SHOWA(R) Blue Nitrile Powder Free Medical Examination Glove
K211003 · Showa Best Glove, Inc. · LZA · Jan 19, 2022 · General Hospital
Device Facts
Record ID
K211003
Device Name
SHOWA(R) Blue Nitrile Powder Free Medical Examination Glove
Applicant
Showa Best Glove, Inc.
Product Code
LZA · General Hospital
Decision Date
Jan 19, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Glove to Permeation by chemotherapy drugs.
Device Story
Disposable, non-sterile, blue nitrile butadiene rubber medical examination glove; worn by clinicians to prevent cross-contamination during patient examinations. Device provides barrier protection; tested for permeation resistance against 29 chemotherapy drugs per ASTM D6978-05. Used in clinical settings; operated by healthcare professionals. Output is physical barrier protection; informs clinical decision-making by enabling safe handling of chemotherapy agents. Benefits include reduced risk of exposure to hazardous drugs and patient-clinician contamination.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing, including ASTM D6319-10 (dimensions, tensile strength, elongation, freedom from holes, residual powder) and ASTM D6978-05 (chemotherapy drug permeation). Biocompatibility testing (ISO 10993-10, ISO 10993-11) confirmed the device is not a primary skin irritant, not a contact sensitizer, and not toxic.
Technological Characteristics
Nitrile butadiene rubber; blue; powder-free; non-sterile. Dimensions and physical properties meet ASTM D6319-10. Permeation resistance tested per ASTM D6978-05. Biocompatibility per ISO 10993-10 and ISO 10993-11. No software or electronic components.
Indications for Use
Indicated for use as a disposable medical device worn on examiner's hands or fingers to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs per ASTM D6978-05.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
January 19, 2022
Showa Best Glove, Inc. Jeffrey Richardson Director of Operations 579 Edison Street Menlo, Georgia 30731
Re: K211003
Trade/Device Name: SHOWA® Blue Nitrile Powder Free Medical Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: December 8, 2021 Received: December 9, 2021
Dear Jeffrey Richardson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211003
Device Name
SHOWA®Blue Nitrile Powder Free Medical Examination Glove
#### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Glove to Permeation by chemotherapy drugs.
Chemotherapy Drugs and Concentration Minimum breakthrough detection time (Min) ug/cm2/min Chemotherapy Drug and concentration Minimum Breakthrough, detection time in minutes, minutes 1) Blenoxane (15.0 mg/ml) 15,000 ppm >240 minutes 2) Busulfan (6.0 mg/ml) 6,000 ppm >240 minutes 3) Carboplatin (10.0 mg/ml) 10,000 ppm >240 minutes 4) Carmustine (3.3 mg/ml) 3,300 ppm 73.7 minutes 5) Cisplatin (1.0 mg/ml) 1,000 ppm >240 minutes 6) Cyclophosphamide (20.0 mg/ml) 20,000 ppm >240 minutes 7) Cytarabine (100.0 mg/ml) 100,000 ppm >240 minutes 8) Dacarbazine (10.0 mg/ml) 10,000 ppm >240 minutes 9) Daunorubicin HCl (5.0 mg/ml) 5,000 ppm >240 minutes 10) Docetaxel (10.0 mg/ml) 10,000 ppm >240 minutes l 1) Doxorubicin HCl (2.0 mg/ml) 2,000 ppm >240 minutes 12) Epirubicin HCl (2.0 mg/ml) 2,000 ppm >240 minutes 13) Etoposide (20.0 mg/ml) 20,000 ppm >240 minutes 14) Fludarabine (25.0 mg/ml) 25.000 ppm >240 minutes 15) Fluorouracil (50.0 mg/ml 50,000 ppm >240 minutes 16) Gemcitabine (38.0 mg/ml) 38,000 ppm >240 minutes 17) Idarubicin HCl (1.0 mg/ml) 1,000 ppm >240 minutes 18) Ifosfamide (50.0 mg/ml) 50,000 ppm >240 minutes 19) Irinotecan (20.0 mg/ml) 20,000 ppm >240 minutes 20) Mechlorethamine HCl (1.0 mg/ml) 1,000 ppm >240 minutes 21) Melphalan (5.0 mg/ml) 5,000 ppm >240 minutes 22) Methotrexate (25.0 mg/ml) 25,000 ppm >240 minutes 23) Mitomycin C (0.5 mg/ml) 500 ppm >240 minutes 24) Mitoxantrone (2.0 mg/ml) 2,000 ppm >240 minutes 25) Paclitaxel (6.0 mg/ml) 6,000 ppm >240 minutes 26) Rituximab (10.0 mg/ml) 10,000 ppm >240 minutes 27) ThioTepa (10.0 mg/ml)10,000 ppm 25.4 minutes 28) Trisenox (1.0 mg/ml) 1,000 ppm >240 minutes 29) Vincristine Sulfate (1.0 mg/ml) 1,000 ppm >240 minutes The maximum testing time is 240 minutes. Warning: Do not use with Carmustine or ThioTepa.
. ..
{4}------------------------------------------------
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary K211003 DATE OF PREPARATION: January 14, 2022
#### I. SUBMITTER
Showa Best Glove, Inc. 579 Edison Street Menlo. GA 30731 Tel: 706-862-6740 Fax: 706-862-6000
#### II. OFFICIAL CORRESPONDENCE/CONTACT PERSON
Jeffrey Richardson Showa Best Glove, Inc. 579 Edison Street Menlo, GA 30731 Tel: 706-862-6740 Fax: 706-862-6000 Email: Bmoseley@showagroup.com
#### III. 510(K) PREPARER
Brian Moseley Showa Best Glove, Inc. 579 Edison Street Menlo, GA 30731 Tel: 706-862-6752 Fax: 706-862-2660 Email: Bmoseley@showagroup.com
#### IV.PROPOSED DEVICE
Trade Name/Proprietary Name: SHOWA® Blue Nitrile Powder Free Medical Examination Glove Common or Usual Name: Nitrile Patient Examination Glove Classification Name: Patient Examination Glove Regulation: 21 CFR 880.6250 Device Classification: Class: I Product Code: LZA, LZC
{6}------------------------------------------------
### V. PREDICATE DEVICE
Device Classification Name: Polymer Patient Examination Glove 510(k) Number: K200581 Product Code: LZA, LZC Applicant: Hartelega NGC SDN.BHD. Kawasan Perindustrian Tanjung Sepang, MY 43900
#### VI.DEVICE DESCRIPTION
The SHOWA® Blue Nitrile Powder Free Medical Examination Glove is a single use, disposable device made from a Nitrile Butadiene Rubber, blue in color, powder free and non-sterile (per 21 CFR Part 880.6250, class I). The device meets all the specifications in ASTM D6319-10. Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs.
The SHOWA® Blue Nitrile Powder Free Medical Examination Glove is designed to be used for medical purposes to be worn on the examiner's hands or fingers to prevent contamination between the patient and the examiner (product code LZA). In addition, these gloves were tested for use with chemotherapy drugs (product code LZC) in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Glove to Permeation by chemotherapy drugs.
#### VII. INDICATIONS FOR USE
A patient examination qlove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Glove to Permeation by chemotherapy drugs.
| Chemotherapy Drug and concentration | Minimum Breakthrough, detection<br>time in minutes, minutes |
|---------------------------------------------|-------------------------------------------------------------|
| 1) Blenoxane (15.0 mg/ml) 15,000 ppm | >240 |
| 2) Busulfan (6.0 mg/ml) 6,000 ppm | >240 |
| 3) Carboplatin (10.0 mg/ml) 10,000 ppm | >240 |
| 4) Carmustine (3.3 mg/ml) 3,300 ppm | 73.7 |
| 5) Cisplatin (1.0 mg/ml) 1,000 ppm | >240 |
| 6) Cyclophosphamide (20.0 mg/ml) 20,000 ppm | >240 |
| 7) Cytarabine (100.0 mg/ml) 100,000 ppm | >240 |
| 8) Dacarbazine (10.0 mg/ml) 10,000 ppm | >240 |
| 9) Daunorubicin HCI (5.0 mg/ml) 5,000 ppm | >240 |
| 10) Docetaxel (10.0 mg/ml) 10,000 ppm | >240 |
| 11) Doxorubicin HCI (2.0 mg/ml) 2,000 ppm | >240 |
| 12) Epirubicin HCI (2.0 mg/ml) 2,000 ppm | >240 |
| 13) Etoposide (20.0 mg/ml) 20,000 ppm | >240 |
| 14) Fludarabine (25.0 mg/ml) 25,000 ppm | >240 |
{7}------------------------------------------------
| 15) Fluorouracil (50.0 mg/ml 50,000 ppm | >240 |
|------------------------------------------|------|
| 16) Gemcitabine (38.0 mg/ml) 38,000 ppm | >240 |
| 17) Idarubicin HCl (1.0 mg/ml) 1,000 ppm | >240 |
| 18) Ifosfamide (50.0 mg/ml) 50,000 ppm | >240 |
{8}------------------------------------------------
| 19) Irinotecan (20.0 mg/ml) 20,000 ppm | >240 |
|-----------------------------------------------|------|
| 20) Mechlorethamine HCl (1.0 mg/ml) 1,000 ppm | >240 |
| 21) Melphalan (5.0 mg/ml) 5,000 ppm | >240 |
| 22) Methotrexate (25.0 mg/ml) 25,000 ppm | >240 |
| 23) Mitomycin C (0.5 mg/ml) 500 ppm | >240 |
| 24) Mitoxantrone (2.0 mg/ml) 2,000 ppm | >240 |
| 25) Paclitaxel (6.0 mg/ml) 6,000 ppm | >240 |
| 26) Rituximab (10.0 mg/ml) 10,000 ppm | >240 |
| 27) ThioTepa (10.0 mg/ml)10,000 ppm | 25.4 |
| 28) Trisenox (1.0 mg/ml) 1,000 ppm | >240 |
| 29) Vincristine Sulfate (1.0 mg/ml) 1,000 ppm | >240 |
The maximum testing time is 240 minutes. Warning: Do not use with Carmustine or ThioTepa.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS VIII.
| | Proposed Device | | Predicate Device | | Comparison |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|------------|
| Characteristics | SHOWA® Blue Nitrile Powder Free Medical<br>Examination Glove K211003 | | Polymer Patient Examination Glove<br>510(k) Number K200581 | | |
| Device<br>Description/<br>Regulation<br>Number | Patient examination glove 21 CFR § 880.6250 | | Patient examination glove 21 CFR §<br>880.6250 | | Same |
| Product Code | LZA, LZC | | LZA, LZC, QDO | | Similar |
| Indications for<br>Use | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands or fingers to<br>prevent contamination<br>between patient and<br>examiner.<br>These gloves were<br>tested for use with<br>chemotherapy drugs in<br>accordance with ASTM<br>D6978-05 Standard<br>Practice for Assessment<br>of Medical Glove to<br>Permeation by<br>chemotherapy drugs.<br>Chemotherapy Drugs and<br>Concentration<br>1) Blenoxane (15.0 mg/ml)<br>15,000 ppm<br>2) Busulfan (6.0 mg/ml) 6,000<br>ppm<br>3) Carboplatin (10.0 mg/ml)<br>10,000 ppm<br>4) Carmustine (3.3 mg/ml)<br>3,300 ppm | Minimum<br>Breakthrough,<br>detection time in<br>minutes, minutes<br>>240 minutes<br>>240 minutes<br>>240 minutes<br>73.7 minutes | Biodegradable Nitrite Powder Free<br>Examination Gloves Tested for Use with<br>Chemotherapy Drugs and Fentanyl<br>Citrate (Blue) is a non-sterile disposable<br>device intended for medical purpose that<br>is worn on the examiner's hand to prevent<br>contamination between patient and<br>examiner. It is also tested to be used<br>against Chemotherapy Drugs and<br>Fentanyl Citrate, These gloves were<br>tested for use with chemotherapy drugs<br>and Fentanyl Citrate as per ASTM<br>D6978-05 (Reapproved 2013) Standard<br>Practice for Assessment of Medical<br>Gloves to Permeation by Chemotherapy<br>Drugs.<br>Chemotherapy Drug and<br>Concentration<br>Carmustine (3.3 mg/ml)<br>Cisplatin (1.0 mg/ml)<br>Cyclophosphamide (20.0<br>mg/ml)<br>Dacarbazine (10.0 mg/ml)<br>Doxorubicin Hydrochloride<br>(2.0 mg/ml)<br>Etoposide (20.0 mg/ml)<br>Fluorouracil (50.0 mg/ml)<br>Methotrexate (25.0 mg/ml) | Minimum<br>Breakthrough<br>Detection<br>Time in<br>Minutes<br>21.4<br>>240<br>>240<br>>240<br>>240<br>>240<br>>240<br>>240 | Similar |
{9}------------------------------------------------
| 5) Cisplatin (1.0 mg/ml) 1,000 ppm | >240 minutes | Paclitaxel (6.0 mg/ml) | >240 |
|---------------------------------------------|--------------|---------------------------------|------|
| 6) Cyclophosphamide (20.0 mg/ml) 20,000 ppm | >240 minutes | Thiotepa (10.0 mg/ml) | 67.2 |
| 7) Cytarabine (100.0 mg/ml) | >240 minutes | Vincristine Sulfate (1.0 mg/ml) | >240 |
| | | Azacytidine (25.0 mg/ml) | >240 |
{10}------------------------------------------------
| | 100,000 ppm | | | | | | | | |
|-----------------------------------|----------------------------------------------------------------------------------------------|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|--------------------------------------------------------------------|--|------|--|--|
| | 8) Dacarbazine (10.0 mg/ml)<br>10,000 ppm | >240 minutes | Carboplatin (10.0 mg/ml) | >240 | | | | | |
| | 9) Daunorubicin HCI (5.0 mg/ml) 5,000 ppm | >240 minutes | Docetaxel (10 mg/ml) | >240 | | | | | |
| | 10) Docetaxel (10.0 mg/ml) 10,000 ppm | >240 minutes | Epirubicin (2.0 mg/ml) | >240 | | | | | |
| | 11) Doxorubicin HCI (2.0 mg/ml) 2,000 ppm | >240 minutes | Gemcitabine (38 mg/ml) | >240 | | | | | |
| | 12) Epirubicin HCI (2.0 mg/ml) 2,000 ppm | >240 minutes | Ifosfamide (50 mg/ml) | >240 | | | | | |
| | 13) Etoposide (20.0 mg/ml) 20,000 ppm | >240 minutes | Irinotecan (20 mg/ml) | >240 | | | | | |
| | 14) Fludarabine (25.0 mg/ml) 25,000 ppm | >240 minutes | Mitoxantrone (2.0 mg/ml) | >240 | | | | | |
| | 15) Fluorouracil (50.0 mg/ml 50,000 ppm | >240 minutes | Oncovin (1.0 mg/ml) | >240 | | | | | |
| | 16) Gemcitabine (38.0 mg/ml) 38,000 ppm | >240 minutes | Oxaliplatin (5 mg/ml) | >240 | | | | | |
| | 17) Idarubicin HCI (1.0 mg/ml) 1,000 ppm | >240 minutes | Vinorelbine () 0 mg/ml) | >240 | | | | | |
| | 18) Ifosfamide (50.0 mg/ml) 50,000 ppm | >240 minutes | Please note that Carmustine and Thiotepa have extremely low penetration times of 21.4 minutes and 67.2 minutes respectively.<br>Warning: Do not use with Carmustine | | | | | | |
| | 19) Irinotecan (20.0 mg/ml) 20,000 ppm | >240 minutes | | | | | | | |
| | 20) Mechlorethamine HCI (1.0 mg/ml) 1,000 ppm | >240 minutes | | | | | | | |
| | 21) Melphalan (5.0 mg/ml) 5,000 ppm | >240 minutes | | | | | | | |
| | 22) Methotrexate (25.0 mg/ml) 25,000 ppm | >240 minutes | | | | | | | |
| | 23) Mitomycin C (0.5 mg/ml) 500 ppm | >240 minutes | Fentanyl Citrate and<br>Concentration | Minimum<br>Breakthrough<br>Detection Time<br>in Minutes | | | | | |
| | 24) Mitoxantrone (2.0 mg/ml) 2,000 ppm | >240 minutes | | | | | | | |
| | 25) Paclitaxel (6.0 mg/ml) 6,000 ppm | >240 minutes | | | | | | | |
| | 26) Rituximab (10.0 mg/ml) 10,000 ppm | >240 minutes | Fentanyl Citrate Injection<br>(100 mcg/2ml) | >240 | | | | | |
| | 27) ThioTepa (10.0 mg/ml)10,000 ppm | 25.4 minutes | | | | | | | |
| | 28) Trisenox (1.0 mg/ml) 1,000 ppm | >240 minutes | | | | | | | |
| | 29) Vincristine Sulfate (1.0 mg/ml) 1,000 ppm | >240 minutes | | | | | | | |
| | The maximum testing time is 240 minutes.<br>Warning: Do not use with Carmustine or ThioTepa. | | | | | | | | |
| Material Use | Nitrile | | Nitrile | | Same | | | | |
| Color | Blue | | Blue | | Same | | | | |
| Sterility | Non-Sterile | | Non-Sterile | | Same | | | | |
| Dimensions | Meets ASTM D6319-10 | | Meets ASTM D6319-10 | | Same | | | | |
| Physical<br>Properties | Meets ASTM D6319-10 | | Meets ASTM D6319-10 | | Same | | | | |
| Freedom from<br>Holes<br>(D 5151) | AQL 2.5 Inspection Meets ASTM D5151-06 | | AQL 2.5 Inspection Meets ASTM D5151-06 | | Same | | | | |
| Residual Powder<br>(D 6124) | Meets ASTM D6124-06 | | Meets ASTM D6124-06 | | Same | | | | |
| Biocompatibility | Biocompatibility<br>test – Primary<br>Skin Irritation Test<br>(ISO 10993-10) | | Under the conditions of the study, not a<br>primary skin irritant. | | Under the conditions of the study, not a<br>primary skin irritant. | | Same | | |
{11}------------------------------------------------
| Biocompatibility<br>test – Dermal<br>Sensitization<br>Assay (ISO<br>10993-10) | Under conditions of the study, not a contact<br>sensitizer | Under conditions of the study, not a<br>contact sensitizer | Same |
|-------------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|------|
| Biocompatibility<br>test – Acute | Under conditions of the ISO Acute Systemic<br>Injection test, not toxic | Under conditions of the ISO Acute<br>Systemic Injection test, not toxic | Same |
{12}------------------------------------------------
| systemic toxicity<br>Study (ISO<br>10993-11) | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Resistance<br>against<br>Chemotherapy<br>Drugs Standards<br>Practice for<br>Assessment of<br>resistance of<br>Medical Glove to<br>Permeation by<br>Chemotherapy<br>drugs<br>ASTM D6978-5<br>(2013) | Breakthrough greater than 240 min.<br>• Blenoxane (15.0 mg/ml)<br>• Busulfan (6.0 mg/ml)<br>• Carboplatin (10.0 mg/ml)<br>• Cisplatin (1.0 mg/ml)<br>• Cyclophosphamide (20.0 mg/ml)<br>• Cytarabine (100.0 mg/ml)<br>• Dacarbazine (10.0 mg/ml)<br>• Daunorubicin HCI (5.0 mg/ml)<br>• Docetaxel (10.0 mg/ml)<br>• Doxorubicin HCI (2.0 mg/ml)<br>• Epirubicin HCI (2.0 mg/ml)<br>• Etoposide (20.0 mg/ml)<br>• Fludarabine (25.0 mg/ml)<br>• Fluorouracil (50.0 mg/ml)<br>• Gemcitabine (38.0 mg/ml)<br>• Idarubicin HCI (1.0 mg/ml)<br>• Ifosfamide (50.0 mg/ml)<br>• Irinotecan (20.0 mg/ml)<br>• Mechlorethamine HCI (1.0 mg/ml)<br>• Melphalan (5.0 mg/ml)<br>• Methotrexate (25.0 mg/ml)<br>• Mitomycin C (0.5 mg/ml)<br>• Mitoxantrone (2.0 mg/ml)<br>• Paclitaxel (6.0 mg/ml)<br>• Rituximab (10.0 mg/ml)<br>• Trisenox (1.0 mg/ml)<br>• Vincristine Sulfate (1.0 mg/ml)<br><br>Breakthrough at 73.7 min.<br>Carmustine (3.3 mg/ml)<br><br>Breakthrough at 25.4 min.<br>ThioTepa (10.0 mg/ml) | Breakthrough greater than 240 min.<br>• Cisplatin (1.0 mg/ml)<br>• Cyclophosphamide (20.0 mg/ml)<br>• Dacarbazine (10.0 mg/ml)<br>• Doxorubicin Hydrochloride (2.0<br>mg/ml)<br>• Etoposide (20.0 mg/ml)<br>• Fluorouracil (50.0 mg/ml)<br>• Methotrexate (25.0 mg/ml)<br>• Mitomycin C (0.5 mg/ml)<br>• Paclitaxel (6.0 mg/ml)<br>• Vincristine Sulfate (1.0 mg/ml)<br>• Azacytidine (25.0 mg/m1)<br>• Carboplatin (10.0 mg/ml)<br>• Docetaxel (10 mg/ml)<br>• Epirubicin (2.0 mg/ml)<br>• Gemcitabine (38 mg/ml)<br>• Ifosfamide (50 mg/ml)<br>• Irinotecan (20 mg/ml)<br>• Mitoxantrone (2.0 mg/ml)<br>• Oncovin (1.0 mg/ml)<br>• Oxaliplatin (5 mg/ml)<br>• Vinorelbine (10 mg/ml)<br>• Fentanyl Citrate (100 mcg/2ml)<br><br>Breakthrough at 21.4 min.<br>Carmustine (3.3 mg/ml)<br><br>Breakthrough at 67.2 min.<br>Thiotepa (10.0 mg/ml) | Similar |
# IX. SUMMARY OF NONCLINICAL TESTING
| Test Method | Purpose | Acceptance Criteria | Results |
|---------------|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| ASTM D6319-10 | Demonstrate<br>accurate sizing | Dimensions<br>Overall Length (mm) =<br>220 mm (sizes XS – S) and<br>230 mm (sizes M - XL)<br>Width (± 10 mm)<br>Size XS = 70 mm Size S = 80 mm Size M = 95 mm<br>Size L = 110 mm Size XL = 120 mm<br>Thickness at Palm (mm) = 0.05 min.<br>Thickness at Finger Tip (mm) = 0.05 min. | Pass |
{13}------------------------------------------------
| ASTM D6319-10 | Demonstrate<br>tensile<br>conforming<br>properties | Before Aging (D 412):<br>Tensile Strength (MPa) = 14 min<br>Ultimate Elongation (%) = 500 min. | Pass |
|---------------|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|------|
| | | After Aging (D 573) (70°C ± 2°C for 166 hrs ± 2 hrs.):<br>Tensile Strength (MPa) = 14 min<br>Ultimate Elongation (%) = 400 min | |
{14}------------------------------------------------
| ASTM D6319-10 | Demonstrate<br>glove integrity | Freedom from Holes (D 5151)<br>AQL 2.5 Inspection Level G-1 | Pass |
|--------------------------------------------------|----------------------------------------------------------------|-------------------------------------------------------------------------|------|
| ASTM D6319-10 | Demonstrate<br>biocompatibility:<br>Residual powder | Residual Powder (D 6124)<br>≤ 2.0 mg/pc | Pass |
| Primary Skin<br>Irritation Test<br>ISO 10993-10 | Demonstrate<br>biocompatibility:<br>Skin irritation | Under the conditions of the study, Not a primary skin<br>irritant. | Pass |
| Dermal<br>Sensitization<br>Assay<br>ISO 10993-10 | Demonstrate<br>biocompatibility:<br>Skin sensitivity | Under the conditions of the study, Not a contact<br>sensitizer. | Pass |
| Acute Systemic<br>toxicity Study<br>ISO 10993-11 | Demonstrate<br>biocompatibility:<br>acute systemic<br>toxicity | Under conditions of the ISO Acute Systemic Injection<br>test, not toxic | Pass |
| Open box testing<br>AAMI 11737-1 | Demonstrate<br>acceptable open<br>box bioburden | Acceptable bioburden levels after 30 days open box | Pass |
# X. SUMMARY OF CLINICAL TESTING
No clinical testing was performed.
### XI. CONCLUSION
The conclusions drawn from the non-clinical testing demonstrates that the SHOWA® Blue Nitrile Powder Free Medical Examination Glove is as safe, as effective, and performs as well or better than the legally marketed device (K200581).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.