K130038 · Hl Rubber Industries Sdn Bhd · LZA · Feb 21, 2013 · General Hospital
Device Facts
Record ID
K130038
Device Name
POWDER-FREE NITRILE EXAMINATION GLOVES, BLUE
Applicant
Hl Rubber Industries Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Feb 21, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05. Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Device Story
Powder-free nitrile examination glove; disposable; worn on examiner's hand/finger. Primary function: barrier protection against contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings by healthcare professionals. Provides protection during handling of chemotherapy agents; breakthrough times vary by drug (e.g., Carmustine 3.0 min, Thiotepa 75.0 min, others >240 min).
Clinical Evidence
No clinical data. Bench testing only per ASTM D6978-05 for chemotherapy drug permeation.
Technological Characteristics
Nitrile material; powder-free; blue color. Tested per ASTM D6978-05 for chemotherapy drug permeation resistance.
Indications for Use
Indicated for use as a disposable patient examination glove to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K220353 — Disposal Medical Nitrile Examination Gloves-Tested for Use with Chemotherapy Drugs · Niujian Technology Co., Ltd. · Mar 31, 2022
K160560 — Nitrile Powder Free Examination Glove with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs (Blue) · Hartalega Sdn Bhd · Aug 4, 2016
K202003 — Nitrile Powder Free Examination Glove, Blue Tested for Use with 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim · Gmp Medicare Sdn. Bhd. · Feb 24, 2021
K151997 — Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs-Violet Blue (VBLU) · Hartalega NGC Sdn. Bhd. · Mar 31, 2016
K180476 — Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs · Comfort Rubber Gloves Industries Sdn. Bhd. · May 15, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## February 21, 2013
Ms. Noorzaliza Ahmad Quality Assurance Executive HL Rubber Industries Sdn Bhd Lot 10, Kawasan Perindustrian Dioh Kuala Pilah Negeri Sembilan, Malaysia 72000
Re: K130038
Trade/Device Name: Powder-Free Nitrile Examination Gloves, Blue, Tested for Use With Chemotherapy Drugs
Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: January 7, 2013 Received: January 29, 2013
### Dear Ms. Ahmad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Image /page/1/Picture/7 description: The image contains a stylized drawing with text and geometric shapes. The text appears to be a signature or inscription, with some words more legible than others. The geometric shapes include a series of stacked lines forming a triangular or mountain-like structure on the right side of the image. The overall style is abstract and appears to be hand-drawn.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and . Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K130038
## Device Name: Powder-Free Nitrile Examination Gloves, Blue, Tested for Use With Chemotherapy Drugs
## Indications For Use: A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05. Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
| No. | Chemotherapy Drug and Concentration | Minimum Breakthrough<br>Detection Time<br>(Minutes) |
|-----|------------------------------------------|-----------------------------------------------------|
| l | Carboplatin ( 10.0 mg/ml) | >240 |
| 2 | Carmustine (BCNU) (3.3 mg/ml) | 3.0 |
| 3 | Cisplatin (1.0 mg/ml) | >240 |
| 4 | Cyclophosphamide (Cytoxan) (20.0 mg/ml). | >240 |
| 5 | Cytarabine (100.0 mg/ml) | >240 |
| б | Dacarbazine (DTIC) (10.0 mg/ml) | >240 |
| 7 | Docetaxel (10.0 mg/ml) | >240 |
| 8 | Doxorubicin Hydrochloride (2.0 mg/ml) | >240 |
| ਰੇ | Etoposide (Toposar) (20.0 mg/ml) | >240 |
| 10 | Fluorouracil (50.0 mg/ml) | >240 |
| l I | Mechlorethamine HCI (1.0 mg/ml) | >240 |
| 12 | Methotrexate (25.0 mg/ml) | >240 |
| 13 | Paclitaxel (Taxol) (6.0 mg/ml) | >240 |
| 14 | Thiotepa (10.0 mg/ml) | 75.0 |
Please note that the following drugs have extremely low permeation time: Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 3.0 minute and Thiotepa (10.0 mg/ml) has a minimum breakthrough time of 75.0 minute.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ √ (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
| Elizabeth Laverie | |
|-----------------------------|--|
| 2013.02.20 18:21:46 -05'00' | |
Concurrence of CDRH, Office of Device Evaluation (ODE)(Division Sign-Off) Division Sign-Off)
Division of Anesthestology, General Hospital Infection Control, Dental Devices
| S10(k) Number: | k130038 |
|----------------|---------|
|----------------|---------|
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.