SYNTHETIC POWDER FREE (YELLOW) VINYL PATIENT EXAMINATION GLOVES

K992861 · Shijiazhuang Great Eagle Plastic Products Co., Ltd. · LYZ · Oct 6, 1999 · General Hospital

Device Facts

Record IDK992861
Device NameSYNTHETIC POWDER FREE (YELLOW) VINYL PATIENT EXAMINATION GLOVES
ApplicantShijiazhuang Great Eagle Plastic Products Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateOct 6, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

Device Story

Synthetic powder-free vinyl patient examination gloves; disposable; worn on examiner's hand or finger. Used in clinical settings to provide a barrier against contamination between healthcare personnel and patients. Manufactured to meet ASTM D5250-92 standards. Device is non-sterile and intended for over-the-counter use.

Clinical Evidence

Bench testing only. Testing included physical and dimension measurements (ASTM D5250-92, AQL 4.0), FDA 1000 ml water fill test (AQL 2.5), and biocompatibility testing (primary skin irritation and sensitization). USP Iodine test confirmed powder-free status.

Technological Characteristics

Material: Synthetic vinyl. Standard: ASTM D5250-92. Form factor: Disposable patient examination glove. Powder-free (verified via USP Iodine test).

Indications for Use

Indicated for use as a disposable medical device worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: K99286 ## 1. Submitter's Identification: Mr. Fang Yi Liu Shijiazhuang Great Eagle Plastics Products Co., Ltd. 685 He Ping Road New Development Zone Shijiazhuang, Hebei Province 050031 P.R. China Date Summary Prepared: Aug 5, 1999 #### Name of the Device: 2. Shijiazhuang Great Eagle Plastics Products Co., Ltd. Synthetic Powder-Free (Yellow) Vinyl Patient Examination Gloves ## 3. Device Description: Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free (yellow) Vinyl Patient Examination Glove, 80LYZ, and meets all requirements of ASTM Standard D5250-92. #### 4. Intended Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment. ## 5. Comparison to Predicate Devices: Shijiazhuang Great Eagle Plastics Products Co., Ltd. Synthetic Powder-Free Vinyl Patient Examination Gloves, is substantially equivalent in safety and effectiveness to the Sunmax Enterprise Shanghai Co., Ltd. Powder-Free Vinyl Patient Examination Gloves and the Cheer & Merit Powder-Free Vinyl Patient Examination Gloves. {1}------------------------------------------------ ## 6. Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence are as follows: The standards used for Shijiazhuang Great Eagle Plastics Products Co., Ltd. glove production are based on ASTM-D-5250-92. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0. The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, Inspection Level S-4, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels. A USP Iodine Test for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free' claims. We adhere to all USP Iodine test methodogy and testing conducted revealed passing results. ## 7. Discussion of Clinical Tests Performed: Not Applicable - There is no hypoallergenic Claim. ## 8. Conclusions: Shijiazhuang Great Eagle Plastics Co., Ltd. Synthetic Vinyl Patient Examination gloves (Powder-Free) conform fully to ASTM-D-5250-92 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data submitted to FDA. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and recognizable, conveying the department's authority and commitment to public service. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 6 1999 Shijiazhuang Great Eagle Plastics Products Co., Ltd. C/O Ms. Susan D. Goldstein-Falk Official Correspondent Basic Medical Industries, Incorporated 12390 East End Avenue Chino, California 91710 Re : K992861 Synthetic Powder-Free (Yellow) Vinyl Patient Trade Name: Examination Gloves Regulatory Class: I Product Code: LYZ Dated: August 23, 1999 Received: August 23, 1999 Dear Ms. Susan D. Goldstein-Falk We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your {3}------------------------------------------------ Page 2 - Ms. Susan D. Goldstein-Falk premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Attachment A Page 1 of 1 K992861 DEVICE NAME: INDICATIONS FOR USE: 510(k) Number (if known): Shijiazhuang Great Eagle Plastics Products Co., Ltd. Synthetic Powder-Free (Yellow) Vinyl Patient Examination Gloves-Powder-Free A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | | | |--------------------------------------------------------|----------------------------------------|--------------------------|---| | <div><img alt="signature" src="signature.png"/></div> | | | | | (Division Sign-Off) | | | | | | Division of Dental, Infection Control, | | | | | and General Hospital Devices | | | | 510(k) Number | K992861 | | | | Prescription Use | | Over-The-Counter Use | X | | (Per 21 CFR 801.109) | OR | (Optional Format 1-2-96) | |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...