DERMAGRIP POWDER FREE BLUE NITRILE PATIENT EXAMINATION GLOVES, NON-STERILE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS
K161422 · Wrp Asia Pacific Sdn. Bhd. · LZA · Oct 20, 2016 · General Hospital
Device Facts
Record ID
K161422
Device Name
DERMAGRIP POWDER FREE BLUE NITRILE PATIENT EXAMINATION GLOVES, NON-STERILE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS
Applicant
Wrp Asia Pacific Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Oct 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile, powder-free blue nitrile patient examination glove. Worn by clinicians to prevent cross-contamination between patient and examiner. Manufactured from synthetic rubber latex with a surface treatment for easy donning. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings; ambidextrous design. Provides a barrier against patient fluids and chemotherapy agents, though specific drugs (Carmustine, Thiotepa) have low permeation times requiring caution.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D6319-10 (physical properties), ASTM D6124-06 (powder content), ASTM D5151-06 (watertightness), ASTM D6978-05 (chemotherapy drug permeation), and ISO 10993-10:2010 (biocompatibility/irritation/sensitization).
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs per ASTM D6978-05.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Dermagrip Powder Free Blue Nitrile Patient Examination Gloves (K141982)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 20, 2016
WRP Asia Pacific Sdn. Bhd. Sarala Devi Jayaraman Regulatory Affairs Manager Lot 1. Jalan 3. Kaw Perusahaan Bandar Baru Salak Tinggi, Sepang, 43900 MY
Re: K161422
Trade/Device Name: Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-sterile, Tested For Use With Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: September 15, 2016 Received: September 21, 2016
Dear Sarala Devi Javaraman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Ryan -S
for Tina Kiang, Ph.D.
Acting Division Director Division of Anesthesiology, General Hospital, Respiratory Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
### 510(k) Number (if known)
### Device Name
Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-Sterile, Tested for use with Chemotherapy Drugs
### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
### Chemotherapy Drug Permeation
The following chemicals have been tested with these gloves.
| Chemotherapy Drug | Concentration | Breakthrough Detection Time in Minutes |
|----------------------------------------|---------------|----------------------------------------|
| Flourouracil (Adrucil) | 50.0mg/ml | > 240 |
| Etopside (Toposar) | 20.0mg/ml | > 240 |
| Cyclophosphamide (Cytoxan) | 20.0mg/ml | > 240 |
| *Carmustine (BCNU) | 3.3mg/ml | 15.0 |
| *Thiotepa | 10.0mg/ml | 2.0 |
| Paclitaxel (Taxol) | 6.0mg/ml | > 240 |
| Doxorubicin Hydrochloride (Adriamycin) | 2.0mg/ml | > 240 |
| Dacarbazine | 10.0mg/ml | > 240 |
| Cisplatin | 1.0mg/ml | > 240 |
| Ifosfamide | 50.0mg/ml | > 240 |
| Mitoxantrone | 2.0mg/ml | > 240 |
| Vincristine Sulfate | 1.0mg/ml | > 240 |
| Methotrexate | 25mg/ml | > 240 |
| Mitomycin C | 0.5mg/ml | > 240 |
* Please note that the-following drugs have extremely low permeation times: Carmustine (BCNU) : 15 minutes and Thiotepa : 2 minutes
| | | | Type of Use (Select one or both, as applicable) | |
|--|--|--|-------------------------------------------------|--|
|--|--|--|-------------------------------------------------|--|
] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the pata needed and complete and review the collection of information. Send comments regarding this burden estimate or any other spect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 1.0 Submitter:
| Name: | Sarala Devi Jayaraman |
|------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | WRP Asia Pacific Sdn Bhd<br>Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak Tinggi, 43900<br>Sepang, Selangor Darul Ehsan, MALAYSIA |
| Phone No.: | +60 3 8706 1486 |
| Fax No.: | +60 3 8706 1485 |
Date of Summary Prepared: 17 October 2016
### 2.0 Name of the modified device Identification:
Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-Sterile, Tested for use with Chemotherapy Drugs Common Name: Exam Gloves Classification Name: Patient Examination Gloves Specialty (21 CFR 880.6250 product code LZC) Patient Examination Gloves (21 CFR 880.6250 product code LZA)
### 3.0 Identification of The Legally Marketed Devices that equivalency is claimed:
| | Predicate |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | WRP Asia Pacific Sdn Bhd |
| Device name | Dermagrip Powder Free Blue Nitrile Patient Examination<br>Gloves, Non-sterile, Tested for use with Chemotherapy<br>Drugs (and various brandnames) |
| 510(k) Number | K141982 |
| MDL | D236305 |
| Regulatory Class | I |
| Product Code | LZC, LZA |
#### 4.0 Description of The Device:
Powder Free Blue Nitrile Patient Examination Glove, Non-Sterile the meets all requirements of ASTM standard D6319-10, D6978-05 and FDA 21 CFR 880.6250.
The powder free nitrile examination qlove is manufactured from synthetic rubber latex. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e. can be worn on right hand or left hand. The physical properties of glove i.e. tensile strength meet ASTM standard D6319-10.
{5}------------------------------------------------
#### 5.0 Indications of Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Chemotherapy Drug Permeation The following chemicals have been tested with these gloves.
| Chemotherapy Drug | Concentration | Breakthrough Detection Time<br>in Minutes |
|----------------------------------------|---------------|-------------------------------------------|
| Flourouracil (Adrucil) | 50.0mg/ml | > 240 |
| Etopside (Toposar) | 20.0mg/ml | > 240 |
| Cyclophosphamide (Cytoxan) | 20.0mg/ml | > 240 |
| *Carmustine (BCNU) | 3.3mg/ml | 15.0 |
| *Thiotepa | 10.0mg/ml | 2.0 |
| Paclitaxel (Taxol) | 6.0mg/ml | > 240 |
| Doxorubicin Hydrochloride (Adriamycin) | 2.0mg/ml | > 240 |
| Dacarbazine | 10.0mg/ml | > 240 |
| Cisplatin | 1.0mg/ml | > 240 |
| Ifosfamide | 50.0mg/ml | > 240 |
| Mitoxantrone | 2.0mg/ml | > 240 |
| Vincristine Sulfate | 1.0mg/ml | > 240 |
| Methotrexate | 25 mg/ml | > 240 |
| Mitomycin C | 0.5 mg/ml | > 240 |
* Please note that the following drugs have extremely low permeation times: Carmustine (BCNU) : 15 minutes and Thiotepa : 2 minutes
{6}------------------------------------------------
### Summary of the Technological Characteristics of the Device: 6.0
The Powder Free Blue Nitrile Examination Gloves, Non-Sterile are summarized with the following technological characteristics compared to ASTM D6319 or equivalent standards as shown in Table 1.
Chemotherapy claim is similar to Predicate, which has a glove thickness below 0.10mm and is shorter than 270mm but compliant with the ASTM standards.
## Table 1
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | |
|-------------------------------------------|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | |
| | | Predicate | Current |
| Manufacturer(s) | | WRP Asia Pacific<br>Sdn Bhd | WRP Asia Pacific<br>Sdn Bhd |
| 510(k) Number | | K141982 | K161422 |
| Dimensions | ASTM D6319-10 | Min 240mm | Min 240mm |
| Physical Properties | ASTM D6319-10 | Meets | Meets |
| Thickness - Finger | ASTM D6319-10 | 0.07 - 0.10mm | 0.07 - 0.10mm |
| - Palm | | 0.07 - 0.09mm | 0.07 - 0.09mm |
| - Cuff | | 0.06 - 0.08mm | 0.06 - 0.08mm |
| Powder Free | ASTM D6124-06<br>(≤ 2 mg/glove) | Meets | Meets |
| Chemotherapy Drug<br>Permeation Test | ASTM D6978-05 | | |
| Test Chemotherapy Drug | Concentration | | Minimum Breakthrough Detection<br>Time (min) |
| Flourouracil (Adrucil) | 50.0mg/ml | > 240 | > 240 |
| Etopside (Toposar) | 20.0mg/ml | > 240 | > 240 |
| Cyclophosphamide (Cytoxan) | 20.0mg/ml | > 240 | > 240 |
| *Carmustine (BCNU) | 3.3mg/ml | 15.0 | 15.0 |
| *Thiotepa | 10.0mg/ml | 2.0 | 2.0 |
| Paclitaxel (Taxol) | 6.0mg/ml | > 240 | > 240 |
| Doxorubicin Hydrochloride<br>(Adriamycin) | 2.0mg/ml | > 240 | > 240 |
| Dacarbazine | 10.0mg/ml | > 240 | > 240 |
| Cisplatin | 1.0mg/ml | > 240 | > 240 |
| Ifosfamide | 50.0mg/ml | > 240 | > 240 |
| Mitoxantrone | 2.0mg/ml | > 240 | > 240 |
| Vincristine Sulfate | 1.0mg/ml | > 240 | > 240 |
| Methotrexate | 25 mg/ml | - | > 240 |
| Mitomycin C | 0.5 mg/ml | - | > 240 |
| Warning Statement | | * WARNING: Please<br>note that the<br>following drugs<br>have extremely low<br>permeation times:<br>Carmustine (BCNU)<br>: 15 minutes and<br>Thiotepa : 2<br>minutes | * WARNING: Please<br>note that the<br>following drugs have<br>extremely low<br>permeation times:<br>Carmustine (BCNU)<br>: 15 minutes and<br>Thiotepa : 2<br>minutes |
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | |
| | | Predicate | Current |
| Biocompatibility | Primary Skin Irritation<br>- | Passes | Passes |
| | ISO 10993-<br>10:2010(E) | Not a primary skin<br>irritant under the<br>conditions of the<br>study. | Not a primary skin<br>irritant under the<br>conditions of the<br>study. |
| | Dermal Sensitization -<br>ISO 10993-<br>10:2010(E) | Passes<br>Not a contact<br>sensitizer under the<br>conditions of the<br>study. | Passes<br>Not a contact<br>sensitizer under the<br>conditions of the<br>study. |
| Watertight (1000ml) | ASTM D5151-06 | Passes | Passes |
| Intended use | - | A patient<br>examination glove<br>is a disposable<br>device intended for<br>medical purposes<br>that is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient<br>and examiner. | A patient<br>examination glove<br>is a disposable<br>device intended for<br>medical purposes<br>that is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient<br>and examiner. |
| Material | ASTM D6319-10 | Nitrile | Nitrile |
| Color | - | Blue | Blue |
| Texture | - | Finger textured | Finger textured |
| Size | ASTM D6319-10 | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large |
| Single Use | Medical Glove<br>Guidance Manual -<br>Labeling | Single use | Single use |
{7}------------------------------------------------
{8}------------------------------------------------
### Substantial Equivalent Based on Assessment of Non-Clinical Performance 7.0 Data
The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above (ASTM Requirements).
The device and the predicate share the same intended use, same material, same specifications for thickness and length, similar permeation rates for chemotherapy drugs, similar labeling according to the glove guidance, and same compliance with standards for physical properties, powder free, biocompatibility and water tightness. Thus, the device is substantial equivalent to the predicate.
#### 8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data
Clinical data was not needed for this device.
#### 9.0 Conclusion
Based on intended uses, technological characteristics and non-clinical performance data, the subject device is substantially equivalent to the predicate device K141982.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.