Optilite High Sensitivity C-Reactive Protein Kit

K180099 · The Binding Site Group , Ltd. · DCN · Oct 12, 2018 · Immunology

Device Facts

Record IDK180099
Device NameOptilite High Sensitivity C-Reactive Protein Kit
ApplicantThe Binding Site Group , Ltd.
Product CodeDCN · Immunology
Decision DateOct 12, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5270
Device ClassClass 2

Indications for Use

The Optilite High Sensitivity C-Reactive Protein Kit is intended for the quantitative in vitro measurement of C-Reactive Protein in serum using the Binding Site Optilite analyser for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. This test should be used in conjunction with other laboratory and clinical findings.

Device Story

The Optilite High Sensitivity C-Reactive Protein Kit is an in vitro diagnostic assay used on the Binding Site Optilite analyser. It measures C-Reactive Protein (CRP) in serum samples via latex-enhanced immunoturbidimetry. The device consists of latex reagent, calibrator, controls, and reaction buffer. In operation, the analyser mixes the sample with specific anti-CRP antibodies, forming insoluble complexes. Light transmission through the suspension is measured by a photodiode; the amount of transmitted light is inversely proportional to the CRP concentration. The analyser automatically calculates concentrations using a stored calibration curve. Used in clinical laboratory settings by trained personnel, the output provides quantitative CRP levels to assist clinicians in evaluating inflammatory conditions. The device benefits patients by providing standardized, high-sensitivity measurements to support clinical decision-making when used alongside other laboratory and clinical findings.

Clinical Evidence

Bench testing only. Performance validated via precision studies (CLSI EP5-A3) showing total precision %CV <10% across 5 samples (0.98-8.50 mg/L). Linearity confirmed (0.44-12.16 mg/L). Method comparison with predicate (n=391 samples, 302 within range) yielded Passing & Bablok regression y=0.99x+0.09 (95% CI slope 0.97-1.01). Analytical sensitivity: LoB 0.15 mg/L, LoD 0.17 mg/L, LoQ 0.5 mg/L. Interference testing (CLSI EP7-A2) showed no significant impact from hemoglobin, bilirubin, rheumatoid factor, lipids, or common therapeutic drugs.

Technological Characteristics

Latex-enhanced immunoturbidimetric assay. Reagents contain 0.099% sodium azide. Utilizes sheep anti-human CRP antibodies. Measuring range 0.5-10 mg/L. Calibration is 6-point, traceable to ERM-DA474. Designed for use on the Binding Site Optilite analyser. Standards referenced: CLSI EP17-A2, EP7-A2, EP6-A, EP5-A3.

Indications for Use

Indicated for the quantitative in vitro measurement of C-Reactive Protein in serum for the evaluation of conditions associated with inflammation in otherwise healthy individuals. Prescription use only.

Regulatory Classification

Identification

A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number k180099 B. Purpose for Submission: New Device C. Measurand: High Sensitivity C-Reactive Protein D. Type of Test: Quantitative Immunoturbidimetry Assay E. Applicant: The Binding Site Group Ltd. F. Proprietary and Established Names: Optilite® High Sensitivity C-Reactive Protein Kit G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | DCN | II | 866.5270, C-reactive protein immunological test system | 82- Immunology | H. Intended Use: 1. Intended use(s): See indications for use below. {1} 2. **Indication(s) for use:** The Optilite High Sensitivity C-Reactive Protein Kit is intended for the quantitative in vitro measurement of C-Reactive Protein in serum using the Binding Site Optilite analyser for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. This test should be used in conjunction with other laboratory and clinical findings. 3. **Special conditions for use statement(s):** For prescription use only. 4. **Special instrument requirements:** The Binding Site Optilite analyser I. **Device Description:** The Optilite High Sensitivity C-Reactive Protein Kit consists of the following: latex reagent, single calibrator, controls (high and low) and reaction buffer in liquid form. The reagents contain 0.099% sodium azide as preservative. J. **Substantial Equivalence Information:** 1. **Predicate device name(s) and 510(k) numbers:** C-Reactive Protein High Sensitive Test System for Cobas Integra Instruments, k053603. 2. **Comparison with predicate:** | Item | Candidate Device: Optilite® High Sensitivity C-Reactive Protein Kit on the Binding Site Optilite analyzer | Predicate Device: C-Reactive Protein High Sensitive Test System for Cobas Integra Instruments (k053603) | | --- | --- | --- | | Similarities | | | | Intended Use | Quantitative in vitro measurement of C-reactive protein (CRP) | Same | | Reagent Type | Latex enhanced | Same | | Differences | | | | Measuring range | 0.5-10mg/L | 0.1-20.0mg/L 1.5-300mg/L (1/15 dilution) | | Analyzer | Optilite | Integra | | Traceability | DA474 | ERM-DA470 (CRM470) | {2} 3 K. Standard/Guidance Document Referenced (if applicable): Guidance for Industry - Review Criteria for Assessment of C - Reactive Protein (CRP), High Sensitivity C-Reactive Protein (hsCRP) and Cardiac C-Reactive Protein (cCRP) Assays Guidance for Industry and FDA Staff. Clinical and Laboratory Standards Institute (CLSI) EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-Third Edition. CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. CLSI EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition. CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline. L. Test Principle: The determination of soluble antigen (CRP) concentration by turbidimetric methods involves the reaction with specific antiserum (anti-CRP) to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The precision studies were performed following the recommendations in the CLSI EP05-A3 guideline. The precision samples consisted of 5 serum samples at the following concentrations: 0.863 mg/L, 1.65 mg/L, 3.02 mg/L, 5.06 mg/L and 8.54 mg/L. The samples were run in duplicate with 2 runs per day, using one instruments and one reagent batch, over the course of 20 days. The results from a representative lot are summarized in the table below: {3} | Level | N | Mean (mg/L) | Within Run | | Between Run | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | | 1 | 80 | 0.863 | 0.017 | 2.0% | 0.010 | 1.2% | 0.020 | 2.3% | | 2 | 80 | 1.647 | 0.009 | 0.5% | 0.010 | 0.6% | 0.016 | 1.0% | | 3 | 80 | 3.019 | 0.022 | 0.7% | 0.014 | 0.4% | 0.036 | 1.2% | | 4 | 80 | 5.063 | 0.024 | 0.5% | 0.067 | 1.3% | 0.078 | 1.5% | | 5 | 80 | 8.546 | 0.109 | 1.3% | 0.000 | 0.0% | 0.131 | 1.5% | b. Linearity/assay reportable range: Linearity studies were performed following the recommendations in the CLSI EP06-A guideline. A high pool was prepared using a serum sample with high CRP activity. The low pool was pooled normal serum. The high and low pools were inter-diluted to achieve concentrations spanning $10\%$ beyond the limits of the claimed measuring range $(0.44\mathrm{mg / L}$ to $12.16\mathrm{mg / L})$ . The results were analyzed using weighted linear regression analysis and the results showed that the deviation from linearity did not exceed $10\%$ . The linearity studies support the claimed measuring range of 0.5 to $10\mathrm{mg / L}$ c. Traceability, Stability, Expected values (controls, calibrators, or methods): The assay is traceable to ERM-DA474, which is traceable to CRM 470. d. Detection limit: Detection limit studies were performed following the recommendations in the CLSI EP17-A2 guideline. Limit of blank (LoB): The LoB was carried out using 4 depleted serum pools and 3 reagent lots using 1 instrument. Samples were run five times each per day, over 3 days to give a total of 60 replicates per reagent lot. For each reagent lot, the LoB results were ranked from lowest to highest concentration, and the LoB was estimated to be the result equivalent to the concentration at the 95th percentile. Limit of detection (LoD): 4 serum sample pools were used to calculate the LoD using 3 reagent lots and 1 instrument. Samples were run five times each per day, over 3 days on each reagent lot. The LoD was calculated for each reagent lot using the parametric approach described in the guideline. Limit of quantitation (LoQ): 4 serum sample pools were used to calculate the LoQ using 3 reagent lots and 1 instrument. Samples were run five times each per day, over 3 days on each reagent lot. The LoQ was determined using the lowest concentration that met the defined performance goals of imprecision $(\% \mathrm{CV}) &lt; 20\%$ . {4} Based on the results of the studies, the sponsor claims the following detection limits: | LoB | LoD | LoQ | | --- | --- | --- | | 0.15 mg/L | 0.17 mg/L | 0.5 mg/L | # e. Analytical specificity: Interference studies were performed following the recommendations in the CLSI EP07-A2 guideline. 4 pool levels of serum samples (with CRP concentrations around 1, 2, 3 and $5\mathrm{mg / L}$ ) were used. Each sample was spiked with interfering substances and tested. Samples were tested in replicates of twenty at three analyte concentrations using a single lot of reagents. The results of the spiked samples were compared to the results of the non-spiked samples. At the following concentrations, the percentage difference between the results of the spiked samples compared to the results of the non-spiked samples was less than $\pm 10\%$ . | Potential Interferent | Highest Concentration Without Significant Interference | | --- | --- | | Hemoglobin | 5 g/L | | Intralipid | 2000 mg/dL | | Ibuprofen | 2425 μmol/L | | Triglyceride | 1000 mg/dL | | Warfarin | 32.5 μmol/L | | Acetaminophen | 1324 μmol/L | | Acetylsalicylic Acid | 3.62 mmol/L | | Conjugated Bilirubin | 200 mg/L | | Caffeine | 308 μmol/L | | Enalapril Maleate | 0.86 μmol/L | | Unconjugated Bilirubin | 200 mg/L | | Rheumatoid Factor | 25 IU/mL | The sponsor includes the following information in the package insert. "Turbidimetric assays are not suitable for measurement of highly lipoemic or haemolysed samples or samples containing high levels of circulating immune complexes (CICs) due to the unpredictable degree of non-specific scatter these sample types may generate. Unexpected results should be confirmed using an alternative assay method." {5} "Samples at selected CRP concentrations were spiked with decreasing concentrations of rheumatoid factor. The following results were obtained. | RF IU/mL | 0.8mg/L CRP | | 1.5mg/L CRP | | 2.6mg/L CRP | | | --- | --- | --- | --- | --- | --- | --- | | | mg/L difference | % difference | mg/L difference | % difference | mg/L difference | % difference | | 500 | 0.338 | 40.44% | 0.394 | 26.70% | 0.345 | 13.27% | | 250 | 0.484 | 62.50% | 0.432 | 27.77% | 0.37 | 13.79% | | 100 | 0.162 | 18.80% | 0.197 | 12.02% | 0.183 | 6.33% | | 50 | 0.145 | 17.40% | 0.103 | 6.01% | | | | 25 | 0.043 | 4.33% | | | | | A sample at 3.8mg/L CRP showed &lt;10% interference by 500 IU/mL RF. Any sample with an RF concentration greater than 25 IU/mL should be re-tested with an alternative method." ## CRP prozone study: A study was performed to verify the CRP prozone performance on the Optilite analyzer. 28 samples with CRP concentrations ranging from 0.8, to 1161.6 mg/L were evaluated. The study demonstrated that up to 829.3 mg/L the results were reported as above the measuring range and flagged as being out of the measuring range (&gt;10 mg/L). However, due to antigen excess, samples with CRP concentrations above 893 mg/L will be reported erroneously low with a prozone flag. For example, a sample with a CRP concentration of 1161 mg/L is reported as 91.3 mg/L with a prozone flag. The sponsor includes the following limitations in the labeling regarding the prozone effect: "The reportable range of this assay is approximately 0.5 – 10mg/L, depending on lot-specific calibrator values. Any result obtained above this range must be considered invalid and the sample should be re-tested using a conventional CRP assay. For example, in a study samples between 893mg/L and 1161 mg/L returned falsely low concentrations (99.46 mg/L and 91.32 mg/L), however antigen excess was correctly detected by the analyser and results were accompanied by a "High activity" flag in all cases" "Potential occurrences of antigen excess cannot be completely excluded. Samples detected as being in excess are flagged as "High activity" and are automatically remeasured at a higher sample dilution to confirm antigen excess. Any sample with a concentration above 10 mg/L should be considered invalid and re-tested with a conventional CRP assay." {6} f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: A method comparison study was performed using a total of 391 human serum samples. All samples were tested in singlicate using both the candidate and predicate device. 298 samples were within the measuring range of the candidate device. Data analysis was performed by Passing Bablok. The regression analysis is summarized in the table below | N | Slope | Intercept | 95% CI Slope | 95% CI Intercept | | --- | --- | --- | --- | --- | | 298 | 0.99 | 0.09 | 0.97-1.01 | 0.05-0.11 | b. Matrix comparison: Not applicable. The assay is only intended to be used with serum samples. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. {7} 5. Expected values/Reference interval: The sponsor performed studies following the CLSI EP28-A3c guideline to verify the reference range of the analyte. A total of 50 serum samples were tested to verify the CRP reference interval. The verification studies support the following reference interval¹. Consensus reference interval for adults: &lt;3 mg/L ¹Macy EM, Hayes TE, Tracy RP. Variability in the measurement of C-Reactive protein in healthy subjects; implications for reference intervals and epidemiological applications. Clin Chem 1997; 43: 52-8. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 8
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...