Self Sealing Sterilization Pouches

K162258 · Tianchang Jiarui Packaging Material Co., Ltd. · FRG · Apr 28, 2017 · General Hospital

Device Facts

Record IDK162258
Device NameSelf Sealing Sterilization Pouches
ApplicantTianchang Jiarui Packaging Material Co., Ltd.
Product CodeFRG · General Hospital
Decision DateApr 28, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Intended Use

The Self Sealing Sterilization Pouches are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Self Sealing Sterilization Pouches are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below: Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time. Ethylene oxide: 1 hours at 55 ℃; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. The Self Sealing sterilization Pouches are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 540g or 1.19 lbs. The sterilization pouch maintains the enclosed devices up until 6 months post sterilization. The pouches are available in the following sizes: Model(s): 57 x 102mm, 57 x 130 mm, 70 x 255 mm, 83 x 165 mm, 13 x 191 mm, 133 x 279 mm, 133 x 290 mm, 140 x 280 mm, 140 x 330 mm, 190 x 330 mm, 190 x 360 mm, 255 x 380 mm, 279 x 406 mm, 300 x 400 mm, 300 x 474 mm, 305 x 457 mm;

Device Story

Self-sealing sterilization pouches; medical-grade paper and CPP/PET plastic film construction; heat-sealed on three sides with adhesive strip on fourth side. Used by healthcare providers to enclose dental instruments for sterilization. External chemical ink indicators (EtO: Pink to Yellow; Steam: Blue to Dark Green) provide visual confirmation of exposure to sterilization process. Maintains sterility for up to 6 months. Not for use with complex devices or lumens. Bench testing confirms microbial barrier integrity, biocompatibility, and sterilization efficacy.

Clinical Evidence

No clinical data. Bench testing only. Testing included microbial barrier properties (ASTM F1608), seal integrity (ASTM F1929), biocompatibility (ISO 10993-5, -7, -10), physical properties (tensile, tear, air permeance), and sterilization process validation for steam and EtO cycles.

Technological Characteristics

Materials: Medical porous paper, CPP/PET plastic film, PE/paper/PE release strip. Sensing: Chemical ink indicators (Class 1). Sterilization: Prevacuum steam (132°C, 4 min) and EtO (55°C, 740 mg/L). Standards: ISO 14937, ISO 11140-1, ASTM D638, ASTM F2251, ASTM D1922, ISO 5636-3, ASTM F1140, ASTM F1608, ASTM F1980, ASTM F1929.

Indications for Use

Indicated for use by healthcare providers to enclose dental instruments for sterilization via prevacuum steam or ethylene oxide (EtO) and to maintain sterility post-processing. Contraindicated for complex devices, including endoscopes and instruments with lumens or channels. Maximum load weight 540g.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 28, 2017 Tianchang Jiarui Packaging Material Co., Ltd. % Ray Wang General Manager Beijing Believe Technology Service Co., Ltd. 5-1206, Build 332, DaFangJu, No. 25 BanBiDian Rd., TongZhou District Beijing, 101121 CN Re: K162258 Trade/Device Name: Self Sealing Sterilization Pouches Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: March 8, 2017 Received: March 13, 2017 Dear Ray Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K162258 Device Name Self Sealing Sterilization Pouches The Self Sealing Sterilization Pouches are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Self Sealing Sterilization Pouches are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cvcles are listed below: Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time. Ethylene oxide: 1 hours at 55 ℃; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. The Self Sealing sterilization Pouches are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 540g or 1.19 lbs. The sterilization pouch maintains the enclosed devices up until 6 months post sterilization. The pouches are available in the following sizes: Model(s): 57 x 102mm, 57 x 130 mm, 70 x 255 mm, 83 x 165 mm, 13 x 191 mm, 133 x 279 mm, 133 x 290 mm, 140 x 280 mm, 140 x 330 mm, 190 x 330 mm, 190 x 360 mm, 255 x 380 mm, 279 x 406 mm, 300 x 400 mm, 300 x 474 mm, 305 x 457 mm; Type of Use (Select one or both, as applicable) ] Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K162258 - 1. Date of Preparation: 2017-4-24 - 2. Sponsor Identification Tianchang Jiarui Packaging Material Co., Ltd. Yeshan Industria Zone, Tianchang City, 239300, Anhui Province of China Contact Person: Zhang Ruiqing Position: General Manager Tel: +86 550 7981778 Fax: +86 550 7323988 Email: sales@cncarate.com - Designated Submission Correspondent 3. Mr. Ray Wang # Beijing Believe Technology Service Co., Ltd. 5-1206, Build 332, DaFangJu, No.25 BanBiDian Rd., LiYuan Town, TongZhou District, Beijing, 101121, China Tel: +86-18910677558 Fax: +86-10-52214696 Email: ray.wang@believe-med.com {4}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: Self Sealing Sterilization Pouches Common Name: Sterilization Pouches Model(s): 57 x 102mm, 57 x 130 mm, 70 x 255 mm, 83 x 165 mm, 90 x 260 mm, 133 x 191 mm, 133 x 279 mm, 133 x 290 mm, 140 x 280 mm, 140 x 330 mm, 190 x 330 mm, 190 x 360 mm, 255 x 380 mm, 279 x 406 mm, 300 x 400 mm, 300 x 474 mm, 305 x 457 mm; # Regulatory Information Classification Name: Wrap, Sterilization/Indicator, Physical/Chemical Sterilization Process Classification: 2 Product Code: FRG/JOJ Regulation Number: 21 CFR 880.6850/ 21 CFR 880.2800 Review Panel: General Hospital ### Intended Use Statement: The Self Sealing Sterilization Pouches are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Self Sealing Sterilization Pouches are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels). The intended sterilization cycles are listed below: Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time. Ethylene oxide: 1 hours at 55 ℃; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃. The pouch's external chemical ink indicators are designed to the user that the pouch has undergone either a steam or EtO sterilization process. The Self Sealing sterilization Pouches are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 540g or 1.19 lbs. The sterilization pouch maintains the enclosed devices up until 6 months post sterilization. The pouches are available in the following sizes: Model(s): 57 x 102mm, 57 x 130 mm, 70 x 255 mm, 83 x 165 mm, 90 x 260 mm, 133 x 191 mm, 133 x 279 mm, 133 x 290 mm, 140 x 280 mm, 140 x 330 mm, 190 x 330 mm, 190 x 360 mm. 255 x 380 mm. 279 x 406 mm. 300 x 400 mm. 300 x 474 mm. 305 x 457 mm: {5}------------------------------------------------ # Device Description There are 18 models of the Self Sealing Sterilization Pouches in this application with different physical specification. These pouches are made from a medical grade paper and plastic (CPP/PET) film that are heat sealed on three sides. The forth side has an adhesive strip that is used to seal the pouch. Release paper used in the pouch is a laminated sheet with composing structure of PE/paper/PE. It is a strip to cover the adhesive area and is released before seal the pouch. The medical grade paper conforms to recognized material standards and can be sterilized by steam or ethylene oxide gas. The indicators printed on the medical grade paper will exhibit a color change (EtO- Pink to Yellow/Steam- Blue to Dark Green) after the pouch is exposed to steam or ethylene oxide gas. The validated maintenance of sterility period is 6 months. - Identification of Predicate Device(s) న్. Predicate Device 510(k) Number: K143637 Product Name: Winner Self Seal Sterilization Pouch Model Name: U&U Medical Technology Co., Ltd. The comparison table is shown in subsection 8. - Non-Clinical Test Conclusion 6. The proposed device Non clinical tests were conducted to subject device. The test results demonstrated that the subject device met the acceptance criteria, and the conducted tests listed as below: - ISO 14937:2009 Sterilization of health care products General requirements or A characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical device - > ASTM D638-14 Standard Test Method For Tensile Properties of Plastics; - > ASTM F2251-03 Standard Test Method for thickness measurement of flexible packaging material; - > ASTM D1922-03 Standard Test Method for Propagation tear resistance of plastic film and thin sheeting by pendulum method; - A ISO 5636-3:2013 Paper and board - Determination of air permeance (medium range) - Part 3: Bendtsen method; - A ASTM F1140/f1140M-13 Standard Test Methods for internal pressurization failure resistance of unrestrained package; - ASTM F1608-00 Standard test methods for Microbial Ranking of Porous packaging materials; > {6}------------------------------------------------ - > ISO 10993-7:2009 Biological Evaluation of Medical Device - Part 7: Ethylene Oxide Sterilization residuals: - > ISO 11140-1:2009 Sterilization of Health Care Products - Chemical Indicators - Part 1: General Requirements; - > ASTM F1980-07 Standard guide for accelerated aging of sterile barrier systems for medical device: - > ASTM F1929-12 Standard test methods for detecting seal leaks in porous medical packaging by dye penetration. - > ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity; - > ISO 10993-10:2010 Biological Evaluation of Medical Devices – Part 10: Tests for irritation and skin sensitization; - > Shelf Life Validation Test, validate the shelf life performance to the proposed device as real time aging method. - > Sterilization Process Validation Test of Self Sealing Sterilization Pouch for EO and Steam sterilization process - > Verification Test of Self Sealing Sterilization Pouch for EO and Steam Sterilization Process - Clinical Test Conclusion 7. No clinical study is included in this submission. {7}------------------------------------------------ #### Substantially Equivalent (SE) Comparison 8. | ITEM | Proposed Device | Predicate Device | Remark | |--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Indication For Use | The Self Sealing Sterilization Pouches are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.<br>The Self Sealing Sterilization Pouches are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).<br>The intended sterilization cycles are listed below:<br>Prevacuum steam; 4 minutes at 132 °C; 10 minute dry time.<br>Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C.<br>The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.<br>The Self Sealing sterilization Pouches are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for each pouch is 540g or 1.19 lbs. The sterilization pouch maintains the enclosed devices up until 6 months post sterilization.<br>The pouches are available in the following sizes:<br>Model(s): 57 x 102mm, 57 x 130 mm, 70 | The U&U sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam and Ethylene Oxide (EtO).<br>The recommended gravity steam sterilization cycle parameters are 30 minutes at 121 °C.<br>The recommended EtO sterilization cycle is 4 hours at 55 °C with a relative humidity between 50%- 85% and a sterilant concentration of 600 mg/L. Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 90Days post sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process. | Similar | # Table 1 General Comparison {8}------------------------------------------------ | | x 255 mm, 83 x 165 mm, 90 x 260 mm,<br>133 x 191 mm, 133 x 279 mm, 133 x 290<br>mm, 140 x 280 mm, 140 x 330 mm, 180 x<br>330 mm, 190 x 330 mm, 190 x 360 mm,<br>255 x 380 mm, 279 x 406 mm, 300 x 400<br>mm, 300 x 474 mm, 305 x 457 mm; | | | |-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Material<br>Compostion | Top Web - Medical Porous Paper<br>Bottom Web - Medical Plastic film(CPP)<br>Bottom Web - Medical two-sided adhesive<br>tape<br>EtO gas indicator ink-Process Indicators<br>Steam indicator ink-Process Indicators | Top Web - Medical Porous Paper<br>Bottom Web - Medical Plastic film(CPP)<br>Bottom Web - Medical two-sided adhesive<br>tape<br>EtO gas indicator ink-Process Indicators<br>class 1<br>Steam indicator ink-Process Indicators<br>class 1 | SE | | Sterilization<br>Cycles | The recommended gravity steam<br>sterilization cycle parameters is<br>Prevacuum steam 4 minutes at 132 °C; 10<br>minute dry time.<br>The recommended EtO sterilization cycle<br>is 1 hours at 55 °C; relative humidity<br>between 40%-80%; 100% ethylene oxide<br>at a concentration of 740 mg/L, 7 day<br>aeration time at 20°C. | The recommended gravity steam<br>sterilization cycle parameters are 30<br>minutes at 121 °C. The<br>recommended EtO sterilization cycle is 4<br>hours at 55 °C with a relative humidity<br>between 50%-85% and a sterilant<br>concentration of 600 mg/L. | Similar | | Configuration/<br>Dimension | Min. 57 x 102 mm<br>Max. 305 x 457 mm | Width: ±0.1"" Length ± 0.2"" | Similar | | Air Permeance | The maximum equivalent pore size<br>diameter shall not exceed 50um. | The maximum equivalent pore size<br>diameter shall not exceed 50um. | Similar | | Microbial Barrier<br>Properties<br>(Packaging<br>Integrity) | Use ASTM 1608 method, and met the<br>acceptance criteria | Use ASTM 1608 method, and met the<br>acceptance criteria | SE | | Material<br>Compatibility | After sterilization, the materials were not<br>degraded | After sterilization, the materials were not<br>degraded | SE | | Biocompatibility | Meet ISO 10993-1 | Meet ISO10993-1 | SE | | Maintenance of<br>Sterility | 6 months | 90 Days | Similar | | Shelf Life | 2 years | 18 months | Similar | | Drying Time | 10 minutes | 25 minutes | Similar | | Aeration Time | 7 days at 20°C | 8 hours at 60°C | Similar | | Chemical<br>Indicator Efficacy | Changed color EtO- Pink to Yellow;<br>Steam- Blue to Dark Green | Changed color EtO- YELLOW to<br>COCOA;<br>Steam- GREEN to PURPLE | Similar | {9}------------------------------------------------ Analysis for difference The noted differences above do not change the intended use of the device or raise new safety and effectiveness concerns. - 9. Substantially Equivalent (SE) Conclusion Based on the comparison and analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...