Powder Free Nitrile Patient Examination Glove, Blue Colored, Non Sterile, Tested for Use with Chemotherapy Drugs

K171873 · Zhonghong Pulin Medical Products Co., Ltd. · LZA · Nov 15, 2017 · General Hospital

Device Facts

Record IDK171873
Device NamePowder Free Nitrile Patient Examination Glove, Blue Colored, Non Sterile, Tested for Use with Chemotherapy Drugs
ApplicantZhonghong Pulin Medical Products Co., Ltd.
Product CodeLZA · General Hospital
Decision DateNov 15, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Story

Disposable, non-sterile, blue nitrile patient examination glove; provides barrier protection for examiner's hands; prevents cross-contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings by healthcare professionals. Output is physical barrier protection; informs clinical decision-making regarding safe handling of hazardous chemotherapy agents. Benefits include reduced risk of exposure to patient pathogens and hazardous drugs.

Clinical Evidence

No clinical data. Bench testing only. Compliance verified via ASTM D6319-10 (physical/dimensional), ASTM D5151-06 (pinholes), ASTM D6124-06 (powder residual), and ASTM D6978-05 (chemotherapy permeation). Biocompatibility testing per ISO 10993-10:2010 confirmed non-irritating and non-sensitizing.

Technological Characteristics

Nitrile compound; powder-free; non-sterile; blue color. Dimensions and physical properties meet ASTM D6319-10. Barrier protection verified via ASTM D5151-11. Biocompatibility per ISO 10993-10:2010. Minimum thickness 0.05mm; minimum length 230mm.

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. November 15, 2017 Zhonghong Pulin Medical Products Co., Ltd. % Da Shi Manager Zhonghong Pulin Medical Products Co.,Ltd. 800 - 10 Four Seasons Place Toronto, M9B 6H7 Ca Re: K171873 Trade/Device Name: Powder Free Nitrile Patient Examination Glove, Blue Colored, Non Sterile, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: October 13, 2017 Received: October 13, 2017 Dear Da Shi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K171873 Device Name Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs Indications for Use (Describe) A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. | Drug and Concentration | Minimum Breakthrough Detection Time in Minute | |-------------------------------------|-----------------------------------------------| | Carmustine (BCNU) 3.3 mg/ml | 14.7 | | Cisplatin 1.0 mq/ml | >240 | | Cyclophosphamide (Cytoxan) 20 mg/ml | >240 | | Cytarabine 100 mq/ml | >240 | | Dacarbazine (DTIC) 10.0 mg/ml | >240 | | Doxorubicin Hydrochloride 2.0 mq/ml | >240 | | Etoposide (Toposar) 20.0 mg/ml | >240 | | Fluorouracil 50.0 mq/ml | >240 | | Ifosfamide 50.0 mq/ml | >240 | | Methotrexate 25 mq/ml | >240 | | Mitomycin C 0.5 mq/ml | >240 | | Mitoxantrone 2.0 mq/ml | >240 | | Paclitaxel (Taxol) 6.0 mq/ml | >240 | | Thiotepa 10.0 mq/ml | 58.8 | | Vincristine Sulfate 1.0 mq/ml | >240 | Please note the following drugs have extremely low permeation times: Carmustine (BCNU) (3.3mg/ml) 14.7 minutes and Thiotepa (10 mg/ml) 58.8 minutes. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Zhong Hong Pulin. The logo consists of a red square with the letters "Z" and "Z" separated by a vertical line inside. To the right of the square are the words "ZHONG HONG" in a larger font, with "PULIN" underneath in a smaller font. The text is black. # 510(k) SUMMARY Preparation Date: Nov. 8th, 2017 #### Submitter: | Company Name: | Zhonghong Pulin Medical Products Co., Ltd. | |-----------------|----------------------------------------------------------------------| | Address: | Pachigang Industrial Park, Luannan County<br>Tangshan, Hebei, 063502 | | Country: | China | | Phone No.: | +86 315-4165760 | | Fax No.: | +86 315-4167664 | | E-mail Address: | RA@zhonghongpulin.com | #### Contact Person: | Contact Person : | Li, Jinge (Regulatory Affairs Manager) | |------------------|----------------------------------------| | Contact Email: | ljg@zhonghongpulin.cn | #### Name of the Device: Device trade or proprietary name: Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs Device common or usual name: Patient Examination Glove Device Classification Name: LZA - Polymer Patient Examination Glove Device Classification Name: LZC - Patient Examination Glove, Specialty Regulation Number: 21 CFR 88.6250 FDA Device Class: Class 1 Product Code: LZA, LZC #### Predicate Device: Class I patient Examination glove and tested for use with Chemotherapy Drugs, Powder Free, LZC, which meets all the requirement of ASTM D 6319-10 and FDA 21 CFR 880.6250. Device Name: Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs Copmany Name: Kossan International Sdn. Bhd. 510(K) Number: K151750 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains a logo for Zhong Hong Pulin. The logo consists of a red square with the letters "Z" and "Z" separated by a vertical line. To the right of the square are the words "ZHONG HONG" in a larger font, with the word "PULIN" underneath in a smaller font. #### Device Description: The subject device in this 510(k) Notification is a Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs The subject device is a patient examination glove made from nitrile compound, blue in color, powder free and non-sterile (as per 21 CFR 880.6250, class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves. ### Intended use of the Device (Indication for use) A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. | Drug and Concentration | Minimum Breakthrough Detection Time in Minutes | |-------------------------------------|------------------------------------------------| | Carmustine (BCNU) 3.3 mg/ml | 14.7 | | Cisplatin 1.0 mg/ml | >240 | | Cyclophosphamide (Cytoxan) 20 mg/ml | >240 | | Cytarabine 100 mg/ml | >240 | | Dacarbazine (DTIC) 10.0 mg/ml | >240 | | Doxorubicin Hydrochloride 2.0 mg/ml | >240 | | Etoposide (Toposar) 20.0 mg/ml | >240 | | Fluorouracil 50.0 mg/ml | >240 | | Ifosfamide 50.0 mg/ml | >240 | | Methotrexate 25 mg/ml | >240 | | Mitomycin C 0.5 mg/ml | >240 | | Mitoxantrone 2.0 mg/ml | >240 | | Paclitaxel (Taxol) 6.0 mg/ml | >240 | | Thiotepa 10.0 mg/ml | 58.8 | | Vincristine Sulfate 1.0 mg/ml | >240 | Please note that the following drugs have extremely low permeation times: Carmustine (BCNU) 14.7 minutes and Thiotepa 58.8 minutes. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains a logo for "ZHONG HONG PULIN". The logo consists of a red square with the letters "Z" and "Z" inside, separated by a vertical line. To the right of the square, the words "ZHONG HONG" are written in a larger font, and below that, the word "PULIN" is written in a smaller font. #### Summary of the Technological Characteristics of the Device: The subject device is summarized with the following technological characteristics compared to ASTM or equivalent standard. | Characteristics | Standard | Device performance | |-----------------------|-------------------------------------------------------------|------------------------------------------------------------------| | Dimension | ASTM standard D 6319-10 | Meets | | Physical Properties | ASTM standard D 6319-10 | Meets | | Freedom from pinholes | 21 CFR 800.20 | Meets | | | ASTM D5151-11 | | | Powder Residual | ASTM D6319-10 and D6124-<br>06(Reapproved 2011) | Meets | | | | | | Biocompatibility | Primary Skin Irritation ISO<br>10993-10:2010 | Not a primary skin irritant under<br>the conditions of the study | | | Dermal sensitization in the<br>guinea pig ISO 10993-10:2010 | Not a contact sensitizer under<br>the conditions of the study | #### Substantial Equivalence Based on Assessment of Non-Clinical Performance Data Bench tests were conducted to verify that the proposed device met all specifications and the proposed device is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards: ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical application. ASTM D5151-06 (2011), Standard Test Method for Detection of Holes in Medical Gloves. ASTM D6124-06 (2011). Standard Test Method for Residual Powder on Medical Gloves. ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs The subject device meets the same test standards conducted by the predicate device. Both device has the same sizes, color, physical and dimensional characteristics. The minimum breakthrough detection time of Carmustine for the subject device is below 30 minutes, similar with predicate K151750 (Blue). The minimum breakthrough detection time of Thiotepa for the subject device is at 58.8 minutes. The subject device is having longer permeation time than predicate K151750 (Blue). Warning statement (Do Not Use with Carmustine and Thiotepa) for subject device is included in Labeling, similar with predicate device. The subject device has similar thickness with predicate K151750 at palm, and similar length with predicate K151750 (Blue). {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains a logo with the text "ZHONG HONG" on the top line and "PULIN" on the bottom line. To the left of the text is a red square containing two white letters, "Z" and "Z", separated by a vertical white line. The logo appears to be for a company or organization named Zhong Hong Pulin. The subject device is having identical specification with predicate K151750 (Blue) with thickness at minimum 0.05mm and length at minimum 230mm. The difference in labeling with additional drugs tested do not affect the safety and effectiveness of the subject device. The subject device and the predicate device K151750 (Blue) share the same intended use, same Nitrile material, same compliance with ASTM standards. There is no difference between the subject device and the predicate device K151750 (Blue) with respect to intended use, nonclinical performance data and technological characteristics. Consequently, the gloves that are the subject of this submission are substantially equivalent to a legally marketed glove K151750 (Blue). #### Substantial Equivalent Based on Assessment of Clinical Performance Data Clinical data is not needed for this submission. #### Biocompatibility Biocompatibility tests indicated that under the conditions of the studies, the gloves were nonsensitizing and non-irritating. #### Legally Marketed Device to which Substantial Equivalence is Claimed The legally marketed predicate device in scope is as follows: K151750 - Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...