Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim

K233405 · Eco Medi Glove Sdn. Bhd. · LZA · Dec 28, 2023 · General Hospital

Device Facts

Record IDK233405
Device NameBlue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim
ApplicantEco Medi Glove Sdn. Bhd.
Product CodeLZA · General Hospital
Decision DateDec 28, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim is a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use. In addition, these gloves were tested for use with Chemotherapy drugs and Fentanyl Test claim in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs and Fentanyl Test claim.

Device Story

Disposable, non-sterile, powder-free nitrile examination glove; blue color; ambidextrous; finger-textured surface; rolled beading cuff. Device worn by examiners to prevent cross-contamination between patient and examiner. Tested for permeation resistance against 15 chemotherapy drugs and fentanyl citrate per ASTM D6978-05. Output is a physical barrier; healthcare providers use the glove during patient examinations or handling of chemotherapy agents. Benefits include protection against biological contaminants and specific chemical permeation. Warning: Not for use with Carmustine or Thiotepa.

Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing (ISO 10993-23, ISO 10993-10, ISO 10993-11) confirmed the device is a non-irritant, non-sensitizer, and does not induce acute systemic toxicity. Permeation testing per ASTM D6978-05 performed for 16 substances.

Technological Characteristics

Material: Nitrile latex compound. Design: Ambidextrous, finger-textured, rolled beading. Dimensions: 230mm length, 0.05mm min thickness (palm/finger). Standards: ASTM D6319-19 (physical properties), ASTM D5151-19 (pinholes), ASTM D6124-06 (residual powder), ASTM D6978-05 (chemotherapy/fentanyl permeation). Biocompatibility: ISO 10993 series. Non-sterile.

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent contamination between patient and examiner. Suitable for over-the-counter use. Tested for permeation resistance against specific chemotherapy drugs and fentanyl citrate.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION December 28, 2023 Eco Medi Glove Sdn. Bhd. Suresh Kumar Official Correspondent Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate Taiping, Perak Darul Ridzuan 34600 Malaysia Re: K233405 Trade/Device Name: Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, QDO, OPJ Dated: October 2, 2023 Received: October 5, 2023 Dear Suresh Kumar: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K233405 - Suresh Kumar Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K233405 - Suresh Kumar Page 3 Sincerely, Bifeng Qian -S Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K233405 Device Name Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim Indications for Use (Describe) Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim is a disposable device intended for medical purposes that are worn on the examiner's hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use. In addition, these gloves were tested for use with Chemotherapy drugs and Fentanyl Test claim in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs and Fentanyl Test claim Chemotherapy Drug and Fentanyl Test Claim Concentration Minimum Breakthrough detection time in Minutes, ug/cm2/minute 1) Carmustine (3.3mg/ml or 3000ppm), 24.0 minutes 2) Cyclophosphamide (20mg/ml or 20,000ppm), >240 minutes 3) Cisplatin (1.0mg/ml or 1,000ppm), >240 minutes 4) Doxorubicin Hydrochloride (2.0mg/ml or 2000ppm), >240 minutes 5) Etoposide (20mg/ml or 20,000ppm), >240 minutes 6) Flourouracil (50mg/ml or 50,000), >240 minutes 7) Methorexate (25mg/ml or 25,000ppm), >240 minutes 8) Paclitaxel (6mg/ml or 6,000ppm), >240 minutes 9) Thiotepa (10mg/ml or 10,000ppm), 56.9 minutes 10) Ifosfamide (50mg/ml) >240 minutes 11) Mitoxantrone (2mg/ml), >240 minutes 12) Vincristine Sulfate (1mg/ml), >240 minutes 13) Carboplatin (10.0mg/ml), >240 minutes 14) Dacarbazine (DTIC) (10.0mg/ml) >240 minutes 15) Mitomicin C (0.5mg/ml) >240 minutes 16) Fentanyl Citrate Injection 100mg/2ml >240 minutes The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time. Carmustine (BCNU) (3.3mg/ml)- Minimum Breakthrough detection time 24.0 ug/cm2/minute. Thiotepa (10ug/ml) - Minimum Breakthrough detection time 56.9 ug/cm2/minute. WARNING: Not for use with Carmustine and ThioTepa Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. {4} This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5} ECO Medi Glove Sdn. Bhd. (815262-D) Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA Plant 2 : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA E-mail : sales@ecomediglove.com.my Website: www.ecomediglove.com.my TEL +60-5-806 2316 FAX +60-5-895 7668 SST NO. : A11-1808-21015730 510(K) Summary K233405 Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim 1.0 Submitter: Company Name : ECO Medi Glove Sdn. Bhd. Company Address : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping, Perak Darul Ridzuan, Malaysia. Contact Person : Mr. Suresh Kumar Telephone : +605-806 2316 Fax : +605-806 2315 Email : qa1@riverstone.com.my 2.0 Preparation Date : 25th December 2023 3.0 Name of the Device Trade Name / Proprietary Name: Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim Device Name: Nitrile Patient Examination gloves. Device Classification Name: Non-Powdered Patient Examination gloves (21 CFR 880.6250). Device Class: Class I. Product Code: LZA, LZC, OPJ, and QDO Page 1 of 9 {6} ECO Medi Glove Sdn. Bhd. (815262-D) Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA Plant 2 : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my TEL +60-5-806 2316 FAX +60-5-895 7668 SST NO.: A11-1808-21015730 ## 4.0 Identification of The Legally Marketed Device: Predicate Device: K171339, EMG Blue Nitrile Examination Glove Powder Free with Chemotherapy Drugs. ## 5.0 Device Description The subject device in this 510(k) Notification is EMG Blue Powder Free Nitrile Examination Gloves with Chemotherapy Drugs and Fentanyl Test Claim. The subject device is a patient examination glove made from nitrile latex compound, blue color, powder-free, and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-19, the Standard Specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs and fentanyl test claims. The Blue Nitrile Medical Examination Gloves, Powder Free, is a disposable device intended for medical purposes that are worn on the examiner's hand or finger to prevent contamination between patient and examiner (product Code LZA) and are used with chemotherapy drugs (Product code LZC and OPJ) and fentanyl test (Product code QDO). Page 2 of 9 {7} ECO Medi Glove Sdn. Bhd. (815262-D) Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my TEL +60-5-806 2316 FAX +60-5-895 7668 SST NO.: A11-1808-21015730 ## 6.0 Intended Use/Indications for Use Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim is a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use. In addition, these gloves were tested for use with Chemotherapy drugs and Fentanyl Test claim in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs and Fentanyl Test claim. | | Chemotherapy Drug and Fentanyl Test Claim Concentration | Minimum Breakthrough detection time in Minutes, ug/cm²/minute | | --- | --- | --- | | 1 | Carboplatin (10mg/ml or 10,000ppm) | >240 minutes | | 2 | Carmustine (3.3mg/ml or 3000ppm), | 24.0 minutes | | 3 | Cisplatin (1.0mg/ml or 1,000ppm), | >240 minutes | | 4 | Cyclophosphamide (20mg/ml or 20,000ppm), | >240 minutes | | 5 | Dacarbazine (DTIC) (10.0mg/ml) | >240 minutes | | 6 | Doxorubicin Hydrochloride (2.0mg/ml or 2000ppm), | >240 minutes | | 7 | Etoposide (20mg/ml or 20,000ppm), | >240 minutes | | 8 | Fluorouracil (50mg/ml or 50,000), | >240 minutes | | 9 | Ifosfamide (50mg/ml) | >240 minutes | | 10 | Methorexate (25mg/ml or 25,000ppm), | >240 minutes | | 11 | Mitomycin (0.5mg/ml or 500ppm) | >240 minutes | | 12 | Mitoxantrone (2mg/ml or 2,000ppm), | >240 minutes | | 13 | Paclitaxel (6mg/ml or 6,000ppm), | >240 minutes | | 14 | Thiotepa (10mg/ml or 10,000ppm), | 56.9 minutes | | 15 | Vincristine Sulfate (1mg/ml), | >240 minutes | | 16 | Fentanyl Citrate Injection 100mg/2ml | >240 minutes | The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time. Carmustine (BCNU) (3.3mg/ml)- Minimum Breakthrough detection time 24.0 ug/cm²/minute. Thiotepa (10ug/ml) - Minimum Breakthrough detection time 56.9 ug/cm²/minute. WARNING: Not for use with Carmustine and ThioTepa Page 3 of 9 {8} ECO Medi Glove Sdn. Bhd. (815262-D) Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA Plant 2 : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my TEL +60-5-806 2316 FAX +60-5-895 7668 SST NO.: A11-1808-21015730 ## 7.0 Specification for the subject Nitrile gloves: ## 7.1 Dimension and Thickness of Gloves | Dimension | Size XXS | Size XS | Size S | Size M | Size L | Size XL | Size XXL | Size XXXL | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Overall Length (mm) | 230min | 230min | 230min | 230min | 230min | 230min | 230min | 230min | | Width (± 10mm) | 65 | 75 | 85 | 95 | 105 | 115 | 125 | 135 | | Thickness at Palm (mm) | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | | Thickness at Finger Tip (mm) | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | ## 7.2 Gloves Physical Properties and Holes | Measurement | Before Ageing | After Aging at 70°C for 168 hrs @ 100°C for 22 hrs | | --- | --- | --- | | Tensile Strength (MPa) | 14min | 14 Min | | Ultimate Elongation (%) | 500min | 400min | | Pin-hole Level | AQL 2.5 Inspection Level G-1 | AQL 2.5 Inspection Level G-1 | Gloves meet all the specifications listed in ASTM D 6319-19 Page 4 of 9 {9} ECO Medi Glove Sdn. Bhd. (815262-D) Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my TEL +60-5-806 2316 FAX +60-5-895 7668 SST NO.: A11-1808-21015730 # Technological Characteristics Comparison of the Proposed and Predicate Devices | Characteristics | Acceptance Criteria | Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim K233405 | Predicate Device EMG Blue Nitrile Examination Glove Powder-Free with Chemotherapy Drugs K171339 | Assessment Similarities and Differences | | --- | --- | --- | --- | --- | | Product Code | | LZA, LZC, OPJ, and QDO | LZA, LZC | Similar | | Intended use/Indications for Use | | A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use. | A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use. | same | | Material use | Nitrile compound | Nitrile compound | Nitrile compound | Same | | Color | | Blue | Blue | Same | | Sterility | | Non sterile | Non sterile | Same | | Surface | Finger Textured | Finger Textured | Finger Textured | Same | | Cuffing Beading | Rolled Beading | Rolled Beading | Rolled Beading | Same | | Design | Ambidextrous | Ambidextrous | Ambidextrous | Same | | Dimensions | Overall Length (mm) Min 230mm Width (± 10mm) Size XXS = 65mm Size XS = 75mm Size S = 85mm Size M = 95mm Size L = 105mm Size XL = 115mm Size XXL=125mm Size XXXL=135mm Thickness at Palm (mm) Min; 0.05 mm Thickness at Finger Tip (mm) Min 0.05 mm | Overall Length (mm) Min 230mm Width (± 10mm) Size XXS = 65mm Size XS = 75mm Size S = 85mm Size M = 95mm Size L = 105mm Size XL = 115mm Size XXL=125mm Size XXXL=135mm Thickness at Palm (mm) Min; 0.05 mm Thickness at Finger Tip (mm) Min 0.05 mm | Overall Length (mm) Min 230 mm Width (± 5mm) Size S = 85mm Size M = 95mm Size L = 105mm Size XL = 115mm Thickness at Palm (mm) Min; 0.05 mm Thickness at Finger Tip (mm) Min 0.05 mm | Similar additional sizes were added to the subject device | Page 5 of 9 {10} ECO Medi Glove Sdn. Bhd. (815262-D) Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA Plant 2 : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my TEL +60-5-806 2316 FAX +60-5-895 7668 SST NO.: A11-1808-21015730 | Characteristics | Acceptance Criteria | Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim, | Predicate Device EMG Blue Nitrile Examination Glove Powder-Free with Chemotherapy Drugs K171339 | Assessment Similarities and Differences | | --- | --- | --- | --- | --- | | Physical properties | **Before Ageing** Tensile Strength (MPa) = 14min Ultimate Elongation (%) = 500min **After Aging at 70°C for 168 hrs @ 100°C for 22 hrs** Tensile Strength (MPa) = 14min Ultimate Elongation (%) = 400min | Meets ASTM D6319-19 | Meets ASTM D6319-19 | same | | Freedom from pinholes | AQL 2.5 Inspection Level G-1 | Meets ASTM D5151-19 | Meets ASTM D5151-19 | same | | Residual Powder | ≤ 2.0 mg/pc | Meets ASTM D6124-06 | Meets ASTM D6124-06 | same | | Biocompatibility | ISO 10993-23: 2021 Biological evaluation of medical devices- Part 23: Test for irritation | Under the conditions of this study, the test article was a non-irritant. | Under the conditions of this study, the test article was a non-irritant. | same | | | ISO 10993-10: 2021 Biological Evaluation on Medical Device – Part 10: Test for Skin Sensitization | Under the conditions of this study, the test article was a non-sensitizer. | Under the conditions of this study, the test article was a non-sensitizer. | same | | | ISO 10993-11 Biological evaluation on medical device Part 11 – Test for systemic toxicity | Under the conditions of this study, not induce any acute systemic toxicity | Under the conditions of this study, not induce any acute systemic toxicity | Same | Page 6 of 9 {11} ECO Medi Glove Sdn. Bhd. (815262-D) Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my TEL +60-5-806 2316 FAX +60-5-895 7668 SST NO.: A11-1808-21015730 | Characteristics | Acceptance Criteria | Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim, | Predicate Device EMG Blue Nitrile Examination Glove Powder-Free with Chemotherapy Drugs K171339 | Assessment Similarities and Differences | | --- | --- | --- | --- | --- | | Resistance against Chemotherapy Drugs | Standards Practice for Assessment of resistance of Medical Glove to Permeation by Chemotherapy drugs ASM D6978-05 | 1) Carboplatin (10mg/ml or 10,000ppm) >240 minutes 2) Carmustine (3.3mg/ml or 3000ppm), 24.0 minutes 3) Cisplatin (1.0mg/ml or 1,000ppm), >240 minutes 4) Cyclophosphamide (20mg/ml or 20,000ppm), >240 minutes 5) Dacarbazine (DTIC) (10.0mg/ml) >240 minutes 6) Doxorubicin Hydrochloride (2.0mg/ml or 2000ppm), >240 minutes 7) Etoposide (20mg/ml or 20,000ppm), >240 minutes 8) Fluorouracil (50mg/ml or 50,000), >240 minutes 9) Ifosfamide (50mg/ml) >240 minutes 10) Methorexate (25mg/ml or 25,000ppm), >240 minutes 11) Mitomicin (0.5mg/ml or 500ppm) >240 minutes 12) Mitoxantrone (2mg/ml or 2,000ppm), >240 minutes 13) Paclitaxel (6mg/ml or 6,000ppm), >240 minutes 14) Thiotepa (10mg/ml or 10,000ppm), 56.9 minutes 15) Vincristine Sulfate (1mg/ml), >240 minutes 16) Fentanyl Citrate Injection (100mg/2ml) >240 minutes Warning: Not for use with Carmustine or Thio Tepa | 1) Carmustine (BCNU) 3.3mg/ml 24.0 minutes 2) Cisplatin 1.0mg/mL, >240 minutes 3) Cyclophosphamide (Cytoxan) (20mg/ml) >240 minutes 4) Dacarbazine (10mg/ml) >240 minutes 5) Doxorubicin HCI (2.0mg/ml) >240 minutes 6) Etoposide (20.0mg/ml >240 minutes 7) Fluorouracil (50.0mg/ml) >240 minutes 8) Methotrexate (25mg/ml) >240 minutes 9) Paclitaxel (6.0mg/ml) >240 minutes 10) ThioTepa (10.0mg/ml) 56.9 minutes 11) Vincristine Sulfate 1.0mg/ml, >240 minutes 12) Ifosfamide (50mg/ml) 13) Mitoxantrone (2mg/ml or 2,000ppm), 14) Carboplatin (10mg/ml or 10,000ppm) 15) Vincristine Sulfate (1mg/ml), Warning: Not for use with Carmustine or Thio Tepa | Similar. Additional Fentanyl test is conducted for Device sample | Page 7 of 9 {12} ECO Medi Glove Sdn. Bhd. (815262-D) Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my TEL +60-5-806 2316 FAX +60-5-895 7668 SST NO.: A11-1808-21015730 # 8.0 Summary of Non-Clinical Performance test data | Characteristics | Test Standard | Acceptance Criteria | Test Result | Remark | | --- | --- | --- | --- | --- | | Freedom from Pin holes | ASTM D5151 -19 (Re-approved 2011) | AQL 2.5 Inspection Level G-1 | Meets ASTM D5151-19 | Per Lot Size 500,000pcs | | Dimensions | ASTM D6319 -19 | Length Min 230mm | Meets ASTM D6319-19 | Lot Size 500,000pcs | | | ASTM D6319 -19 | Width Size XXS = 65mm±10 Size XS = 75mm±10 Size S = 85mm±10 Size M = 95mm±10 Size L = 105mm±10 Size XL = 115mm±10 SizeXXL =125mm±10 Size XXXL = 135mm±10 | ISO 2859-1 / S2/AQL 4.0 | | | | ASTM D6319 -19 | Thickness at Palm (mm) Min; 0.05 mm Thickness at Finger Tip (mm) Min 0.05 mm | ISO 2859-1 / S2/AQL 4.0 | | | Physical properties | ASTM D6319 -19 ASTM D412-16(2021) | Before Ageing Tensile Strength (MPa) = 14min Ultimate Elongation (%) = 500min | Meets ASTM D6319-19 | | | | ASTM D6319 -19 and ASTM D412-06(2021) | After Aging at 70°C for 168 hrs @ 100°C for 22 hrs Tensile Strength (MPa) = 14min Ultimate Elongation (%) = 400min | | | | Powder-free residue | ASTM D6124-06 | ≤ 2.0 mg/pc | Meets ASTM D6124-06 (2022) | Once every PO | | Biocompatibility | ISO 10993-23: 2021 Biological evaluation of medical devices- Part 23: Test for irritation | The Device shall be non-irritant. | Under the conditions of this study, the test article was a non-irritant | As per Device with same color | | | ISO 10993-10: 2021 Biological Evaluation on Medical Device – Part 10: Test for Skin Sensitization | The device shall be non-sensitizer. | Under the conditions of this study, the test article was a non-sensitizer. | As per Device with same color | | | ISO 10993-11 Biological evaluation on medical device Part 11 – Test for systemic toxicity | The device shall not induce any acute systemic toxicity | Under the conditions of this study, not induce any acute systemic toxicity | As per Device with same color | {13} ECO Medi Glove Sdn. Bhd. (815262-D) Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my TEL +60-5-806 2316 FAX +60-5-895 7668 SST NO.: A11-1808-21015730 Summary of Clinical Testing: Not applicable. ## 9.0 Conclusion The Conclusion drawn from the non-Clinical test demonstrates that the subject device in 510(K) Submission, Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim is as safe, as effective, and performs as well as or better than the legally marketed Predicate device cleared under K171339.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...