Powder Free Nitrile Patient Examination Gloves, Blue Color

K211860 · Anhui Powerguard Technology Co., Ltd. · LZA · Aug 13, 2021 · General Hospital

Device Facts

Record IDK211860
Device NamePowder Free Nitrile Patient Examination Gloves, Blue Color
ApplicantAnhui Powerguard Technology Co., Ltd.
Product CodeLZA · General Hospital
Decision DateAug 13, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Device is a non-sterile, blue, powder-free nitrile patient examination glove. Used by healthcare professionals in clinical settings to provide a barrier against contamination between the examiner and the patient. Device is a single-use, disposable physical barrier. No electronic, software, or algorithmic components. Benefit is the reduction of cross-contamination risk during medical examinations.

Clinical Evidence

Bench testing only. No clinical data. Testing included physical property verification (ASTM D6319-10), residual powder testing (ASTM D6124-06), freedom from pinholes (ASTM D5151-19), and biocompatibility (ISO 10993-5 cytotoxicity, ISO 10993-10 irritation and sensitization). All tests met acceptance criteria.

Technological Characteristics

Material: Nitrile. Form factor: Disposable, non-sterile, blue, ambidextrous glove. Standards: ASTM D6319-10 (Reapproved 2015), ASTM D6124-06 (Reapproved 2017), ASTM D5151-19, ISO 10993-5, ISO 10993-10. No energy source, connectivity, or software.

Indications for Use

Indicated for use as a non-sterile, disposable medical glove worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. August 13, 2021 Anhui Powerguard Technology Co., Ltd. % Chu Xiaoan Official Correspondent Beijing Easy-Link Company Rm. F302 Bldg., 41, Jing Cheng Ya Ju, Courtyard 6 of Southern Dou Ge Zhuang, Chaoyang District Beijing, 100121 China Re: K211860 Trade/Device Name: Powder Free Nitrile Patient Examination Gloves, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: April 16, 2021 Received: June 16, 2021 Dear Chu Xiaoan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K211860 Device Name Powder Free Nitrile Patient Examination Gloves, Blue Color Indications for Use (Describe) Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(K) Summary "The assigned 510(k)number is: K211860 This summary of 510(k) is being submitted in accordance with 21 CFR 807. Premarket Notification [510(k)] Summary #### 1.0 Submitter: | Submitter's name : | Anhui Powerguard Technology Co., Ltd. | |--------------------|---------------------------------------| |--------------------|---------------------------------------| Submitter's address : Lingbi Economic Development Zone (North), Suzhou City, Anhui Province, 234200,China Name of contact person: Mr. Nick Dai Submitter information 0086-557-6868999 Date of preparation : 2021-08-09 #### 2.0 Name of the Device Proprietary/,Trade name:Powder Free Nitrile Patient Examination Gloves, Blue Color | Common Name: | Patient Examination gloves | |------------------------|----------------------------------------| | Classification Name: | Non-powdered Patient examination glove | | Device Classification: | I | | Regulation Number: | 21 CFR 880.6250 | | Panel: | General Hospital | | Product Code: | LZA | #### 3.0 Predicate device | Device Name: | Powder Free<br>Glove, Blue Color | Nitrile Patient Examination | |----------------|--------------------------------------------|-----------------------------| | Company name: | Tangshan Zhonghong Pulin Plastic Co., Ltd. | | | 510(K) Number: | K120970 | | ### 4.0 Device Description: The proposed device is Powder Free Nitrile Examination Gloves. The proposed device is blue. The proposed device is non-sterile. ### 5.0 Indications for Use Statement: Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. {4}------------------------------------------------ | Features & Description | Predicate Device (K120970) | Subject Device (K211860) | Comparison | Test Methodology | Purpose | Acceptance Criteria | Results | | |------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|---------------------------------------------------------------------------------------------|--------------------------|--------------------------------------------------------------------------------------------------|-----------|-------------------------------| | Product name | Powder Free Nitrile Patient Exa mination Glove, Blue Color | Powder Free Nitrile Patient Examination Gloves, Blue Color | Same | ASTM D 6319-<br>06(Reapproved<br>2015). | Dimension | Length | ≥230mm | | | Regulation Number | 21CFR880.6250 | 21CFR880.6250 | Same | | | Width | Small | 75-90 mm | | Product Code | LZA | LZA | Same | | | | Medium | 87-102 mm | | Color | Blue | Blue | Same | | | | Large | 107-119mm | | Size | Small/ Medium/ Large/X large | Small/ Medium/ Large/X large | Same | | | | X large | 114-128 mm | | Indications for Use | Powder Free Nitrile Patient Examination Glove, Blue Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Powder Free Nitrile Patient Examination Gloves, Blue Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Same | | | Thickness | Fingertip | ≥0.08mm | | Device Description and Specifications | Meets ASTM D6319-10 | Meets ASTM D6319-10 (Reapproved 2015) | Same | | | | Palm | ≥0.08mm | | Dimensions --Length<br>ILS-2<br>AQL4.0 | Meets ASTM D6319-10<br>>=230mm min | 232 mm min for all sizes | Similar | ASTM D 6319-<br>06(Reapproved<br>2015). | Physical<br>Properties | Tensile strength<br>(Before & After<br>aging) | ≥14MPa | 17-25 | | Dimensions -- Width<br>IL S-2<br>AQL4.0 | Meets ASTM D6319-10<br>Small 70-90 mm<br>Medium 85-105mm<br>Large 100-120mm<br>X large 110-130 mm | Small 75-90 mm<br>Medium 87-102 mm<br>Large 107-119mm<br>X large 114-128 mm | Similar | | | Before aging<br>Elongation | ≥500% | 560-610 | | Dimensions --Thickness<br>IL S-2<br>AQL4.0 | Meets ASTM D6319-10<br>Finger 0.05mm min.<br>Palm 0.05mm min. | Thickness (mm) min.<br>Finger 0.08<br>Palm 0.08 | Similar | | | After aging<br>Elongation | ≥400% | 460-570 | | Physical Properties<br>IL S-2<br>AQL4.0 | Meets ASTMDD6319-10 | Aging Before After | Similar | | | | | | | | Before aging/after aging<br>Tensile Strength≥ 14MPa<br>Before aging<br>Elongation≥500%<br>After aging<br>Elongation ≥400% | Elongation (% 560-610 460-570<br>Tensile Strength (MPa)<br>19-25 17-23 | | ● 21 CFR 800.20<br>● ASTM D 6319-<br>06(Reapproved<br>2015).<br>● ASTM D5151-19 | Freedom from<br>pinholes | Water leakage test:<br>Inspection LevelI, AQL2.5, and<br>Accept/Reject criteria of 10/11. | | 5noncompliance<br>is allowed. | | Freedom from Pinholes<br>Inspection Level I<br>AQL2.5 | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10 | 1) Inspection Levell AQL2.5,and Accept/Reject criteria of 10/11<br>2) Water leakage test:5 noncomplianceis allowed. | Similar | ● ASTM D6319-<br>10(Reapproved<br>2015) | Powder<br>Residual | Meets<br><2mg/glove | | Mean:<br>0.1mg/pcs | | Residual Powder | Meets ASTM D 6124-06 (Reaffirmation 2011) | 1) Checked on 5pcs sub-samples (N=5). | Similar | ● ASTM D6124-<br>06 (Reapproved<br>2017), | | | | Pass | | | below 2mg of residual<br>powder | Mean: 0.1mg/pcs | | Primary Skin<br>Irritation in rabbits<br>ISO 10993-10:<br>2010-08-01 | Biocompatibility | Under the conditions of the study,<br>the subject device is not a<br>primary skin irritant. | | Passes | | Materials used to<br>fabricate the devices | Nitrile | Nitrile | Same | Dermal<br>sensitization in the<br>guinea pig ISO<br>10993-10: 2010-<br>08-01 | | Under the conditions of the study,<br>the subject device is not a skin<br>sensitizer. | | Passes | | Single Patient Use | Single Patient Use | Single Patient Use | Same | The test article was<br>added to L929<br>cells measured by<br>MTT assay ISO<br>10993-5:2009 | | Under the conditions of this<br>study, the test article was non-<br>cytotoxicity to L-929 cells. | | Pass | | Biocompatibility | Under the conditions of<br>this study, the test<br>article was a non-<br>irritant or non-<br>sensitizer<br><br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10:2002/Amd.1:<br>2006 | Under the conditions of this<br>study, the test article was a<br>non- irritant or non-<br>sensitizer and non-potential<br>toxicity to L-929 cells.<br><br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION STUDIES<br>Meets ISO 10993-10: Third<br>Edition 2010-08-01.<br><br>Cytotoxicity study meets<br>ISO 10993-5 Third edition<br>2009-06-01 | Similar | | | | | | | Labeling for the<br>legally marketed<br>device to which<br>substantial<br>equivalence is<br>claimed. | -Powder Free<br>-Patient Examination<br>Glove<br>-Single Use Only<br>-Manufactured For:<br>-Lot<br>-Blue color<br>-Non sterile | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>-Manufactured For:<br>-Lot<br>-Blue color<br>-Non sterile | Same | | | | | | # 6.0 Technological Characteristic Comparison: {5}------------------------------------------------ ### 7.0 Discussion of Non-clinical and Performance Testing Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards: ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. ASTM D6319-10(Reapproved 2015), Standard Specification for Nitrile Examination Gloves for Medical Application. {6}------------------------------------------------ # 8.0 Discussion of Clinical and Performance Testing Clinical testing is not needed for this device. # 9.0 Conclusion The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...