Turbett Surgical Container

K153025 · Turbett Surgical, LLC · FRG · Jan 14, 2016 · General Hospital

Device Facts

Record IDK153025
Device NameTurbett Surgical Container
ApplicantTurbett Surgical, LLC
Product CodeFRG · General Hospital
Decision DateJan 14, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

The Turbett Surgical Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical filters. The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270°F (132°C) and exposure time of 4 minutes. The container may be loaded with up to 15 trays and up to 25 lbs. per tray. Validation was done using three trays per level and a maximum instrument load of 25 lbs. per tray. The validation load included six 1mm x 500mm lumens and six 3mm x 400mm lumens. The total weight of instruments and trays validated is 375 lbs. Use only uncovered, perforated or wire mesh general delivery trays within the Turbett Surgical Container.

Device Story

Rigid sterilization container; holds multiple open surgical instrument trays; used in steam autoclaves. Input: surgical instruments in perforated/wire mesh trays. Operation: steam sterilization cycle; single-use disposable filter cartridge in fenestrated door replaces traditional reusable gaskets to reduce contamination risk. Output: sterilized instruments stored for up to 30 days. Used in hospital sterile processing departments; loaded via dedicated transfer carriage. Benefits: high-capacity sterilization (up to 375 lbs); improved steam penetration via high vent-to-volume ratio (905 holes); eliminates gasket failure risks.

Clinical Evidence

Bench testing only. Validated per ANSI/AAMI ST77 (thermal profile, sterilization efficacy, drying time, microbial aerosol challenge, 30-day shelf-life), AAMI TIR 30 (cleaning methods), and ANSI/AAMI HE75 (human factors usability). Validated for 375 lb load including 1mm x 500mm and 3mm x 400mm lumens.

Technological Characteristics

Rigid container; 304 stainless steel sheet metal on stainless steel frame; anodized aluminum filter housing/door. Single-use filter (2 layers paper + compressible gasket). Dimensions: 34"x24"x22". Sterilization: Pre-vacuum steam (270°F/132°C). Connectivity: None. Standards: ANSI/AAMI ST77, AAMI TIR 30, ANSI/AAMI HE75.

Indications for Use

Indicated for enclosing medical devices for sterilization by healthcare providers in pre-vacuum steam sterilizers (270°F/132°C, 4-minute exposure). Supports up to 15 trays (max 25 lbs/tray, 375 lbs total). Validated for lumens down to 1mm diameter x 500mm length.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 14, 2016 Turbett Surgical LLC % David Furr Principal Consultant FDC Services 8708 Capehart Cove Austin, Texas 78733 Re: K153025 Trade/Device Name: Turbett Surgical Container Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: December 14, 2015 Received: December 17, 2015 Dear Mr. Furr: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {2}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, *Tejashri Purohit-Sheth, M.D.* Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K153025 Device Name Turbett Surgical Container #### Indications for Use (Describe) The Turbett Surgical Container is indicated for enclosing other medical devices that are to be sterilized by a healtheare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical filters. The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 2709 (132℃) and exposure time of 4 minutes. The container may be loaded with up to 15 trays and up to 25 lbs. per trav. Validation was done using three trays per level and a maximum instrument load of 25 lbs. per tray. The validation load included six 1mm x 500mm lumens and six 3mm x 400mm lumens. The total weight of instruments and trays validated is 375 lbs. Use only uncovered, perforated or wire mesh general delivery trays within the Turbett Surgical Container. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary Pursuant to 21 CFR 807.92 ## Date: December 14, 2015 ## K153025 - 1. Submitted By: Turbett Surgical LLC. 125 Tech Park Drive Suite 2108 Rochester, NY 14623 585-755-0133 - 2. Contact: David C. Furr FDC Services, LLC 8708 Capehart Cove Austin, Texas 78733 512-906-9654 3. Turbett Surgical Container Product: - Product code: FRG Class II (21 CFR 880.6850) - 4. Common/Trade Name: Sterilization wrap/container ### Description: The Turbett Surgical Container is a rigid sterilization container with a fenestrated door holding a single-use filter. The container is designed to be used in a steam autoclave and hold multiple open trays containing surgical instruments. Trays within the container are separated by dividers to ensure separation and maximum steam exposure. The container has been validated to sterilize 375 lbs. of instruments along with the dividers. The validation was conducted with 15 instrument trays at 25 lbs. each to represent the most challenging case. Actual expected loads in hospital settings are likely to be less. Sterilized instruments can be stored for up to 30 days within the closed container. The Turbett Surgical Container is loaded into the autoclave with a dedicated transfer carriage. The container is constructed of 304 stainless steel with an anodized aluminum filter housing/door. The use of a single-use disposable filter cartridge installed in the fenestrated door eliminates the need for a sealed gasket found on other rigid containers. The omission of a reusable gasket eliminates contamination risks due to failed reusable gaskets. {5}------------------------------------------------ # Intended Use:. The Turbett Surgical Container is indicated for enclosing other medical devices that are to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed materials and maintain sterility for up to 30 days until used. The unit is intended to be used in pre-vacuum steam sterilizers. The unit must be used with Turbett Surgical filters. The unit is intended to be used in prevacuum steam sterilizers with a prevacuum cycle of 270°F (132°C) and exposure time of 4 minutes. The container may be loaded with up to 15 trays and up to 25 lbs. per tray. Validation was done using three trays per level and a maximum instrument load of 25 lbs. per tray. The validation load included six 1mm x 500mm lumens and six 3mm x 400mm lumens. The total weight of instruments and trays validated is 375 lbs. Use only uncovered, perforated or wire mesh general delivery trays within the Turbett Surgical Container. # Technological Characteristics: The Turbett Surgical Container has been validated to sterilize 375 Ibs. of instruments along with the dividers. The validation was conducted with 15 instrument trays at 25 lbs. each to represent the most challenging case. The validations included thermal profile, sterilization efficacy, and drying in a prevacuum steam sterilizer. Sterilized instruments can be stored for up to 30 days within the closed container. The container is constructed of 304 stainless steel sheet metal on a rigid stainless steel frame. The container opens from the side for easy placement and retrieval of surgical trays. A filter comprised of 2 layers of filter paper and a compressible gasket is placed in the outer door. After use, the filter/gasket is discarded. The following testing was conducted to establish substantial equivalence and efficacy: ANSI/AAMI ST77: Containment Devices for Reusable Medical Device Sterilization testing: - Pre-vacuum thermal profile - Steam pre-vacuum sterilization efficacy ● - Pre-vacuum drv time ● - Microbial aerosol challenge - 30 day shelf-life study o {6}------------------------------------------------ ### 510(k) Premarket Notification Turbett Surgical Container AAMI TIR 30: A Compendium of Processes, Materials, Test Methods and Acceptance Criteria for Cleaning Reusable Medical Devices testing: - Protein analysis and Total Organic Carbon Manual cleaning methods . ANSI/AAMI HE75: Human Factors Engineering- Design of Medical Devices testing: - Human factors usability study . # Substantial Equivalence: The Turbett Surgical Container is substantially equivalent to the SCORES Unit Sterilization Container (K110898). Both products have the same indications for use and perform in a similar manner. {7}------------------------------------------------ | Element of<br>Comparison | 510(k) Device:<br>Turbett Surgical Container<br>K153025 | Predicate Device:<br>SCORES Unit<br>K110898 | Explanation of Differences | |--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Regulation and<br>Product<br>Classification Code | 21 CFR 880.6850<br>FRG | 21 CFR 880.6850<br>FRG | None | | Indications for Use | The Turbett Surgical<br>Container is indicated for<br>enclosing other medical<br>devices that are to be<br>sterilized by a healthcare<br>provider. It is intended to<br>allow sterilization of the<br>enclosed materials and<br>maintain sterility for up to<br>30 days until used. The unit<br>is intended to be used in<br>pre-vacuum steam<br>sterilizers. The unit must be<br>used with Turbett Surgical<br>filters. The unit is intended<br>to be used in prevacuum<br>steam sterilizers with a<br>prevacuum cycle of 270°F<br>(132°C) and exposure time<br>of 4 minutes. The container<br>may be loaded with up to<br>15 trays and up to 25 lbs.<br>per tray. Validation was<br>done using three trays per<br>level and a maximum<br>instrument load of 25 lbs.<br>per tray. The validation load<br>included six 1mm x 500mm<br>lumens and six 3mm x<br>400mm lumens. The total<br>weight of instruments and<br>trays validated is 375 lbs.<br>Use only uncovered,<br>perforated or wire mesh<br>general delivery trays<br>within the Turbett Surgical<br>Container. | The SCORES Sterilization Container<br>is indicated for enclosing other<br>medical devices that are to be<br>sterilized by a healthcare provider. It<br>is intended to allow sterilization of the<br>enclosed materials and maintain<br>sterility for up to 30 days until used.<br>The unit must be used with the<br>SCORES transfer cart, SCORES<br>filters and integrity locks.<br>The unit is intended to be used in<br>prevacuum steam sterilizers with a<br>prevacuum cycle of 270 degrees and<br>exposure time of 4 minutes. Use no<br>more than 3 trays per shelf or 25 lbs.<br>per tray.<br>The SCORES Unit was tested and<br>validated with rigid instruments<br>containing lumens with an inner<br>diameter of 3.8mm and an overall<br>length of 370mm. Do not use with<br>instruments containing lumens with an<br>inner diameter smaller than 3.8mm<br>and an overall length longer than<br>370mm.<br>Use only uncovered, perforated or<br>wire mesh general delivery trays<br>within the SCORES Sterilization<br>Container. | No significant differences<br>other than lumen limitation on<br>SCORES unit.<br>Turbett Surgical Container has<br>been validated with smaller<br>diameter/longer length lumens. | | Principal Material<br>of Construction | Stainless Steel and<br>aluminum | Stainless Steel | Aluminum is used for<br>dividers and filter panels for<br>lighter weight | | Overall Size | 34"x24"x22" | 33"x36"x21" | Turbett device is more<br>compact. Validated at this<br>size. | | Weight of Empty<br>Container | 136 lbs. | 150 lbs. (approximated) | The Turbett device is not as<br>heavy as the comparison. | | Presentation of<br>Device | Very Large Container with<br>transfer cart | Very Large Container with transfer<br>cart | None | | Sterilization Method | Steam autoclave | Steam autoclave | None | | Sterilization Cycle | Prevacuum Steam 4<br>minute cycle | Prevacuum Steam 4 minute cycle | None | | Temperature of<br>Sterilization Cycle | 270°F (132°C) | 270°F (132°C) | None | | Load | Up to 15 trays at 25 lb<br>each | Up to 12 trays at 25 lb each | Turbett container holds<br>slightly more trays. System<br>has been validated at this<br>load. | | Number of<br>Perforations | 905 holes 0.74 inches<br>diameter (398.2 square<br>inches) | 144 holes 1 inch diameter (113.76<br>square inches) | The vent to volume ratio for<br>the Turbett device allows for<br>better steam penetration. | | Storage | Up to 30 days | Up to 30- days | None | # Predicate Device Comparison Table {8}------------------------------------------------ ## 510(k) Premarket Notification Turbett Surgical Container # Conclusion: The performance testing data for the subject device (Turbett Surgical Container) demonstrates it is substantially equivalent to the predicate device (K110898). Testing and performance evaluations demonstrate that the product meets specifications and validation requirements.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...