDX is a diagnostic electromyograph system used to acquire, display, and analyze electrophysiological signals from the nervous and muscular systems. It consists of a console base, Viking control panel, amplifiers (2 or 8-channel), PC, and Viking software. The system uses electrical, auditory, and visual stimulators (LED goggles, photic strobe) to elicit responses, which are recorded via electrodes and processed by the amplifier. The system is operated by qualified healthcare providers in clinical or surgical settings (e.g., OR, ICU). Output is displayed on a monitor for clinical interpretation to support diagnosis of neuromuscular conditions or intra-operative neural structure localization. The device benefits patients by providing objective data on nerve and muscle function, aiding in surgical safety and diagnostic accuracy.
Clinical Evidence
No clinical or animal data were required to demonstrate safety and effectiveness. Substantial equivalence was established through bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing against the applicable standard for electromyographs and evoked response equipment (IEC 60601-2-40).
Technological Characteristics
System includes console base, Viking control panel, 2/8-channel amplifiers, PC, and Viking software. Features 24-bit A/D conversion, 48kHz sampling rate, >110 dB CMRR, and >1000 MΩ input impedance. Connectivity via USB. Power via mains (100-240 VAC) with isolation transformer. Sterilization not applicable (non-invasive). Software is moderate level of concern.
Indications for Use
Indicated for qualified healthcare providers to acquire, display, analyze, and manage electrophysiological data (NCS, EMG, EP, EEG, autonomic responses) to detect nervous system function, locate neural structures during surgery, and support diagnosis of neuromuscular diseases in adult and pediatric patients.
Regulatory Classification
Identification
An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.
{0}------------------------------------------------
# 510(k) Summary CareFusion Nicolet EDX with Viking Software
[A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92]
#### l. Submitter / 510(k) Holder
2.
3.
·
| Name | CareFusion 209, Inc. |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 1850 Deming Way, Middleton WI 53562 |
| Establishment Reg. No. | 3008289288 |
| Contact person | Mr. Curtis Truesdale |
| Phone number | 608-829-8684 |
| Fax number | 608-829-8769 |
| Email | Curtis.truesdale@carefusion.com |
| Date prepared | March 9, 2012 |
| Device Name | |
| Proprietary name | CareFusion Nicolet EDX with Viking Software |
| Common name | Diagnostic Electromyograph |
| Device Class | Class II |
| Classification name | Evoked Response Electrical Stimulator<br>Diagnostic Electromyograph<br>Nerve Conduction Velocity Measurement Device<br>Non-normalizing Quantitative Electroencephalograph Software<br>Evoked Response Auditory Stimulator<br>Evoked Response Photic Stimulator<br>Evoked Response Mechanical Stimulator |
| Product Code, Regulation | GWF 21 CFR §882.1870<br>IKN 21 CFR §890.1375<br>JXE 21 CFR §882.1550<br>OLT 21 CFR §882.1400<br>GWJ 21 CRF §882.1900<br>GWE 21 CFR §882.1890<br>GZP 21 CFR §882.1880 |
| Predicate Devices | |
| 510(K) Number | K890495<br>K965065<br>K984052<br>K991054 |
| Device Name | CareFusion Viking II System<br>CareFusion Synergy Mobile System<br>CareFusion Synergy IOM System<br>CareFusion Bravo Multi-Modality System |
| K102610 | TeleEMG Focus EMG |
| K924723 | Cadwell Sierra Wave System |
| K010590 | Nihon Kohden Micro System |
510(k) K112052, CareFusion Nicolet EDX with Viking Software
Page 1 of 19
{1}------------------------------------------------
#### Device Description র্ব.
The CareFusion Nicolet EDX system with Viking Software (Viking EDX) is designed for the acquisition, display, analysis, reporting, and management of electrophysiological information from the human nervous and muscular systems. The system is designed to perform nerve conduction studies (NCS), needle electromyography (EMG) testing, evoked potential (EP) testing, and intra-operative monitoring (IOM). The system can also perform multi-modality recording through Multi-mode programs (MMP). Viking EDX provides a variety of tests spanning the various modalities.
The Viking EDX consists of the following major components:
- Nicolet EDX console base unit; ●
- Viking control panel;
- Nicolet amplifier (there are two types available: 2 channel (AT2) with two non-switched . amplifier channels and an 8 channel (AT2 + 6) amplifier with two non-switched and six switched amplifier channels;
- Desktop or laptop computer with a keyboard and mouse; .
- Display monitor; and .
- Viking Software .
The Viking EDX optional accessories/components consists of the following:
- . Nicolet HB6 or HB7 Head Box
- . Stimulator probes (RS 10 probe, WR 50 Probe, S403 probe)
- . SP1/SP2 electrical stimulator switching units
- . Footswitches (single and triple)
- . LED goggles
- Patient response button
- . Photic strobe
- . Headphones or other auditory transducers
- . Cart
- . Isolation transformer
- Printer ◆
#### Description of Major Components
#### Nicolet EDX Console Base:
The console base is an AC mains powered electrical device that houses the core Nicolet EDX hardware, and provides interconnection capability with the rest of the Nicolet EDX hardware and PC. The PC contains the software.
#### Viking Control Panel:
The control panel along with the mouse is the primary user interface for the Nicolet EDX system. The control panel contains a variety of controls that allow the user to access and use the Nicolet EDX system from the touch of a button or knob.
{2}------------------------------------------------
#### Nicolet Amplifiers (AT2 and AT2 + 6):
The Nicolet amplifiers are DC powered electrical devices that record, amplify and transmit responses from the nerve and/or muscle to the Nicolet EDX console base. The AT2 amplifier collects up to two channels of neurophysiological information and the AT2 + 6 amplifier collects up to 8 channels of neurophysiological information.
## Personal Computer (PC) with Keyboard and Mouse:
The Nicolet EDX system hardware is used in conjunction with a personal computer (PC), which is offered in either a desktop or laptop configuration. The Viking software is supplied preloaded onto the PC.
#### Display Monitor:
The desktop PC version requires a display monitor which is provided with the Nicolet EDX system hardware. The laptop version does not require a separate display monitor beyond the built-in laptop display; however, an external display monitor may be used if desired. The additional display monitor is facilitated through the connection to an isolation transformer that is included with a desktop PC and available as an optional accessory for laptop based systems.
#### Description of Optional Accessory Components
#### HB6 and HB7 Head Boxes:
The HB6 and HB7 Head Boxes are passive devices. They can be connected to the AT2+6 Amplifier via a cable to allow the electrodes' receptacles to be in closer proximity to the patient.
#### Electrical Stimulator Probes:
There are three types of electrical stimulator probes available for use with the Nicolet EDX System Hardware: (1) the WR 50 comfort Plus Probe, (2) the RS 10 Comfort Probe and (3) the S403 Probe. The probes connect to the Nicole EDX console base via a cable. Each probe contains a tip that facilitates direct stimulus contact to the patient when activated. The probes can deliver a stimulus ranging from 0 - 400 V / 0 - 100 mA.
#### SP-1/SP-2 Electrical Stimulus Switching Units:
The SP-1 and SP-2 Stimulus Switching Units allow the user to connect multiple sets of electrodes on the patient for stimulation at different locations.
#### Footswitch and Triple Footswitch:
The footswitches allow the user to activate defined functions such as electrical stimulation and initiate acquisition of trace data. Pressing the footswitch activates or deactivates the user defined function.
#### Patient Response Button:
The patient response button allows the patient to respond to a rare stimulus during specific EP testing. The patient presses the button when the rare stimulus is detected. The Nicolet EDX senses the button press and increments a count.
#### Visual Stimulators:
- (1) LED goggles are available to provide visual stimulation to the patient when the Visual Evoked Potential (VEP) software option is in use.
- (2) A photic strobe is also a visual stimulus option. It can be mounted on an optional stand Similar to the other visual stimulus; it is used with the VEP option. All three visual
### 510(k) K112052, CareFusion Nicolet EDX with Viking Software
Page 3 of 19
{3}------------------------------------------------
stimulators have no controls or indicators and are connected to the Nicolet EDX Console Base Unit. This photic strobe was previously cleared under K921927 and K991054.
## Headphones or other Auditory Transducers:
Headphones are available, or other auditory transducers may be used to provide auditory stimulation to the patient through the transducer when the Auditory Evoked Potential (AEP) is in use. Headphones and transducers have no controls or indicators and are connected to the Nicolet EDX Console Base unit auditory stimulator output connectors.
#### Cart and Printer:
The metal cart provides a convenient way to contain all of the components of the Nicolet EDX System Hardware into one mobile location. The cart has lockable wheels and a convenient handle to facilitate movement of the cart. Two articulating arms are provided for mounting the display monitor and amplifier. There are multiple shelves present to accommodate the placement and storage of various Nicolet EDX System Hardware components and supplies. A retractable keyboard shelf and mouse pad is available. In terms of printing capabilities, a printer equivalent to a DeskJet or Laser printer is available for connection to the system to print reports or screen copies.
#### 5. Indications for Use:
The CareFusion Nicolet EDX is intended for the acquisition, display, analysis, storage, reporting, and management of electrophysiological information from the human nervous and muscular systems including Nerve Conduction (NCS), Electromyography (EMG), Evoked Potentials (EP), Autonomic Responses and Intra-Operative Monitoring including Electroencephalography (EEG).
Evoked Potential (EP) includes Visual Evoked Potentials (VEP), Auditory Evoked Potentials (AEP), Somatosensory Evoked Potentials (SEP), Electroretinography (ERG), Electrooculography (EOG), P300, Motor Evoked Potentials (MEP). · The Nicolet EDX with Viking Software may be used to determine autonomic responses to physiologic stimuli by measuring the change in electrical resistance between two electrodes (Galvanic Skin Response and Sympathetic Skin Response). Autonomic testing also includes assessment of RR Interval variability. The Nicolet EDX with Viking Software is used to detect the physiologic function of the nervous system, for the location of neural structures during surgery, and to support the diagnosis of neuromuscular disease or condition.
The listed modalities do include overlap in functionality. In general, Nerve Conduction Studies measure the electrical responses of the nerve; Electromyography measures the electrical activity of the muscle and Evoked Potentials measure electrical activity from the Central Nervous System.
The Nicolet EDX with Viking Software is intended to be used by a qualified healthcare provider.
#### Summary of Technical Characteristics Compared to the Predicate Devices 6 .
The CareFusion Nicolet EDX System with Viking Software is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristic to its predicate devices through comparison in areas that include intended use,
{4}------------------------------------------------
design, functionality, principle of operation, specifications, material composition and performance.
:
.
・
:
:
.
.
:
. .
.
## 510(k) K112052, CareFusion Nicolet EDX with Viking Software
.
.
:
Page 5 of 19
、・
.
{5}------------------------------------------------
. . .
General
uropack, Neuropac
lihon Koden Micr d prognosis of
uromuscular diseas
MG). The device is a
tended to measure ne MEB-9100A Ser ndications for Use
nclude the notor conduction dwell Cascad erve, spinal co unctionality an neasurements
equiring multip ifferent types
neasurement ascade Pr ascade Elit k924723 imulators of ntegrity; 2,
dectro-neura ollowing: 1,
evaluation of nd cortica ensory and oncurren nachine can also
ise electrical The Focus is
ntended for use leEMG, LLC onductions ar y a healthca EMG studies : ound stimul ocus EMI <102610 erform ner atients wit valuation an aid in th rovider to iseases of muscle an erves. The timulus o display data fro reFusion 209, l
vo Multi-Moda d-party monito
ices. It is inten critical care uro intensive c ), and to impr id the diagno ensive care ur ergency room K991054 ihe Bravo Mult System ich provides
ilities for EMG, EE signed to provide
ta recording, disp dord Instrumen (now owned by
eFusion), System
IOM System
IOM System
oked Potentials,
G and NCV testing
be used for thera
The used for thera
Terative Monitor d reporting of
ctrophysiological
ormation from t The Synergy 10M
tystem is a 2, 5, or marı nervous a
scular system edical Syster ctromyograp hergy 10M is nannel dford instrumen The Millennium is :
r 5 channel ctrophysiologics
ormation from t otentials testing
range of clinical
pplications. ow owned b וניצחק מונים
מנויות מקום שניירות שנויות שנייר
esigned to enab
eliable recording cumentation o edical Syster lectromyograp
thich provides
c1lities for EM illennium is nd Evoked splay and ctrophysiologica
ormation from human nervo functional sta reFusion 209 Nicolet Vikin uisition, displ
lysis, reporti
managemen tem in order t ect changes i colet Viking
(K890495) em, or for t tion of neur ctures durir he nervou l muscular for data nd Intra-Operative
Ionitoring including
ectroencephalograp reFusion 209, Inc
colet EDX System otentials (EP),
utonomic Responsi This Submissio man nervous ar uscular system ectromyograp
MG), Evoked cluding Nerve
onduction (NC Characteri
1.1
Indication for Use
Pa
0(k) K l 12052, CareFusion Nicolet EDX with Viking Softw
age 6 of
f he device may use vise
timurical stimulus stimulus stimulus stimulus stimuricas
Turnuslus, stimuslus studios submissiones and series
The asuries may be useos in t
enerated by the
lectrical, auditor
i single device,
whereas each
redicate device has
allow manual
betitive sequence
be automated, ar
nat it contains
hultiple modalitic
itures are provided
thin the 10M system
edicate devices
fers from th
vous system, or
location of neu
ctures. Additio
itencies
tudies
ervous system;
neasurement neasurement easurement voked poked politudes anditudes anditudes anditudes anditudes and
ncluding those o
he peripheral
rerves, spinal co
tudies
atencies
for evoked
potentials (El
d monitoring of
tential disorder
nd centra
tem and muscl
e central and
ripheral nervo
anges in the
actional state of t
ler to detec
man nervous an
iuscular system
inical environm
EEG)
שנייה
e intended us
voked Potential (EP)
cludes Visual Evoke
cludes Visual Evoke
otentials (VER)
uditory Evolked
ted here.
matosensory Evok
tentials (SEP),
tentials (AEP
lectroretinograph
{6}------------------------------------------------
| Characteristic | CareFusion 209, Inc.<br>Nicolet EDX System<br>(This Submission) | CareFusion 209, Inc.<br>Nicolet Viking II<br>(K890495) | Oxford Instruments<br>Medical Systems<br>(now owned by<br>CareFusion),<br>Millennium (now<br>called Synergy<br>Mobile) System<br>(K965065) | Oxford Instruments<br>Medical Systems<br>(now owned by<br>CareFusion), Synergy<br>IOM System<br>(K984052) | CareFusion 209, Inc.<br>Bravo Multi-Modality<br>System<br>(K991054) | TeleEMG, LLC<br>Focus EMG<br>(K102610) | Cadwell Cascade,<br>Cascade Elite,<br>Cascade Pro<br>(K924723) | Nihon Koden Micro,<br>Neuropack, Neuropack S1<br>(K010590) | Characteristic | | 1.2 Warnings | 1.3 Contra-indications |
|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (ERG),<br>Electrooculography<br>(EOG), P300, Motor<br>Evoked Potentials<br>(MEP). The Nicolet EDX<br>with Viking Software<br>may be used to<br>determine autonomic<br>responses to<br>physiologic stimuli by<br>measuring the change<br>in electrical resistance<br>between two electrodes<br>(Galvanic Skin Response<br>and Sympathetic Skin<br>Response). Autonomic<br>testing also includes<br>assessment of RR<br>Interval variability.<br>The listed modalities do<br>include overlap in<br>functionality. In<br>general, Nerve<br>Conduction Studies<br>measure the electrical<br>responses of the nerve;<br>Electromyography<br>measures the electrical<br>activity of the muscle<br>and Evoked Potentials<br>measure electrical<br>activity from the Central | | | to provide automatic<br>warnings when<br>neurological<br>parameters exceed<br>user defined limits. | very limited number of<br>modalities.<br>Nevertheless, like the<br>predicate devices, the<br>Bravo Multi-Modality<br>System is intended to<br>involve competent<br>human intervention<br>before any impact on<br>human health occurs<br>(i.e., clinical judgment<br>and experience must<br>be used to check and<br>interpret the system's<br>output). | | and visual<br>stimulators<br>provided with the<br>Cascade; 5,<br>localization of<br>nerves, nerve<br>roots, and other<br>neutrally active<br>tissue by<br>observing the<br>response of<br>muscle, nerve,<br>spinal cord or<br>brain to electrical<br>or mechanical<br>stimulation of the<br>tissue; 6, use of<br>anesthetic<br>patients; 7,<br>Assessment of<br>myoneural<br>junction blockade. | determine autonomic<br>responses as psychological<br>indicators by measuring the<br>electrical resistance of the<br>skin and the tissue path<br>between two electrodes<br>applied to the skin. The<br>device may also measure<br>and record the electrical<br>activity of the patient's<br>brain obtained by placing<br>two or more electrodes on<br>the head (EEG).<br>The device is intended for<br>use by medical personnel<br>within a medical facility or<br>outside of a medical facility<br>under direct supervision of<br>a medical professional. The<br>device may also be placed in<br>the intensive care unit or<br>operating room for<br>continuous recording.<br>The device is available for<br>use on any patient as<br>determined by the medical<br>professional including<br>adults and children. | CareFusion 209, Inc.<br>Nicolet EDX System<br>(This Submission) | Nervous System.<br>The Nicolet EDX with<br>Viking Software is<br>intended to be used by<br>a qualified healthcare<br>provider. | Items related to off-<br>label use or misuse. | Items related to design<br>and indicated use<br>limitations, such as, not<br>for use in the presence<br>of flammable<br>anesthetics or in<br>conjunction with<br>defibrillation<br>equipment |
| CareFusion 209, Inc.<br>Nicolet Viking II<br>(K890495) | Oxford Instruments<br>Medical Systems<br>(now owned by<br>CareFusion),<br>Millennium (now<br>called Synergy<br>Mobile) System<br>(K965065) | Items related to off-<br>label use or misuse. | Items related to design<br>and indicated use<br>limitations, such as, not<br>for use in the presence<br>of flammable<br>anesthetics or in<br>conjunction with<br>defibrillation<br>equipment | | | | | | | | | |
| CareFusion 209, Inc.<br>Oxford Instruments<br>Medical Systems<br>(now owned by<br>CareFusion), Synergy<br>IOM System<br>(K984052) | CareFusion 209, Inc.<br>Bravo Multi-Modality<br>System<br>(K991054) | Items related to off-<br>label use or misuse. | Items related to design<br>and indicated use<br>limitations, such as, not<br>for use in the presence<br>of flammable<br>anesthetics or in<br>conjunction with<br>defibrillation<br>equipment | | | | | | | | | |
| TeleEMG, LLC<br>Focus EMG<br>(K102610) | Cadwell Cascade,<br>Cascade Elite,<br>Cascade Pro<br>(K924723) | Items related to<br>off-label use. | Items related to<br>design and<br>indicated use<br>limitations, such<br>as, not for use in<br>the presence of<br>flammable<br>anesthetics or in<br>conjunction with<br>defibrillation<br>equipment | | | | | | | | | |
| Nihon Koden Micro,<br>Neuropack, Neuropack S1<br>(K010590) | | Items related to off-label<br>use. | Unknown | | | | | | | | | |
·
Page 7 of 19
・
.
{7}------------------------------------------------
.
Page 8 of 19
.
510(k) K112052, CareFusion Nicolet EDX with Viking Software
: :
{8}------------------------------------------------
| General<br>Characteristic | CareFusion 209, Inc.<br>Nicolet EDX System<br>(this submission) | CareFusion 209, Inc.<br>Nicolet Viking II<br>(K890495) | Oxford<br>Instruments<br>Medical Systems<br>(now owned by<br>CareFusion),<br>Millennium (now<br>called Synergy<br>Mobile) System<br>(K965065) | Oxford<br>Instruments<br>Medical Systems<br>(now owned by<br>CareFusion),<br>Millennium (now<br>called Synergy<br>Mobile) System<br>(K984052) | CareFusion 209, Inc.<br>Bravo Multi-<br>Modality System<br>(K991054) | TeleEMG, LLC<br>Focus EMG<br>(K102610) | Cadwell Cascade,<br>Cascade Elite, Cascade<br>Pro<br>(K924723) | Nihon Koden Micro,<br>Neuropack,<br>Neuropack S1<br>(K010590) |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2.1 General systems<br>approach | Computer based<br>equipment with<br>dedicated hardware<br>peripherals /<br>components. | Computer based<br>equipment with<br>dedicated hardware<br>peripherals /<br>components. | Computer based<br>equipment with<br>dedicated hardware<br>peripherals /<br>components. | Computer based<br>equipment with<br>dedicated hardware<br>peripherals /<br>components. | Computer based<br>equipment with<br>dedicated hardware<br>peripherals /<br>components. | Computer based<br>equipment with<br>dedicated hardware<br>peripherals /<br>components. | Computer based<br>equipment with<br>dedicated hardware<br>peripherals /<br>components. | Computer based<br>equipment with<br>dedicated hardware<br>peripherals /<br>components. |
| 2.2 User input device | Window<br>mouse/keyboard<br>driven graphic<br>interface with<br>dedicated control<br>panel. | Window<br>mouse/keyboard<br>driven graphic<br>interface with<br>dedicated control<br>panel. | Window<br>mouse/keyboard<br>driven graphic<br>interface with<br>dedicated control<br>panel. | Window<br>mouse/keyboard<br>driven graphic<br>interface with<br>dedicated control<br>panel. | Window<br>mouse/keyboard<br>driven graphic<br>interface with<br>dedicated control<br>panel. | Window<br>mouse/keyboard<br>driven graphic<br>interface with<br>built-in keyboard. | Window<br>mouse/keyboard<br>driven graphic<br>interface with<br>dedicated control<br>panel. | Window<br>mouse/keyboard<br>driven graphic<br>interface with<br>dedicated control<br>panel. |
| 2.3 User output<br>device | Digital color display<br>and commercial<br>printers | Digital color display<br>and commercial<br>printers | Digital color<br>display and<br>commercial<br>printers | Digital color<br>display and<br>commercial<br>printers | Digital color display<br>and commercial<br>printers | Digital color<br>display and<br>commercial<br>printers | Digital color display<br>and commercial<br>printers | Digital color display<br>and commercial<br>printers |
| 2.4 Patient inputs | 2 to 8 channel<br>amplifier, isolated | 2 to 8 channel<br>amplifier, isolated | 2 to 10 channel<br>amplifier, isolated | 2 to 10 channel<br>amplifier, isolated | 2 to 10 channel<br>amplifier, isolated | 2 channel<br>amplifier, isolated | 2 to 10 channel<br>amplifier, isolated | 2 to 4 channel<br>amplifier, isolated;<br>newer version up to<br>16 channels |
| 2.5 Signal acquisition | Analog to digital<br>conversion at 48kHz<br>sample rate | Analog to digital<br>conversion at 48kHz<br>sample rate | Analog to digital<br>conversion at<br>variable sample<br>rate | Analog to digital<br>conversion at<br>variable sample<br>rate | Analog to digital<br>conversion at<br>variable sample rate | Analog to digital<br>conversion at<br>variable sample<br>rate | Analog to digital<br>conversion at variable<br>sample rate | Analog to digital<br>conversion, unknown<br>rate |
| 2.6 Trigger input<br>(synchronization to<br>external events) | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| 2.7 Trigger output<br>(synchronization for<br>external devices) | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| Characteristic | CareFusion 209, Inc.<br>Nicolet EDX System<br>(this submission) | CareFusion 209, Inc.<br>Nicolet Viking II<br>(K890495) | Oxford<br>Instruments<br>Medical Systems<br>(now owned by<br>CareFusion),<br>Millennium (now<br>called Synergy<br>Mobile) System<br>(K965065) | Oxford<br>Instruments<br>Medical Systems<br>(now owned by<br>CareFusion),<br>Millennium (now<br>called Synergy<br>Mobile) System<br>(K984052) | CareFusion 209, Inc.<br>Bravo Multi-<br>Modality System<br>(K991054) | TeleEMG, LLC<br>Focus EMG<br>(K102610) | Cadwell Cascade.<br>Cascade Elite, Cascade<br>Pro<br>(K924723) | Nihon Koden Micro,<br>Neuropack,<br>Neuropack S1<br>(K010590) |
| 2.8 Footswitch for<br>hands-free operation | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
| 2.9 Use of standard<br>software platform<br>(Operating System) | Yes Microsoft<br>Windows | Yes Intel IRMX | Yes Microsoft<br>Windows | Yes Microsoft<br>Windows | Yes Microsoft<br>Windows | Yes Microsoft<br>Windows | Yes Microsoft<br>Windows | Yes Microsoft<br>Windows |
| 2.10 Customization<br>of protocols | Via storage /<br>retrieval of user<br>defined settings | Via storage /<br>retrieval of user<br>defined settings | Via storage /<br>retrieval of user<br>defined settings | Via storage /<br>retrieval of user<br>defined settings | Via storage /<br>retrieval of user<br>defined settings | Via storage /<br>retrieval of user<br>defined settings | Via storage /<br>retrieval of user<br>defined settings | Via storage / retrieval<br>of user defined<br>settings |
| 2.11 Application<br>flexibility /<br>expandability | Via software update | Via software update | Via software update | Via software<br>update | Via software upd | Via software<br>update | Via software<br>update | Via software update |
| 2.12 Safety Standards | EN/IEC 60601-1:1998<br>+ A1:1991+A2:1995, | UL 544, | IEC 60601-1, | IEC 60601-1, | IEC 60601-1, | IEC 60601-1-<br>1:2000, | UL 2601-1 | IEC 60601-1 (1988-12),<br>Amendment 1 (1991-<br>11), |
| | IEC 60601-1-1:2000 | IEC 60601-1, | IEC 60601-1-1, | IEC 60601-1-1, | IEC 60601-1-1, | IEC 60601-1-<br>2:2001, | CSA601-1 | Amendment 2 (1995-<br>03); IEC 60601-1-1<br>(1992-06), |
| | EN/IEC 60601-1-2:<br>Ed. 2.0+A1:2004 | IEC 60606-1-2, | IEC 60601-1-2, | IEC 60601-1-2, | IEC 60601-1-2, | IEC 60601-1-2-<br>40:1998, | | Amendment 1 (1995-<br>10); EN 60601-1-2<br>(1993-05), |
| | IEC 60601-2-40:<br>1998, Ed:1, | CSA-C22.2 No 125-<br>M1984 | IEC 60601-2-40, | IEC 60601-2-40, | IEC 60601-2-26, | | | IEC 60601-2-40 (1998) |
| | UL 60601-1: 2003-<br>04-25 ED1<br>Rev:2003/06/30, | | European<br>Community (CE<br>Mark) | European<br>Community (CE<br>Mark) | IEC 60601-2-40, | IEC 62471:2006 | | |
| | CAN/CSA-C22.2 no.<br>601.1-M90<br>Issue:1990/01/01<br>Rev:2003/11 | | | | European<br>Community (CE<br>Mark) | | | |
| Characteristic | CareFusion 209, Inc.<br>Nicolet EDX System<br>(this submission) | CareFusion 209, Inc.<br>Nicolet Viking II<br>(K890495) | Oxford Instruments<br>Medical Systems<br>(now owned by CareFusion),<br>Millennium (now called Synergy Mobile) System<br>(K965065) | Oxford Instruments<br>Medical Systems<br>(now owned by CareFusion),<br>Millennium (now called Synergy Mobile) System<br>(K984052) | CareFusion 209, Inc.<br>Bravo Multi-<br>Modality System<br>(K991054) | TeleEMG, LLC<br>Focus EMG<br>(K102610) | Cadwell Cascade.<br>Cascade Elite, Cascade<br>Pro<br>(K924723) | Nihon Koden Micro,<br>Neuropack,<br>Neuropack S1<br>(K010590) |
| 2.13 Patient circuitry Isolation | European<br>Community (CE<br>Mark) | Optic/transformer | Optic/transformer | Optic/transformer | Optic/transformer | Optic/transformer | Optic/transformer | Optic/transformer |
| 2.14 System components | EDX base console<br>including 2 electrical<br>stimulators, auditory<br>stimulator, trigger<br>input/output, LED<br>goggle interface;<br><br>Control panel;<br>Amplifier;<br>Computer, monitor,<br>keyboard, mouse,<br>printer | Base unit including 2<br>electrical stimulators,<br>auditory stimulator,<br>trigger input/output,<br>LED goggle interface;<br><br>Control panel;<br>Amplifier;<br>Computer,…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.