Optilite C-Reactive Protein Reagent,Optilite C-Reactive Protein Calibrator,Optilite C-Reactive Protein Controls

K172868 · The Binding Site Group , Ltd. · DCN · Feb 28, 2018 · Immunology

Device Facts

Record IDK172868
Device NameOptilite C-Reactive Protein Reagent,Optilite C-Reactive Protein Calibrator,Optilite C-Reactive Protein Controls
ApplicantThe Binding Site Group , Ltd.
Product CodeDCN · Immunology
Decision DateFeb 28, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5270
Device ClassClass 2

Intended Use

The Optilite C-Reactive Protein Reagent is intended for the quantitative in vitro determination of C-reactive protein (CRP) concentration in serum using the Binding Site Optilite analyser. Measurement of C-Reactive Protein aids in evaluation of the amount of injurv to body tissues and for evaluation of infection, tissue injurv, and inflammatory disorders. This test should be used in conjunction with other laboratory and clinical findings. The Optilite C-Reactive Protein Calibrator is intended for the calibration of the Optilite C-Reactive Protein Reagent on the Optilite analyser. The Optilite C-Reactive Protein Controls are intended for use in quality control by monitoring accuracy and precision for the Optilite C-Reactive Protein Reagent.

Device Story

The Optilite C-Reactive Protein (CRP) system is an in vitro diagnostic test used in clinical laboratories. It utilizes serum samples to measure CRP concentration via immunoturbidimetry. The device consists of a reagent (antiserum and reaction buffer), calibrator, and controls, all designed for use on the Binding Site Optilite analyser. During operation, the sample reacts with specific anti-CRP antiserum to form insoluble complexes. The analyser passes light through the suspension and measures transmitted light via a photodiode; the intensity of transmitted light is inversely proportional to the CRP concentration. The system automatically calculates concentrations using a stored calibration curve. Healthcare providers use these results to assess tissue injury, infection, and inflammatory states in conjunction with other clinical findings. The system benefits patients by providing quantitative data to support diagnostic and monitoring decisions for inflammatory conditions.

Clinical Evidence

Bench testing only. Performance evaluated per CLSI guidelines (EP05-A3, EP06-A, EP17-A2, EP07-A2). Precision: total CV% 2.6–6.2%. Linearity: R=1.00 over 2.70–316.03 mg/L. LoQ: 5.00 mg/L. No interference observed for common endogenous substances or 14 therapeutic drugs. Method comparison (n=113) against predicate showed Passing-Bablok regression y = 1.00x + 5.56 (r=1.00). Reference interval verified against literature consensus (<5 mg/L).

Technological Characteristics

Immunoturbidimetric assay; goat anti-human CRP antiserum; reaction buffer (TRIS pH 7.5, 0.099% sodium azide). Photometric detection at 340 nm. Traceable to ERM-DA474. Automated processing on Optilite analyzer. Reagent stability: 5 months shelf-life, 3 months open-vial, 30 days on-board.

Indications for Use

Indicated for the quantitative determination of C-reactive protein (CRP) in serum to aid in the evaluation of tissue injury, infection, and inflammatory disorders in patients. Prescription use only.

Regulatory Classification

Identification

A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## Optilite C-Reactive Protein Reagent, Calibrator and Controls 510(k) Summary ## A. 510(k) Number: K172868 ### B. Purpose for Submission: New device #### C. Measurand: C-Reactive Protein #### D. Type of Test: Quantitative immunoturbidimetry #### E. Applicant: The Binding Site #### F. Proprietary and Established Names: Optilite® C-Reactive Protein Reagent, Optilite®C-Reactive Protein Calibrator, Optilite® C-Reactive Protein Controls. #### G. Regulatory Information: - 1. Requlation section: 21 CFR 866.5270, C-reactive protein immunological test system. - 2. Classification: Class II - 3. Product code: DCN, System, Test, C-Reactive Protein - 4. Panel: Immunology (82) {1}------------------------------------------------ #### H. Intended use: - 1. Intended use(s): The Optilite C-Reactive Protein Reagent is intended for the quantitative in vitro determination of C-reactive protein (CRP) concentration in serum using the Binding Site Optilite analyser. Measurement of C-Reactive Protein aids in evaluation of the amount of injurv to body tissues and for evaluation of infection, tissue injurv, and inflammatory disorders. This test should be used in conjunction with other laboratory and clinical findings. The Optilite C-Reactive Protein Calibrator is intended for the calibration of the Optilite C-Reactive Protein Reagent on the Optilite analyser. The Optilite C-Reactive Protein Controls are intended for use in quality control by monitoring accuracy and precision for the Optilite C-Reactive Protein Reagent. #### 2. Indication(s) for use: Same as Intended use(s). - 3. Special conditions for use statement(s): Prescription use only - 4. Special instrument requirements: The Binding Site Optilite analyser #### I. Device Description: The Optilite C-Reactive Protein Reagent is comprised of a dual wedge containing the following: Antiserum: Supplied in stabilised liquid form. Preservatives: 0.099% sodium azide, TRIS pH 8.0. Reaction Buffer: Containing 0.099% sodium azide, TRIS pH 7.5 as preservatives. The Optilite C-Reactive Protein Calibrator is comprised of the following: Pooled human serum, supplied in stabilised liquid form. Containing 0.099% sodium azide, as preservative. The Optilite C-Reactive Protein Controls are comprised of the following: Pooled human serum, supplied in stabilised liquid form. Containing 0.099% sodium azide, as preservative. #### J. Substantial equivalence information: - 1. Predicate device name(s) and 510(k) number(s): Roche Diagnostics Tina-Quant C-Reactive Protein Gen. 3 (K083444) {2}------------------------------------------------ #### 2. Comparison with predicate: #### Similarities | ltem | Device | Predicate | |--------------------|---------------------------------------------------------------------|--------------------------------------------| | Intended use | Quantitative in vitro determination of<br>C-reactive protein (CRP). | Same. | | Method | Turbidimetry | Same. | | Reference Interval | <5mg/L | Same. | | Sample type | Serum | Serum (also Li-heparin and EDTA<br>plasma) | #### Differences | Item | Device | Predicate | |---------------------|---------------------------------------------------|-----------------------------------------------------| | Measuring Range | 5-285mg/L (neat)<br>25-1425mg/L (1/5 dilution) | 0.3-350mg/L (neat)<br>0.6-700mg/L (1/2 dilution) | | Analyser | Optilite | Hitachi 912, 917, Modular P | | Antibody | anti- human CRP (goat) | anti-human CRP (mouse) | | Open vial stability | Three months at 2 to 8°C | Not stated. | | On board stability | 30 days. | 84 days opened and refrigerated<br>on the analyser. | | Reagent | Antiserum | Latex particles coated with anti-CRP<br>(mouse) | | Calibration | 6 point, single calibrator diluted on<br>analyser | 5 point, single calibrator diluted on<br>analyser | | Wavelength | 340nm | 570nm/800nm | | Traceability | DA474 | CRM470 | #### K. Standards and Guidance documents referenced: Guidance for Industry - Review Criteria for Assessment of C - Reactive Protein (CRP), High Sensitivity C-Reactive Protein (hsCRP) and Cardiac C-Reactive Protein (cCRP) Assays Guidance for Industry and FDA Staff CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition CLSI EP7-A2 Interference Testing in Clinical Chemistry, Approved Guideline - Second Edition CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach CLSI EP5-A3 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Third Edition {3}------------------------------------------------ ## L. Test Principle: The determination of soluble antigen (CRP) concentration by turbidimetric methods involves the reaction with specific antiserum (anti-CRP) to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument. ## M. Performance Characteristics (if/when applicable): #### 1. Analytical performance: #### Precision/Reproducibility: a. The studies were based on CLSI EP5-A3, where 4 sera samples (targeted to 6.8mg/L, 9.5mg/L, 21.5mg/L and 65.4mg/L) were tested in 2 runs per day (each of the 2 runs in duplicate) over 21 days using 1 reagent lot on 3 analysers. Results met the acceptance criteria for total precision (%CV<10%), within-run precision (%CV<5%), between-run precision (%CV<8%), and between-day precision (%CV<8%). | Sample | 1 | 2 | 3 | 4 | | |------------------------------|-----|------|------|------|------| | Mean concentration (mg/L) | 6.8 | 9.5 | 21.5 | 65.4 | | | Total precision | SD | 0.42 | 0.41 | 0.55 | 2.03 | | | %CV | 6.2 | 4.3 | 2.6 | 3.1 | | Within-run precision | SD | 0.18 | 0.12 | 0.22 | 0.49 | | | %CV | 2.7 | 1.2 | 1.0 | 0.7 | | Between-run precision | SD | 0.09 | 0.12 | 0.18 | 0.52 | | | %CV | 1.4 | 1.3 | 0.8 | 0.8 | | Between-day precision | SD | 0.37 | 0.37 | 0.47 | 1.90 | | | %CV | 5.4 | 3.9 | 2.2 | 2.9 | | Between-batch precision | SD | 0.06 | 0.41 | 1.67 | 4.24 | | | %CV | 0.6 | 1.9 | 2.6 | 2.9 | | Between-instrument precision | SD | 0.20 | 0.06 | 0.17 | 1.64 | | | %CV | 3.0 | 0.7 | 0.8 | 2.5 | ## b. Linearity/assay reportable range: The studies followed CLSI EP6-A, whereby linearity was assessed across the curve width at the standard sample dilution (1+0). The acceptance criteria were that the %CV for each sample should be ≤8% and the allowable nonlinearity was 0.5mg/L up to 5mg/L, and ± 10% above 5mg/L. In order to maintain matrix integrity, any spiking with pure protein was no greater than 30% of the total volume. {4}------------------------------------------------ A series of 14 samples was tested in 3 replicates. Results met the Acceptance criteria for CV<8%. | % High pool | Mean<br>Result | SD | %CV | |-------------|----------------|-------|------| | 100.0% | 316.03 | 0.231 | 0.1% | | 90.0% | 274.87 | 1.041 | 0.4% | | 80.0% | 254.07 | 1.595 | 0.6% | | 70.0% | 224.27 | 1.457 | 0.6% | | 60.0% | 193.40 | 1.706 | 0.9% | | 50.0% | 163.53 | 0.451 | 0.3% | | 40.0% | 138.37 | 0.764 | 0.6% | | 30.0% | 108.87 | 0.643 | 0.6% | | 20.0% | 73.37 | 0.321 | 0.4% | | 10.0% | 35.80 | 0.608 | 1.7% | | 2.5% | 9.97 | 0.058 | 0.6% | | 1.5% | 6.27 | 0.058 | 0.9% | | 1.0% | 4.93 | 0.058 | 1.2% | | 0.5% | 2.70 | 0.100 | 3.7% | - c. Traceability, Stability, Expected values (controls, calibrators, or methods): #### i) Traceability: The calibration of the assay is traceable to the international reference standard ERM-DA474. #### ii) Kit Stability: Real-time stability - A study to establish shelf-life stability (from the date of manufacture when stored at recommended temperature 2-8°C) of the Optilite CRP Kit is on-going. Currently available data supports a 5-month stability claim. Open-vial stability - The Optilite CRP Kit reagents can be stored, opened at 2 - 8°C for up to 3 months. On-board stability - The Optilite CRP Kit reagents can be stored on-board the Optilite Analyser for at least 30 days. #### d. Detection limit. The analytical sensitivity was determined in accordance with CLSI EP17-A2. The Limit of Blank (LoB) was based on 4 serum samples run 5 times per day, over 3 days, on 2 reagent lots to give a total of 60 results per lot. The LoB was estimated for each lot as the 95% percentile of the distribution. The Limit of Detection (LoD) was based on 4 serum samples run 5 times per day, over 3 days, on 2 reagent lots to give a total of 60 results per lot. The LoD calculation followed parametric analysis. {5}------------------------------------------------ The LoQ was determined from 4 serum samples targeted to be close to the bottom of the measuring range, tested 5 times per day over 3 days on 2 reagent lots. For LoB and LoD, the highest result obtained from the two lots tested was the final result (see below): | LoB | LoD | LoQ | |----------|----------|---------| | 1.35mg/L | 2.66mg/L | 5.0mg/L | e. Analytical specificity: Interferences were assessed according to CLSI EP7-A2 by testing serum samples with CRP concentrations falling at approximately 9mg/L, 60mg/L and 150mg/L. Each sample was spiked with interfering substances and tested in replicates of 3. For non-interference to be claimed, the mean results from the spiked samples must be within 10% of the mean of the control samples. Haemoglobin and Bilirubin results: The data demonstrated that the assay was not affected by high levels of the following substances: haemoglobin (5g/L) and bilirubin (200mg/L). Intralipid and Triglyceride results: The data demonstrated that the assay was not affected by triglyceride (500mg/dL), and intralipid (250mg/dL). Rheumatoid Factor results: The data demonstrated that the assay was not affected by rheumatoid factor (2417IU/mL). Drug Interference results: The data demonstrated that the assay was not affected by the 14 therapeutic drugs tested at the concentrations given below. | Drug | Concentration tested | |----------------------|----------------------| | Acetaminophen | 1324µmol/L | | Acetylsalicylic Acid | 3.63mmol/L | | Amoxicillin | 206µmol/L | | Ascorbic Acid | 342µmol/L | | Caffeine | 308µmol/L | | Cefotaxime | 673µmol/L | | Theophylline | 222µmol/L | | Chloramphenicol | 155µmol/L | | Cimetidine | 79.2µmol/L | | Digoxin | 7.8nmol/L | | Fluconazole | 245µmol/L | | Ibuprofen | 1212.5µmol/L | | Penicillin | 75mg/L | | Phenytoin | 198µmol/L | f. Assay cut-off: Not applicable - 2. Comparison studies: a. Method comparison with predicate device: {6}------------------------------------------------ A total of 193 serum samples spanning the dynamic range were assayed in singlicate by both the Optilite C-Reactive Protein Kit and the Tina-Quant C-Reactive Protein Gen. 3 assay for use on the Modular P analyser. The samples included 83 normal donors and 110 clinical samples. Regression statistics are based on the balance of the paired results. | Regression<br>fit | Regression<br>Equation | Slope<br>(95% CI) | Intercept<br>( 95% CI) | |--------------------|------------------------|-------------------|------------------------| | Weighted<br>Linear | Y = 1.07x + 1.16 | 1.05 to 1.09 | 0.84 to 1.48 | | Passing-<br>Bablok | Y = 1.00x + 5.56 | 0.98 to 1.03 | 4.10 to 6.87 | | Weighted<br>Deming | Y = 1.07x + 1.04 | 1.05 to 1.09 | 0.45 to 1.64 | 98.4% of samples tested gave clinically concordant results on both analysers. - b. Matrix comparison: None - 3. Clinical studies: - a. Clinical Sensitivity: Not applicable - b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable - 4. Clinical cut-off: Not applicable - 5. Expected values/Reference range: The reference interval was verified by testing 50 adult donor samples. Consensus reference interval for adults: <5 mg/L Reference: Dati F, Schumann G, Thomas L et al. Consensus of a group of professional societies and diagnostic companies on guidelines for interim reference ranges for 14 proteins in serum based on the standardization against the IFCC/BCR/CAP reference material (CRM 470). Eur J. Clin Chem Clin Biochem 1996;34:517-520. {7}------------------------------------------------ #### N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. #### O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. {8}------------------------------------------------ # Please wait... If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document. You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download. For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader. Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 28, 2018 The Binding Site Group Ltd. Kirsty Samuels Regulatory Affairs Officer 8 Calthorpe Road, Edgbaston Birmingham, B15 1QT Gb Re: K172868 Trade/Device Name: Optilite C-Reactive Protein Reagent.Optilite C-Reactive Protein Calibrator, Optilite C-Reactive Protein Controls Regulation Number: 21 CFR 866.5270 Regulation Name: C-reactive protein immunological test system Regulatory Class: Class II Product Code: DCN Dated: September 18, 2017 Received: September 20, 2017 Dear Kirsty Samuels: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR {10}------------------------------------------------ 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...