Paper Sterilization Pouch and Roll(Sterilization Pouch Flat, Sterilization Pouch Gusseted, Sterilization Roll Flat and Sterilization Roll Gusseted)
K182184 · Mdk (Shanghai) Medical Packing Co., Ltd. · FRG · Oct 10, 2019 · General Hospital
Device Facts
Record ID
K182184
Device Name
Paper Sterilization Pouch and Roll(Sterilization Pouch Flat, Sterilization Pouch Gusseted, Sterilization Roll Flat and Sterilization Roll Gusseted)
Applicant
Mdk (Shanghai) Medical Packing Co., Ltd.
Product Code
FRG · General Hospital
Decision Date
Oct 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Sterilization Pouch Flat are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Sterilization Pouch Flat are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below: Steam; 30 minutes at 121 ℃; 25 minutes dry time. Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60 ℃. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. The Sterilization Pouch Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 1470 g. The sterilization pouch maintains the enclosed devices up until 6 months post sterilization. The Sterilization Pouch Gusseted are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Sterilization Pouch Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below: Steam; 30 minutes at 121 ℃; 25 minutes dry time. Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60 ℃. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. The Sterilization Pouch Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 1470g. The sterilization pouch maintains the enclosed devices up until 6 months post sterilization. The Sterilization Roll Flat are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Sterilization Roll Flat are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below: Steam; 30 minutes at 121 ℃; 25 minutes dry time. Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60 ℃. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. The Sterilization Roll Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 910g. The sterilization roll maintains the enclosed devices up until 6 months post sterilization. The Sterilization Roll Gusseted are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Sterilization Roll Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below: Steam; 30 minutes at 121 ℃; 25 minutes dry time. Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60 ℃. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. The Sterilization Roll Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 910g. The sterilization roll maintains the enclosed devices up until 6 months post sterilization.
Device Story
Medical-grade paper and plastic (CPP/PET) pouches/rolls; used by healthcare providers to enclose dental instruments for terminal sterilization. Devices inserted, heat-sealed, then processed via gravity steam (121°C, 30 min) or EtO (55°C, 600 mg/L). External chemical ink indicators (printed on pouch) change color to verify exposure to sterilization process; indicators do not signify sterility. Maintains sterility for 6 months post-sterilization. Benefits include protection of instruments and verification of processing exposure.
Clinical Evidence
No clinical data; bench testing only. Validated via ASTM/ISO standards: tensile strength (ASTM D882), thickness (ASTM F2251), tear strength (ASTM D1922), air permeability (ISO 5636-3), burst strength (ASTM F1140), microbial penetration (ASTM F1608), seal strength (ASTM F88), and EO/ECH residuals (ISO 10993-7). Sterilization efficacy validated via half-cycle method (ISO 14937) achieving SAL 10^-6.
Technological Characteristics
Materials: Medical porous paper and CPP/PET plastic film. Sensing: External chemical ink indicators (Class 1). Dimensions: Various sizes/lengths. Connectivity: None. Sterilization: Steam (121°C) or EtO (55°C). Standards: ASTM D882, F2251, D1922, F1140, F1608, F1929, F88; ISO 5636-3, 10993-5/7/10, 11140-1, 14937.
Indications for Use
Indicated for healthcare providers to enclose dental instruments for terminal sterilization via steam or EtO. Excludes complex devices (endoscopes, instruments with lumens/channels). Maximum weight limits apply (1470g for pouches, 910g for rolls). Maintains sterility for up to 6 months.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
K143637 — U&U Sterilization Pouch and Roll · U&U Medical Technology Co, Ltd. · Sep 9, 2015
K172280 — BH Sterilization Pouch · Bh Medical Products Co., Ltd. · Aug 3, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 10, 2019
MDK (ShangHai) Medical Packing Co., Ltd. Ray Wang Official Correspondent No. 233 Fengijan Rd, FengCheng Town, FengXian District ShangHai, 201409 Cn
Re: K182184 Trade/Device Name: Paper Sterilization Pouch and Roll (Sterilization Pouch Flat, Sterilization Pouch Gusseted, Sterilization Roll Flat and Sterilization Roll Gusseted) Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: September 5, 2019 Received: September 9, 2019
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K182184
### Device Name
Paper Sterilization Pouch and Roll (Sterilization Pouch Gat, Sterlization Roll Flat and Sterilization Roll Gusseted)
Indications for Use (Describe)
Sterilization Pouch Flat
The Sterilization Pouch Flat are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Sterilization Pouch Flat are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Gravity Steam; 30 minutes at 121 ℃; 25 minutes dry time.
Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60℃.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Sterilization Pouch Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 1470 g.
The sterilization pouch maintains the enclosed devices up until 6 months post sterilization.
Sterilization Pouch Gusseted
The Sterilization Pouch Gusseted are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterlity of the enclosed device until used.
The Sterilization Pouch Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Gravity Steam; 30 minutes at 121 ℃; 25 minutes dry time.
Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60°C.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Sterilization Pouch Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 1470g.
The sterilization pouch maintains the enclosed devices up until 6 months post sterilization.
Sterilization Roll Flat
The Sterilization Roll Flat are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Sterilization Roll Flat are intended for sterilization of dental instruments, excluding complex devices and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Gravity Steam; 30 minutes at 121 ℃; 25 minutes dry time.
Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60℃.
{3}------------------------------------------------
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Sterilization Roll Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 910g.
The sterilization roll maintains the enclosed devices up until 6 months post sterilization.
### Sterilization Roll Gusseted
The Sterilization Roll Gusseted are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterlity of the enclosed device until used.
The Sterilization Roll Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Gravity Steam; 30 minutes at 121 ℃; 25 minutes dry time.
Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60°C.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Sterilization Roll Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 910g.
The sterilization roll maintains the enclosed devices up until 6 months post sterilization.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This 510(k) Summary of 510(k) is being submitted in accordance with Title 21, CFR Section 807.92.
The assigned 510(k) Number: K182184
- 1. Date of Preparation: 10/7/2018
- 2. Sponsor Identification
### MDK (ShangHai) Medical Packing Co., Ltd.
No .: 233 Fengjian Rd, FengCheng Town, FengXian District, Shanghai, 201409, China
Establishment Registration Number: Not yet registered or the Number not issued yet.
Contact Person: Gong Yaoren Position: Quality Manager Tel: +86 18601673483 Fax: +86 21 64868852 Email: gongyr@126.com
- 3. Designated Submission Correspondent
Mr. Ray Wang Tel: +86-18910677558 Fax: +86-10-56335780 Email: Ray.Wang@believe-med.com
- 4. Identification of Proposed Device
Trade Name: Paper Sterilization Pouch and Roll(Sterilization Pouch Flat, Sterilization Pouch Gusseted, Sterilization Roll Flat and Sterilization Roll Gusseted)
Common Name: Sterilization Pouches
Type(s): Sterilization Pouch Flat/ Sterilization Pouch Gusseted/ Sterilization Roll Flat/ Sterilization Roll Gusseted
Regulatory Information Classification Name: Wrap, Sterilization/Indicator, Physical/Chemical Sterilization Process Classification: 2
Product Code: FRG/JOJ Regulation Number: 21 CFR 880.6850/ 21 CFR 880.2800 Review Panel: General Hospital
Page 1 of 16
{5}------------------------------------------------
- న. Identification of Predicate Device(s)
Predicate Device 510(k) Number: K143637 Product Name: U & U Sterilization Pouch and Roll Model Name: U&U Medical Technology Co., Ltd.
- 6. Device Description
The Paper Sterilization Pouch and Roll is intended to contain medical devices to be terminally sterilized by the EtO or Steam sterilization process. The recommended sterilization cycle parameter is:
Steam; 30 minutes at 121°C; 25 minutes dry time.
Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60 ℃.
The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized for the EtO or Steam Sterilization Process. The heat-sealed pouch/roll are heat sealed prior to sterilization processing. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened.
The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 6 months post sterilization.
The Pouch/Roll is printed with a chemical indicator bar that changes from Red/Pink to Yellow/Brown (EtO) or Blue to Brown/Black (Steam) when exposed to EtO gas or Steam vapor during process. The EtO and Steam Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The EtO and Steam Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the EtO or Steam.
The Paper Sterilization Pouch and Roll is offered in the follow 4 types:
- * Sterilization Pouch Flat;
- * Sterilization Pouch Gusseted;
- * Sterilization Pouch Roll Flat;
- * Sterilization Pouch Roll Gusseted.
The defining characteristics of the 4 types as follows:
Sterilization Pouch Flat: These pouches are made from a medical grade paper and plastic (CPP/PET) film that is heat sealed on three sides, the forth side is left opened and will be heat-sealed when using. The Process Indicators Ink printed on the Pouch will exhibit a color change after the pouch is exposed to EtO gas or Steam.
{6}------------------------------------------------
Sterilization Pouch Gusseted: These rolls are the same as the Sterilization Pouch Flat, with the Heat Sealing except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
Sterilization Roll Flat: These rolls are made from a medical grade paper and plastic (CPP/PET) film that are heat sealed on opposite two sides. It will be cut into the suitable length and the opened sides will be heat-sealed. The indicators printed on it are the same as the sterilization pouches.
Sterilization Roll Gusseted: These rolls are the same with the flat sterilization roll, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
| Type | Model | Dimension in S.I |
|--------------------------|--------|------------------|
| Sterilization Pouch Flat | PFP001 | 55mm*150mm |
| | PFP002 | 55mm*200mm |
| | PFP003 | 55mm*250mm |
| | PFP004 | 55mm*300mm |
| | PFP005 | 75mm*150mm |
| | PFP006 | 75mm*200mm |
| | PFP007 | 75mm*250mm |
| | PFP008 | 75mm*300mm |
| | PFP009 | 100mm*150mm |
| | PFP010 | 100mm*200mm |
| | PFP011 | 100mm*250mm |
| | PFP012 | 100mm*300mm |
| | PFP013 | 100mm*350mm |
| | PFP014 | 100mm*400mm |
| | PFP015 | 125mm*150mm |
| | PFP016 | 125mm*200mm |
| | PFP017 | 125mm*250mm |
| | PFP018 | 125mm*300mm |
| | PFP019 | 125mm*350mm |
| | PFP020 | 125mm*400mm |
| | PFP021 | 150mm*200mm |
| | PFP022 | 150mm*250mm |
| | PFP023 | 150mm*300mm |
| | PFP024 | 150mm*350mm |
| | PFP025 | 150mm*400mm |
| | PFP026 | 200mm*250mm |
| | PFP027 | 200mm*300mm |
| | PFP028 | 200mm*350mm |
| | PFP029 | 200mm*400mm |
| | PFP030 | 200mm*450mm |
{7}------------------------------------------------
| PFP031 | 250mm*300mm |
|--------|-------------|
| PFP032 | 250mm*350mm |
| PFP033 | 250mm*400mm |
| PFP034 | 250mm*450mm |
| PFP035 | 300mm*300mm |
| Type | Model | Dimension in S.I |
|------------------------------|--------|--------------------|
| Sterilization Pouch Gusseted | PFP036 | 300mm*350mm |
| | PFP037 | 300mm*400mm |
| | PFP038 | 300mm*450mm |
| | PFP039 | 350mm*450mm |
| | PFP040 | 450mm*550mm |
| | PGP001 | 75mm*25mm*150mm |
| | PGP002 | 100mm*50mm*300mm |
| | PGP003 | 150mm*50mm *350mm |
| | PGP004 | 150mm *50mm *450mm |
| | PGP005 | 200mm*50mm *350mm |
| | PGP006 | 200mm*50mm *450mm |
| Sterilization Roll Flat | PGP007 | 250mm*55mm*400mm |
| | PGP008 | 300mm*60mm *400mm |
| | PGP009 | 350mm*65mm*450mm |
| | PGP010 | 400mm*75mm*500mm |
| | PFR001 | 55mm *200m |
| | PFR002 | 75mm *200m |
| | PFR003 | 100mm *200m |
| | PFR004 | 125mm *200m |
| | PFR005 | 150mm *200m |
| | PFR006 | 175mm *200m |
| | PFR007 | 200mm *200m |
| Sterilization Roll Flat | PFR008 | 250mm*200m |
| | PFR009 | 300mm *200m |
| | PFR010 | 350mm *200m |
| | PFR011 | 400mm*200m |
| | PFR012 | 450mm *200m |
| | PGR001 | 75mm*25mm*100m |
| | PGR002 | 100mm*50mm*100m |
| | PGR003 | 150mm*50mm*100m |
| | PGR004 | 200mm*50mm*100m |
| | PGR005 | 250mm*55mm*100m |
| Sterilization Roll Gusseted | PGR006 | 300mm*60mm*100m |
| | PGR007 | 350mm*65mm*100m |
| | PGR008 | 400mm*75mm*100m |
{8}------------------------------------------------
#### 7. Indication Use Statement:
#### Sterilization Pouch Flat
The Sterilization Pouch Flat are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Sterilization Pouch Flat are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below:
Steam; 30 minutes at 121 ℃; 25 minutes dry time.
Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60 ℃.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. The Sterilization Pouch Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 1470 g. The sterilization pouch maintains the enclosed devices up until 6 months post sterilization.
#### Sterilization Pouch Gusseted
The Sterilization Pouch Gusseted are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Sterilization Pouch Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below:
Steam; 30 minutes at 121 ℃; 25 minutes dry time.
Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60 ℃.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. The Sterilization Pouch Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 1470g. The sterilization pouch maintains the enclosed devices up until 6 months post sterilization.
#### Sterilization Roll Flat
The Sterilization Roll Flat are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Sterilization Roll Flat are intended for sterilization of dental instruments, excluding complexdevices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below:
### Steam; 30 minutes at 121 ℃; 25 minutes dry time.
Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%-85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60 ℃.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone Page 5 of 16
{9}------------------------------------------------
either a steam or EtO sterilization process. The Sterilization Roll Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 910g. The sterilization roll maintains the enclosed devices up until 6 months post sterilization.
### Sterilization Roll Gusseted
The Sterilization Roll Gusseted are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Sterilization Roll Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below:
Steam; 30 minutes at 121 ℃; 25 minutes dry time.
Ethylene oxide: 4 hours at 55 ℃; relative humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60 ℃.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. The Sterilization Roll Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 910g. The sterilization roll maintains the enclosed devices up until 6 months post sterilization.
{10}------------------------------------------------
#### 8. Technological Characteristics Comparison Table
Table 1 General Comparison
| ITEM | Proposed Device | Predicate Device K143637 | Remark |
|--------------------|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|---------|
| Indication For Use | Sterilization Pouch Flat | The U&U sterilization pouch and roll are intended to provide health care | Similar |
| | The Sterilization Pouch Flat are intended to be used to enclose another medical | workers with an effective method to enclose devices intended for sterilization | |
| | devices that is to be sterilized by a health care provider. It is intended to allow | in steam and Ethylene Oxide (EtO). | |
| | sterilization of the enclosed medical device and also to maintain sterility of the | The recommended gravity steam sterilization cycle parameters are 30 minutes | |
| | enclosed device until used. | at 121 °C. | |
| | The Sterilization Pouch Flat are intended for sterilization of dental instruments, | The recommended EtO sterilization cycle is 4 hours at 55 °C with a relative | |
| | excluding complex devices (endoscopes and instruments with lumen/channels). | humidity between 50%- 85% and a sterilant concentration of 600 mg/L. | |
| | The intended sterilization cycles are listed below: | Furthermore, the sterilization pouch and roll maintain the enclosed devices up | |
| | Steam; 30 minutes at 121 °C; 25 minutes dry time. | until 90Days post sterilization. Lastly, the pouch's external chemical ink | |
| | Ethylene oxide: 4 hours at 55 °C; relative humidity between 50%- 85%; ethylene | indicators are designed to indicate to the user that the pouch has undergone | |
| | oxide concentration is 600 mg/L, 8 hours aeration time at 60°C. | either a steam or EtO sterilization process. | |
| | The pouch's external chemical ink indicators are designed to indicate to the user | | |
| | that the pouch has undergone either a steam or EtO sterilization process. | The recommended gravity steam sterilization cycle parameters | |
| | The Sterilization Pouch Flat are not intended use for any load with | Steam sterilization temperature: 121°C (250°F) | |
| | lumen/channels and complex device. The maximum wrapped package weight for | Steam sterilization time: 30 minutes. | |
| | each pouch is 1470 g. | Drying time: 25 minutes | |
| | The sterilization pouch maintains the enclosed devices up until 6 months post | The recommended EtO sterilization cycle parameters | |
| | sterilization. | EtO sterilization temperature: 55°C (130 °F) | |
| | Sterilization Pouch Gusseted | EtO sterilization time: 4 hours | |
| | The Sterilization Pouch Gusseted are intended to be used to enclose another | EtO sterilization humidity: 50% to 85% RH | |
| | medical devices that is to be sterilized by a health care provider. It is intended to | EtO sterilization concentration:600mg/L | |
{11}------------------------------------------------
| allow sterilization of the enclosed medical device and also to maintain sterility of<br>the enclosed device until used.<br>The <i>Sterilization Pouch Gusseted</i> are intended for sterilization of dental<br>instruments, excluding complex devices (endoscopes and instruments with<br>lumen/channels).<br>The intended sterilization cycles are listed below:<br>Steam; 30 minutes at 121 °C; 25 minutes dry time.<br>Ethylene oxide: 4 hours at 55 °C; relative humidity between 50%-85%; ethylene<br>oxide concentration is 600 mg/L, 8 hours aeration time at 60°C.<br>The pouch's external chemical ink indicators are designed to indicate to the user<br>that the pouch has undergone either a steam or EtO sterilization process.<br>The <i>Sterilization Pouch Gusseted</i> are not intended use for any load with<br>lumen/channels and complex device. The maximum wrapped package weight for<br>each pouch is 1470g.<br>The sterilization pouch maintains the enclosed devices up until 6 months post<br>sterilization.<br><i>Sterilization Roll Flat</i><br>The <i>Sterilization Roll Flat</i> are intended to be used to enclose another medical<br>devices that is to be sterilized by a health care provider. It is intended to allow<br>sterilization of the enclosed medical device and also to maintain sterility of the<br>enclosed device until used.<br>The <i>Sterilization Roll Flat</i> are intended for sterilization of dental instruments,<br>excluding complex devices (endoscopes and instruments with lumen/channels).<br>The intended sterilization cycles are listed below:<br>Steam; 30 minutes at 121 °C; 25 minutes dry time.<br>Ethylene oxide: 4 hours at 55 °C; relative humidity between 50%-85%; ethylene | Aeration time: 8 hours.<br>Aeration Temperature: 60°C<br>Sterilization load claim:<br>Two types of sterilization loads were validated.<br>Load A: Metal medical instruments and Hand-control pen, the total weight is<br>24lbs.<br>Load B: Tubing (Silicone) and Surgical Towels. The total weight is 18lbs. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| oxide concentration is 600 mg/L, 8 hours aeration time at 60°C.<br>The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.<br>The Sterilization Roll Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 910g.<br>The sterilization roll maintains the enclosed devices up until 6 months post sterilization. | |
| Sterilization Roll Gusseted<br>The Sterilization Roll Gusseted are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.<br>The Sterilization Roll Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).<br>The intended sterilization cycles are listed below:<br>Steam; 30 minutes at 121 °C; 25 minutes dry time.<br>Ethylene oxide: 4 hours at 55 °C; relative humidity between 50%-85%; ethylene oxide concentration is 600 mg/L, 8 hours aeration time at 60°C.<br>The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.<br>The Sterilization Roll Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 910g.<br>The sterilization roll maintains the enclosed devices up until 6 months post sterilization. | |
{12}------------------------------------------------
{13}------------------------------------------------
| Material | Top Web - Medical Porous Paper Bottom | Top Web - Medical Porous Paper Bottom | Similar |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Composition | Web - Medical Plastic film(CPP) | Web - Medical Plastic film(CPP) | |
| | EtO gas indicator ink-Process Indicators<br>Steam indicator ink-Process Indicators | EtO gas indicator ink-Process Indicators class 1<br>Steam indicator ink-Process Indicators class 1 | |
| Sterilization | Prevacuum steam; 30 minutes at 121 °C; 25 minutes dry time. | The recommended gravity steam sterilization cycle parameters are 30 minutes<br>at 121 °C. | Similar |
| Cycles | Ethylene oxide: 4 hours at 55 °C; relative humidity between 50%- 85% and a | The recommended EtO sterilization cycle is 4 hours at 55 °C with a | |
| | sterilant concentration of 600 mg/L. | relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. | |
| Configuration/<br>Dimension | Various Size, Heat Sealing | Various Size, Heat Sealing and Self Sealing | Similar |
| Air Permeance | The maximum equivalent pore size diameter shall not exceed 50um. | The maximum equivalent pore size diameter shall not exceed 50um. | Similar |
| Microbial Barrier<br>Properties<br>(Packaging<br>Integrity) | Use ASTM 1608 method, the LRV is more than 4.0 Use ASTM 1929 method, the<br>inspection result is PASS | Use ASTM 1608 method, the LRV is more than 3.5 Use ASTM 1929 method,<br>the inspection result is PASS | Similar |
| Material<br>Compatibility | After sterilization, the materials were not degraded | After sterilization, the materials were not degraded | Similar |
| Biocompatibility | ISO10993-10, Test for Irritation; ISO10993-10, Test for Skin sensitization; Both<br>were non-cytotoxic post sterilization. | ISO10993-10, Test for Irritation; ISO10993-10, Test for Skin sensitization; Both<br>were non-cytotoxic post sterilization. | Similar |
| Maintenance of<br>Sterility | 6 months | 90 Days | Similar |
| Shelf Life | 2 years from date of manufacture for EO and Steam Indicators | 18 months from date of manufacture for EO and steam Indicators | Similar |
| Drying Time | 25 minutes | 25 minutes | Similar |
| Aeration Time | 8 hours at 60°C | 8 hours at 60°C | Similar |
| Chemical | Changed color EtO- Red/Pink to Yellow/Brown; | Changed color EtO- YELLOW to COCOA; | Similar |
| Indicator Efficacy | Steam- Blue to Brown/Black | Steam- GREEN to PURPLE | Similar |
| End point stability<br>testing results | The color of chemical indicator for EO sterilization indicator ink is Red/Pink, and<br>the color of chemical indicator for steam sterilization indicator ink is Blue after 2<br>year shelf life before sterilization. | The color of chemical indicator could maintain stability as claimed shelf life (18<br>months from date of manufacture for EO and steam indicators, 90 days<br>maintenance of sterility after sterilization process) | Similar |
{14}------------------------------------------------
| The color of chemical indicator for EO sterilization indicator ink is Golden<br>Yellow/Brown, and the color of chemical indicator for steam sterilization indicator<br>ink is Blue after EO sterilized and 6 months shelf life. |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The color of chemical indicator for EO sterilization indicator ink is Red/Pink , and<br>the color of chemical indicator for steam sterilization indicator ink is Brown/Black |
{15}------------------------------------------------
### 9. Summary of Non-Clinical Testing
The proposed device Non-clinical tests were conducted to subject device. The test results demonstrated that the subject device met the acceptance criteria, and the conducted tests listed as below:
| Testing Name | Brief Description of Test | Acceptance criteria OR End Point | Test Results |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tensile<br>strength | ASTM D882-12 Standard Test Methods for Tensile<br>Properties of Thin Plastic Sheeting; | ≥4.4 KN/m | The test results of all samples are ≥4.4 KN/m.<br>Pass |
| Thickness | ASTM F2251-03 Standard Test Method for thickness<br>measurement of flexible packaging material; | 138±13 µm | The test results of all samples are in the range<br>of 138±13 µm.<br>Pass |
| Tear Strength | ASTM D1922-03 Standard Test Method for Propagation<br>tear resistance of plastic film and thin sheeting by pendulum<br>method; | ≥ 550 mN/15mm | The test results of all samples are ≥ 550<br>mN/15mm.<br>Pass |
| Air<br>permeability<br>coefficient | ISO 5636-3:2013 Paper and board - Determination of air<br>permeance (medium range) – Part 3: Bendtsen method; | ≥3.4µm/Pa.s | The test results of all samples are ≥3.4µ<br>m/Pa.s.<br>Pass |
| Burst Strength | ASTM F1140/f1140M-13 Standard Test Methods for<br>internal pressurization failure resistance of unrestrained<br>package; | ≥2.0 kPa | The test results of all samples are ≥2.0 kPa.<br>Pass |
| Microorganism<br>Penetration | ASTM F1608-00 Standard test methods for Microbial<br>Ranking of Porous packaging materials; | Microbial Barrier LRV ≥ 2.0<br>Retention Rate (spore%) ≥ 99.0% | The Microbial Barrier LRV test results of all<br>samples are ≥2.0. |
| | | | The Retention Rate test results of all samples are ≥99.0%. |
| | | | Pass |
| Chemical<br>Indicator<br>Performance | ISO 11140-1:2009 Sterilization of Health Care Products -<br>Chemical Indicators - Part 1: General Requirements; | Color changed from Red/Pink to Golden<br>Yellow/Brown for EO sterilization indicator ink at the condition of 4 hours at 55 °C; relative<br>humidity between 50%- 85%; ethylene oxide concentration is 600 mg/L;<br>Color changed from Blue to Brown/Black at the condition of Gravity Steam, 121°C, 30 mins; | Color changed from Red/Pink to Golden<br>Yellow/Brown under indicated condition and no changed under other condition.<br>Pass<br>Color changed from Blue to Brown/Black under indicated condition and no changed under other condition.<br>Pass |
| Dye<br>penetration | ASTM F1929-12 Standard test methods for detecting seal leaks in porous medical packaging by dye penetration. | No dye leakage | No dye leakage/Pass |
| Seal strength | ASTM F88-15 Standard Test Method for Seal Strength of Flexible Barrier Materials | ≥1.5 N/15 mm | The test results of all samples are ≥1.5 N/15 mm.<br>Pass |
| EO residue | ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals | No more than 4 mg per set | The test results of all samples are no more than 4 mg per set.<br>Pass |
| ECH residue | | No more than 9 mg per set | The test results of all samples are no more than 9 mg per set.<br>Pass |
| In Vitro<br>Cytotoxicity | ISO 10993-5:2009 Biological Evaluation of Medical<br>Devices – Part 5: Tests for In Vitro Cytotoxicity; | In Vitro Cytotoxicity | Under the conditions of the study, the test<br>article (pouches and chemical indicator)<br>extract did not show potential toxicity to L-<br>929 cells before and after sterilized. |
| Skin Irritation<br>AND Skin<br>Sensitization | ISO 10993-10:2010 Biological Evaluation of Medical<br>Devices - Part 10: Tests for irritation and skin sensitization; | Skin sensitization; | Under the conditions of the study, the test<br>article (pouches and chemical indicator)<br>extract did not show no significant evidence of<br>causing skin sensitization in the guinea pig<br>before and after sterilized. |
| | | Skin Irritation | Under the conditions of the study, the test<br>article (pouches and chemical indicator)<br>extract did not induce intracutaneous<br>reactivity in rabbit before and after sterilized. |
| Shelf Life<br>Validation | Shelf Life Validation Test;<br>Conducted the shelf life testing as real time aging method. | After the shelf life indicated as following,<br>Shelf Life: 2 Years;<br>Shelf Life after Sterilized: 6 Months;<br>The performance shall be meet the requirements<br>of :<br>Tensile strength: ≥4.4 KN/m;<br>Thickness: 138±13 µm;<br>Tear Strength: ≥ 550 mN/15mm;<br>Air permeability coefficient: ≥3.4µm/Pa.s;<br>Burst Strength: ≥2.0 kPa; | The test results of all samples are meet the<br>requirements.<br>Pass |
| | | Microbial Barrier LRV ≥ 2.0;<br>Retention Rate (spore%) ≥ 99.0%;<br>Dye penetration: No dye leakage;<br>Seal strength: ≥1.5 N/15 mm;<br>Color of indicator shall be meet the specification. | |
| Sterilization<br>Process<br>Validation | EO Sterilization Process Validation Test.<br>Conducted the EO sterilization process validation as the<br>method/principle of ISO 14937:2009.<br>Use the half cycle method to validate the EO sterilization<br>cycle claimed in indication for use is effective. | Ethylene oxide: 4 hours at 55 °C; relative<br>humidity between 50%- 85%; ethylene oxide<br>concentration is 600 mg/L, 8 hours aeration time<br>at 60°C.<br>SAL= $10^{-6}$ . | SAL= $10^{-6}$ .<br>Pass |
| | Steam Sterilization Process Validation Test.<br>Conducted the Steam sterilization process validation as the<br>method/principle of ISO 14937:2009.<br>Use the half cycle method to validate the EO sterilization<br>cycle claimed in indication for use is effective. | Steam; 30 minutes at 121 °C; 25 minutes dry<br>time.<br>SAL= $10^{-6}$ . | SAL= $10^{-6}$ .<br>Pass |
{16}------------------------------------------------
{17}------------------------------------------------
{18}------------------------------------------------
{19}------------------------------------------------
## 10. Clinical Test Conclusion
No clinical study is included in this submission.
## 11. Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.