Turbett Surgical Instrument Pod (TS1500); Turbett Surgical Instrument Pod (TS1200); Turbett Surgical Instrument Pod (TS1000)
Applicant
Turbett Surgical
Product Code
FRG · General Hospital
Decision Date
Apr 4, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Turbett Surgical Instrument Pod is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical Filters. The unit is intended to be used in a pre-vacuum steam sterilizers with a pre-vacuum cycle of 270°F (132°C) and exposure time of 4 minutes. •The TS1500 Container may be loaded to a maximum weight of 375 lbs., not to exceed 25 lbs. in any tray. Minimum drying time for loads up to 140lbs. is 10minutes, for loads up to 375 lbs. is 30 minutes. •The TS1200 Container may be loaded to a maximum weight of 300 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes for loads up to120 lbs. and 30 minutes for loads up to 300 lbs. •The TS1000 Container may be loaded to a maximum weight of 100 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes. •The system was validated with 0.7 mm x 500 mm and 1.0 mm x 850 mm rigid and flexible lumens. Use only uncovered, perforated, or wire mesh general delivery trays within the Turbett Surgical Instrument Pod.
Device Story
Rigid sterilization container; holds multiple open, perforated, or wire mesh surgical instrument trays. Constructed from 304 stainless steel and anodized aluminum. Features fenestrated door with disposable filter (2 layers filter paper, compressible gasket). Loaded into steam autoclave via dedicated transfer mechanism/cart. Steam penetrates container to sterilize instruments; container maintains sterility for 30 days post-sterilization. Used by healthcare providers in clinical settings. Benefits include organized, sterile transport and storage of surgical instruments.
Clinical Evidence
No clinical data. Safety and efficacy established through non-clinical bench testing including sterilization efficacy (SAL 10^-6), thermal profiling, drying time, 30-day shelf-life, cleaning validation, microbial aerosol challenge, and usability testing per ANSI/AAMI standards.
Technological Characteristics
Rigid container; 304 stainless steel sheet metal/bar stock; anodized aluminum dividers/door panels. Single-use filter (paper/gasket). Dimensions vary by model (TS1500, TS1200, TS1000). Steam sterilization (270°F/132°C). No software or electronics.
Indications for Use
Indicated for enclosing medical devices for sterilization by healthcare providers; maintains sterility for up to 30 days. Used in pre-vacuum steam sterilizers (270°F/132°C for 4 minutes). Supports specific load weights (TS1500: 375 lbs; TS1200: 300 lbs; TS1000: 100 lbs) and lumen sizes (0.7mm x 500mm to 1.0mm x 850mm).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 4, 2025
Turbett Surgical
Melissa Conley
Director of Quality and Regulatory Affairs
245 Summit Point Drive
Suite 2A
Henrietta, New York 14467
Re: K250011
Trade/Device Name: Turbett Surgical Instrument Pod (TS1500); Turbett Surgical Instrument Pod (TS1200); Turbett Surgical Instrument Pod (TS1000)
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization Wrap
Regulatory Class: Class II
Product Code: FRG
Dated: January 2, 2025
Received: January 2, 2025
Dear Melissa Conley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K250011 - Melissa Conley
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K250011 - Melissa Conley
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Christopher K. Dugard -S
Christopher K. Dugard, M.S.
Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250011 | |
| Device Name Turbett Surgical Instrument Pod | |
| Indications for Use (Describe) | |
| The Turbett Surgical Instrument Pod is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical Filters. The unit is intended to be used in a pre-vacuum steam sterilizers with a pre-vacuum cycle of 270°F (132°C) and exposure time of 4 minutes. •The TS1500 Container may be loaded to a maximum weight of 375 lbs., not to exceed 25 lbs. in any tray. Minimum drying time for loads up to 140lbs. is 10minutes, for loads up to 375 lbs. is 30 minutes. •The TS1200 Container may be loaded to a maximum weight of 300 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes for loads up to120 lbs. and 30 minutes for loads up to 300 lbs. •The TS1000 Container may be loaded to a maximum weight of 100 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes. •The system was validated with 0.7 mm x 500 mm and 1.0 mm x 850 mm rigid and flexible lumens. Use only uncovered, perforated, or wire mesh general delivery trays within the Turbett Surgical Instrument Pod. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
K250011 510(k) Summary
Page 1 of 7
# K250011 - 510(k) Summary
## Submitter
**510(k) Owner:**
Turbett Surgical
245 Summit Point Drive
Suite 2A
Henrietta, NY 14467
**Official Correspondent:**
Melissa Conley
Director of Quality and Regulatory Affairs
Turbett Surgical, Inc.
(412) 901-6650
**Date of Preparation:**
03/26/2025
## Device
**Proprietary/Trade Name:**
Turbett Surgical Instrument Pod
**Common/ Usual Name:**
Sterilization Wrap/ Container
**Device Classification:**
II
**Classification Name/ Product Code:**
21 CFR 880.6850 / FRG
**Primary Predicate Device:**
Turbett Surgical Instrument Pod, K202593
**Submission Reason:**
Indications expansion
## Device Description
The Turbett Surgical Instrument Pod is a multi-tray rigid sterilization container with a fenestrated door holding a single use filter. The container is designed to be used in a steam autoclave and hold multiple open trays containing surgical instruments. Unwrapped, uncovered, perforated sterilization trays are stacked into the container, separated by anodized aluminum dividers, which allow steam to evenly penetrate and allow condensation to drain.
The container is manufactured from 304 stainless steel sheet metal and bar stock. The container is sealed on all sides except one which is the door / filter panel.
{6}
The single fenestrated door consists of two perforated anodized aluminum panels which cover the exposed container side wall. A custom-made disposable filter consisting of 2 layers of filter paper separated by a layer of compressible single-use gasket material is placed between the two perforated metal panels of the door. Spring loaded latches compress the anodized aluminum door frame with filter assembly to the container for a secure seal.
The Turbett Surgical Instrument Pod is loaded into the auto clave with a dedicated transfer mechanism and cart.
## Indications for Use
The Turbett Surgical Instrument Pod is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical Filters.
The unit is intended to be used in a pre-vacuum steam sterilizers with a pre-vacuum cycle of 270°F (132°C) and exposure time of 4 minutes.
- The TS1500 Container may be loaded to a maximum weight of 375 lbs., not to exceed 25 lbs. in any tray. Minimum drying time for loads up to 140 lbs. is 10 minutes, for loads up to 375 lbs. is 30 minutes.
- The TS1200 Container may be loaded to a maximum weight of 300 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes for loads up to 120 lbs. and 30 minutes for loads up to 300 lbs.
- The TS1000 Container may be loaded to a maximum weight of 100 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes.
- The system was validated with 0.7 mm x 500 mm and 1.0 mm x 850 mm rigid and flexible lumens.
Use only uncovered, perforated, or wire mesh general delivery trays within the Turbett Surgical Instrument Pod.
K250011 510(k) Summary
Page 2 of 7
{7}
Comparison of Technological Characteristics with the Predicate Device
| Characteristic | Subject Device | Predicate Device, K202593 | Comparison |
| --- | --- | --- | --- |
| Regulation and Product Classification Code | 21 CFR 880.6850 FRG | 21 CFR 880.6850 FRG | Same. |
| Indications for Use | The Turbett Surgical Instrument Pod is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical Filters.
The unit is intended to be used in a pre-vacuum steam sterilizers with a pre-vacuum cycle of 270°F (132°C) and exposure time of 4 minutes.
• The TS1500 Container may be loaded to a maximum weight of 375 lbs., not to exceed 25 lbs. in any tray. Minimum drying time for loads up to 140lbs. is 10 minutes, for loads up to 375 lbs. is 30 minutes.
• The TS1200 Container may be loaded to a maximum weight of 300 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes for loads up to 120 lbs. and | The Turbett Surgical Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical Filters.
The unit is intended to be used in pre-vacuum steam sterilizers with a pre-vacuum cycle of 270°F (132°C) and exposure time of 4 minutes.
• The TS1500 Container may be loaded to a maximum weight of 375 lbs., not to exceed 25 lbs. in any tray. Minimum drying time for loads up to 140lbs. is 10 minutes, for loads up to 375 lbs. is 30 minutes.
• The TS1200 Container may be loaded to a maximum weight of 120lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes. | The intended use of the device, allowing the sterilization of enclosed materials and maintaining sterility for up to 30 days until used, remains the exact same between the predicate device and the subject device.
The indications for lumen sizes that are to be sterilized and the maximum weight capacity for the TS1200 model are different between the predicate and the subject device. The subject device proposes the sterilization of thinner and longer, both rigid and flexible lumens and a the TS1200 |
K250011 510(k) Summary
Page 3 of 7
{8}
| | 30 minutes for loads up to 300 lbs.
• The TS1000 Container may be loaded to a maximum weight of 100 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes.
• The system was validated with 0.7 mm x 500 mm and 1.0 mm x 850 mm rigid and flexible lumens.
Use only uncovered, perforated, or wire mesh general delivery trays within the Turbett Surgical Instrument Pod. | • The TS1000 Container may be loaded to a maximum weight of 100 lbs., not to exceed 25 lbs. in any tray. Minimum drying time is 10 minutes.
• The system was validated with 1mm x 500mm lumens.
Use only uncovered, perforated or wire mesh general delivery trays within the Turbett Instrument Pod. | model has been validated for up to 300 lbs. in the subject device, vs. 120 lbs. labeled for the predicate device. |
| --- | --- | --- | --- |
| Principal Material of Construction | Stainless Steel and Aluminum | Stainless Steel and Aluminum | Same. |
| Sterilization Parameters | 270°F (132°C) for 4 minutes | 270°F (132°C) for 4 minutes | Same. |
| Recommended Sterilization Trays | All manufacturers’ trays – uncovered, perforated, or wire mesh general delivery trays. | All manufacturers’ trays – uncovered, perforated, or wire mesh general delivery trays. | Same. |
| Recommended Sterilizers | All makes and models with dimensions compatible with the Turbett Surgical Instrument Pod and that are capable of achieving the required sterilization parameters of 270°F (132°C) for 4 minutes. | All makes and models with dimensions compatible with the Turbett Surgical Instrument Pod and that are capable of achieving the required sterilization parameters of 270°F (132°C) for 4 minutes. | Same. |
| Deployment into Sterilizer | Transfer Mechanism and Cart | Transfer Mechanism and Cart | Same. |
| Load | TS1500 up to 375lbs.
TS1200 up to 300 lbs.
TS1000 up to 100 lbs. | TS1500 up to 375 lbs.
TS1200 up to 120 lbs.
TS1000 up to 100 lbs. | Similar, The TS1200 model has been validated for up to 300 lbs. vs the predicate of 120 lbs. |
| Storage | Up to 30 days | Up to 30 days | Same. |
| Transportation | The TS1500, TS1200 and TS1000 may be transported. | The TS1500 may be transported. | The TS1500 model represents the |
K250011 510(k) Summary
Page 4 of 7
{9}
worst case model to be transported as it is the largest by size and holds the highest maximum load weight.
# Performance Testing Data
# Non-Clinical Tests
The following non-clinical tests were completed and submitted/ referenced as part of this submission:
| Test Name | Method and/ or Standard | Acceptance Criteria | Results (Pass / Fail) |
| --- | --- | --- | --- |
| Sterilization Efficacy | ANSI/AAMI ST77:2013, Containment Devices for Reusable Medical Device Sterilization | Log reduction of 10-6sterility assurance level | Pass |
| Thermal Profile | ANSI/AAMI ST77:2013, Containment Devices for Reusable Medical Device Sterilization | Demonstrate various locations within the container can reach and maintain exposure temperature | Pass |
| Dry Time | ANSI/AAMI ST79:2007, Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities | Pre/Post weight difference less than 0.2% and no visible moisture after specified drying time | Pass |
| 30 Day Event Related Shelf Life | ANSI/AAMI ST77:2013, Containment Devices for Reusable Medical | Contents remain sterile post 30-Day Shelf-life. | Pass |
K250011 510(k) Summary
{10}
| | Device Sterilization | | |
| --- | --- | --- | --- |
| **Cleaning Validation** | TIR 30:2011, A compendium of Processes, Materials, Test Methods and Acceptance Criteria for Cleaning Reusable Medical Devices. | Residual Protein and Total Carbon within test limits (Protein level of less than 6.4 μg/cm² and Total Carbon level of less than 2.2 μg/cm²) | Pass |
| **Microbial Aerosol Challenge** | ANSI/AAMI ST77:2013, Containment Devices for Reusable Medical Device Sterilization | Demonstrate maintenance of sterility by no growth of internal test coupons following exposure to microbial aerosol | Pass |
| **Usability** | ANSI/AAMI HE75:2013, Human Factors Engineering - Design of Medical Devices. | Demonstrate that the device can be used by representative users under simulated use conditions without producing patterns of failures that could result in negative clinical impact to patients or injury to users.
Verify use of the IFU is effective.
Determine whether the use related safety issues associated with using the device have been adequately mitigated. | Pass |
## Clinical Tests
Clinical tests were not required to demonstrate safety and effectiveness of the Turbett Surgical Instrument Pod. Safety and efficacy have been established through non-clinical tests.
## Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device in 510(k) submission K250011, the Turbett Surgical Instrument Pod, is as
K250011 510(k) Summary
Page 6 of 7
{11}
safe, as effective, and performs as well as or better than the legally marketed predicate device cleared in K202593.
K250011 510(k) Summary
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.