SR GLOVE, NU GLOVE, HIGHSTRESS MODEL 801, MODEL 802, MODEL 803
K093696 · Central Medicare Sdn Bhd · LZA · Feb 12, 2010 · General Hospital
Device Facts
Record ID
K093696
Device Name
SR GLOVE, NU GLOVE, HIGHSTRESS MODEL 801, MODEL 802, MODEL 803
Applicant
Central Medicare Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Feb 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
This glove is disposable and intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Blue Nitrile Examination Gloves, Powder Free, are disposable protective barriers worn by healthcare examiners. The gloves function as a physical barrier to prevent cross-contamination between the patient and the examiner during medical procedures. They are intended for use in clinical or medical settings. The device is a passive, non-sterile or sterile (as applicable) physical barrier; it does not involve electronic processing, algorithms, or software. The benefit to the patient and examiner is the reduction of potential pathogen transmission during physical examinations.
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner. Suitable for over-the-counter use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K211860 — Powder Free Nitrile Patient Examination Gloves, Blue Color · Anhui Powerguard Technology Co., Ltd. · Aug 13, 2021
K211182 — H2H Powder Free Examination Blue Glove · Ntl Innovation Co., Ltd. · Aug 1, 2021
K181106 — Powder Free Nitrile Patient Examination Gloves, Blue Color · Jiangsu Dongxin Medical Technology Co., Ltd. · Jan 16, 2019
K121526 — POWER-FREE NITRILE PATIENT EXAMINATION GLOVES · Shandong Junchi Medical Products Co., Ltd. · Aug 3, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
## FEB 1 2 2010
Central Medicare Sdn. Bhd. C/O Mr. Nick Wang Senior Vice President Encompass Medical Supplies, Incorporated 1930 South Brea Canyon Road, # 240 Diamond Bar, California 91765
Re: K093696
Trade/Device Name: Blue Nitrile Examination Gloves, Powder Free Regulation Number: 21CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: November 20, 2009 Received: November 30, 2009
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2- Mr. Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, BS, MS, MBA Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{2}------------------------------------------------
K093696
Image /page/2/Picture/1 description: The image contains a logo for Central Medicare. The words "Central Medicare" are written in a bold, sans-serif font. Below the word "Medicare" is the text "CENTRAL MEDICARE SDN BHD" in a smaller font. To the left of the text is a graphic of a globe.
CENTRAL MEDICARE SDN. BHD. (660896 T)
(FORMERLY KNOWN AS CENTRAL BOILER AND ENERGY SYSTEM SDN. BHD.) PT 2609 - 2620, Bt 8,Jalan Changkat Jong, Mukim Changkat Jong, Teluk Intan, Perak. Tel: +605 623 1220 Fax: +605 623 1230
## INDICATION FOR USE
| Applicant: | Central Medicare Sdn Bhd |
|----------------|----------------------------------------------|
| 510(k) Number: | K093696 |
| Device Name: | Blue Nitrile Examination Gloves, Powder Free |
Indication for Use:
This glove is disposable and intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
Prescription Use __
AND/OR
Over-The-Counter Use_ X
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
Elizabeth F. Clumio-Wills
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
KO93696 510(k) Number:
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.