Retrospective archived clinical specimens were used to supplement the prospective clinical study to ensure sufficient sample sizes for influenza B and RSV, which had low prevalence during the prospective study period.
Clinical performance study (retrospective component); Retrospective analysis of archived clinical specimens; Follow-up/Duration: Not applicable; Study Period: September 2014 to March 2015
Patients with signs and symptoms of respiratory infection; Sample Size: 259; Number of Sites: 12
Comparator NAAT
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) for Influenza A, Influenza B, and RSV
Indications for Use
The Cepheid Xpert® Flu+RSV Xpress Assay, performed on the GeneXpert® Xpress System, is an automated, multiplex real-time, reverse transcriptase polymerase chain reaction (RT-PCR) assay intended for the in vitro qualitative detection and differentiation of influenza A, influenza B, and respiratory syncytial virus (RSV) viral RNA. The Xpert Flu+RSV Xpress Assay uses nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory infection. The Xpert Flu+RSV Xpress Assay is intended as an aid in the diagnosis of influenza and respiratory syncytial virus in conjunction with clinical and epidemiological risk factors. Negative results do not preclude influenza virus or respiratory syncytial virus infection and should not be used as the sole basis for treatment or other patient management decisions. Performance characteristics for influenza A were established during the 2014-2015 influenza season. When other novel influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. The Xpert® Nasopharyngeal Sample Collection Kit is designed to collect, preserve, and transport nasopharyngeal swab specimens containing viruses from patients with signs and symptoms of respiratory infection prior to analysis with the Xpert Flu+RSV Xpress Assay.
Device Story
The Xpert Flu+RSV Xpress Assay is an automated, multiplex real-time RT-PCR test for qualitative detection of influenza A, influenza B, and RSV. It uses nasopharyngeal swab specimens collected in the Xpert Nasopharyngeal Sample Collection Kit. The assay is performed on the GeneXpert Xpress System (GX-I), which integrates sample extraction, purification, amplification, and detection within a single-use, multi-chambered fluidic cartridge. The system uses an ultrasonic horn for cell lysis and an I-CORE thermocycler for real-time PCR. The process is hands-off and takes approximately 60 minutes. Results are automatically generated for clinician review. The device is intended for use in clinical settings, including by operators without clinical laboratory experience in CLIA-waived environments. It aids in diagnosis by providing rapid viral detection, allowing for timely clinical decision-making and patient management.
Clinical Evidence
Clinical study conducted at 12 sites (10 CLIA-waived, 2 high-complexity) with 2,435 NP swab specimens (2,176 prospective, 259 archived). PPA for Flu A was 100% (fresh), Flu B 100% (fresh), and RSV 96.9% (fresh). NPA was 94.8% (Flu A), 99.5% (Flu B), and 99.6% (RSV). Study included analytical precision, LoD, inclusivity, cross-reactivity, and competitive interference testing. No evidence of carry-over contamination observed.
Technological Characteristics
Multiplex real-time RT-PCR assay. Uses single-use disposable cartridges with integrated sample preparation, amplification, and detection. Employs I-CORE thermocycler, ultrasonic lysis, and syringe-driven fluidics. Connectivity includes LIS integration. Software version 4.4.2. Qualitative detection of viral RNA targets (Matrix, Polymerase, Nucleocapsid genes).
Indications for Use
Indicated for patients with signs and symptoms of respiratory infection to aid in the diagnosis of influenza A, influenza B, and respiratory syncytial virus (RSV) infections. Uses nasopharyngeal swab specimens.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K151226
B. Purpose for Submission:
To obtain a Substantial Equivalence Determination for a new 510(k) application for Xpert® Flu+RSV Xpress Assay performed on the Cepheid GeneXpert Xpress System.
C. Measurand:
This assay uses nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection. The viral nucleic acid is extracted from the sample and influenza and/or Respiratory Syncytial Virus (RSV) RNA is amplified and detected through real-time reverse transcription polymerase chain reaction (RT-PCR).
D. Type of Test:
This assay is a multiplex nucleic acid assay that detects and differentiates influenza A, influenza B, and RSV through nucleic acid extraction, amplification, and detection using real-time RT-PCR. All steps of the assay are automated, after sample addition, and performed in a single container.
E. Applicant:
Cepheid
F. Proprietary and Established Names:
Xpert® Flu+RSV Xpress Assay
G. Regulatory Information:
1. Regulation section:
21CFR 866.3980 - Respiratory viral panel multiplex nucleic acid assay
21CFR 866.2390 - Transport culture medium
2. Classification:
Class II
{1}
3. Product codes:
OCC, JSM, OOI
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use(s):
Assay:
The Cepheid Xpert® Flu+RSV Xpress Assay, performed on the GeneXpert® Xpress System, is an automated, multiplex real-time, reverse transcriptase polymerase chain reaction (RT-PCR) assay intended for the in vitro qualitative detection and differentiation of influenza A, influenza B, and respiratory syncytial virus (RSV) viral RNA. The Xpert Flu+RSV Xpress Assay uses nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory infection. The Xpert Flu+RSV Xpress Assay is intended as an aid in the diagnosis of influenza and respiratory syncytial virus in conjunction with clinical and epidemiological risk factors.
Negative results do not preclude influenza virus or respiratory syncytial virus infection and should not be used as the sole basis for treatment or other patient management decisions.
Performance characteristics for influenza A were established during the 2014-2015 influenza season. When other novel influenza A viruses are emerging, performance characteristics may vary.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Sample Collection Kit
The Xpert® Nasopharyngeal Sample Collection Kit is designed to collect, preserve, and transport nasopharyngeal swab specimens and to preserve and transport nasal aspirate/wash specimens containing viruses from patients with signs and symptoms of respiratory infection prior to analysis with the Xpert Flu Assay or the Xpert Flu/RSV XC Assay. The Xpert® Nasopharyngeal Sample Collection Kit is designed to collect, preserve, and transport nasopharyngeal swab specimens containing viruses from patients with signs and symptoms of respiratory infection prior to analysis with the Xpert Flu+RSV Xpress Assay.
2
{2}
2. Indication(s) for use:
Same as intended use.
3. Special conditions for use statement(s):
This is a prescription only test.
4. Special instrument requirements:
The assay is performed using the GeneXpert® Xpress System.
I. Device Description:
This assay uses nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection. Viral nucleic acid is extracted from the sample and the influenza and/or RSV RNA is amplified and detected through real-time reverse transcription polymerase chain reaction (RT-PCR). Detection and differentiation of influenza A, influenza B, and RSV is reported to the user.
The assay uses single use disposable cartridge that has a separate section for specimen loading. The cartridge also contains all PCR reagents and is where the PCR reaction takes place. The GeneXpert Xpress System performs all assay steps from clinical sample to reporting assay results automatically. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge. The SPC is present to control for adequate processing of the target viruses and to monitor the presence of inhibitors in the PCR reaction. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.
The GeneXpert Xpress System, comprised of the GeneXpert Dx System GX-I, has one module that is capable of performing separate sample preparation and real-time PCR and RT-PCR tests. This single module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells, and a proprietary I-CORE® thermocycler for performing real-time PCR and RT-PCR and detection.
Turnaround time for analysis of a sample is approximately 60 minutes. The assay results are automatically generated at the end of the process and provided in a report that can be viewed and printed.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Cepheid Xpert Flu/RSV XC Assay
2. Predicate 510(k) number(s):
{3}
K142045
3. Comparison with predicate:
Comparison of Similarities and Differences of the Xpert Flu+RSV Xpress Assay with the Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate Device |
| | Cepheid Xpert Flu+RSV Xpress Assay | Cepheid Xpert Flu/RSV XC Assay |
| Intended Use | The Cepheid Xpert Flu+RSV Xpress Assay, performed on the GeneXpert Xpress System, is an automated, multiplex real-time, reverse transcriptase polymerase chain reaction (RT-PCR) assay intended for the in vitro qualitative detection and differentiation of influenza A, influenza B, and respiratory syncytial virus (RSV) viral RNA. The Xpert Flu+RSV Xpress Assay uses nasopharyngeal swab specimens collected from patients with signs and symptoms of respiratory infection. The Xpert Flu+RSV Xpress Assay is intended as an aid in the diagnosis of influenza and respiratory syncytial virus in conjunction with clinical and epidemiological risk factors. | The Cepheid Xpert Flu/RSV XC Assay is an automated, multiplex real-time, reverse transcriptase polymerase chain reaction (RT-PCR assay intended for the in vitro qualitative detection and differentiation of influenza A, influenza B, and respiratory syncytial virus (RSV) viral RNA. The Xpert Flu/RSV XC Assay uses nasopharyngeal swab and nasal aspirate/wash specimens collected from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. The Xpert Flu/RSV XC Assay is intended as an aid in the diagnosis of influenza and respiratory syncytial virus. |
| | Negative results do not preclude influenza virus or respiratory syncytial virus infection and should not be used as the sole basis for treatment or other patient management decisions. | Negative results do not preclude influenza virus or respiratory syncytial virus infection and should not be used as the sole basis for treatment or other patient management decisions. |
| | Performance characteristics for influenza A were established during the 2014-2015 influenza season. When other novel influenza A viruses are emerging, | Performance characteristics for influenza A were established during the 2013-2014 influenza season. When other novel influenza A viruses are emerging, |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate Device |
| Intended Use | cepheid Xpert Flu+RSV Xpress Assay | Cepheid Xpert Flu/RSV XC Assay |
| | performance characteristics may vary.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. | performance characteristics may vary.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. |
| Indication for Use | Same | Patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. |
| Technology Principle of Operation | Same | Multiplex real time RT-PCR |
| Assay Targets | Same | Influenza A Virus, Influenza B Virus, and RSV viral RNA |
| Specimen Types | Nasopharyngeal (NP) swab specimens | Nasopharyngeal (NP) swab specimens and Nasal aspirate/wash (NA/W) specimens |
| Nucleic Acid Extraction | Yes | Yes |
| Extraction Methods | Sample preparation integrated in GeneXpert Cartridge and GeneXpert Xpress System | Sample preparation integrated in GeneXpert Cartridge and GeneXpert Instrument System |
| Assay Results | Same | Qualitative |
{5}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate Device |
| | Cepheid Xpert Flu+RSV Xpress Assay | Cepheid Xpert Flu/RSV XC Assay |
| Instrument System | Cepheid GeneXpert Xpress System (instrument model GX-I); same Cepheid I-core technology | Cepheid GeneXpert Instrument Systems (various instrument models including instrument model GX-I); Cepheid I-core technology |
| Assay Controls | Same | Encapsulated (armored) RNA pseudovirus as a sample processing control.
Available but not provided are inactivated virus controls for influenza A/B and RSV as external positive controls, and Coxsackie virus as an external negative control. |
| Time to obtain test results | Approximately 60 minutes for sample preparation and real-time RT-PCR. | Approximately 60 minutes or less for sample preparation and real-time RT-PCR. |
| Primers and probes | Same | Primers and probes to detect the presence of nucleic acid sequences of influenza A, influenza B, and RSV. |
| Differences | | |
| Instrument System | Cepheid GeneXpert Xpress System | Cepheid GeneXpert Dx Systems and GeneXpert Infinity Systems |
| Laboratory Users | Untrained operators with no clinical lab experience in a CLIA-waiver environment. | Operators with no clinical lab experience to experienced clinical laboratory technologists. |
| Combinatorial Assay Selections | No combinatorial assay selections are available. | Yes, user may select combined assay with all targets or a Flu only assay or a RSV only assay. |
| Early assay termination function | No early assay termination function is available. | Yes, on Flu only or RSV only assay selections. |
K. Standard/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay
{6}
# L. Test Principle:
The assay detects viral nucleic acids that have been extracted from a patient respiratory sample. A multiplex Real-time RT-PCR reaction is carried out under optimized conditions generating amplicons for influenza A, influenza B, RSV, and the Sample Process Control (SPC). Identification of influenza A, influenza B, RSV, and the SPC occurs by the use of target-specific primers and fluorescent-labeled probes that hybridize to conserved regions in the genomes.
| Xpert Flu+RSV Xpress Assay Probe Targets | |
| --- | --- |
| Virus | Target |
| Influenza A | Matrix Protein gene, Polymerase B2 Protein gene, Polymerase A Protein gene |
| Influenza B | Matrix Protein gene, Non-Structrural Protein gene |
| RSV A | Nucleocapsid Protein gene |
| RSV B | Nucleocapsid Protein gene |
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
The reproducibility studies were conducted over 10 days and testing was performed at three clinical sites and one in-house laboratory. There were three operators at each site. Operators were blinded to the sample identity. Testing was conducted following the package insert instructions for use. One kit lot was used to perform the study. One strain of each virus, influenza A, influenza B, and RSV was tested at three different concentrations, 2-3x LoD, 1x LoD, and below LoD for the Moderate Positive, Low Positive, and High Negative samples respectively. The following strains were tested: Flu A/Perth/16/09, Flu B/Wisconsin/01/2011, and RSV-A/2/Australia/61.
| Specimen | Concentration | Expected Positivity Rate |
| --- | --- | --- |
| Negative | 0 | 0% |
| Flu A | High negative (below LOD) C5 | 20-80% |
| Flu A | Low positive (~1X LOD) C95 | 95% |
| Flu A | Moderate positive (~2-3X LOD) | 100% |
| Flu B | High negative (below LOD) C5 | 20-80% |
| Flu B | Low positive (~1X LOD) C95 | 95% |
| Flu B | Moderate positive (~2-3X LOD) | 100% |
| RSV | High negative (below LOD) C5 | 20-80% |
| RSV | Low positive (~1X LOD) C95 | 95% |
| RSV | Moderate positive (~2-3X LOD) | 100% |
{7}
The following table presents the summary of results for the reproducibility study. The results show that the assay had appropriate reproducibility at all three sites. The reproducibility for this device is acceptable. The instances where the expected result differed from the actual result are within what is reasonable for $C_5$ (High negative) and $C_{95}$ (Low positive) samples.
| Sample | Site 1 | | | Site 2 | | | Site 3 | | | % Total Agreementa |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Op 1 | Op 2 | Op 3 | Op 1 | Op 2 | Op 3 | Op 1 | Op 2 | Op 3 | |
| Neg | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (9/9)b | 100% (9/9)b | 100% (10/10) | 100% (10/10) | 100% (88/88)b |
| Flu A-High Neg | 0.0% (0/10) | 20.0% (2/10) | 60.0% (6/10) | 20.0% (2/10) | 10.0% (1/10) | 50.0% (5/10) | 75.0% (6/8)b | 20.0% (2/10) | 20.0% (2/10) | 29.5% (26/88)b |
| Flu A-Low Pos | 100% (10/10) | 100% (10/10) | 90.0% (9/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 98.9% (89/90) |
| Flu A-Mod Pos | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (90/90) |
| Flu B-High Neg | 50.0% (5/10) | 0.0% (0/10) | 50.0% (5/10) | 20.0% (2/10) | 10.0% (1/10) | 55.6% (5/9)b | 80.0% (8/10) | 50.0% (5/10) | 20.0% (2/10) | 37.1% (33/89)b |
| Flu B-Low Pos | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 90.0% (9/10) | 100% (10/10) | 100% (10/10) | 98.9% (89/90) |
| Flu B-Mod Pos | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (90/90) |
| RSV-High Neg | 60.0% (6/10) | 30.0% (3/10) | 66.7% (6/9)b | 50.0% (5/10) | 60.0% (6/10) | 50.0% (5/10) | 90.0% (9/10) | 33.3% (3/9)b | 50.0% (5/10) | 54.5% (48/88)b |
| RSV-Low Pos | 88.9% (8/9) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 90.0% (9/10) | 100% (10/10) | 100% (10/10) | 97.8% (87/89)c |
| RSV-Mod Pos | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (10/10) | 100% (90/90) |
aPercent of samples yielding expected results - negative for Neg and High Neg samples; positive for Low Pos and Mod Pos samples.
bSeven samples (2 Neg, 2 Flu A High Neg, 1 Flu B High Neg, and 2 RSV High Neg) were indeterminate upon initial testing and retest.
One RSV Low Pos sample was inadvertently not tested.
All negative samples were correctly classified as negative (88/88).
All moderate positive samples were correctly classified as positive, Flu A (90/90), Flu B (90/90), and RSV (90/90).
The high negative samples were prepared to target a concentration below the LoD. At this concentration we would expect approximately $20 - 80\%$ of samples to be negative. Overall $29.5\%$ (26/88) of the high negative Flu A samples were classified as negative; $37.1\%$ (33/89) of the high negative Flu B samples were classified as negative; and $54.5\%$ (48/88) of the high negative RSV samples were classified as negative.
{8}
The low positive samples were prepared to target a concentration at the limit of detection. At this concentration we would expect approximately 95% of samples to be positive. Overall 98.9% (89/90) of the low positive Flu A samples were classified as positive; 98.9% (89/90) of the low positive Flu B samples were classified as positive; and 97.8% (87/89) of the low positive RSV samples were classified as positive.
## b. Limit of Detection:
The limit of detection (LoD) study was conducted with 10% of the Xpert Flu+RSV Xpress Assay cartridges on the GeneXpert Xpress System and 90% with the Xpert Flu/RSV XC Assay cartridges on the Gene Xpert Dx System and the GeneXpert Infinity Systems. The LoD of the Xpert Flu+RSV Xpress Assay, defined as the lowest TCID₅₀/mL that is detected ≥95% of the time, was determined by testing influenza A H3N2 strains, influenza A 2009 H1N1 strains, influenza B strains, RSV-A strains, RSV-B strains, and an influenza A H7N9 strain in a background matrix of clinical, pooled swab specimens collected in Copan UTM.
### Influenza A pandemic 2009 H1N1 strains
- A/California/7/2009 (H1N1)
- A/Florida/27/2011 (H1N1)
### Influenza A H3N2 strains
- A/Perth/16/2009 (H3N2)
- A/Victoria/361/2011 (H3N2)
### Influenza B strains
- B/Mass/2/2012
- B/Wisconsin/01/2011
### Respiratory Syncytial Virus A strains
- RSV-A/2/Australia/61
- RSV-A/Long/MD/56
### Respiratory Syncytial Virus B strains
- RSV-B/Washington/18537/62
- RSV-B/9320/Massachusetts/77
The LoD was estimated using probit regression analysis and confirmed with twenty replicates for each dilution of each virus. The samples were prepared and tested across 3 testing days using one reagent lot. Testing was conducted according to the package insert. The table below shows the LoD determined for each virus type.
{9}
Confirmed LoD, Point Estimates and 95% Confidence Intervals for Influenza and RSV Targets
| Influenza A 2009 H1N1 | Confirmed LoD (TCID50/mL) [at least 19/20 positive] | LoD Estimate by Probit Regression Analysis (TCID50/mL) | | |
| --- | --- | --- | --- | --- |
| | | LoD Point Estimate | Lower 95% CI | Upper 95% CI |
| A/California/7/2009 | 0.3 (20/20) | 0.1 | 0.06 | 0.11 |
| A/Florida/27/2011 | 16.0 (20/20) | 15.4 | 13.77 | 20.59 |
| Influenza A H3N2 | Confirmed LoD (TCID50/mL) [at least 19/20 positive] | LoD Estimate by Probit Regression Analysis (TCID50/mL) | | |
| | | LoD Point Estimate | Lower 95% CI | Upper 95% CI |
| A/Perth/16/2009 | 0.3 (20/20) | 0.1 | 0.05 | 0.09 |
| A/Victoria/361/2011 | 0.8 (20/20) | 0.8 | 0.65 | 1.21 |
| Influenza B | Confirmed LoD (TCID50/mL) [at least 19/20 positive] | LoD Estimate by Probit Regression Analysis (TCID50/mL) | | |
| | | LoD Point Estimate | Lower 95% CI | Upper 95% CI |
| B/Mass/2/2012 | 0.5 (20/20) | 0.5 | 0.38 | 0.62 |
| B/Wisconsin/01/2011 | 0.6 (20/20) | 0.5 | 0.46 | 0.74 |
| RSV A | Confirmed LoD (TCID50/mL) [at least 19/20 positive] | LoD Estimate by Probit Regression Analysis (TCID50/mL) | | |
| | | LoD Point Estimate | Lower 95% CI | Upper 95% CI |
| RSV A/2/Australia/61 | 1.2 (20/20) | 1.1 | 0.96 | 1.26 |
| RSV A/Long/MD/56 | 1.0 (20/20) | 1.0 | 0.84 | 1.38 |
| RSV B | Confirmed LoD (TCID50/mL) [at least 19/20 positive] | LoD Estimate by Probit Regression Analysis (TCID50/mL) | | |
| | | LoD Point Estimate | Lower 95% CI | Upper 95% CI |
| RSV B/Wash/18537/62 | 1.8 (20/20) | 1.6 | 1.49 | 1.95 |
| RSV B/9320/MA/77 | 2.0 (20/20) | 2.1 | 1.84 | 2.64 |
| Influenza A (H7N9) | Confirmed LoD (TCID50/mL) [at least 19/20 positive] | LoD Estimate by Probit Regression Analysis (TCID50/mL) | | |
| | | LoD Point Estimate | Lower 95% CI | Upper 95% CI |
| A/Anhui/1/2013 (H7N9) | 0.8 (19/20) | 0.8 | 0.7 | 0.9 |
{10}
Analytical Studies: provided in support of K142045 and the current submission
The following study data were originally submitted for the Xpert Flu/RSV XC Assay in K142045 and also included in the current submission. The majority of samples in these studies were tested by the Xpert Flu/RSV XC Assay and a smaller proportion of samples were tested by the Xpert Flu+RSV Xpress Assay. All reagents, buffers, and other components that comprise both assays are exactly the same. The detection method, the signal analysis, and the software algorithm used to calculate the results are also identical.
c. Baseline, Threshold and Cut-off:
The lot specific parameters (LSP) and assay settings (background subtraction, background minimum cycle, background maximum cycle, manual threshold, curve analysis, and valid minimum and maximum Ct values) for the Xpert Flu+RSV XC Assay were determined using pre-clinical and analytical data with the Xpert Flu/RSV XC Assay cleared in K142045. The technology was not significantly changed from the XC Assay to the Xpress Assay; the assay reagents and assay analysis algorithm have not changed. The LSP and assay settings were confirmed for the Xpert Flu +RSV Xpress Assay.
d. Stability:
The Flu/RSV XC Assay was used to determine the stability of reagents in the Xpert Flu+RSV Xpress Assay because all reagents, buffers, and other components that comprise both assays are exactly the same. The detection method, the signal analysis, and algorithm used by the different software to calculate the results are the same.
Kit Stability:
The real-time reagent stability data for the Xpert Flu/RSV XC Assay is currently available for nine months at the claimed storage temperatures of 2°C and 28°C. Results to date show that averaged Ct values are within each assay lot acceptance criteria and testing is ongoing for up to 37 months on three lots.
Specimen Stability:
The study data supports specimen stability at extreme storage temperatures and storage times (7 days at 2°C followed by 24 hours at 30°C followed by testing on day 9) for all three claimed transport media (Becton Dickinson UVTM, Remel M5, and Copan UTM).
e. Expected values (controls, calibrators, or methods):
An internal assay control called the 'sample processing control' (SPC) monitors adequate sample processing and the integrity of the RT-PCR. The SPC is mixed with the sample automatically inside the test cartridge to control for adequate lysis of the target viruses, sample processing, and detection of assay interference.
No external controls are provided with this assay. The recommended external controls are:
- ZeptoMetrix inactivated virus controls: (catalog # NATFLUAB-6C and catalog # NATRSV-6C) are external positive controls required for the Xpert Flu/RSV XC Assay.
{11}
- ZeptoMetrix (catalog # NATCXVA9-6C) is an inactivated Coxsackie virus external negative control required for the Xpert Flu/RSV XC Assay.
# f. Analytical Inclusivity:
The analytical reactivity of the Xpert Flu+RSV Xpress Assay was evaluated against multiple strains of influenza A H1N1 (seasonal pre-2009), influenza A H1N1 (pandemic 2009), influenza A H3N2 (seasonal), avian influenza A (H5N2, H6N2, H7N3, and H7N9, and H9N2), influenza B (representing strains from both Victoria and Yamagata lineages), and respiratory syncytial virus subgroups A and B (RSV A and RSV B) at levels near the analytical LoD. All virus identities and titers were confirmed prior to analysis. Sixty-four (64) strains comprised of 54 influenza viruses and 10 RSV strains were tested in this study with the Xpert Flu+RSV Xpress Assay. Virus dilution and sample preparation was performed with simulated background matrix consisted of $2.5\%$ (w/v) porcine mucin, $1\%$ (v/v) human whole blood in $0.85\%$ sodium chloride (NaCl) formulated in 1X PBS solution with $15\%$ glycerol, which was then diluted in UTM to a final concentration of $16.7\%$ . An aliquot of $300~\mu \mathrm{L}$ of the final inocula was added to the Xpert Flu+RSV Xpress Assay cartridge and 3 replicates were tested for each strain. Due to the biosafety restrictions on viable viral particles, purified nucleic acids at $\leq 1\rho \mathrm{g / mL}$ in simulated background matrix were tested for 9 avian influenza A strains. Three replicates of a no template control were also tested which contained the simulated background matrix only. One replicate each of three external controls (one positive control for Flu A/B, one positive control for RSV, and one negative control) was tested with the Xpert Flu/RSV XC Assay on each day of the study.
| Virus | Strain | Concentration | Result | | |
| --- | --- | --- | --- | --- | --- |
| | | | Flu A | Flu B | RSV |
| No Template Control | | | NEG | NEG | NEG |
| Influenza A H1N1 (pre-2009) | A/swine/Iowa/15/30 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/WS/33 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/PR/8/34 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/Mal/302/54 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/Denver/1/57 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/New Jersey/8/76 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/New Caledonia/20/1999 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/New York/55/2004 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/Soloman Island/3/2006 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/Taiwan/42/06 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/Brisbane/59/2007 | 32.0 TCID50/mL | POS | NEG | NEG |
| Influenza A H1N1 (pdm2009) | A/California/7/2009 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/swine/NY/02/2009 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/Florida/27/2011 | 32.0 TCID50/mL | POS | NEG | NEG |
| | A/Colorado/14/2012 | 32.0 TCID50/mL | POS | NEG | NEG |
{12}
| Virus | Strain | Concentration | Result | | |
| --- | --- | --- | --- | --- | --- |
| | | | Flu A | Flu B | RSV |
| | A/Washington/24/2012 | 80.0aTCID50/mL | POS | NEG | NEG |
| Influenza A H3N2 (Seasonal) | A/Aichi/2/68 | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/HongKong/8/68 | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/Port Chalmers/1/73 | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/Hawaii/15/2001 | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/Wisconsin/67/05 | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/Brisbane/10/2007 | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/Perth/16/2009 | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/Minnesota/11/2010 (H3N2)v | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/Indiana/08/2011 (H3N2)v | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/Victoria/361/2011 | 1.6 TCID50/mL | POS | NEG | NEG |
| | A/Texas/50/2012 | 1.6 TCID50/mL | POS | NEG | NEG |
| Avian influenza A | A/duck/Hunan/795/2002 (H5N1) | 1 g/Lb | POS | NEG | NEG |
| | A/chicken/Hubei/327/2004 (H5N1) | 1 g/Lb | POS | NEG | NEG |
| | A/Anhui/01/2005 (H5N1) | 1 g/Lb | POS | NEG | NEG |
| | A/Japanese white eye/HongKong/1038/2006 (H5N1) | 1 g/Lb | POS | NEG | NEG |
| | A/mallard/WI/34/75 (H5N2) | 1 g/Lb | POS | NEG | NEG |
| | A/chicken/CA431/00 (H6N2) | 1 g/Lb | POS | NEG | NEG |
| | A/duck/LTC-10-82743/1943 (H7N2) | 1 g/Lb | POS | NEG | NEG |
| | A/chicken/NJ/15086-3/94 (H7N3) | 1 g/Lb | POS | NEG | NEG |
| | A/Anhui/1/2013 (H7N9) | N/Ac | POS | NEG | NEG |
| | A/Shanghai/1/2013 (H7N9) | N/Ac | POS | NEG | NEG |
| | A/chicken/Korea/38349-p96323/1996 (H9N2) | 1 g/Lb | POS | NEG | NEG |
| | A/Mallard/NY/6750/78 (H2N2) | 1 g/Lb | POS | NEG | NEG |
| Influenza B | B/Lee/40 | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Allen/45 | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/GL/1739/54 | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Maryland/1/59 | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Panama/45/90d | 3.0 TCID50/mL e | NEG | POS | NEG |
| | B/Florida/07/2004f | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Florida/02/06d | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Florida/04/06f | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Wisconsin/01/2011d | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Massachusetts/2/2012f | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Hong Kong/5/72 | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Wisconsin/01/2010f | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/Malaysia/2506/04d | 1.2 TCID50/mL | NEG | POS | NEG |
| Influenza B/H1N1 (Seasonal) | B/Lee/10 | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/China/1/10 | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/China/1/10 | 1.2 TCID50/mL | NEG | POS | NEG |
| | B/China/1/10 | 1.2 TCID50 | NEG | POS | NEG |
| | B/China/1/10 | 1.2 TCID50 | NEG | POS | NEG |
| | B/China/1/10 | 1.25 | NEG | NEG | NEG |
| | B/China/1/10 | 1.25 | NEG | NEG | NEG |
| | B/China/1/10 | 1.25 | NEG | NEG | NEG |
| | B/China/1/10 | 1.25 | NEG | NEG | NEG |
| | B/China/1/10 | 1.25 | NEG | NEG | NEG |
| | B/China/1/10 | 1.25 | NEG | NEG | NEG |
| | B/China/1/10 | 1.25 | | | |
| | B/China/1/10 | 1.25 | | | |
{13}
| Virus | Strain | Concentration | Result | | |
| --- | --- | --- | --- | --- | --- |
| | | | Flu A | Flu B | RSV |
| RSV A | B/Taiwan/2/62 | 1.2 TCID_{50}/mL | NEG | POS | NEG |
| | B/Brisbane/60/2008^{d} | 1.2 TCID_{50}/mL | NEG | POS | NEG |
| | RSV-A/Long/MD/56 | 2.4 TCID_{50}/mL | NEG | NEG | POS |
| | RSV-A/2/Australia/61 | 2.4 TCID_{50}/mL | NEG | NEG | POS |
| RSV B | RSV-A/NY (Clinical unknown) | 2.4 TCID_{50}/mL | NEG | NEG | POS |
| | RSV-A/WI/629-8-2/2007 | 2.4 TCID_{50}/mL | NEG | NEG | POS |
| | RSV-A/WI/629-11-1/2008 | 2.4 TCID_{50}/mL | NEG | NEG | POS |
| | RSV-B/Wash/18537/62 | 4.0 TCID_{50}/mL | NEG | NEG | POS |
| RSV B | RSV-B/9320/MA/77 | 4.0 TCID_{50}/mL | NEG | NEG | POS |
| | RSV-B/WV14617/85 | 4.0 TCID_{50}/mL | NEG | NEG | POS |
| | RSV-B/CH93(18)-18 | 20.0 TCID_{50}/mL^{g} | NEG | NEG | POS |
| | RSV-B/WI/629-5B/0607 | 4.0 TCID_{50}/mL | NEG | NEG | POS |
aInfluenza A/Washington/24/2012 was tested at 5x LoD (80.0 TCID<sub>50</sub>/mL) to obtain 3 of 3 Flu A POSITIVE result calls.
bPurified viral RNA in simulated background matrix was used for avian influenza A viruses due to biosafety regulations.
cInactivated avian influenza A (H7N9) viruses without viral titer was diluted 100,000 fold in simulated background matrix and tested due to biosafety regulations.
dKnown Victoria lineage.
eInfluenza B/Panama/45/90 was tested at 5x LoD (3.0 TCID<sub>50</sub>/mL) to obtain 3/3 Flu B POSITIVE result calls.
fKnown Yamagata lineage.
gRSV-B/CH93(18)-18 was tested at 10x LoD (20.0 TCID<sub>50</sub>/mL) to obtain 3/3 RSV POSITIVE result calls.
All 64 influenza and RSV strains were correctly identified in the study, 39 as influenza A, 15 as influenza B, and 10 as RSV using the Xpert Flu+RSV Xpress Assay.
g. Analytical Specificity/Cross-Reactivity:
Cross Reactivity:
This study was designed to determine potential cross-reactivity of the assay with viruses, yeast, and bacteria common in the nasopharynx. Bacteria were tested at 1.00E+6 cfu/ml (except one strain which was tested at 1.00E+05 TCID<sub>50</sub>/mL) and viruses were tested at 1.00E+05 TCID<sub>50</sub>/mL. Testing was performed in triplicate. All organisms were diluted in simulated background matrix and testing was performed according to the package insert directions.
{14}
Analytical Specificity/Cross-reactivity of the Cepheid Xpert Flu+RSV Xpress Assay
| Organism | Concentration | Influenza A | Influenza B | RSV |
| --- | --- | --- | --- | --- |
| No Template Control | | 0/3 | 0/3 | 0/3 |
| Adenovirus Type 1 | 1.12x107TCID50/mL | 0/3 | 0/3 | 0/3 |
| Adenovirus Type 7 | 1.87x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human coronavirus OC43 | 2.85x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human coronavirus 229E | 1x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Cytomegalovirus | 7.24x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Echovirus | 3.31x107TCID50/mL | 0/3 | 0/3 | 0/3 |
| Enterovirus | 1x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Epstein Barr Virus | 7.16x107TCID50/mL | 0/3 | 0/3 | 0/3 |
| HSV | 8.9x106TCID50/mL | 0/3 | 0/3 | 0/3 |
| Measles | 6.3x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human metapneumovirus | 3.8x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Mumps virus | 6.31x106TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human parainfluenza Type 1 | 1.15x106TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human parainfluenza Type 2 | 1x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human parainfluenza Type 3 | 3.55x107TCID50/mL | 0/3 | 0/3 | 0/3 |
| Rhinovirus Type 1A | 1.26x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Acinetobacter baumannii | >1x106CFU/mL | 1/26a | 0/26 | 0/26 |
| Burkholderia cepacia | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Candida albicans | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Candida parapsilosis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Bordetella pertussis | 1x108CFU/mL | 0/3 | 0/3 | 0/3 |
{15}
| Organism | Concentration | Influenza A | Influenza B | RSV |
| --- | --- | --- | --- | --- |
| Chlamydia pneumoniae | 3.16x105CFU/mL | 0/3 | 0/3 | 0/3 |
| Citrobacter freundii | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Corynebacterium sp. | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Escherichia coli | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Enterococcus faecalis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Hemophilus influenzae | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Lactobacillus sp. | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Legionella spp. | 1x108CFU/mL | 0/3 | 0/3 | 0/3 |
| Moraxella catarrhalis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Mycobacterium tuberculosis (avirulent) | 1.15x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Mycoplasma pneumoniae | 1x107CFU/mL | 0/3 | 0/3 | 0/3 |
| Neisseria meningitides | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Neisseria mucosa | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Propionibacterium acnes | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Pseudomonas aeruginosa | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Staphylococcus aureus (protein A producer) | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Staphylococcus epidermidis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Staphylococcus haemolyticus | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus agalactiae | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus pneumoniae | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus pyogenes | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus salivarius | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus sanguinis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
a. For Acinetobacter baumannii, upon initial testing 1 of 3 replicates was positive for Flu A with a Ct of 39.4 (cut-off = 40). An additional 23 replicates were tested at >1x106 CFU/mL; 23 of 23 replicates were correctly reported as “Flu A NEGATIVE; Flu B NEGATIVE; RSV NEGATIVE”.
Under the conditions of this study, 43 of the 44 non-influenza isolates tested were correctly reported as “Flu A NEGATIVE; Flu B NEGATIVE; RSV NEGATIVE” by the Xpert Flu+RSV Xpress Assay. For Acinetobacter baumannii, 1 of the 3 replicates initially tested was reported as
{16}
"Flu A POSITIVE; FLU B NEGATIVE; RSV NEGATIVE" with a Flu A Ct value of 39.4 (cutoff = 40). A Ct value of 39.4 is not typical of positive Cts for Flu A and the false positive result was likely caused by contamination. An additional twenty-three (23) replicates of Acinetobacter baumannii were tested; all 23 replicates were correctly reported as "Flu A NEGATIVE; Flu B NEGATIVE; RSV NEGATIVE". Based on the results of this study, the Xpert Flu+RSV Xpress Assay showed no cross-reactivity with any of the organisms tested.
## Microbial Interference:
The analytical specificity of the Xpert Flu+RSV Xpress Assay was evaluated by testing a panel of forty-four (44) microorganisms consisting of 16 viral, 26 bacterial, and 2 yeast strains representing common respiratory pathogens or those potentially encountered in the nasopharynx. One replicate each of three external controls (one positive control for Flu A/B, one positive control for RSV, and one negative control) was tested with the Xpert Flu/RSV XC Assay on each day of the study.
All bacterial strains were tested in triplicate at concentrations of $\geq 10^{6}$ CFU/mL with the exception of one strain which was tested at $10^{5}$ CFU/mL (Chlamydia pneumoniae). All viral strains were tested in triplicate at concentrations of $\geq 10^{5}$ TCID$_{50}$/mL. Three replicates of a no template control (background matrix diluted in background matrix) were also tested.
All the strains were diluted into a simulated background matrix. The simulated background matrix consisted of $2.5\%$ (w/v) porcine mucin, $1\%$ (v/v) human whole blood in $0.85\%$ sodium chloride (NaCl) formulated in 1X PBS solution with $15\%$ glycerol, which was then diluted in UTM to a final concentration of $16.7\%$. Negative specimens consisted of the diluted simulated background matrix only. Dilutions were prepared daily and kept on ice prior to testing.
Results are summarized below. The results demonstrate that the Xpert Flu+RSV Xpress Assay does not cross-react with any of the 44 common respiratory microorganisms tested in this study.
Analytical Specificity/Cross-reactivity of the Cepheid Xpert Flu+RSV Xpress Assay
| Organism | Concentration | Influenza A | Influenza B | RSV |
| --- | --- | --- | --- | --- |
| No Template Control | | 0/3 | 0/3 | 0/3 |
| Adenovirus Type 1 | 1.12x10^{7} TCID_{50}/mL | 0/3 | 0/3 | 0/3 |
| Adenovirus Type 7 | 1.87x10^{5} TCID_{50}/mL | 0/3 | 0/3 | 0/3 |
| Human coronavirus OC43 | 2.85x10^{5} TCID_{50}/mL | 0/3 | 0/3 | 0/3 |
| Human coronavirus 229E | 1x10^{5} TCID_{50}/mL | 0/3 | 0/3 | 0/3 |
{17}
| Organism | Concentration | Influenza A | Influenza B | RSV |
| --- | --- | --- | --- | --- |
| Cytomegalovirus | 7.24x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Echovirus | 3.31x107TCID50/mL | 0/3 | 0/3 | 0/3 |
| Enterovirus | 1x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Epstein Barr Virus | 7.16x107TCID50/mL | 0/3 | 0/3 | 0/3 |
| HSV | 8.9x106TCID50/mL | 0/3 | 0/3 | 0/3 |
| Measles | 6.3x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human metapneumovirus | 3.8x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Mumps virus | 6.31x106TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human parainfluenza Type 1 | 1.15x106TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human parainfluenza Type 2 | 1x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Human parainfluenza Type 3 | 3.55x107TCID50/mL | 0/3 | 0/3 | 0/3 |
| Rhinovirus Type 1A | 1.26x105TCID50/mL | 0/3 | 0/3 | 0/3 |
| Acinetobacter baumannii | >1x106CFU/mL | 1/26a | 0/26 | 0/26 |
| Burkholderia cepacia | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Candida albicans | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Candida parapsilosis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Bordetella pertussis | 1x108CFU/mL | 0/3 | 0/3 | 0/3 |
| Chlamydia pneumoniae | 3.16x105CFU/mL | 0/3 | 0/3 | 0/3 |
| Citrobacter freundii | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Corynebacterium sp. | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Escherichia coli | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Enterococcus faecalis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Hemophilus influenzae | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Lactobacillus sp. | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Lactobacillus rhamnosus | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Lactobacillus rhamnosus | 3.16x105CFU/mL | 0/3 | 0/3 | 0/3 |
| Lactobacillus rhamnosus | 3.16x105CFU/mL | 0/3 | 0/3 | 0/3 |
| Lactobacillus rhamnosus | 3.16x105CFU/mL | 0/3 | 0/3 | 0/3 |
| Lactobacillus rhamnosus | 3.16x105CFU/mL | 0/3 | 0/3 | 0/3 |
| L. plantarum | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| L. plantarum | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| L. plantarum | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| L. plantarum | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| L. ruminantium | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| L. ruminantium | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| L. ruminantium | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| L. ruminantium | 1x106CFU/mL | 0/3 | 0/3 | 0/3 |
{18}
| Organism | Concentration | Influenza A | Influenza B | RSV |
| --- | --- | --- | --- | --- |
| Legionella spp. | 1x108CFU/mL | 0/3 | 0/3 | 0/3 |
| Moraxella catarrhalis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Mycobacterium tuberculosis (avirulent) | 1.15x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Mycoplasma pneumoniae | 1x107CFU/mL | 0/3 | 0/3 | 0/3 |
| Neisseria meningitides | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Neisseria mucosa | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Propionibacterium acnes | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Pseudomonas aeruginosa | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Staphylococcus aureus (protein A producer) | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Staphylococcus epidermidis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Staphylococcus haemolyticus | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus agalactiae | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus pneumoniae | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus pyogenes | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus salivarius | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
| Streptococcus sanguinis | >1x106CFU/mL | 0/3 | 0/3 | 0/3 |
a1 out of the 3 original replicates tested as "Flu A Positive". The Ct score associated with the false positive was 39.4; near the cut-off of 40.0 for a positive call. An additional 23 replicates were tested and all 23 were negative for all three analyte channels.
# Competitive Interference:
The purpose of this study is to determine whether competitive interference exists between analytes when more than one analyte is present in a clinical sample. Competitive interference caused by clinically relevant co-infections of the targets in the Xpert Flu+RSV Xpress Assay was evaluated by testing individual influenza and RSV strains at LoD and spiking in different influenza or RSV strains at a higher concentration in a simulated background matrix. The concentration of each strain at LoD ranged from 0.57 TCID50/mL to 2.0 TCID50/mL and the concentration of the competitive strains ranged from $10^{3}$ TCID50/mL to $10^{4}$ TCID50/mL. Analytical competitive interference was assessed using one (1) seasonal Flu A H3 strain (H3/Victoria/361/2011), one (1) Flu B strain (Flu B/Mass/2/2012), one (1) RSV A strain (RSV-A/2/Australia/61), and one (1) RSV B strain (RSV-B/Wash/18537/62).
{19}
Replicates of 20 were tested for each target strain and each competitive strain combination. The normal binomial distribution with 20 replicate samples at LoD is between 17 and 20 positive results based on the binomial distribution with $N = 20$ , $p = .95$ (X~Bin(20,0.95)). Therefore, sets of 20 with 16 or less positives would be rare and an indication of a competitive inhibitory effect due to high levels of a competing analyte. One replicate each of three external controls (one Flu A/B positive, one RSV positive and one negative) was tested with the Xpert Flu+RSV Xpress Assay on each day of this study.
Under the conditions of the study using the normal binomial distribution with 20 replicate samples at LoD of between 17 and 20 positive results, no competitive inhibitory effects were observed at the analytical LoD for each of the strains tested in the presence of another analyte. The results for each analyte at LoD in combination with another competitive analyte are shown in the following four tables.
Note: The average Ct values for each strain tested at LoD without a competitive analyte are taken from the Limit of Detection study.
Competitive Interference for Flu A/Victoria/361/2011 at LoD
| Target Strain Titer (TCID50/mL) | Competitive Strain | Competitive Strain Titer (TCID50/mL) | Positives/ 20 replicates | Average Ct | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | Flu A1 | Flu B | RSV | SPC |
| 0.8 | n/a | n/a | 20/20 | 33.3 | - | - | 30.8 |
| 1.0 (Upper 95% CI of LoD) | B/Wisconsin/01/11 | 1.00E+03 | 18/20 | 36.2 | 22.7 | - | 31.2 |
| | RSV- A/Long/MD/56 | 1.00E+04 | 18/20 | 36.3 | - | 19.4 | 31.2 |
| | RSV- B/Wash/18537/62 | 1.00E+04 | 19/20 | 35.0 | - | 20.3 | 31.0 |
Competitive Interference for Flu B/Mass/2/2012 at LoD
| Target Strain Titer (TCID50/mL) | Competitive Strain | Competitive Strain Titer (TCID50/mL) | Positives/ 20 replicates | Average Ct | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | Flu A1 | Flu B | RSV | SPC |
| 0.5 | n/a | n/a | 20/20 | - | 32.4 | - | 30.7 |
| 0.57 (Upper 95% CI of LoD) | H3/Victoria/361/2011 | 1.00E+03 | 17/20 | 23.4 | 37.1 | - | 31.4 |
| | RSV-A/Long/MD/56 | 1.00E+03 | 19/20 | - | 34.9 | 23.1 | 31.4 |
| | RSV- B/Wash/18537/62 | 1.00E+03 | 19/20 | - | 33.5 | 24.0 | 31.3 |
Competitive Interference for RSV-A/2/Australia/61 at LoD
| Target Strain Titer (TCID50/mL) | Competitive Strain | Competitive Strain Titer (TCID50/mL) | Positives/ 20 replicates | Average Ct | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | Flu A1 | Flu B | RSV | SPC |
| 1.2 | n/a | n/a | 20/20 | - | - | 35.4 | 30.7 |
| 1.3 (Upper 95% CI of LoD) | H3/Victoria/361/2011 | 1.00E+03 | 18/20 | 23.7 | - | 37.2 | 31.7 |
| | B/Wisconsin/01/11 | 1.00E+03 | 18/20 | - | 22.9 | 36.4 | 31.5 |
{20}
Competitive Interference for RSV-B/Wash/18537/62 at LoD
| Target Strain Titer (TCID50/mL) | Competitive Strain | Competitive Strain Titer (TCID50/mL) | Positives/ 20 replicates | Average Ct | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | Flu A1 | Flu B | RSV | SPC |
| 1.8 | n/a | n/a | 20/20 | - | - | 34.1 | 30.8 |
| 2.0 (Upper 95% CI of LoD) | H3/Victoria/361/2011 | 1.00E+03 | 19/20 | 23.3 | - | 35.4 | 31.6 |
| | B/Wisconsin/01/11 | 1.00E+03 | 19/20 | - | 22.9 | 34.6 | 31.5 |
h. Interfering Substances:
In a non-clinical study, potentially interfering substances that may be present in the nasopharynx were evaluated relative to the performance of the Xpert Flu+RSV Xpress Assay. Negative samples $(n = 8)$ were tested per each substance to determine the effect on the performance of the sample processing control (SPC). Positive samples $(n = 8)$ were tested per substance spiked at $2x$ the analytical LoD determined for each of the following six influenza viruses, two 2009 H1N1 strains (A/California/7/2009 and A/Florida/27/2011), two Flu A H3N2 strains (H3/Perth/16/09 and H3/Victoria/361/2011, and two Flu B stains (Flu B/Wisconsin/01/11 and Flu B/Mass/02/2012), and four RSV viruses, two RSV A strains (RSV-A/Long/MD/56 and RSV-A/2/Australia/61) and two RSV B strains (RSV-B/Wash/18537/62 and RSV-B/9320/MA/77). All results were compared to positive and negative Universal Transport Medium (UTM) controls. The evaluated substances are below with active ingredients and concentrations tested shown. There was no assay interference in the presence of the substances at the concentrations tested in this study. All positive and negative replicates were correctly identified using the Xpert Flu+RSV Xpress Assay.
FluMist vaccine samples were correctly reported as FLU A POSITIVE; FLU B POSITIVE; RSV NEGATIVE as expected. Samples containing FluMist may cause false positive results.
Interfering Substances Tested
| Substance ID | Substance/Class | Substance/Active Ingredient | Concentration Tested |
| --- | --- | --- | --- |
| C | Control | UTM | 100% (v/v) |
| Albuterol Sulfate | Beta-adrenergic bronchodilator | Albuterol Sulfate | 0.83 mg/mL |
| Blood | Blood | Blood (human) | 2% (v/v) |
| BD | Transport Media | n/a | 100% (v/v) |
| M4 | Transport Media | n/a | 100% (v/v) |
| M4RT | Transport Media | n/a | 100% (v/v) |
| M5 | Transport Media | n/a | 100% (v/v) |
| Menthol | Oral anesthetic and analgesic | Benzocaine, Menthol | 1.7 mg/mL |
| Mucin | Mucin | Purified Mucin protein | 2.5% (w/v) |
| Mupirocin | Antibiotic nasal ointment | Mupirocin | 10 mg/mL |
{21}
Interfering Substances: Agreement with Expected Results
| Substance | Negative | Flu A H3N2 | | Flu A 2009 H1N1 | | Flu B | | RSV A | | RSV B | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Strain 1 | Strain 2 | Strain 1 | Strain 2 | Strain 1 | Strain 2 | Strain 1 | Strain 2 | Strain 1 | Strain 2 |
| UTM | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Albuterol | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Blood | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| BD UVT transport media | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| M4 transport media | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| M4RT transport media | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| M5 transport media | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Menthol | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Mucin | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Mupirocin | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| NaCl | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Oxymetazoline | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Phenylephrine | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Tamiflu | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Tobramycin | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Zicam | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| FluMist* | ND | ND | ND | ND | ND | ND | ND | ND | ND | ND | ND |
| UTM control for Fluticasone Propionate | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
| Fluticasone Proprionate | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 | 8/8 |
* The FluMist vaccine was tested as part of the interfering substance study. FluMist contains three live attenuated influenza vaccine virus strains: one influenza A (H1N1) strain, one influenza A (H3N2) strain, and one influenza B
{22}
strain. Negative samples of this substance cannot be tested to observe its impact on SPC Ct. The FluMist vaccine samples tested with the Xpert Flu+RSV Xpress Assay were correctly reported as “Flu A POSITIVE; Flu B POSITIVE; RSV NEGATIVE” as expected. Samples containing FluMist may cause false positives. This is addressed in the Limitations Section of the Package Insert.
i. Carry-Over/Cross-Contamination:
The purpose of this study was to demonstrate whether single use, self-contained GeneXpert cartridges prevent specimen and amplicon carry-over contamination into negative samples when these are run following very high positive samples in the same GeneXpert module. GeneXpert cartridges are designed such that patient specimens and all assay reagents are confined within the self-contained GeneXpert cartridge and do not come into contact with the instrument and/or its moving parts.
This study consisted of a negative sample (16.7% simulated background matrix only) processed in the same module of the GeneXpert Xpress System (GX-I) immediately following a high positive sample comprised of a Flu A strain (A/Canada/6294/2009, 1x10⁶ TCID₅₀/mL) or a RSV A virus (A/2/Australia/61, 1x10⁴ TCID₅₀/mL) spiked into a 16.7% simulated background matrix. This process was repeated 20 times in a single GeneXpert module for a total of 41 runs resulting in 20 positives and 21 negatives. The simulated background matrix consisted of 2.5% (w/v) porcine mucin, 1% (v/v) human whole blood in 0.85% sodium chloride (NaCl) formulated in 1X PBS solution with 15% glycerol, which was then diluted in UTM to a final concentration of 16.7%.
All of the 42 negative replicates were correctly reported as “Flu A NEGATIVE; Flu B NEGATIVE; RSV NEGATIVE”. All the negative sample Cts were zero.
All 20 high Flu A positive replicates were correctly reported as “Flu A POSITIVE; Flu B NEGATIVE; RSV NEGATIVE”. All 20 high RSV positive replicates were correctly reported as “Flu A NEGATIVE; Flu B NEGATIVE; RSV POSITIVE”.
Of the total 121 runs, two runs provided indeterminate GeneXpert results (2 ERROR). Both runs were repeated such that 21 negative and 20 positive replicates were tested for each virus strain. Under the conditions of this study, no evidence of specimen or amplicon carry-over contamination was observed in the GeneXpert Dx GX-IV module.
2. Comparison studies:
a. Method comparison with predicate device: N/A
b. Matrix comparison:
Clinical vs. simulated matrix comparison data was originally submitted for the Xpert Flu/RSV XC Assay in K142045.
c. Fresh vs. Frozen sample comparison:
{23}
Fresh vs. Frozen study data was originally submitted for the Xpert Flu/RSV XC Assay in K142045. The Flu/RSV XC Assay was used to determine the stability of the Xpert Flu+RSV Xpress Assay because all reagents, buffers and other components that comprise both assays are exactly the same. The detection method, the signal analysis, and algorithm used by the different software to calculate the results are the same.
# 3. Clinical studies:
A total of 12 sites, 10 CLIA waived and 2 High Complexity CLIA laboratories participated in the study. Five sites were emergency departments, five sites were walk-in clinics or student health centers, and two sites were hospital laboratories. The study was conducted from September 2014 to March 2015.
One nasopharyngeal swab was collected from each subject and placed into UTM after collection. This sample was used for immediate testing with Xpert Flu+ RSV Xpress Assay and the Comparator NAAT. Due to the low prevalence of Flu B and RSV during the sampling time frame of this study, supplementation with pre-selected archived NP swab specimens known to be positive for Flu B or RSV was required to meet the sample size for these targets. A total of 2435 NP swab specimens were tested for influenza A, influenza B, and RSV by the Xpert Flu+RSV Xpress Assay and the comparator assay. Of the 2435 NP swab specimens 2176 were fresh, prospectively collected and 259 were pre-selected frozen, archived specimens.
Of the Xpert Flu+RSV Xpress Assay runs performed with eligible specimens, $95.0\%$ (2335/2459) of these specimens were successful on the first attempt. The initial invalid rate was $5.0\%$ (95% CI 4.2-6.0%). Invalid results were returned on the first attempt for 124 specimens (121 NO RESULT-REPEAT TEST and 3 INSTRUMENT ERROR). One-hundred eighteen of the 124 specimens were retested, of which 107 yielded valid results after a single retest. There were 17 NP swab specimens with invalid results upon retest which were excluded from the analyses.
The tables below show the performance of the Xpert Flu+RSV Xpress Assay, the data are stratified by analyte, Influenza A, Influenza B, and RSV and by sample collection, prospective or retrospective.
Xpert Flu+RSV Xpress Assay Performance on NP Swab Specimens
| Specimen Type | Target | n | TP | FP | TN | FN | PPA % (95 CI) | NPA % (95 CI) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Fresh | Flu A | 2176 | 250 | 101a | 1825 | 0 | 100 (98.5-100) | 94.8 (93.7-95.7) |
| | Flu B | 2176 | 63 | 10b | 2103 | 0 | 100 (94.3-100) | 99.5 (99.1-99.8) |
| | RSV | 2176 | 125 | 8c | 2039 | 4d | 96.9 (92.3-99.1) | 99.6 (99.2-99.8) |
| | | | | | | | | |
| Pre-selected Frozen | Flu A | 259 | 0 | 4e | 255 | 0 | NA | 98.5 (96.1-99.6) |
| | Flu B | 259 | 100 | 2f | 157 | 0 | 100 | 98.7 |
{24}
25
| | | | | | | | (96.4-100) | (95.5-99.8) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | RSV | 259 | 40 | 1^{g} | 217 | 1^{h} | 97.6
(87.1-99.9) | 99.5
(97.5-100) |
a. Testing results by sequencing: 92 of 101 were Flu A positive; 8 of 101 failed to sequence; 1 of 101 had insufficient remaining volume for sequencing.
b. Testing results by sequencing: 9 of 10 were Flu B positive; 1 of 10 had insufficient remaining volume for sequencing.
c. Testing results by sequencing: 7 of 8 were RSV positive; 1 of 8 was RSV negative.
d. Testing results by sequencing: 3 of 4 were RSV positive; 1 of 4 was RSV negative.
e. Testing results by sequencing: 4 of 4 had insufficient remaining volume for sequencing.
f. Testing results by sequencing: 2 of 2 had insufficient remaining volume for sequencing.
g. Testing results by sequencing: 1 of 1 had insufficient remaining volume for sequencing.
h. Testing results by sequencing: 1 of 1 had insufficient remaining volume for sequencing.
## 4. Clinical cut-off:
Please refer to the “Baseline, Threshold, Cut-off” sections of this document.
## 5. Expected values/Reference range:
The Xpert Flu+RSV Xpress clinical study included a total of 2176 prospectively collected fresh NP swab specimens. The number and percentage of cases positive for one or more viruses of influenza A, influenza B, and RSV, as determined by the Xpert Flu+RSV Xpress Assay are shown by age category below.
Age Group Flu A, Flu B and RSV Positive by Xpert Flu+RSV Xpress Assay – NP Swabsᵃ
| Age Group | Number of Patients | Flu A | | Flu B | | RSV | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Number of Positives | Positivity | Number of Positives | Positivity | Number of Positives | Positivity |
| ≤5 years | 347 | 44 | 12.7% | 4 | 1.2% | 73 | 21.0% |
| 6-21 years | 382 | 90 | 23.6% | 8 | 2.1% | 15 | 3.9% |
| 22-59 years | 1222 | 175 | 14.3% | 53 | 4.3% | 33 | 2.7% |
| ≥60 years | 225 | 42 | 18.7% | 8 | 3.6% | 12 | 5.3% |
| Total | 2176 | 351 | 16.1% | 73 | 3.4% | 133 | 6.1% |
a. Five subjects had dual infections (Flu A & RSV) by Xpert Flu+RSV Xpress Assay (all were RSV POS only by comparator assay) and are therefore counted twice in this table.
## N. Instrument Name:
GeneXpert Xpress System using Cepheid GeneXpert Xpress software version 4.4 or higher.
{25}
26
O. System Descriptions:
The GeneXpert Xpress System automates and integrates sample purification, nucleic acid amplification, and detection of the target sequences in simple or complex samples using real-time PCR and RT-PCR. The system consists of a GeneXpert GX-I instrument (also referred to as GeneXpert I), touchscreen computer, and preloaded software for running the tests and viewing the results. The GeneXpert Xpress System requires the use of single-use disposable assay-specific cartridges (the Xpert Flu+RSV Xpress Assay cartridge) that hold the PCR reagents and host the PCR process. In this platform, sample preparation, amplification, and real-time detection are fully-automated and completely integrated.
The GeneXpert Xpress System platform is the same platform as the GeneXpert Dx System (model GX-I) that is used with Cepheid’s previously cleared Xpert assays (including K142025). The detection method, the signal analysis, and algorithm used by the systems to calculate the results are the same. The differences between the systems lie in the user interface of the system and the reporting of results: 1) the Xpress system has an ATM-like user interface and is simplified to utilize pictures and videos to guide the user through ordering and running a test, and 2) the test results for INVALID and ERROR results are simplified.
1. Modes of Operation:
The Xpert Flu+RSV Xpress Assay can be used with GeneXpert Xpress software version 4.4.2 and GeneXpert Dx software version 4.4. Directions for operating the system are included in the package insert.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
Level of Concern:
Moderate
Software Description:
The GeneXpert Xpress software utilizes the same GeneXpert module control (optics, thermal, fluidics, etc.), data collection and data analysis as the GeneXpert Dx software using the GeneXpert GX-I instrument (identical instrument as the GeneXpert Dx GX-I instrument). Two major differences are: 1) Xpress software user interface is simplified to utilize pictures and videos to guide the user through ordering and running a test, and 2) the test result interpretations for INVALID and ERROR results are simplified. The following table outlines the similarities and differences:
{26}
Comparison of GeneXpert Xpress Versions 4.4.2 and GeneXpert Dx 4.4 Software Component or Feature For Use With the GeneXpert Dx GX-I
| Component or Feature | GeneXpert Xpress version 4.4.2 | GeneXpert Dx version 4.4 |
| --- | --- | --- |
| Communication with Module | Yes - 1 module (for use only with the GeneXpert Dx GX-I) | Same (and for use with the GeneXpert Dx family of instruments) |
| Graphic User Interface (GUI) | Yes - Simplified user interface (touchpad and keyboard integrated laptop and touchscreen) adopted for use in a CLIA waived environment by untrained users | Yes - Interface (keyboard and mouse with laptop or desktop PC or keyboard integrated laptop) adopted for use in a CLIA moderate complexity environment by untrained or experienced users |
| Test Results | Same except the test results for INVALID and ERROR results are simplified (NO RESULT - REPEAT TEST or INSTRUMENT ERROR) | Same standard test results including INVALID, ERROR, and NO RESULT. |
| Reports | No | Yes (Individual, Patient, Specimen, and External Control) |
| Detailed Raw Data Results | No | Yes, according to user access privileges (Basic or Administrator user). Cycle threshold data is only visible to the user with Administrator privileges and access to data files. |
| Data Management | No | Yes (Archive of data and back up data) |
| Supports computer resolution | Screen resolution of 1366 x 768 | Screen resolution of 1366 x 768 or 1024 X 768 |
| Operating System | Windows 7 | Windows XP or Windows 7 |
| 2D Barcode Scanner Configuration | Yes (hand held) | Same |
| Assay Definition File (ADF) | Supports qualitative ADFs | Same |
| Handling of Lot Specific Parameters embedded in barcode | Yes | Same |
{27}
| Cartridge handling including loading, staging, removal and disposal | Manual | Same |
| --- | --- | --- |
| Fluidics handling | Yes | Same |
| Ultrasonic handling | Yes | Same |
| Thermal control | Yes | Same |
| Optical Detection - 6-color | Yes | Same |
| Data Acquisition and Analyses | Yes | Same |
| Data Storage | MS SQL 2005 database | Same |
| Internal Control data analysis | Yes | Same |
| External Control data analysis | Yes | Same |
| Viewing Test Results | Yes | Same |
| Laboratory Information System connectivity: uploading results | Yes | Same |
| Provide alerts for low memory and for modules nearing re-calibration | Yes | Same |
# 3. Specimen Identification:
Specimen identification is described in the "Test Principle" and "Baseline, Threshold, Cutoff" sections of this document.
# 4. Calibration:
N/A
# 6. Quality Control:
- Sample Processing Control (SPC): Ensures the sample was processed correctly.
The SPC is an Armored RNA® that is included in each cartridge to verify adequate processing of the sample. The SPC verifies that release of RNA from the Flu or RSV virus has occurred if the organism is present and verifies that the specimen processing is adequate. Additionally, this control detects specimen-associated inhibition of the RT-PCR and PCR reactions. The SPC should be positive in a negative sample and can be negative or positive in a positive sample. The SPC passes if it meets the validated acceptance criteria.
{28}
- Probe Check Control (PCC): Before the start of the PCR reaction, the GeneXpert Xpress System measures the fluorescence signal from the probes to monitor bead rehydration, reaction tube filling, probe integrity, and dye stability. The PCC passes if it meets the validated acceptance criteria.
P. Other Supportive Instrument Performance Characteristics Data Not Covered in the "Performance Characteristics" Section Above:
N/A
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
29
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.