Turbett Instrument Pod

K202593 · Turbett Surgical, Inc. · FRG · May 19, 2021 · General Hospital

Device Facts

Record IDK202593
Device NameTurbett Instrument Pod
ApplicantTurbett Surgical, Inc.
Product CodeFRG · General Hospital
Decision DateMay 19, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

The Turbett Instrument Pod is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with the Turbett Surgical filters. The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270F (132C) and exposure time of 4 minutes. The TS1500 Container may be loaded to a maximum weight of 375 lbs., not to exceed 25 lbs. in any tray. Minimum drying time for loads up to 140 lbs. is 10 minutes, for loads up to 375 lbs. 30 minutes. The TS1200 Container may be loaded to a maximum weight of 120 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes. The TS1000 Container may be loaded to a maximum weight of 100 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes. The system was validated with 1mm x 500mm lumens. Do not use with instruments containing lumens with an inner diameter of smaller than 1mm and an overall length longer that 500mm. Use only uncovered, perforated or wire mesh general delivery trays within the Turbett Instrument Pod.

Device Story

Rigid sterilization container system; holds multiple open, perforated/wire mesh instrument trays; utilizes single-use disposable filter cartridge in fenestrated door; eliminates need for reusable gaskets. Available in three sizes (TS1500, TS1200, TS1000). Loaded into steam autoclave via dedicated transfer carriage. Principle of operation: steam penetration through filter for sterilization; container maintains sterile barrier post-process. Used in clinical sterile processing departments. Benefits: high-capacity instrument sterilization; reduced contamination risk from failed gaskets; 30-day sterile storage.

Clinical Evidence

Bench testing only. Validated per AAMI ST77 and ISO 17665-1 for sterilization efficacy (10^-6 SAL) and thermal profiling (minimum 270°F maintained). Transportation and shelf-life studies conducted per modified ASTM D4169-14, demonstrating no growth after 30 days post-sterilization and transport.

Technological Characteristics

Materials: 304 stainless steel, anodized aluminum. Form factor: Rigid container with fenestrated door and single-use filter. Sterilization: Pre-vacuum steam (270°F/132°C). Connectivity: None. Software: None.

Indications for Use

Indicated for healthcare providers to enclose medical devices for sterilization in pre-vacuum steam sterilizers (270°F/132°C, 4-minute cycle) and maintain sterility for up to 30 days. Compatible with TS1500 (max 375 lbs), TS1200 (max 120 lbs), and TS1000 (max 100 lbs) containers. Contraindicated for lumens with inner diameter <1mm and length >500mm.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 19, 2021 Turbett Surgical Inc. % David Furr Consultant FDC Services 8708 Capehart Cove Austin, Texas 78733 Re: K202593 Trade/Device Name: Turbett Instrument Pod Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: April 13, 2021 Received: April 22, 2021 Dear David Furr: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202593 Device Name Turbett Instrument Pod #### Indications for Use (Describe) The Turbett Instrument Pod is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with the Turgical filters. The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cvcle of 270F (132C) and exposure time of 4 minutes. . The TS1500 Container may be loaded to a maximum weight of 375 lbs. in any tray. Minimum drying time for loads up to 140 lbs. is 10 minutes, for loads up to 375 lbs. 30 minutes. . The TS1200 Container may be loaded to a maximum weight of 120 lbs. in any tray. Minimum drving time is 10 minutes. The TS1000 Container may be loaded to a maximum weight of 100 lbs. not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes. The system was validated with 1mm x 500mm lumens. . Use only uncovered, perforated or wire mesh general delivery trays within the Turbett Instrument Pod. Type of Use (Select one or both, as applicable) | | Prescription Use (Part 21 CFR 801 Subpart D) | |--|----------------------------------------------| | | Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K202593 ### Date: May 16, 2021 | 1. | Submitted By: | Turbett Surgical, Inc.<br>125 Tech Park Drive<br>Rochester, NY 14625<br>585-755-0133 | |----|--------------------------------|---------------------------------------------------------------------------------------------------| | 2. | Contact: | David C. Furr<br>FDC Services, LLC<br>8708 Capehart Cove<br>Austin, Texas 78733<br>512-906-9654 | | 3. | Product: | Turbett Instrument Pod – TS1500, TS1200, TS1000<br>Product code: FRG - Class II (21 CFR 880.6850) | | 4. | Common/Classification<br>Name: | Sterilization wrap/container;<br>Predicate device - Turbett Surgical Container K200240 | ### Description: The Turbett Instrument Pod is a rigid sterilization container with a fenestrated door holding a single-use filter. The container is designed to be used in a steam autoclave and hold multiple open trays containing surgical instruments. Trays within the container are separated by dividers to ensure separation and maximum steam exposure. The container is available in three sizes (TS1500 34"x24"x22"; TS1200 34" x17" x22"; TS1000 23" x17" x 22"). The original container (TS1500) has been validated to sterilize 375 lbs. of instruments along with the dividers. The validation was conducted with 15 instrument trays at 25 lbs. each to represent the most challenging case. Additional containers include the TS1200 and TS1000 sizes which can hold 120 lbs. of instruments. Sterilized instruments can be transported and stored for up to 30 days within the closed container. The Turbett Instrument Pod is loaded into the autoclave with a dedicated transfer carriage. The container is constructed of 304 stainless steel with an anodized aluminum filter housing/door. The use of a single-use disposable filter cartridge installed in the fenestrated door eliminates the need for a sealed gasket found on other rigid containers. The omission of a reusable gasket eliminates contamination risks due to failed reusable gaskets. {4}------------------------------------------------ ## Intended Use: The Turbett Instrument Pod is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with the Turbett Surgical filters. The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270F (132C) and exposure time of 4 minutes. - The TS1500 Container may be loaded to a maximum weight of 375 lbs., not . to exceed 25 lbs. in any tray. Minimum drying time for loads up to 140 lbs. is 10 minutes, for loads up to 375 lbs. 30 minutes. - . The TS1200 Container may be loaded to a maximum weight of 120 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes. - . The TS1000 Container may be loaded to a maximum weight of 100 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes. - . The system was validated with 1mm x 500mm lumens. Do not use with instruments containing lumens with an inner diameter of smaller than 1mm and an overall length longer that 500mm. Use only uncovered, perforated or wire mesh general delivery trays within the Turbett Instrument Pod. ## Technological Characteristics Table: Shown below is the technological characteristics comparison table for the subject device with the predicate device. {5}------------------------------------------------ | Element of<br>Comparison | 510(k) Device:<br>Turbett Instrument Pod<br>K202593 | Predicate Device:<br>Turbett Instrument Pod<br>K200240 | Explanation of<br>Differences | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------| | Regulation and<br>Product<br>Classification<br>Code | 21 CFR 880.6850<br>FRG | 21 CFR 880.6850<br>FRG | None | | Indications for<br>Use | The Turbett Instrument Pod is<br>indicated for enclosing other medical<br>devices that are to be sterilized by a<br>healthcare provider. It is intended to<br>allow sterilization of the enclosed<br>materials and maintain sterility for up<br>to 30 days until used. The unit is<br>intended to be used in pre-vacuum<br>steam sterilizers. The unit must be<br>used with the Turbett Surgical filters.<br>The unit is intended to be used in<br>prevacuum steam sterilizers with a<br>prevacuum cycle of 270F (132C) and<br>exposure time of 4 minutes.<br>The TS1500 Container may be<br>loaded to a maximum weight of 375<br>lbs., not to exceed 25 lbs. in any tray.<br>Minimum drying time for loads up to<br>140 lbs. is 10 minutes, for loads up to<br>375 lbs. 30 minutes. The TS1200 Container may be<br>loaded to a maximum weight of 120<br>lbs., not to exceed 25 lbs. in any tray.<br>Minimum drying time is 10 minutes. The TS1000 Container may be<br>loaded to a maximum weight of 100<br>lbs., not to exceed 25 lbs. in any tray.<br>Minimum drying time is 10 minutes. The system was validated with<br>1mm x 500mm lumens. Do not use<br>with instruments containing lumens<br>with an inner diameter of smaller than<br>0.5mm and an overall length longer<br>that 500mm. Use only uncovered, perforated or<br>wire mesh general delivery trays<br>within the Turbett Instrument Pod. | The Turbett Instrument Pod is indicated<br>for enclosing other medical devices that<br>are to be sterilized by a healthcare<br>provider. It is intended to allow<br>sterilization of the enclosed materials and<br>maintain sterility for up to 30 days until<br>used. The unit is intended to be used in<br>pre-vacuum steam sterilizers. The unit<br>must be used with the Turbett Surgical<br>filters.<br>The unit is intended to be used in<br>prevacuum steam sterilizers with a<br>prevacuum cycle of 270F (132C) and<br>exposure time of 4 minutes.<br>The TS1500 Container may be loaded<br>to a maximum weight of 375 lbs., not to<br>exceed 25 lbs. in any tray. Minimum<br>drying time for loads up to 140 lbs. is 10<br>minutes, for loads up to 375 lbs. 30<br>minutes. The TS1200 Container may be loaded<br>to a maximum weight of 120 lbs., not to<br>exceed 25 lbs. in any tray. Minimum<br>drying time is 10 minutes. The TS1000 Container may be loaded<br>to a maximum weight of 100 lbs., not to<br>exceed 25 lbs. in any tray. Minimum<br>drying time is 10 minutes. The system was validated with 1mm x<br>500mm lumens. Do not use with<br>instruments containing lumens with an<br>inner diameter of smaller than 1mm and<br>an overall length longer that 500mm. Use only uncovered, perforated or wire<br>mesh general delivery trays within the<br>Turbett Instrument Pod. | Indications are<br>identical | | Principal<br>Material of<br>Construction | Stainless Steel and aluminum | Stainless Steel and aluminum | Same | | Overall Size | Approximate size<br>TS1500 34"x24"x22"<br>TS1200 34" x17" x22"<br>TS1000 23" x17" x 22" | Approximate size<br>TS1500 34"x24"x22"<br>TS1200 34" x17" x22"<br>TS1000 23" x17" x 22" | All pods are<br>identical | | Presentation of Device | Very Large Container with transfer cart, 3 sizes. | Very Large Container with transfer cart, 3 sizes. | All pods are identical | | Sterilization Cycle | Prevacuum Steam 4 minute cycle 132°C | Prevacuum Steam 4 minute cycle 132°C | None | | Load | TS1500 up to 375 lbs.<br>TS1200 up to 120 lbs.<br>TS1000 up to 100 lbs. | TS1500 up to 375 lbs.<br>TS1200 up to 120 lbs.<br>TS1000 up to 100 lbs. | All pods are identical | | Storage | Up to 30 days | Up to 30 days | None | | Transportation | TS1500 may be transported | None | Transport added | {6}------------------------------------------------ # Summary of Non-Clinical Testing: Provided below is the non-clinical testing performed with the subject device to demonstrate that the device met the acceptance criteria with the standard listed below. | Type of Testing | Purpose | Acceptance Criteria | Result | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|-------------------------------------------------------------------------------|--------| | Pre-vacuum sterilization<br>efficacy<br>AAMI ST77<br>Containment Devices<br>for Reusable Medical<br>Device Sterilization<br>ISO 17665-1 Sterilization<br>of Health Care Products –<br>Moist Heat – Part 1<br>Requirements for the<br>Development, Validation,<br>and Routine Control of a<br>Sterilization Process for<br>Medical Devices | Demonstrate Sterilization<br>Capabilities | 10-6 SAL | PASSED | | Thermal Profile<br>AAMI ST77 Containment<br>Devices for Reusable<br>Medical Device<br>Sterilization | Demonstrate Heat<br>Penetration in Load | Minimum temperature of 270° F<br>achieved within 2 minutes and<br>maintained. | PASSED | {7}------------------------------------------------ | Transportation Shelf life<br>Study<br><br>AAMI ST77 Containment<br>Devices for Reusable<br>Medical Device<br>Sterilization<br><br>Modified ASTM D4169-14<br>- Standard Practice for<br>performance testing of<br>shipping containers and<br>systems | The purpose was to simulate<br>transportation of the subject<br>device followed by a shelf life<br>study. | No growth 30 days following<br>sterilization loading, unloading<br>and transport. | PASSED | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|--------| |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|--------| ## Conclusion: The conclusions drawn from the non-clinical tests demonstrate that the proposed device is as safe, as effective, and performed as well as or better than the legally marketed predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...