K212338 · Sterileright Packaging Mfg., Inc. · FRG · Nov 1, 2021 · General Hospital
Device Facts
Record ID
K212338
Device Name
SterileRight sterilization pouch and roll
Applicant
Sterileright Packaging Mfg., Inc.
Product Code
FRG · General Hospital
Decision Date
Nov 1, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The SterileRight Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows: - · Gravity steam at 121°C (250°F) for 30 minutes; Drying time of 25 minutes. - · Pre-vacuum steam at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. - · Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F). The SterileRight provides the sterilization pouch and roll made with Paper/Film or Tyvek®/Film. The SterileRight sterilization pouch and roll which are made with Paper maintains the sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. The SterileRight sterilization pouch and roll which are made with Tyvek® is for EO gas sterilization only. It also maintains the sterility of the enclosed devices for up to 6 months post EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture.
Device Story
Sterilization packaging system consisting of pouches and rolls; materials include medical-grade paper/film or Tyvek®/film. Used by healthcare workers to enclose medical devices prior to steam or EO sterilization. Features external chemical ink indicators (Steam: Pink to Brown/Black; EO: Blue to Yellow/Brown) to verify exposure to sterilization process. Self-seal variants include adhesive strips; others heat-sealed. Maintains sterility of enclosed devices for up to 6 months post-sterilization. Provides physical barrier to maintain sterility; protects contents during storage and handling. Benefits include effective containment and visual confirmation of sterilization exposure.
Clinical Evidence
Bench testing only. Validation studies included sterilant penetration (SAL 10^-6), drying time, aeration, biocompatibility (ISO 10993-10), and package integrity (ASTM F2251, F1140, F1929, F88, F1886, F2096, DIN 58953-6). Chemical indicator efficacy tested per ISO 11140-1. Accelerated aging confirmed 3-year shelf life and 6-month post-sterilization sterility maintenance.
Technological Characteristics
Materials: Medical grade paper, CPP, PET, Tyvek®, PE, PU adhesive. Sensing: Chemical indicator ink (Steam: Pink to Brown/Black; EO: Blue to Yellow/Brown). Form factor: Flat rolls, gusseted rolls, self-seal pouches, heat-seal pouches. Connectivity: None. Sterilization: Steam (gravity/pre-vacuum) or EO gas. Standards: ASTM F2251, F1140, F1929, F88, F1886, F2096, ISO 11607-1, ISO 10993-10, ISO 11140-1.
Indications for Use
Indicated for healthcare workers to enclose medical devices for sterilization via steam or ethylene oxide (EO) gas. Paper/film pouches are compatible with steam and EO; Tyvek®/film pouches are for EO only. Maximum validated load is 3.3 lbs (1.49 kg).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
November 1, 2021
SterileRight Packaging Mfg., Inc. % Uta Shih Regulatory Affairs Manager Sen Mu Technology Co., LTD. 15F-2, No.1, Lane 26, Mincyuan 1st Rd., Lingya District Kaohsiung City, 802 Taiwan
Re: K212338
Trade/Device Name: SterileRight sterilization pouch and roll Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: July 28, 2021 Received: August 3, 2021
Dear Uta Shih:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K212338
#### Device Name
SterileRight Sterilization Pouch and Roll
#### Indications for Use (Describe)
The SterileRight Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
- · Gravity steam at 121°C (250°F) for 30 minutes; Drying time of 25 minutes.
- · Pre-vacuum steam at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
- · Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F).
The SterileRight provides the sterilization pouch and roll made with Paper/Film or Tyvek®/Film.
The SterileRight sterilization pouch and roll which are made with Paper maintains the sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process.
The SterileRight sterilization pouch and roll which are made with Tyvek® is for EO gas sterilization only. It also maintains the sterility of the enclosed devices for up to 6 months post EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Type | Size | Model |
|-------------------------------------------|-------------------------------------|-------|
| Paper/Film Sterilization Flat Roll | W: 50mm ~ 400mm<br>L: ~ 200M | P01 |
| Paper/Film Sterilization Gusseted Roll | W: 75mm ~ 400mm<br>L: ~ 100M | P02 |
| Paper/Film Self-seal Sterilization Pouch | W: 57mm ~ 300mm<br>L: 133mm ~ 474mm | P03 |
| Paper/Film Sterilization Pouch | W: 50mm ~ 400mm<br>L: 100mm ~ 600mm | P04 |
| Tyvek®/Film Sterilization Flat Roll | W: 50mm ~ 400mm<br>L: ~ 200M | T01 |
| Tyvek®/Film Sterilization Gusseted Roll | W: 75mm ~ 400mm<br>L: ~ 100M | T02 |
| Tyvek®/Film Self-seal Sterilization Pouch | W: 57mm ~ 300mm<br>L: 133mm ~ 474mm | T03 |
| Tyvek®/Film Sterilization Pouch | W: 50mm ~ 400mm<br>L: 100mm ~ 600mm | T04 |
The maximum validated pouch load is 3.3 pounds (1.49 kg).
{4}------------------------------------------------
# 510 (k) Summary for K212338
# Submitter's Information
| Name: | Sterileright Packaging Mfg., Inc. |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Address: | 1F, No.33-7, Dahua 2nd Rd., Qidu Dist., Keelung<br>City 206,Taiwan |
| Establishment<br>Registration Number: | 1st submission |
| Contact: | Laney Huang<br>Sales Director<br>Factory: +886-2-2455-5498<br>Office: +886-2-2697-5499<br>email: laney@sterileright.com.tw |
| | Or<br>Thomas Chou<br>RD Director<br>Factory: +886-2-2455-5498<br>Office: +886-2-2697-5499<br>email: thomas@sterileright.com.tw |
| Date Prepared: | July 22, 2021 |
| Device Name | |
| Trade Name:<br>Common/usual Name: | SterileRight Sterilization Pouch and Roll<br>SterileRight Sterilization Pouch and Roll |
| Device Classification Names: | 1) Wrap, Sterilization.<br>2) Indicator, Physical/Chemical Sterilization Pro-<br>cess |
| Panel: | General Hospital |
| Classification Product Code:<br>Subsequent Product Code: | 1) FRG<br>2) JOJ |
Device Classification: Class II Regulation Number: 1) 21 CFR 880.6850 2) 21 CFR 880.2800
## Predicate Devices
The predicate device [510(k) Notification K102158, cleared August 02, 2011] is the SIGMA Sterilization Pouch and Roll.
{5}------------------------------------------------
# Indications for Use
The SterileRight Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
- · Gravity steam at 121°C (250°F) for 30 minutes; Drying time of 25 minutes.
- Pre-vacuum steam at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
- · Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F).
The SterileRight provide sterilization pouch and roll made with Paper/Film or Tyvek® /Film.
The SterileRight sterilization pouch and roll which are made with Paper maintains the sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture. The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process.
The SterileRight sterilization pouch and roll which are made with Tvyek® is for EO gas sterilization only. It also maintains the sterility of the enclosed devices for up to 6 months post EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture.
The SterileRight sterilization pouch and roll is offered 8 types in the following:
- · Paper/Film Sterilization Flat Roll
- · Paper/Film Sterilization Gusseted Roll
- · Paper/Film Self-seal Sterilization Pouch
- · Paper/Film Sterilization Pouch
- · Tyvek® /Film Sterilization Flat Roll
- · Tyvek® /Film Sterilization Gusseted Roll
- · Tyvek® /Film Self-seal Sterilization Pouch
- · Tyvek® /Film Sterilization Pouch
The following table (Table 1) lists the model numbers of the SterileRight sterilization pouch and roll by type, model, dimensions, and content/max. load (lbs.):
{6}------------------------------------------------
| Type | Description | Model | Dimension | Content / Max Load (lbs.) | | |
|------|-----------------------------------------------------------------------------------------|---------------|---------------|---------------------------|---------|-------------------|
| | | | | Metal | Plastic | Gauze / Linens |
| P01 | Paper/Film Sterilization Flat Roll<br>(with EO, Steam Indicators) | P01-050030 | 50mm x 30m | 0.04 | 0.03 | 0.01 |
| | | P01-075030 | 75mm x 30m | 0.06 | 0.05 | 0.02 |
| | | P01-100030 | 100mm x 30m | 0.22 | 0.19 | 0.09 |
| | | P01-150030 | 150mm x 30m | 0.40 | 0.35 | 0.18 |
| | | P01-200030 | 200mm x 30m | 0.63 | 0.57 | 0.28 |
| | | P01-250030 | 250mm x 30m | 2.22 | 2.08 | 1.04 |
| | | P01-300030 | 300mm x 30m | 2.64 | 2.64 | 1.32 |
| | | P01-350030 | 350mm x 30m | 2.64 | 2.64 | 1.32 |
| | | P01-400030 | 400mm x 30m | 2.64 | 2.64 | 1.32 |
| | | P01-050200 | 50mm x 200m | 0.04 | 0.03 | 0.01 |
| | | P01-075200 | 75mm x 200m | 0.06 | 0.05 | 0.02 |
| | | P01-100200 | 100mm x 200m | 0.22 | 0.19 | 0.09 |
| | | P01-150200 | 150mm x 200m | 0.40 | 0.35 | 0.18 |
| | | P01-200200 | 200mm x 200m | 0.63 | 0.57 | 0.28 |
| | | P01-250200 | 250mm x 200m | 2.22 | 2.08 | 1.04 |
| | | P01-300200 | 300mm x 200m | 2.64 | 2.64 | 1.32 |
| | | P01-350200 | 350mm x 200m | 2.64 | 2.64 | 1.32 |
| | | P01-400200 | 400mm x 200m | 2.64 | 2.64 | 1.32 |
| P02 | Paper/Film Sterilization Gusseted Roll<br>(with EO, Steam Indicators) | P02-075030 | 75mm x 30m | 0.08 | 0.06 | 0.03 |
| | | P02-100030 | 100mm x 30m | 0.13 | 0.09 | 0.05 |
| | | P02-150030 | 150mm x 30m | 0.67 | 0.43 | 0.22 |
| | | P02-200030 | 200mm x 30m | 1.13 | 0.71 | 0.36 |
| | | P02-250030 | 250mm x 30m | 1.78 | 1.09 | 0.55 |
| | | P02-300030 | 300mm x 30m | 3.30 | 3.30 | 1.65 |
| | | P02-350030 | 350mm x 30m | 3.30 | 3.30 | 1.65 |
| | | P02-400030 | 400mm x 30m | 3.30 | 3.30 | 1.65 |
| | | P02-075100 | 75mm x 100m | 0.08 | 0.06 | 0.03 |
| | | P02-100100 | 100mm x 100m | 0.13 | 0.09 | 0.05 |
| | | P02-150100 | 150mm x 100m | 0.67 | 0.43 | 0.22 |
| | | P02-200100 | 200mm x 100m | 1.13 | 0.71 | 0.36 |
| | | P02-250100 | 250mm x 100m | 1.78 | 1.09 | 0.55 |
| | | P02-300100 | 300mm x 100m | 3.30 | 3.30 | 1.65 |
| | | P02-350100 | 350mm x 100m | 3.30 | 3.30 | 1.65 |
| | | P02-400100 | 400mm x 100m | 3.30 | 3.30 | 1.65 |
| Type | Description | Model | Dimension | Content / Max Load (lbs.) | | |
| | | | | Metal | Plastic | Gauze /<br>Linens |
| P03 | Paper/Film<br>Self-seal Steri-<br>lization Pouch<br>(with EO,<br>Steam Indica-<br>tors) | P03-057133 | 57mm x 133mm | 0.04 | 0.03 | 0.01 |
| | | P03-070257 | 70mm x 257mm | 0.11 | 0.09 | 0.04 |
| | | P03-090162 | 90mm x 162mm | 0.08 | 0.07 | 0.03 |
| | | P03-090257 | 90mm x 257mm | 0.15 | 0.13 | 0.06 |
| | | P03-090594 | 90mm x 594mm | 0.53 | 0.47 | 0.23 |
| | | P03-133391 | 133mmX391mm | 0.50 | 0.45 | 0.22 |
| | | P03-135283 | 135mm x 283mm | 0.31 | 0.27 | 0.14 |
| | | P03-180335 | 180mm x 335mm | 0.64 | 0.57 | 0.29 |
| | | P03-190358 | 190mm x 358mm | 0.78 | 0.70 | 0.35 |
| | | P03-200435 | 200mm x 435mm | 1.18 | 1.08 | 0.54 |
| | | P03-300380 | 300mm x 380mm | 1.88 | 1.76 | 0.88 |
| | | P03-300474 | 300mm x 474mm | 2.64 | 2.63 | 1.32 |
| P04 | Paper/Film<br>Sterilization<br>Pouch (with<br>EO, Steam In-<br>dicators) | P04-050100 | 50mm x 100mm | 0.02 | 0.02 | 0.01 |
| | | P04-075150 | 75mm x 150mm | 0.06 | 0.05 | 0.02 |
| | | P04-075200 | 75mm x 200mm | 0.08 | 0.07 | 0.03 |
| | | P04-100200 | 100mm x 200mm | 0.12 | 0.10 | 0.05 |
| | | P04-100550 | 100mm x 550mm | 0.55 | 0.49 | 0.25 |
| | | P04-120250 | 120mm x 250mm | 0.22 | 0.19 | 0.09 |
| | | P04-120280 | 120mm x 280mm | 0.26 | 0.22 | 0.11 |
| | | P04-150200 | 150mm x 200mm | 0.22 | 0.19 | 0.09 |
| | | P04-150250 | 150mm x 250mm | 0.30 | 0.26 | 0.13 |
| | | P04-150380 | 150mm x 380mm | 0.58 | 0.52 | 0.26 |
| | | P04-150550 | 150mm x 550mm | 1.08 | 0.99 | 0.49 |
| | | P04-180300 | 180mm x 300mm | 0.54 | 0.48 | 0.24 |
| | | P04-200330 | 200mm x 330mm | 0.74 | 0.67 | 0.33 |
| | | P04-200600 | 200mm x 600mm | 2.06 | 1.93 | 0.96 |
| | | P04-205400 | 205mm x 400mm | 1.07 | 0.98 | 0.49 |
| | | P04-250300 | 250mm x 300mm | 0.92 | 0.83 | 0.42 |
| | | P04-300380 | 300mm x 380mm | 1.88 | 1.76 | 0.88 |
| | P04-300460 | 300mm x 460mm | 2.64 | 2.49 | 1.25 | |
| | P04-400600 | 400mm x 600mm | 2.64 | 2.64 | 1.32 | |
| | | | Dimension | Content / Max Load (lbs.) | | |
| Type | Description | Model | | Metal | Plastic | Gauze /<br>Linens |
| T01 | Tyvek®/Film<br>Sterilization Flat<br>Roll | T01-050060 | 50mm x 60m | 0.04 | 0.03 | 0.01 |
| | | T01-075060 | 75mm x 60m | 0.06 | 0.05 | 0.02 |
| | | T01-100060 | 100mm x 60m | 0.22 | 0.19 | 0.09 |
| | | T01-150060 | 150mm x 60m | 0.40 | 0.35 | 0.18 |
| | | T01-200060 | 200mm x 60m | 0.63 | 0.57 | 0.28 |
| | | T01-250060 | 250mm x 60m | 2.22 | 2.08 | 1.04 |
| | | T01-300060 | 300mm x 60m | 2.64 | 2.64 | 1.32 |
| | | T01-350060 | 350mm x 60m | 2.64 | 2.64 | 1.32 |
| | | T01-400060 | 400mm x 60m | 2.64 | 2.64 | 1.32 |
| | | T01-050200 | 50mm x 200m | 0.04 | 0.03 | 0.01 |
| | | T01-075200 | 75mm x 200m | 0.06 | 0.05 | 0.02 |
| | | T01-100200 | 100mm x 200m | 0.22 | 0.19 | 0.09 |
| | | T01-150200 | 150mm x 200m | 0.40 | 0.35 | 0.18 |
| | | T01-200200 | 200mm x 200m | 0.63 | 0.57 | 0.28 |
| | | T01-250200 | 250mm x 200m | 2.22 | 2.08 | 1.04 |
| | | T01-300200 | 300mm x 200m | 2.64 | 2.64 | 1.32 |
| | | T01-350200 | 350mm x 200m | 2.64 | 2.64 | 1.32 |
| | | T01-400200 | 400mm x 200m | 2.64 | 2.64 | 1.32 |
| T02 | Tyvek®/Film<br>Sterilization<br>Gusseted Roll | T02-075030 | 75mm x 30m | 0.08 | 0.06 | 0.03 |
| | | T02-100030 | 100mm x 30m | 0.13 | 0.09 | 0.05 |
| | | T02-150030 | 150mm x 30m | 0.67 | 0.43 | 0.22 |
| | | T02-200030 | 200mm x 30m | 1.13 | 0.71 | 0.36 |
| | | T02-250030 | 250mm x 30m | 1.78 | 1.09 | 0.55 |
| | | T02-300030 | 300mm x 30m | 3.30 | 3.30 | 1.65 |
| | | T02-350030 | 350mm x 30m | 3.30 | 3.30 | 1.65 |
| | | T02-400030 | 400mm x 30m | 3.30 | 3.30 | 1.65 |
| | | T02-075100 | 75mm x 100m | 0.08 | 0.06 | 0.03 |
| | | T02-100100 | 100mm x 100m | 0.13 | 0.09 | 0.05 |
| | | T02-150100 | 150mm x 100m | 0.67 | 0.43 | 0.22 |
| | | T02-200100 | 200mm x 100m | 1.13 | 0.71 | 0.36 |
| | | T02-250100 | 250mm x 100m | 1.78 | 1.09 | 0.55 |
| | | T02-300100 | 300mm x 100m | 3.30 | 3.30 | 1.65 |
| | | T02-350100 | 350mm x 100m | 3.30 | 3.30 | 1.65 |
| | | T02-400100 | 400mm x 100m | 3.30 | 3.30 | 1.65 |
| | | | | | | |
| Type | Description | Model | Dimension | Content / Max Load (lbs.) | | |
| | | | | Metal | Plastic | Gauze /<br>Linens |
| T03 | Tyvek®/Film<br>Self-seal Sterilization Pouch | T03-057133 | 57mm x 133mm | 0.04 | 0.03 | 0.01 |
| | | T03-070257 | 70mm x 257mm | 0.11 | 0.09 | 0.04 |
| | | T03-090162 | 90mm x 162mm | 0.08 | 0.07 | 0.03 |
| | | T03-090257 | 90mm x 257mm | 0.15 | 0.13 | 0.06 |
| | | T03-135283 | 135mm x 283mm | 0.31 | 0.27 | 0.14 |
| | | T03-180335 | 180mm x 335mm | 0.64 | 0.57 | 0.29 |
| | | T03-190358 | 190mm x 358mm | 0.78 | 0.70 | 0.35 |
| | | T03-254408 | 254mm x 408mm | 1.59 | 1.48 | 0.74 |
| | | T03-300380 | 300mm x 380mm | 1.88 | 1.76 | 0.88 |
| | | T03-300474 | 300mm x 474mm | 2.64 | 2.63 | 1.32 |
| T04 | Tyvek®/Film<br>Sterilization Pouch | T04-050100 | 50mm x 100mm | 0.02 | 0.02 | 0.01 |
| | | T04-100200 | 100mm x 200mm | 0.12 | 0.10 | 0.05 |
| | | T04-150255 | 150mm x 255mm | 0.31 | 0.27 | 0.14 |
| | | T04-150380 | 150mm x 380mm | 0.58 | 0.52 | 0.26 |
| | | T04-190330 | 190mm x 330mm | 0.68 | 0.61 | 0.31 |
| | | T04-200330 | 200mm x 330mm | 0.74 | 0.67 | 0.33 |
| | | T04-250300 | 250mm x 300mm | 0.92 | 0.83 | 0.42 |
| | | T04-300380 | 300mm x 380mm | 1.88 | 1.76 | 0.88 |
| | | T04-300460 | 300mm x 460mm | 2.64 | 2.49 | 1.25 |
| | | T04-400600 | 400mm x 600mm | 2.64 | 2.64 | 1.32 |
Table 1. The model numbers of SterileRight sterilization pouch and roll (Type, Model and Dimension and Content/Max. Load)
{7}------------------------------------------------
Table 1. (Continued) The model numbers of SterileRight sterilization pouch and roll (Type, Model and Dimension and Content/Max. Load)
{8}------------------------------------------------
Table 1. (Continued) The model numbers of SterileRight sterilization pouch and roll (Type, Model and Dimension and Content/Max. Load)
{9}------------------------------------------------
Table 1. (Continued) The model numbers of SterileRight sterilization pouch and roll (Type, Model and Dimension and Content/Max. Load)
# Device Descrintion
The medical devices are inserted into the Sterilization Pouch/Roll, sealed, and then sterilized in the Sterilization System. After completion of the sterilization process, the Pouch/Roll maintains the sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 6 months post Steam or EO gas sterilization.
The Self-seal pouch permits the sealing of the pouch without the need of heat-sealing equipment, while the heat-sealed pouches and rolls are heat-sealed prior to processing in the steam/or EO Sterilization.
The chemical indicators ink printed on the "medical grade paper" will exhibit a color change after the pouch is exposed to steam or ethylene oxide gas. The color of the Chemical Indicator changes from Pink to Brown/Black when exposed to Steam. And the color changes from Blue to Yellow/Brown, when exposed to EO gas.
{10}------------------------------------------------
The Chemical Indicator offers an additional way to verify processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to Steam/or EO gas.
The SterileRight sterilization pouch and roll made with Tyvek® is for EO gas sterilization only. It also maintains the sterility of the enclosed devices for up to 6 months post EO gas sterilization.
## Description of Technological Characteristics Comparison
A summary of the technological characteristics of the device subject of this premarket notification in comparison to those of the predicate devices is included in Table 2.
| Feature | Proposed device | | Predicate device | Comparison | |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>name | SterileRight sterilization pouch and roll | | SIGMA sterilization pouch and roll<br>(K102158) | | |
| Material<br>Composi-<br>tion | Medical Grade Paper, CPP, PET, adhesive, EO and Steam Process Indicator, Print Ink. Tyvek®, PET, PE, adhesive. | | Medical Grade Paper, CPP, PET, PU adhesive, EO and Steam Process Indicator, Print Ink. | Similar | The Pro-<br>posed de-<br>vice pro-<br>vides extra<br>Tyvek®<br>material. |
| Intended<br>use | The SterileRight Sterilization Pouch and Roll are<br>intended to provide health care workers with an ef-<br>fective method to enclose devices intended for ster-<br>ilization in the Steam or via Ethylene Oxide (EO).<br>The recommended sterilization cycles are as fol-<br>lows:<br>•Gravity steam at 121°C (250°F) for 30 minutes;<br>Drying time of 25 minutes.<br>•Pre-vacuum steam at 132°C (270°F) for 4<br>minutes; Drying time of 20 minutes.<br>•Ethylene Oxide (EO) with a concentration of 735<br>mg/L at 55°C (131°F) and 50% to 80% relative<br>humidity for 60 minutes. Aeration time of 8 hours<br>at 60°C (140°F).<br>The SterileRight provides the sterilization pouch<br>and roll made with Paper/Film or Tyvek® /Film.<br>The SterileRight sterilization pouch and roll which<br>are made with Paper maintains the sterility of the<br>enclosed devices for up to 6 months post Steam or<br>EO gas sterilization, and before sterilization has a<br>maximum shelf life of 3 years from the date of<br>manufacture. The pouch's external chemical ink in-<br>dicators are designed to indicate to the user that the<br>pouch has undergone either a steam or EO steriliza-<br>tion process.<br>The SterileRight sterilization pouch and roll which<br>are made with Tyvek® is for EO gas sterilization<br>only. It also maintains the sterility of the enclosed<br>devices for up to 6 months post EO gas sterilization,<br>and before sterilization has a maximum shelf life of<br>3 years from the date of manufacture. | | The sterilization pouch and roll are intended<br>to provide health care workers with an effec-<br>tive method to enclose devices intended for<br>sterilization in steam auto claves and via Eth-<br>ylene Oxide (EO). The recommended steam<br>sterilization cycle parameters are 30 minutes<br>at 121°C. The recommended EO sterilization<br>cycle is 4 hours at 55°C with a relative hu-<br>midity between 50%-85% and a sterilant con-<br>centration of 600 mg/L. Furthermore, the ster-<br>ilization pouch and roll maintains the en-<br>closed devices up until 3 years post steriliza-<br>tion. Lastly, the pouch's external chemical<br>ink indicators are designed to indicate to the<br>user that the pouch has undergone either a<br>steam or EO sterilization process. | Similar | The Steam<br>sterilization<br>of the pro-<br>posed de-<br>vice in-<br>crease by<br>one cycle<br>(Pre-vac-<br>uum 132°C/<br>4 min).<br>The param-<br>eters of EO<br>gas sterili-<br>zation are<br>similar. |
{11}------------------------------------------------
| Feature | Proposed device: SterileRight sterilization pouch and roll | | | Predicate device: SIGMA sterilization pouch and roll (K102158) | | Comparison | Note | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------|--|------------|-------------------------------------------------------------------------------------|--|--|-----|--|-----------|----------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|---------------------------------------------|--|--|--|--|--|--|
| Pouch<br>Types | Paper/Film Sterilization Flat Roll Paper/Film Sterilization Gusseted Roll Paper/Film Self-seal Sterilization Pouch Paper/Film Sterilization Pouch Tyvek® /Film Sterilization Flat Roll Tyvek® /Film Sterilization Gusseted Roll Tyvek® /Film Self-seal Sterilization Pouch Tyvek® /Film Sterilization Pouch | | | Self-sealing sterilization pouches Sterilization pouches, Flat Sterilization pouches, Gusseted Sterilization rolls, Flat Sterilization rolls, Gusseted | | Similar | The Proposed device provides extra types Tyvek® /Film Sterilization pouch and roll. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Device<br>models<br>(Configu-<br>ra-<br>tions/Di-<br>mensions) | Paper/Film Sterilization Flat Roll Size Model W: 50mm ~ 400mm<br>L: ~ 200M P01 Paper/Film Sterilization Gusseted Roll Size Model W: 75mm ~ 400mm<br>L: ~ 100M P02 Paper/Film Self-seal Sterilization Pouch Size Model W: 57mm ~ 300mm<br>L: 133mm ~ 474mm P03 Paper/Film Sterilization Pouch Size Model W: 50mm ~ 400mm<br>L: 100mm ~ 600mm P04 | | | | | | | | | | | | | | | | | | | | | | | | | | | Sterilization rolls, Flat Size Model W: 50 mm ~ 400 mm<br>L: ~ 200 M SMFR Sterilization rolls, Gusseted Size Model W: 75 mm ~ 400 mm<br>L: ~ 100M<br>H: 35 mm ~ 80 mm SMGR Self-Sealing Sterilization Pouches Size Model W: 57 mm ~ 300 mm<br>L: 133 mm ~ 474 mm SMSE Sterilization Pouches, Flat Size Model W: 75 mm ~ 300 mm<br>L: 200 mm ~ 500 mm SMFP Sterilization Pouches Gusseted Size Model W: 100 mm ~ 300 mm<br>L: 30 mm ~ 500 mm<br>H:40 mm ~ 70 mm SMGP | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Same Similar Same Similar Different | | | | | | |
| | N/A | | | | | | The Predicate device provides one extra type. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Tyvek® /Film Sterilization Flat Roll Size Model W: 50mm ~ 400mm<br>L: ~ 200M T01 | | | | | | | | | N/A | | Different | The Proposed device provides one extra type. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Tyvek® /Film Sterilization Gusseted Roll Size Model W: 75mm ~ 400mm<br>L: ~ 100M T02 | | | | | | | | | N/A | | Different | The Proposed device provides one extra type. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Tyvek® /Film Self-seal Sterilization Pouch Size Model W: 57mm ~ 300mm<br>L: 133mm ~ 474mm T03 | | | | | | | | | N/A | | Different | The Proposed device provides one extra type. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Tyvek® /Film Sterilization Pouch Size Model W: 50mm ~ 400mm<br>L: 100mm ~ 600mm T04 | | | | | | | | | N/A | | Different | The Proposed device provides one extra type. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
Table 2. (Continued) Summary of the Proposed and Predicate Devices Technological Characteristics
{12}------------------------------------------------
| Feature | Proposed Device: | Predicate device: | Comparison | Note |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--------------------------------------------------------------------------------------------------------------------------|
| | SterileRight sterilization pouch and roll | SIGMA sterilization pouch and roll<br>(K102158)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.