DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES, WHITE COLOR

K081535 · Rich Mountain Medical Products, Inc. · LYZ · Aug 14, 2008 · General Hospital

Device Facts

Record IDK081535
Device NameDISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES, WHITE COLOR
ApplicantRich Mountain Medical Products, Inc.
Product CodeLYZ · General Hospital
Decision DateAug 14, 2008
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Disposable powder-free vinyl synthetic examination glove; worn by healthcare examiners on hands or fingers. Acts as a physical barrier to prevent contamination between patient and examiner. Used in clinical settings. Device is non-sterile and intended for over-the-counter use.

Clinical Evidence

No clinical data. Bench testing only: physical properties and dimensions (ASTM D5250-06), watertight/pinhole testing (ASTM D5151-06, AQL 2.5), residual powder testing (ASTM D6124-06, <2mg/glove), and biocompatibility (primary skin irritation and sensitization).

Technological Characteristics

Vinyl synthetic material; powder-free. Complies with ASTM D5250-06. Testing includes physical dimensions, watertight integrity (ASTM D5151-06), and residual powder content (ASTM D6124-06).

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hand or fingers to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K081535 # REVISED 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS AUG 1 4 2008 This summary of 510(K) safety and effectiveness information is being submitted in Accordance with requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(K) number is: ## 1. Submitter's Identification: Michelle Shih Rich Mountain Medical Products Inc. No. 8, Jiang Yin Road, Yitang Town Pizhou County, Jiangsu Province China Date Summary: April 25, 2008 # 2. Name of the Device: Rich Mountain Medical Products Inc. Disposable Powder Free Vinyl Synthetic Exam Gloves, White Color ### 3. Predicate Device Information: #### MDM ENTERPRISE CORP. Disposable Powder Free Vinyl Synthetic Examination Glove, White Color (K033835) ## 4. Device description: Classified by FDA's General and Plastic Surgery Device panel as Class 1, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirements of ASTM Standard D5250-06. # 5. Intended Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. {1}------------------------------------------------ # 6.Comparison to Predicate Devices: Rich Mountain Medical Products Inc's Disposable Powder Free Vinyl Synthetic Exam Gloves, White Color is substantially equivalent to the device manufactured by MDM ENTERPRISE CORP. (K033835). # 7. Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence are as follows: The standards used for Rich Mountain Medical Products Inc's glove production are Based on ASTM D5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection level S-2, AQL 4.0. The FDA 1000 ml Watertight Test based on ASTM D-5151-06 was also conducted with samplings of AQL 2.5, Inspection level 1, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels. A residual Powder test that based on ASTM D6124-06 for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per gloves) #### 8. Discussion of Clinic Tests Performed: Not applicable - There is no Hypoallergenic Claim. #### 9. Conclusions: 彩 Rich Mountain Medical Products Inc's Disposable Powder Free Vinyl Synthetic Exam Gloves, White Color conform fully to ASTM D-5250-06 standard as well as applicable 21CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # AUG 1 4 2008 Ms. Michelle Shih Coordinator Rich Mountain Medical Products Incorporated No. 8, Jiang Yin Road, Yitang Town Pizhou County, Jiangsu CHINA Re: K081535 Trade/Device Name: Disposable Powder Free Vinyl Synthetic Exam Glove, White Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LYZ Dated: July 31, 2008 Received: August 4, 2008 Dear Ms. Shih: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Ms. Shih Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Clues Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use #### RICH MOUNTIAN MEDICAL PRODUCTS INC. Applicant: 510(k) Number (if known): APPLIED Device Name: Disposable Powder Free Vinyl Synthetic Exam Glove, White Color Indications for Use: A PATIENT EXAMINATION GLOVES IS A DISPOSABLE DEVICE INTENDED FOR MEDICAL PURPOSE THAT IS WORN ON THE EXAMINER'S HAND OR FINGERS TO PREVENT CONTAMINATION BETWEEN PATIENT AND EXAMINER. Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Shil A. Mungur Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign Off) ్లాల (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K081535 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...