Synthetic Nitrile Patient Exam Gloves, Powder Free, Blue, Tested for Use w/Chemotherapy Drug
K200093 · Anhui Lntco Medical Products Co., Ltd. · LZA · Aug 20, 2020 · General Hospital
Device Facts
Record ID
K200093
Device Name
Synthetic Nitrile Patient Exam Gloves, Powder Free, Blue, Tested for Use w/Chemotherapy Drug
Applicant
Anhui Lntco Medical Products Co., Ltd.
Product Code
LZA · General Hospital
Decision Date
Aug 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The synthetic nitrile patient examination gloves, powder free, blue color and tested for use with chemotherapy drugs are disposable devices intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner. The glove was tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Synthetic nitrile patient examination gloves; powder-free; blue color. Designed as disposable barrier protection for medical examiners to prevent cross-contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings by healthcare professionals. Provides physical barrier; protects examiner from patient contaminants and chemotherapy agents. Warning: Do not use with Carmustine.
Clinical Evidence
Bench testing only. Device tested for permeation resistance to chemotherapy drugs per ASTM D6978-05 (Reapproved 2019). Breakthrough detection times provided for 13 chemotherapy agents; results indicate >240 minutes for most, with exceptions for Thiotepa (46.9 min) and Carmustine (22.6 min).
Technological Characteristics
Material: Synthetic nitrile. Form factor: Powder-free, blue, disposable patient examination glove. Testing standard: ASTM D6978-05 (Reapproved 2019) for chemotherapy drug permeation.
Indications for Use
Indicated for use as a disposable medical glove worn on examiner's hands to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K170136 — Synthetic Nitrile Patient Examination Gloves, Powder Free, Blue Color, and Tested for Use with Chemotherapy Drugs · Shandong Feiyang Professional Gloves Co., Ltd. · Aug 25, 2017
K181066 — Powder Free Nitrile Examination Glove (Blue) with Low Dermatitis claim and with tested for Use with Chemotherapy Drugs Claims · Eco Medi Glove Sdn. Bhd. · Sep 26, 2018
K214110 — Nitrile Powder Free Blue Patient Examination Gloves Non-Sterile, Tested For Use With Chemotherapy Drugs · Mah Sing Healthcare Sdn. Bhd. · Feb 19, 2022
K211390 — Nitrile Blue Powder-free Examination Glove · Cardinal Health200, LLC · Aug 28, 2021
K212532 — Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs) · Crdlight Optoelectronic Technology Co., Ltd. · Nov 11, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 20, 2020
Anhui Intco Medical Products Co., Ltd. % Jason Ji Official Correspondent lntco Medical Industries, Inc. 805 Barrington Ave Ontario, California 91764
Re: K200093
Trade/Device Name: Synthetic Nitrile Patient Exam Gloves, Powder Free, Blue, Tested for Use w/Chemotherapy Drug Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: May 16, 2020 Received: Jun 02, 2020
Dear Jason Ji:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200093
### Device Name
Synthetic Nitrile Patient Examination Gloves, Powder Free, Blue Color, and Tested for Use with Chemotherapy Drugs
### Indications for Use (Describe)
The synthetic nitrile patient examination gloves, powder free, blue color and tested for use with chemotherapy drugs are disposable devices intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner. The glove was tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug and Concentration | Breakthrough Detection Time<br>in Minutes |
|----------------------------------------------|-------------------------------------------|
| Blenoxane (Bleomycin), 15 mg/ml (15,000 ppm) | >240 |
| Carmustine (BCNU), 3.3 mg/ml (3,300 ppm) | 22.6 |
| Cyclophosphamide, 20 mg/ml (20,000 ppm) | >240 |
| Cytarabine, 100 mg/ml (100,000 ppm) | >240 |
| Etoposide (Toposar), 20 mg/ml (20,000 ppm) | >240 |
| Fluorouracil (5 Flu), 50 mg/ml (50,000 ppm) | >240 |
| Idarubicin, 1 mg/ml (1,000 ppm) | >240 |
| Mesna, 100 mg/ml (100,000 ppm) | >240 |
| Mitomycin C, 0.5 mg/ml (500 ppm) | >240 |
| Paclitaxel, 6 mg/ml (6,000 ppm) | >240 |
| Thiotepa, 10 mg/ml (10,000 ppm) | 46.9 |
| Trisenox, 1 mg/ml (1,000 ppm) | >240 |
| Vincristine Sulfate, 1 mg/ml (1,000 ppm) | >240 |
1. CAUTION: Testing showed an average breakthrough time of 46.9 minutes with Thiotepa
and 22.6 minutes with Carmustine
2. WARNING: Do not use with Carmustine.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions. search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.