ACCESS (POWERCLAVE)
Device Facts
| Record ID | K954736 |
|---|---|
| Device Name | ACCESS (POWERCLAVE) |
| Applicant | Mdt Biologic Co. |
| Product Code | FLE · General Hospital |
| Decision Date | Aug 15, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.6880 |
| Device Class | Class 2 |
Indications for Use
Access (Powerclave) Steam Sterilizer Controls are to be used with "Castle" brand sterilizers manufactured by MDT Biologic Company for use in industrial, laboratory, and health care facility environments. These sterilizers are intended to be used to sterilize wrapped and unwrapped surgical instruments, and liquids.
Device Story
Microprocessor-based control system for hospital steam sterilizers; upgrade to existing control systems. Inputs: operator commands via interactive graphic touch screen. Outputs: automated control of sterilization cycles. Used in industrial, laboratory, and healthcare settings; operated by facility staff. Device manages sterilization parameters for wrapped/unwrapped instruments and liquids. Chamber vessel and piping remain unchanged from existing sterilizer hardware. Benefits: modern interface and updated microprocessor control for sterilization process management.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Microprocessor-based control system; interactive graphic touch screen interface. Designed to ANSI/AAMI ST-8-1994, UL-544, CSA C22.2, IEC 601.1. Pressure vessels constructed to ASME Section VIII.
Indications for Use
Indicated for use in industrial, laboratory, and health care facilities to control steam sterilizers for the sterilization of wrapped and unwrapped surgical instruments and liquids.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
Predicate Devices
- Castle Microcomputer Control (K820783)
Related Devices
- K970907 — CASTLE SERIES 100 STEAM STERILIZERS (STRAIGHTLINE) · Getinge/Castle, Inc. · Nov 12, 1997
- K994314 — CASTLE SERIES 100HC STEAM STERILIZERS · Getinge/Castle, Inc. · Feb 16, 2000
- K032192 — PRE-VACUUM AUTOCLAVE WITH STEAM GENERATOR AND VERTICAL SLIDING DOOR, MODELS 4472 EP-1V AND 5596 EP-1V · Tuttnauer U.S.A Co, Ltd. · Aug 12, 2004
- K020747 — AMSCO CENTURY MEDIUM STERILIZER 26 X 26 · STERIS Corporation · May 17, 2002
- K012573 — CASTLE 400HC/500HC SERIES STEAM STERILIZER, MODELS 422HC, 433HC, 522HC, AND 533HC · Getinge/Castle, Inc. · Sep 5, 2001