Nitrile Powder Free Blue Patient Examination Gloves Non-Sterile, Tested For Use With Chemotherapy Drugs
K214110 · Mah Sing Healthcare Sdn. Bhd. · LZA · Feb 19, 2022 · General Hospital
Device Facts
Record ID
K214110
Device Name
Nitrile Powder Free Blue Patient Examination Gloves Non-Sterile, Tested For Use With Chemotherapy Drugs
Applicant
Mah Sing Healthcare Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Feb 19, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile, ambidextrous nitrile (acrylonitrile-butadiene) patient examination glove; blue color; powder-free. Used by healthcare professionals in clinical settings to prevent cross-contamination. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05 (2019). Provides a physical barrier; protects both patient and examiner. Not for use with Carmustine or Thiotepa.
Clinical Evidence
No clinical data. Bench testing only, including physical properties (tensile strength, elongation), dimensions, freedom from holes (ASTM D5151-19), residual powder (ASTM D6124-06), and chemotherapy drug permeation (ASTM D6978-05). Biocompatibility testing performed per ISO 10993-10 and ISO 10993-11.
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Tested for use with chemotherapy drugs per ASTM D6978-05 (2019).
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-sterile, Tested For Use With Chemotherapy Drugs (K161422)
Related Devices
K212532 — Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs) · Crdlight Optoelectronic Technology Co., Ltd. · Nov 11, 2021
K181066 — Powder Free Nitrile Examination Glove (Blue) with Low Dermatitis claim and with tested for Use with Chemotherapy Drugs Claims · Eco Medi Glove Sdn. Bhd. · Sep 26, 2018
K212827 — Jr Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs · Jr Engineering & Medical Technologies (M) Sdn. Bhd. · Dec 6, 2021
K161422 — DERMAGRIP POWDER FREE BLUE NITRILE PATIENT EXAMINATION GLOVES, NON-STERILE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS · Wrp Asia Pacific Sdn. Bhd. · Oct 20, 2016
K233405 — Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim · Eco Medi Glove Sdn. Bhd. · Dec 28, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
February 19, 2022
Mah Sing Healthcare Sdn Bhd Sazalinda Musa Senior RA Executive Wisma Mah Sing, Penthouse Suite 1, 163 Jalan Sungai Besi Kuala Lumpur, Kuala Lumpur 57100 Malaysia
Re: K214110
Trade/Device Name: Nitrile Powder Free Blue Patient Examination Gloves Non-Sterile, Tested For Use With Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: December 17, 2021 Received: December 29, 2021
Dear Sazalinda Musa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K214110
Device Name
Nitrile Powder Free Blue Patient Examination Gloves Non-Sterile, Tested for use with Chemotherapy Drugs
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 (2019), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Chemotherapy Drugs Permeation The following chemicals have been tested with these gloves.
Test Chemotherapy Drugs Concentration Breakthrough Detection Time in minutes *Carmustine 3.3mg/ml 24.8 Cisplatin 1 mg/ml >240 Cyclophosphamide 20 mg/ml >240 Dacarbazine 10 mg/ml >240 Doxorubicin, HCl 2 mg/ml >240 Etoposide 20 mg/ml >240 Fluorouracil 50 mg/ml >240 Methotrexate 25 mg/ml >240 Mitomycin C 0.5 mg/ml >240 Oxaliplatin 5 mg/ml >240 Paclitaxel 6 mg/ml >240 *Thiotepa 10 mg/ml 38.4 Vincristine 1 mg/ml >240
Warning- Not for use with Carmustine and Thiotepa.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Mah Sing Healthcare. The words "MahSing" are in a bold, red font on the top line. Below that, the word "HEALTHCARE" is in a smaller, black font.
# Premarket Notification [510(k)] No: K214110
| | 510 (K) SUMMARY |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.0 Device Name | Nitrile Powder Free Blue Patient Examination Gloves, Non-Sterile,<br>Tested For Use With Chemotherapy Drugs. |
| 2.0 Submitter name /<br>Contact details | Mah Sing Healthcare Sdn. Bhd<br>Wisma Mah Sing, Penthouse Suite 1,<br>163 Jalan Sungai Besi,<br>Kuala Lumpur<br>57100<br>MALAYSIA<br><br>Contact Person Details:<br>Sazalinda Musa (Ms)<br>E-mail: sazalinda.m@mshealthcare.com<br>Tel: +60-3-3396 2288, Extn: 2282<br>Fax: +60-3-3396 2299 |
| 3.0 Summary<br>Preparation Date | December 17, 2021 |
| 4.0 Device Name &<br>Classification | Trade Name: Nitrile Powder Free Blue Patient Examination Gloves,<br>Non-sterile, Tested For Use With Chemotherapy Drugs.<br><br>Common Name: Nitrile Powder Free Patient Examination Glove<br><br>Classification Name: Patient Examination Gloves Specialty (code LZC),<br>Polymer Patient Examination Gloves (code LZA).<br><br>Device Classification: I<br><br>Regulation Number: 21 CFR 880.6250<br><br>Panel: General Hospital<br><br>Product Code: LZC, LZA |
| 5.0 Identification of<br>The Legally Marketed<br>Device | Predicate Device Name: Dermagrip Powder Free Blue Nitrile Patient<br>Examination Gloves, Non-sterile, Tested For Use<br>With Chemotherapy Drugs<br><br>Predicate 510(K) Number: K161422<br><br>Manufacture's Name: WRP Asia Pacific Sdn Bhd. |
MAH SING HEALTHCARE SDN BHD Lot 6478, Lorong Sungai Puloh / KU6, Kawasan Industri Sungai Puloh, 42100 Klang, Selangor. MALAYSIA
T 603 3396 2288 F 603 3396 2299 mshealthcare.com
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Mah Sing Healthcare. The word "MahSing" is in red, with the "M", "S", and "g" being larger than the other letters. Below "MahSing" is the word "HEALTHCARE" in black, block letters. The logo is simple and professional, and the colors are eye-catching.
**6.0 Description of Device**
Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs meets all the requirements of ASTM standards D6319-19, D6978-05 (2019) and FDA 21 CFR 880.6250.
Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs are Class I patient examination gloves bear the product code Nitrile – LZA (21CFR880.6250).
The gloves are made from acrylonitrile-butadiene copolymer dispersion. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. These gloves are blue in color and are powder free. The gloves are ambidextrous i.e., can be worn on right hand or left hand, single use disposable devices that come in four sizes (S, M, L and XL). The physical properties of glove, i.e., tensile strength meet ASTM D 6319-19.
**7.0 Indications for Use**
A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 (2019), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Chemotherapy Drugs Permeation
The following chemicals have been tested with these gloves.
| Test Chemotherapy Drugs | Concentration | Breakthrough Detection Time in minutes |
|-------------------------|---------------|----------------------------------------|
| *Carmustine | 3.3mg/ml | 24.8 |
| Cisplatin | 1 mg/ml | >240 |
| Cyclophosphamide | 20 mg/ml | >240 |
| Dacarbazine | 10 mg/ml | >240 |
| Doxorubicin, HCl | 2 mg/ml | >240 |
| Etoposide | 20 mg/ml | >240 |
| Fluorouracil | 50 mg/ml | >240 |
| Methotrexate | 25 mg/ml | >240 |
| Mitomycin C | 0.5 mg/ml | >240 |
| Oxaliplatin | 5 mg/ml | >240 |
| Paclitaxel | 6 mg/ml | >240 |
| *Thiotepa | 10 mg/ml | 38.4 |
| Vincristine | 1 mg/ml | >240 |
***Note:**
Please note that the following drugs have low permeation times:
(1) Carmustine – 24.8 minutes
(2) Thiotepa – 38.4 minutes.
HOW DOES HEALTHCARE SPENDING COMPARE?
MAH SING HEALTHCARE SDN BHD Lot 6478, Lorong Sungai Puloh / KU6, Kawasan Industri Sungai Puloh, 42100 Klang, Selangor. MALAYSIA
T 603 3396 2288 F 603 3396 2299 mshealthcare.com
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Mah Sing Healthcare. The words "MahSing" are in red, with the "M" being the largest letter. Below that, the word "HEALTHCARE" is in black, with all letters being the same size. The logo is simple and modern.
8.0 Summary of the Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs meets all the requirements of ASTM Technological Characteristic of standards D6319-19, D6978-05 (2019) and FDA 21 CFR 880.6250. the Device
Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use With Chemotherapy Drugs are Class I patient examination gloves bear the product code Nitrile - LZA (21CFR880.6250).
The gloves are made from acrylonitrile-butadiene copolymer dispersion. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. These gloves are blue in color and are powder free. The gloves are ambidextrous i.e., can be worn on right hand or left hand, single use disposable devices that come in four sizes (S, M, L and XL). The physical properties of glove, i.e., tensile strength meet ASTM D 6319-19.
| Characteristics and<br>Parameters | Standard | Proposed Device | Predicate device | Comparison<br>Analysis |
|-------------------------------------------------|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| 510(k) Number | - | - | K161422 | - |
| Name of device | - | Nitrile Powder Free Blue<br>Patient Examination<br>Gloves, Non-Sterile | Powder Free Nitrile<br>Examination Gloves<br>Non- Sterile | Similar |
| Device Classification<br>Name/Regulation Number | Patient Examination<br>Glove,<br>21 CFR Part 880.6250 | Patient Examination<br>Glove,<br>21 CFR Part 880.6250 | Patient Examination<br>Glove,<br>21 CFR Part 880.6250 | Similar |
| Product Code | - | LZA, LZC | LZA, LZC | Similar |
| Intended Use | - | A patient examination<br>glove is a disposable<br>device intended for<br>medical purpose that is<br>worn on the examiner's<br>hand or finger to prevent<br>contamination between<br>patient and examiner | A patient examination<br>glove is a disposable<br>device intended for<br>medical purpose that is<br>worn on the examiner's<br>hand or finger to prevent<br>contamination between<br>patient and examiner. | Same intended<br>use |
| Classification | - | Class 1 | Class 1 | Same Class |
| Raw Rubber Material | ASTM D 6319-19 | Nitrile (Acrylonitrile-<br>butadiene) | Nitrile (Acrylonitrile-<br>butadiene) | Same synthetic<br>rubber material |
| Design, Color and<br>Surface Appearance | - | 1. Ambidextrous<br>2. Blue<br>3. Powder Free<br>4. Finger Textured | 1. Ambidextrous<br>2. Blue<br>3. Powder Free<br>4. Finger Textured | Same ambidextrous<br>Design, same color,<br>same features and<br>Same textured<br>area |
### Table 1
MAH SING HEALTHCARE SDN BHD Lot 6478, Lorong Sungai Puloh / KU6,
Kawasan Industri Sungai Puloh,
42100 Klang, Selangor.
MALAYSIA
T 603 3396 2288 F 603 3396 2299 mshealthcare.com
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Mah Sing Healthcare. The word "MahSing" is in red, with the "M" and "S" capitalized and slightly larger than the other letters. Below "MahSing" is the word "HEALTHCARE" in black, with all letters capitalized.
| Characteristics and<br>Parameters | Standard | Proposed Device | Predicate device | Comparison<br>Analysis |
|--------------------------------------|---------------|-------------------------------------------|------------------|--------------------------------------------------------------------------------------|
| Chemotherapy Drug<br>Permeation Test | ASTM D6897-05 | | | |
| Test Chemotherapy<br>Drugs | Concentration | Minimum Breakthrough Detection Time (min) | | |
| Fluorouracil | 50 mg/ml | >240 | >240 | Similar |
| Etoposide | 20 mg/ml | >240 | >240 | Similar |
| Cyclophosphamide | 20 mg/ml | >240 | >240 | Similar |
| Fluorouracil | 50 mg/ml | >240 | >240 | Similar |
| *Carmustine | 3.3mg/ml | 24.8 | 15.0 | Similar<br>Below 240 minutes<br>permeation time,<br>similar with<br>predicate device |
| *Thiotepa | 10 mg/ml | 38.4 | 2.0 | Similar<br>Below 240 minutes<br>permeation time,<br>similar with<br>predicate device |
| Paclitaxel | 6 mg/ml | >240 | >240 | Similar |
| Doxorubicin, HCl | 2 mg/ml | >240 | >240 | Similar |
| Dacarbazine | 10 mg/ml | >240 | >240 | Similar |
| Cisplatin | 1 mg/ml | >240 | >240 | Similar |
| Ifosfamide | 50 mg/ml | Not tested | >240 | Optional, Predicate<br>device perform<br>additional<br>Chemotherapy drug<br>test |
| Mitoxantrone | 2.0 mg/ml | Not tested | >240 | Optional, Predicate<br>device perform<br>additional<br>Chemotherapy drug<br>test |
| Vincristine | 1 mg/ml | >240 | >240 | Similar |
| Methotrexate | 25 mg/ml | >240 | >240 | Similar |
| Mitomycin C | 0.5 mg/ml | >240 | >240 | Similar |
| Oxaliplatin | 5 mg/ml | >240 | Not tested | Optional, Subject<br>device perform<br>additional<br>Chemotherapy drug<br>test |
#### MAH SING HEALTHCARE SDN BHD
DrivingInnovation. Protecting Lives.
Lot 6478, Lorong Sungai Puloh / KU6, Kawasan Industri Sungai Puloh, 42100 Klang, Selangor. MALAYSIA
T 603 3396 2288 F 603 3396 2299 mshealthcare.com
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Mah Sing Healthcare. The words "Mah Sing" are in a large, bold, red font. Below that, the word "HEALTHCARE" is in a smaller, black, sans-serif font. The logo is simple and professional.
| Characteristics and<br>Parameters | Standard | Proposed Device | Predicate device | Comparison<br>Analysis |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Overall Length (Minimum<br>230mm) | ASTM D 6319-19 | 245 - 255 mm | Min. 240 mm | Similar |
| Width<br>S: 75mm - 95mm<br>M: 85mm - 105mm<br>L: 100mm – 120mm<br>XL:110mm - 130mm | ASTM D 6319-19 | S: 83 - 85mm<br>M: 93 - 96mm<br>L: 105 - 109mm<br>XL: 116 - 118mm | Not stated by predicate<br>device | Similar, subject<br>device meet<br>requirements<br>of ASTM<br>D6319 |
| Cuff Thickness<br>(Minimum 0.05mm) | ASTM D 6319-19 | 0.05 - 0.06mm | 0.06 - 0.08mm | Similar |
| Palm Thickness<br>(Minimum 0.05mm) | ASTM D 6319-19 | 0.06 - 0.07mm | 0.07 - 0.10mm | Similar |
| Finger Thickness<br>(Minimum 0.05mm) | ASTM D 6319-19 | 0.08 - 0.11mm | 0.07 - 0.09mm | Similar |
| Tensile Strength<br>(Before aging) Minimum<br>14 MPa | ASTM D 6319-19 | Average: 21.96MPa | Meets | Similar |
| Tensile Strength<br>(After accelerated aging)<br>Minimum 14 MPa | ASTM D 6319-19 | Average: 28.30MPa | Meets | Similar |
| Ultimate Elongation<br>(before aging) Minimum<br>500% | ASTM D 6319-19 | Average: 537% | Meets | Similar |
| Ultimate Elongation<br>(after accelerated aging)<br>Minimum 400% | ASTM D 6319-19 | Average: 449% | Meets | Similar |
| Freedom of Holes Meet<br>AQL 2.5 at G1 | ASTM D 5151-19 | Meet AQL 1.5 with G1 | Passes | Similar |
| Residual powder test<br>(Less than 2mg/glove) | ASTM D 6124-06 | Average powder residue<br>for each size.<br>S: 0.32 mg /glove<br>M: 0.28 mg /glove<br>L: 0.32 mg /glove<br>XL: 0.30 mg/glove | Average powder residue<br>Meets | Similar |
| Animal Irritation Test | ISO 10993-10<br>Biological evaluation<br>of medical devices -<br>Part 10: Tests for<br>irritation and skin<br>sensitization | Passed.<br>Under the conditions of<br>study, not an irritant | Passed<br>Not a primary skin irritant<br>under the conditions of the<br>study | Similar |
| Dermal Sensitization | ISO 10993-10<br>Biological evaluation<br>of medical devices -<br>Part 10: Tests for<br>irritation and skin<br>sensitization | Passed.<br>Under the conditions of<br>study, not a sensitizer | Passed<br>Not a contact sensitizer<br>under the conditions of the<br>study | Similar |
| Acute Systemic Toxicity | ISO 10993-11<br>Biological evaluation<br>of medical devices -<br>Part 11: Tests for<br>systemic toxicity | Not induce systemic<br>toxicity | Predicate device has not<br>performed test. | Subject devices<br>meet requirements<br>of ISO 10993-11 |
| Characteristics and<br>Parameters | Standard | Proposed Device | Predicate device | Comparison<br>Analysis |
| Expiration Date | ASTM D 7160-16<br>Standard Practice for<br>Determination of<br>Expiration Dating for<br>Medical Gloves | 3 years from date of<br>manufactured | Predicate device has not<br>stated. | - |
| Manufacturer | - | Mah Sing Healthcare Sdn.<br>Bhd. | WRP Asia Pacific Sdn.<br>Bhd. | - |
MAH SING HEALTHCARE SDN BHD Lot 6478, Lorong Sungai Puloh / KU6, Kawasan Industri Sungai Puloh, 42100 Klang, Selangor. MALAYSIA
T 603 3396 2288 F 603 3396 2299 mshealthcare.com
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Mah Sing Healthcare. The word "MahSing" is in red, and the word "Healthcare" is in black. The words are stacked on top of each other, with "MahSing" on top and "Healthcare" on the bottom.
# 9.0 Summary of Non-Clinical Testing
# Table 2 -Performance Testing
| Non-Clinical Testing | | | |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Test Method | Purpose | Acceptance Criteria | Result |
| ASTM D6124-06<br>(Reapproved 2017)<br>Standard Test Method for<br>Residual Powder on Medical<br>Gloves. | To determine the<br>residual powder in<br>the gloves | Less than 2mg / glove | Size S 0.32mg /glove<br>Size M 0.28mg /glove<br>Size L 0.32mg /glove<br>Size XL 0.30mg /glove |
| ASTM D5151-19<br>Standard Test Method for<br>Detection of Holes in<br>Medical Gloves. | To determine the<br>holes in the gloves | Inspection level, G-I<br>AQL 2.5<br>(In accordance with ASTM<br>D6319-19) | Passed G-I, AQL 1.5 |
| ASTM D6319-19<br>Standard Specification<br>for Nitrile Examination<br>Gloves for Medical<br>Application. | To determine the<br>length of the gloves | Size S 220mm, min<br>Size M 230mm, min<br>Size L 230mm, min<br>Size XL 230mm, min | Size S 250 - 255mm<br>Size M 250 - 255mm<br>Size L 245 - 252mm<br>Size XL 245 - 252mm |
| ASTM D6319-19<br>Standard Specification for<br>Nitrile Examination Gloves<br>for Medical Application. | To determine the<br>width of the gloves | Size S 80 ± 10mm<br>Size M 95 ± 10mm<br>Size L 110 ± 10mm<br>Size XL 120 ± 10mm | Size S S: 83 - 85mm<br>Size M M: 93 - 96mm<br>Size L L: 105 - 109mm<br>Size XL XL: 116 - 118mm |
| Non-Clinical Testing (Cont'd) | | | |
| Test Method | Purpose | Acceptance Criteria | Result |
| ASTM D6319-19<br>Standard Specification<br>for Nitrile Examination<br>Gloves for Medical<br>Application. | To determine the<br>thickness of the<br>gloves | Measured in single wall at cuff edge area<br>Cuff 0.05mm, min | Size S 0.05 - 0.06mm<br>Size M 0.05 - 0.06mm<br>Size L 0.05 - 0.06mm<br>Size XL 0.05 - 0.06mm |
| | | Measured in single wall at approximate center of palm area<br>Palm 0.05mm, min | Size S 0.06 - 0.07mm<br>Size M 0.06 - 0.06mm<br>Size L 0.06 - 0.07mm<br>Size XL 0.06 - 0.07mm |
| | | Measured in single wall at 13±3mm from the tip of middle finger<br>Finger 0.05mm, min | Size S 0.08 - 0.09mm<br>Size M 0.09 - 0.10mm<br>Size L 0.10 - 0.11mm<br>Size XL 0.10 - 0.11mm |
| ASTM D6319-19<br>Standard Specification<br>for Nitrile Examination<br>Gloves for Medical<br>Application. | To determine the<br>physical properties-<br>Tensile strength | Before Ageing<br>Tensile Strength<br>14Mpa, min for all sizes | Size S 21.94 MPa, average<br>Size M 22.08 MPa, average<br>Size L 21.76 MPa, average<br>Size XL 22.05 MPa, average |
| | | After Ageing<br>Tensile Strength<br>14Mpa, min for all sizes | Size S 29.53 MPa, average<br>Size M 29.12 MPa, average<br>Size L 24.90 MPa, average<br>Size XL 29.63 MPa, average |
| | To determine the<br>physical properties-<br>Ultimate Elongation | Before Ageing<br>Ultimate Elongation<br>500%, min for all sizes | Size S 537%, average<br>Size M 550%, average<br>Size L 526%, average<br>Size XL 536%, average |
| | | After Ageing<br>Ultimate Elongation 400%,<br>min for all sizes | Size S 452%, average<br>Size M 450%, average<br>Size L 436%, average<br>Size XL 458%, average |
MAH SING HEALTHCARE SDN BHD Lot 6478, Lorong Sungai Puloh / KU6, Kawasan Industri Sungai Puloh, 42100 Klang, Selangor. MALAYSIA
T 603 3396 2288 F 603 3396 2299 mshealthcare.com
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Mah Sing Healthcare. The words "MahSing" are in red, with the word "Healthcare" in black underneath. The font is sans-serif and bolded.
MAH SING HEALTHCARE SDN BHD Lot 6478, Lorong Sungai Puloh / KU6, Kawasan Industri Sungai Puloh, 42100 Klang, Selangor. MALAYSIA
T 603 3396 2288 F 603 3396 2299 mshealthcare.com
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Mah Sing Healthcare. The words "Mah Sing" are in red, and the word "Healthcare" is in black. The word "Healthcare" is directly below "Mah Sing".
| Biocompatibility Testing | | | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Test Method | Purpose | Acceptance Criteria | Result | | |
| ISO 10993-10<br>Biological evaluation of<br>medical devices - Part<br>10: Tests for irritation<br>and skin sensitization<br>(Animal Irritation<br>Test) | To determine the<br>potential of the<br>material under test<br>to produce dermal<br>irritation in Rabbits | Under the condition<br>of study not an<br>irritant. | There was no observable irreversible<br>alteration on the skin at the sites of contact<br>with the test material. The Primary<br>Irritation Index (PII) was "0". The test<br>material was not corrosive, and the<br>Primary Irritation Response Category is<br>therefore "negligible", thereof met the<br>requirement. | | |
| ISO 10993-10<br>Biological evaluation of<br>medical devices -<br>Part 10: Tests for<br>irritation and skin<br>sensitization<br>(Dermal Sensitization<br>Assay Test) | To determine the<br>skin sensitization<br>potential of the<br>material both in<br>terms of induction<br>and elicitation in<br>Guinea pig | Under the condition<br>of the study not a<br>sensitizer. | There was no sensitization induced by the<br>application of the test material on the<br>albino guinea pigs under the condition of<br>this test, thereof met the requirement. | | |
| ISO 10993-11<br>Biological evaluation of<br>medical devices - Part 11:<br>Tests for systemic toxicity<br>(Acute Systemic Toxicity) | To provide<br>information on<br>health hazards<br>likely to arise from<br>a short-term<br>exposure to the<br>extracts of test<br>material by<br>intravenous and<br>intraperitoneal<br>injection in mice | Not induce systemic<br>toxicity | Under the condition of this study, the<br>single dose acute systemic toxicity of<br>extracts from test material using both<br>normal saline and sesame oil, shown non-<br>toxic effects, thereof met the requirement. | | |
Non-clinical tests were carried out to demonstrate product performance conformity with standards referenced.
The following bench tests were performed:
Non-clinical tests
- Residual Powder Content
- Physical Properties
- Physical Dimension
- Freedom from Holes
### Biocompatibility Testing
- Animal Irritation Test
- Dermal Sensitization Assay
- Acute Systemic Toxicity
The results from these performance evaluations demonstrated that the Nitrile Powder Free Blue Patient Examination Gloves, Non-Sterile, Tested For Use With Chemotherapy Drugs met the acceptance criteria defined in standards referenced.
MAH SING HEALTHCARE SDN BHD Lot 6478, Lorong Sungai Puloh / KU6, Kawasan Industri Sungai Puloh, 42100 Klang, Selangor. MALAYSIA
T 603 3396 2288 F 603 3396 2299 mshealthcare.com
{11}------------------------------------------------
| 10.0 Summary of Clinical Testing: | Clinical Testing is not needed for this device. |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 11.0 Conclusion | The conclusion drawn from the non-clinical test demonstrate that the<br>subject device is as safe, as effective, and performs as well as or better<br>than the legally marketed predicate device K161422. |
DrivingInnovation. Protecting Lives.
MAH SING HEALTHCARE SDN BHD Lot 6478, Lorong Sungai Puloh / KU6, Kawasan Industri Sungai Puloh, 42100 Klang, Selangor. MALAYSIA
T 603 3396 2288 F 603 3396 2299 mshealthcare.com
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.