K143637 · U&U Medical Technology Co, Ltd. · FRG · Sep 9, 2015 · General Hospital
Device Facts
Record ID
K143637
Device Name
U&U Sterilization Pouch and Roll
Applicant
U&U Medical Technology Co, Ltd.
Product Code
FRG · General Hospital
Decision Date
Sep 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The U&U sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam and Ethylene Oxide (EtO). The recommended gravity steam sterilization cycle parameters are 30 minutes at 121°C. The recommended EtO sterilization cycle is 4 hours at 55°C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 90 Days post sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
Device Story
U&U Sterilization Pouch and Roll consist of medical porous paper and plastic film (CPP) heat-sealed on three sides; fourth side sealed via adhesive tape or heat-sealing machine. Used by healthcare workers to enclose medical instruments for gravity steam (121°C) or EtO (55°C, 600mg/L) sterilization. External chemical ink indicators (Class 1) change color (Steam: Green to Purple; EtO: Yellow to Cocoa) to confirm exposure to sterilization process. Pouch maintains sterility for 90 days post-sterilization. Protects instruments during storage and transport; ensures sterility maintenance until point of use.
Clinical Evidence
Bench testing only. Validated via physical testing of sterilized finished devices: seal strength, package burst, dye migration, sterilant penetration, package integrity, material compatibility, and biocompatibility (ISO 10993-4, ISO 10993-10). Microbial barrier properties confirmed via ASTM 1608 (LRV > 3.5). Chemical indicator efficacy and shelf-life testing performed.
Technological Characteristics
Materials: Medical porous paper (top), medical plastic film (CPP, bottom), two-sided adhesive tape. Class 1 chemical process indicators (steam/EtO). Standards: AAMI/ANSI ST77, AAMI TIR22, ISO 11607-1/2, ISO 11140-1. Dimensions: Various pouch/roll sizes. Sterilization: Gravity steam (121°C) or EtO (55°C).
Indications for Use
Indicated for healthcare workers to enclose medical devices for steam or EtO sterilization. Compatible with reusable surgical instruments (clamps, scissors, retractors, pliers), endoscopes, reusable containers, elastic products (rubber, latex, silicone, plastics), HF-cables, handpieces, power tools, and textile materials (surgical towels).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 9, 2015
U&U Medical Technology Co., Ltd Mr. Garfield Wang General Manager Dongzhou Village, Hengshanqiao RM EE1092 1/F Building 1, No 1755 Hongmei Road Changzhou, Jiangsu China
Re: K143637
Trade/Device Name: U&U Sterilization Pouch and Roll Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG, JOJ Dated: August 2, 2015 Received: August 5, 2015
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Tejashri Purohit-Sheth, M.D.**
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143637
Device Name U&U Sterilization Pouch and Roll
#### Indications for Use (Describe)
| Form Approved: OMB No. 0910-0120 |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement below. |
The U&U sterilization pouch and roll are intended to provide with an effective method to enclose devices intended for sterlization in steam and Ethylene Oxide (EtO). The recommended g r a vi t y steam sterilization vole parameters at 121°C. The recommended EiO sterlization cycle is 4 hours at 5°C with a relative humidity between 50%-55% and a sterilant of 600 mgL. Furthernon, the sterlization pouch and roll maintains the enclosed devices up until 90Days post sterial chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
STERILIZATION CYCLE:
The recommended g r a v i t y steam sterilization cycle parameters Steam sterilization temperature: 121°C (250 °F) Steam sterilization time: 30 minutes.
Drying time: 25 minutes The recommended EtO sterilization cycle parameters EtO sterilization temperature: 55℃ (130 °F) EtO sterilization time: 4 hour EtO sterilization humidity: 50% to 85% RH EtO sterilization concentration: 600mg/L Aeration time: 8 hours. Aeration Temperature: 60°C
Sterilization load claim:
Two types of sterilization loads were validated. Load A: Metal medical instruments and Hand-control pen, the total weight is 24lbs. Load B: Tubing (Silicone) and Surgical Towels. The total weight is 181bs.
Type of Use (Select one or both, as applicable)
□Prescription Use (Part 21 CFR 801 Subpart D)
|X} Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### Load Types:
The following matrix is the types of medical instruments used for steam and EtO sterilization.
| Medical Device | Steam | EtO Gas |
|------------------------------------------------------------------------------------------------------------|-------|---------|
| Reusable surgical instruments such as clamps, scissors, retractors, pliers | X | X |
| Endoscopes (thermostable) | X | X |
| Endoscopes<br>(thermolabile) | - | X |
| Reusable containers | X | - |
| Elastic Products such as Rubber or latex (also combined), siliconelastomer, plastics, closed hollow bodies | X | X |
| HF-cable and handpieces | X | X |
| Powertools | X | X |
| Textile Material such as surgical towels | X | X |
{5}------------------------------------------------
U&U ChangZhou Medical Packaging Technology Co., Ltd Dongzhou Village,Hengshangiao,Changzhou,Jiangsu,China U&U (Shanghai) Medical Technology Co., Limited RM EE1092 1/F Building 1, No 1755, HONGMEI Road, Shanghai, China [Sterilization Pouch]
Rev 1.00 09/09/15
# Sec 005_510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92
Date Prepared: 2015-09-09 Submission Numbers for Pre-Submission: K143637
#### 1. Submitter Name and Address:
| Name: | U&U Medical Technology Co., Ltd |
|-----------------|---------------------------------------------------------------------------------------------------------------------------|
| Address: | Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China<br>RM EE1092 1/F Building 1, No 1755, HONGMEI Road, Shanghai, China |
| Contactor Name: | Garfield Wang |
| TEL: | +86-13902471751 |
| E-mail: | Wangxuebo_11@hotmail.com |
#### US Agent:
| Name: | CARELIFE (USA) INC. |
|------------------|----------------------------------------------|
| Address: | 1580 Boggs Rd, Suite 500/600 Duluth GA 30096 |
| TEL: | 404 6612228 |
| Contact person : | Ms. LI QIAN liqian@shanghaicarelife.com |
#### 2. Submission Devices Information:
Trade/Proprietary Name: U&U Sterilization Pouch and Roll Common Name: Sterilization Pouch Classification name: 1) Sterilization wrap 2) Indicator, Physical/Chemical Sterilization Process 21 CFR 880.6850 & 21 CFR 880.2800 Class: Il FDA review panel code: General Hospital Product code: 1) FRG 2) JOJ
#### 3. Predicate Devices Information:
| Trade Name: | SIGMA Sterilization Pouch and Roll |
|----------------|------------------------------------|
| 510(K) Number: | K102158 |
| Manufacturer: | SIGMA Medical Supplies Corp. |
## 4. Devices Description:
The U&U Sterilization Pouch and Roll are manufactured from paper and medical plastic film that are heat sealed on three sides. The fourth side has an adhesive tape that is used to seal the pouch or heat-sealed by the heat-seal machine. The paper conforms to recognized material standards and can be sterilized by steam and Eto. The Sterilization Pouch has the same intended use, Essential Component, Raw material, Sterilization method, manufacturing methods and same technological characteristics as the predicate device. Substantial equivalent to the predicate device was established by physical testing of the sterilized finished devices, as well as, performance of these finished devices (seal strength, package burst, and dye migration).
{6}------------------------------------------------
U&U ChangZhou Medical Packaging Technology Co., Ltd Dongzhou Village,Hengshangiao,Changzhou,Jiangsu,China U&U (Shanghai) Medical Technology Co., Limited RM EE1092 1/F Building 1, No 1755, HONGMEI Road, Shanghai, China [Sterilization Pouch]
Rev 1.00 09/09/15
### 5. Model numbers of the pouches:
| Ref<br>Number | Model<br>Number | Description | Size |
|---------------|-----------------|---------------------------------------------|--------------|
| UUPP0001 | UUPP | STERILIZATION Pouch-Heat-Seal and Self-Seal | 3" X 8" |
| UUPP0002 | UUPP | STERILIZATION Pouch-Heat-Seal and Self-Seal | 4" X 10.25" |
| UUPP0003 | UUPP | STERILIZATION Pouch-Heat-Seal and Self-Seal | 6" X 12.5" |
| UUPP0004 | UUPP | STERILIZATION Pouch-Heat-Seal and Self-Seal | 4" X 14" |
| UUPP0005 | UUPP | STERILIZATION Pouch-Heat-Seal and Self-Seal | 8" X 16" |
| UUPP0006 | UUPP | STERILIZATION Pouch-Heat-Seal and Self-Seal | 6" X 16.5" |
| UUPP0007 | UUPP | STERILIZATION Pouch-Heat-Seal and Self-Seal | 10" X 19" |
| UUPP0008 | UUPP | STERILIZATION Pouch-Heat-Seal and Self-Seal | 12.5" X 22" |
| UUPR0001 | UUPR | STERILIZATION Roll-Heat-seal | 3" X 228' |
| UUPR0002 | UUPR | STERILIZATION Roll-Heat-seal | 4" X 228' |
| UUPR0003 | UUPR | STERILIZATION Roll-Heat-seal | 6" X 228' |
| UUPR0004 | UUPR | STERILIZATION Roll-Heat-seal | 14" X 228' |
| UUPR0005 | UUPR | STERILIZATION Roll-Heat-seal | 16.5" X 228' |
| UUPR0006 | UUPR | STERILIZATION Roll-Heat-seal | 20" X 228' |
#### 5. Intended Use:
The U&U sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam and Ethylene Oxide (EtO). The recommended gravity steam sterilization cycle parameters are 30 minutes at 121°C. The recommended EtO sterilization cycle is 4 hours at 55°C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 90 Davs post sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
#### STERILIZATION CYCLE:
The recommended gravity steam sterilization cycle parameters Steam sterilization temperature: 121°C (250 °F) Steam sterilization time: 30 minutes. Drying time: 25 minutes The recommended EtO sterilization cycle parameters EtO sterilization temperature: 55°C (130 ° F) EtO sterilization time: 4 hour EtO sterilization humidity: 50% to 85%RH EtO sterilization concentration: 600mg/L Aeration time: 8 hours. Aeration Temperature: 60°C
#### Sterilization load claim:
Two types of sterilization loads were validated
Load A: Metal medical instruments and Hand-control pen, the total weight are 24lb. The double pouched devices are placed into the baskets. Load B: Tubing (Silicone) and Surgical Towels. The total weight is 18lbs.
ISEC 0051
{7}------------------------------------------------
#### Rev 1.00 09/09/15
#### Load Types:
The following matrix is the types of medical instruments used to the sterilization:
| Medical Device | Steam | EtO Gas |
|------------------------------------------------------------------------------------------------------------------|-------|---------|
| Reusable surgical instruments such as<br>clamps, scissors, retractors, pliers | X | X |
| Reusable containers | X | - |
| Elastic Products such as Rubber or latex<br>(also combined), siliconelastomer,<br>plastics, closed hollow bodies | X | X |
| HF-cable and handpieces | X | X |
| Powertools | X | X |
| Textile Material such as surgical towels | X | X |
#### 6. Technological Characteristics:
All testing followed the following standards.
AAMI/ANSI ST77 Containment devices for reusable medical device sterilization AAMI TIR22 Guidance for ANSI/AAMI/ISO 11607-Packaging for terminally sterilized medical devices-Part 1 and Part 2:2006
AAMI/ANS//ISO 11140-1 Sterilization of health care products - Chemical indicators - Part 1: General requirements
### 7. Performance Testing:
Performance testing was conducted to show that the U&U Sterilization Pouch and Roll maintain sterility until the seal of the Pouch/Roll is opened. Performance Tests
| Element | Results |
|-----------------------------------------|---------|
| Sterilant Penetration | PASSED |
| Package Integrity | PASSED |
| Maintenance of Package Integrity | PASSED |
| Material Compatibility | PASSED |
| Shelf-Life | PASSED |
| Biocompatibility | PASSED |
| Labeling | PASSED |
| Configurations /Dimensions | PASSED |
| Air permeance | PASSED |
| Maintenance of Sterility | PASSED |
| Endpoint stability of process indicator | PASSED |
| Shelf Life of Process Indicator | PASSED |
| Chemical Indicator Efficacy | PASSED |
| Seal strength | PASSED |
| Peel-open characteristic | PASSED |
| Seal width | PASSED |
| Self seal strength | PASSED |
| Visual inspection | PASSED |
{8}------------------------------------------------
| U&U ChangZhou Medical Packaging Technology Co., Ltd<br>Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China<br>U&U (Shanghai) Medical Technology Co., Limited<br>RM EE1092 1/F Building 1, No 1755, HONGMEI Road, Shanghai, China | 510(k) Submission | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|----------|--|
| [Sterilization Pouch] | Rev 1.00 | 09/09/15 | |
| Ethylene oxide residuals | PASSED | | |
# 8. SUBSTANTIAL EQUIVALENCE DISCUSSION:
Our Sterilization Pouch and the predicate device are same in intended use, components, materials, sterilization, and function.
Side by side testing was conducted on the U&U Sterilization Pouch and Roll and SIGMA Sterilization pouch and roll - K102158 to determine substantial equivalence. Sterilant Penetration, Biocompatibility, Package Integrity, Material Compatibility, Sterility Maintenance were the parameters used to determine substantial equivalence and validate the safety and efficacy of the device.
| Element of<br>Comparison | Submission Device | Predicate Device<br>K102158 | | | | | | | | | | |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|
| Intended Use | The U&U sterilization pouch and roll are<br>intended to provide health care workers<br>with an effective method to enclose<br>devices intended for sterilization in<br>steam and Ethylene Oxide (EtO). The<br>recommended gravity steam sterilization<br>cycle parameters are 30 minutes at<br>121 °C. The recommended Eto<br>sterilization cycle is 4 hours at 55 °C<br>with a relative humidity between 50%-<br>85% and a sterilant concentration of 600<br>mg/L. Furthermore, the sterilization<br>pouch and roll maintains the enclosed<br>devices up until 90Days post sterilization.<br>Lastly, the pouch's external chemical ink<br>indicators are designed to indicate to the<br>user that the pouch has undergone either<br>a steam or Eto sterilization process<br><br>STERILIZATION CYCLE:<br>The recommended gravity steam sterilization cycle<br>parameters<br>Steam sterilization temperature: 121°C (250 °F)<br>Steam sterilization time: 30 minutes.<br>Drying time: 25 minutes<br>The recommended Eto sterilization cycle parameters<br>Eto sterilization temperature: 55°C (130 ° F)<br>Eto sterilization time: 4 hour<br>Eto sterilization humidity: 50% to 85%RH<br>Eto sterilization concentration: 600mg/L<br>Aeration time: 8 hours.<br>Aeration Temperature: 60°C<br><br>Sterilization load claim:<br>Two types of sterilization loads were validated.<br><br>Load A: Metal medical instruments and Hand-control pen,<br>the total weight are 24lb. The double pouched devices are<br>placed into the baskets.<br>Load B: Tubing (Silicone) and Surgical Towels. The total<br>weight is 18lbs. | The SIGMA sterilization pouch and roll<br>are intended to provide health care<br>workers with an effective method to<br>enclose devices intended for sterilization<br>in steam and Ethylene Oxide (EtO). The<br>recommended steam sterilization cycle<br>parameters are 30 minutes at 121 °C.<br>The recommended Eto sterilization cycle<br>is 4 hours at 55 °C with a relative<br>humidity between 50%-85% and a<br>sterilant concentration of 600 mg/L.<br>Furthermore, the sterilization pouch and<br>roll maintains the enclosed devices up<br>until 2 years post sterilization. Lastly, the<br>pouch's external chemical ink indicators<br>are designed to indicate to the user that<br>the pouch has undergone either a steam<br>or EtO sterilization process | | | | | | | | | | |
| | | | | | | | | | | | | |
| | Medical Device Steam EtO Gas Reusable<br>surgical<br>instruments<br>such as clamps,<br>scissors,<br>retractors, pliers X X Reusable<br>containers X - | | | | | | | | | | | |
{9}------------------------------------------------
U&U ChangZhou Medical Packaging Technology Co., Ltd Dongzhou Village,Hengshanqiao,Changzhou,Jiangsu,China (Shanghai) Medical Technology Co., Limited
U&U (Shanghai) Medical Technology Co., Limited
RM EEE0092 1/F Building 1, No 1755, HONGMEl Road, Shanghai, China [Sterilization Pouch]
510(k) Submission
09/09/15 Rev 1.00
| | Elastic Products<br>such as Rubber<br>or latex (also<br>combined).<br>siliconelastomer<br>, plastics,<br>closed hollow<br>bodies<br>HF-cable and | × | X | |
|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | handpieces | × | X | |
| | Powertools | × | X | |
| | Textile Material<br>such as surgical<br>towels | × | X | |
| Material<br>Composition | Top Web - Medical Porous Paper<br>Bottom Web - Medical Plastic film(CPP)<br>Bottom Web - Medical two-sided adhesive tape<br>EtO gas indicator ink-Process Indicators class 1 | | | Top Web - Medical Porous Paper<br>Bottom Web - Medical Plastic film(CPP)<br>Bottom Web -Medical two-sided adhesive tape<br>EtO gas indicator ink-Process Indicators class 1 |
| | Steam indicator ink-Process Indicators class 1 | | | Steam indicator ink-Process Indicators class 1 |
| Sterilization<br>Cycles | The recommended gravity steam sterilization cycle<br>parameters are 30 minutes at 121 ℃. The<br>recommended EtO sterilization cycle is 4 hours at<br>55 ℃ with a relative humidity between 50%-85% and<br>a sterilant concentration of 600 mg/L. | | The recommended steam sterilization cycle parameters<br>are 30 minutes at 121 ℃. The recommended EtO<br>sterilization cycle is 4 hours at 55 ℃ with a relative<br>humidity between 50%-85% and a sterilant<br>concentration of 600 mg/L. | |
| Confiqurations<br>/Dimensions | For Pouch:<br>Width: ±0.1"" Length ± 0.2""<br>For Roll:<br>Width: ±0.1"" Length ± 3% | | | For Pouch:<br>Width: ±0.1"" Length ± 0.2""<br>For Roll:<br>Width: ±0.1"" Length ± 3% |
| Air permeance | The maximum equivalent pore size diameter shall not<br>exceed 50um. | | | The maximum equivalent pore size diameter shall not<br>exceed 50um. |
| Sterilant<br>Penetration | Use the EN 868-2:1999 Annex C: Method for the<br>determination of the pore size, the pore size diameter is<br>40 to 44 um<br>The data is a little higher than the predicated device, The<br>difference can be accepted. | | Use the EN 868-2:1999 Annex C: Method for the<br>determination of the pore size, the pore size diameter is<br>38 to 43 um | |
| Microbial Barrier<br>Properties<br>(Packaging<br>Integrity) | Use ASTM 1608 method, the LRV is more than 3.5<br>Use ASTM 1929 method, the inspection result is PASS<br>The data is a little higher than the predicated device, The<br>difference can be accepted. | | Use ASTM 1608 method, the LRV is 3.3<br>Use ASTM 1929 method, the inspection result is PASS | |
| Material<br>Compatibility | After sterilization, the materials were not degraded | | After sterilization, the materials were not degraded | |
| Toxicological<br>Properties(Bioco<br>mpatibility,<br>including<br>Sterilant Residue<br>Limits) | Meet ISO10993-4.Haemolysis Test (Direct-contract<br>Method), ISO10993-10, Test for Irritation (Intractaneous<br>Reactivity Test) and ISO10993-10, Test for Skin<br>sensitization (Maximization test) | | Meet ISO10993-4,Haemolysis Test (Direct-contract<br>Method), ISO10993-10, Test for Irritation (Intractaneous<br>Reactivity Test) and ISO10993-10, Test for Skin<br>sensitization (Maximization test) | |
| Maintenance of<br>Sterility | 90 Days | | | 2 years |
| Endpoint stability<br>of process<br>indicator | 90 Days | | 2 years | |
| Shelf Life of<br>Process Indicator | 18 months | | 2 years | |
| Drying time | 25 minutes | | 25 minutes | |
| Aeration time | 8 hours at 60°C | | 8 hours at 60°C | |
| Chemical<br>Indicator Efficacy | Changed color<br>EtO- YELLOW to COCOA; Steam- GREEN to PURPLE | | Changed color | |
| | | | | |
# 9. Conclusion:
The performance testing data for the U&U Sterilization Pouch and Roll demonstrates substantial equivalence to the SIGMA Sterilization Pouch and Roll (K102158).
END
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.