Powder Free Nitrile Patient Examination Gloves, Blue Color

K181106 · Jiangsu Dongxin Medical Technology Co., Ltd. · LZA · Jan 16, 2019 · General Hospital

Device Facts

Record IDK181106
Device NamePowder Free Nitrile Patient Examination Gloves, Blue Color
ApplicantJiangsu Dongxin Medical Technology Co., Ltd.
Product CodeLZA · General Hospital
Decision DateJan 16, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Device is a non-sterile, blue, nitrile patient examination glove. Used in clinical settings by healthcare professionals to provide a barrier against contamination between the examiner and the patient. The device is a single-use, disposable physical barrier. It functions by covering the hand or fingers of the examiner. No electronic, software, or algorithmic components are involved. Benefit is the reduction of cross-contamination risk during medical examinations.

Clinical Evidence

No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-10, ISO 10993-5), residual powder (ASTM D6124-06), pinhole detection (ASTM D5151-06), and physical properties/dimensions (ASTM D6319-10).

Technological Characteristics

Material: Nitrile. Form factor: Disposable, non-sterile, blue examination glove. Standards: ASTM D6319-10 (specifications), ASTM D6124-06 (residual powder), ASTM D5151-06 (pinholes), ISO 10993-5/10 (biocompatibility). No energy source, connectivity, or software.

Indications for Use

Indicated for use as a non-sterile, disposable, patient examination glove worn by medical examiners to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue. January 16, 2019 JiangSu DongXin Medical Technology Co., Ltd % Chu Xiaoan Official Correspondent Beijing Easylink Co., Ltd Rm. F302 Bldg., 41, Jing Cheng Ya Ju, Courtyard 6 of Southern Dou Ge Zhuang Beijing, 100021 Cn Re: K181106 Trade/Device Name: Powder Free Nitrile Patient Examination Gloves, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: November 27, 2018 Received: December 11, 2018 Dear Chu Xiaoan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Clarence W. Murray lii III -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K181106 ### Device Name Powder Free Nitrile Patient Examination Gloves, Blue Color Indications for Use (Describe) Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Type of Use (Select one or both, as applicable) ] Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # K181106 This summary of 510(k) is being submitted in accordance with 21 CFR 807.92 #### 1.0 Submitter: | Submitter's name : | JiangSu DongXin Medical Technology Co.,Ltd | |--------------------------|---------------------------------------------------------------------------------------------------------| | Submitter's address : | LongJin Road,The Economic Development Zone<br>of SuCheng,SuQian City,JiangSu Province,<br>223800 China. | | Name of contact person : | Deng Yujie | | Date of preparation : | 2019-01-14 | ### Designated Submission Correspondent | Company's name | Beijing Easy-Link Company | |--------------------|----------------------------------------------------------------------------------------------------------------------------------| | Company 's address | Rm. F302 Bldg., 41, Jing Cheng Ya Ju,<br>Courtyard 6 of Southern Dou Ge Zhuang,<br>Chaoyang District, Beijing 100121, P.R. China | | Contact person | Chu Xiaoan | #### 2.0 Name of the Device | Proprietary/Trade name: | Powder Free Nitrile Patient Examination<br>Gloves, Blue Color | |-------------------------|---------------------------------------------------------------| | Common Name: | Patient Examination gloves | | Classification Name: | Non-powdered Patient examination glove | | Device Classification: | I | | Regulation Number: | 21 CFR 880.6250 | | Panel: | General Hospital | | Product Code: | LZA | ## 3.0 Predicate device | Device Name: | Powder Free Nitrile Patient Examination<br>Glove, Blue Color | |----------------|--------------------------------------------------------------| | Company name: | Tangshan Zhonghong Pulin Plastic Co., Ltd. | | 510(K) Number: | K120970 | {4}------------------------------------------------ ### 4.0 Device Description: The proposed device is Powder Free Nitrile Examination Gloves. The proposed device is blue. The proposed device is non-sterile. ### 5.0 Indications for Use Statement: Powder Free Nitrile Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. #### Technological Characteristic Comparison Table: 6.0 | Features & Description | Predicate Device (K120970) | Subject Device (K181106) | Comparison | |-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Product name | Powder Free Nitrile Patient Examination Glove, Blue Color | Powder Free Nitrile Patient Examination Gloves, Blue Color | Same | | Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | same | | Product Code | LZA | LZA | same | | Color | Blue | Blue | same | | Size | Small/ Medium/ Large/X large | Small/ Medium/ Large/X large | same | | Indications for Use | Powder Free Nitrile Patient Examination Glove, Blue Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Powder Free Nitrile Patient Examination Gloves, Blue Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | same | | Device Description and Specifications | Meets ASTM D6319-10 | Meets ASTM D6319-10 | same | | Dimensions --Length<br>ILS-2 AQL4.0 | Meets ASTM D6319-10<br>≥230mm min | 232 mm min for all sizes | similar | | Dimensions -- Width<br>IL S-2 AQL4.0 | Meets ASTM D6319-10<br>Small 70-90 mm<br>Medium 85-105mm<br>Large 100-120mm<br>X large 110-130 mm | Small 76-90 mm<br>Medium 89-102 mm<br>Large 108-119mm<br>X large 115-128 mm | similar | | Dimensions --Thickness<br>IL S-2 AQL4.0 | Meets ASTM D6319-10<br>Finger 0.05mm min.<br>Palm 0.05mm min. | Thickness (mm) min.<br>Finger 0.08<br>Palm 0.08 | similar | {5}------------------------------------------------ | Physical Properties<br>IL S-2<br>AQL4.0 | | Aging<br>Elongation<br>(%) | Before | After | similar | |------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------|---------| | | Meets ASTM D D6319-10 | | 550-600 | 450-570 | | | | Before aging/after aging<br>Tensile Strength≥<br>14MPa | Tensile<br>Strength<br>(MPa) | 18-25 | 17-22 | | | | Before aging<br>Elongation ≥500% | | | | | | | After aging<br>Elongation ≥400% | | | | | | Freedom from<br>Pinholes<br>Inspection Level I<br>AQL2.5 | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10 | 1) Inspection Level I AQL2.5,and Accept/Reject criteria of 10/11<br>2) Water leakage test: 5 noncompliance is allowed. | | | similar | | Residual Powder | Meets ASTM<br>D 6124-06<br>(Reaffirmation 2011)<br><br>below 2mg of residual<br>powder | 1) Checked on 5pcs sub-samples (N=5).<br>2) Result as following:<br>Mean: 0.1mg/pcs | | | similar | | Materials used to<br>fabricate the devices | Nitrile | Nitrile | | | same | | Single Patient Use<br>Biocompatibility | Single Patient Use<br>Under the conditions<br>of this study, the test<br>article was a non-<br>irritant or non-<br>sensitizer | Single Patient Use<br>Under the conditions of<br>this study, the test<br>article was a non-<br>irritant or non-<br>sensitizer<br><br>Under the conditions of<br>this study, the test<br>article was non-<br>cytotoxicity to L-929<br>cells. | | | similar | | Labeling for the<br>legally marketed<br>device to which<br>substantial<br>equivalence is<br>claimed. | -Powder Free<br>-Patient Examination<br>Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot<br>-Blue color | -Powder Free<br>-Patient Examination<br>Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot<br>-Blue color | | | same | ### 7.0 Discussion of Non-Clinical Performance Testing Non-clinical tests were conducted to verify that the proposed device will meet acceptance crtiteria for each test. The test results demonstrated that the proposed device met the {6}------------------------------------------------ acceptance criteria found in the following standards below: ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves ASTM D5151-06(Reapproved2015), Standard Test Method for Detection of Holes in Medical Gloves. ASTM D6319-10(Reapproved 2015), Standard Specification for Nitrile Examination Gloves for Medical Application. ### 8.0 Discussion of Clinical and Performance Testing Clinical testing is not needed for this device. ### 9.0 Conclusion The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...