Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim
K233405 · Eco Medi Glove Sdn. Bhd. · LZA · Dec 28, 2023 · General Hospital
Device Facts
Record ID
K233405
Device Name
Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim
Applicant
Eco Medi Glove Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Dec 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim is a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use. In addition, these gloves were tested for use with Chemotherapy drugs and Fentanyl Test claim in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs and Fentanyl Test claim.
Device Story
Disposable, non-sterile, powder-free nitrile examination glove; blue color; ambidextrous; finger-textured surface; rolled beading cuff. Device worn by examiners to prevent cross-contamination between patient and examiner. Tested for permeation resistance against 15 chemotherapy drugs and fentanyl citrate per ASTM D6978-05. Output is a physical barrier; healthcare providers use the glove during patient examinations or handling of chemotherapy agents. Benefits include protection against biological contaminants and specific chemical permeation. Warning: Not for use with Carmustine or Thiotepa.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing (ISO 10993-23, ISO 10993-10, ISO 10993-11) confirmed the device is a non-irritant, non-sensitizer, and does not induce acute systemic toxicity. Permeation testing per ASTM D6978-05 performed for 16 substances.
Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent contamination between patient and examiner. Suitable for over-the-counter use. Tested for permeation resistance against specific chemotherapy drugs and fentanyl citrate.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
EMG Blue Nitrile Examination Glove Powder Free with Chemotherapy Drugs (K171339)
Related Devices
K213408 — BPG ChemoPlus Nitrile Patient Examination Gloves, Powder Free · Meditech Gloves Sdn Bhd · Apr 17, 2022
K231349 — Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs and Fentanyl) · Nephron Nitrile, LLC · Jun 6, 2023
K210944 — Powder Free Nitrile Examination Glove, Tested for Use with Chemoterapy Drugs (Blue) · Harbour Health, LLC · Apr 29, 2021
K212532 — Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs) · Crdlight Optoelectronic Technology Co., Ltd. · Nov 11, 2021
K214110 — Nitrile Powder Free Blue Patient Examination Gloves Non-Sterile, Tested For Use With Chemotherapy Drugs · Mah Sing Healthcare Sdn. Bhd. · Feb 19, 2022
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 28, 2023
Eco Medi Glove Sdn. Bhd.
Suresh Kumar
Official Correspondent
Lot 23826, Jalan Tembaga Kuning, Kamunting Raya
Industrial Estate
Taiping, Perak Darul Ridzuan 34600
Malaysia
Re: K233405
Trade/Device Name: Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim
Regulation Number: 21 CFR 880.6250
Regulation Name: Non-Powdered Patient Examination Glove
Regulatory Class: Class I, reserved
Product Code: LZA, LZC, QDO, OPJ
Dated: October 2, 2023
Received: October 5, 2023
Dear Suresh Kumar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K233405 - Suresh Kumar
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K233405 - Suresh Kumar
Page 3
Sincerely,
Bifeng Qian -S
Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4B: Division of Infection Control
and Plastic and Reconstructive Surgery Devices
OHT4: Office of Surgical
and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K233405
Device Name
Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim
Indications for Use (Describe)
Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim is a disposable device intended for medical purposes that are worn on the examiner's hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use.
In addition, these gloves were tested for use with Chemotherapy drugs and Fentanyl Test claim in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs and Fentanyl Test claim
Chemotherapy Drug and Fentanyl Test Claim Concentration Minimum Breakthrough detection time in Minutes, ug/cm2/minute
1) Carmustine (3.3mg/ml or 3000ppm), 24.0 minutes
2) Cyclophosphamide (20mg/ml or 20,000ppm), >240 minutes
3) Cisplatin (1.0mg/ml or 1,000ppm), >240 minutes
4) Doxorubicin Hydrochloride (2.0mg/ml or 2000ppm), >240 minutes
5) Etoposide (20mg/ml or 20,000ppm), >240 minutes
6) Flourouracil (50mg/ml or 50,000), >240 minutes
7) Methorexate (25mg/ml or 25,000ppm), >240 minutes
8) Paclitaxel (6mg/ml or 6,000ppm), >240 minutes
9) Thiotepa (10mg/ml or 10,000ppm), 56.9 minutes
10) Ifosfamide (50mg/ml) >240 minutes
11) Mitoxantrone (2mg/ml), >240 minutes
12) Vincristine Sulfate (1mg/ml), >240 minutes
13) Carboplatin (10.0mg/ml), >240 minutes
14) Dacarbazine (DTIC) (10.0mg/ml) >240 minutes
15) Mitomicin C (0.5mg/ml) >240 minutes
16) Fentanyl Citrate Injection 100mg/2ml >240 minutes
The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time.
Carmustine (BCNU) (3.3mg/ml)- Minimum Breakthrough detection time 24.0 ug/cm2/minute.
Thiotepa (10ug/ml) - Minimum Breakthrough detection time 56.9 ug/cm2/minute.
WARNING: Not for use with Carmustine and ThioTepa
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}
ECO Medi Glove Sdn. Bhd. (815262-D)
Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2 : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
E-mail : sales@ecomediglove.com.my Website: www.ecomediglove.com.my
TEL +60-5-806 2316
FAX +60-5-895 7668
SST NO. : A11-1808-21015730
510(K) Summary K233405
Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim
1.0 Submitter:
Company Name : ECO Medi Glove Sdn. Bhd.
Company Address : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping, Perak Darul Ridzuan, Malaysia.
Contact Person : Mr. Suresh Kumar
Telephone : +605-806 2316
Fax : +605-806 2315
Email : qa1@riverstone.com.my
2.0 Preparation Date : 25th December 2023
3.0 Name of the Device
Trade Name / Proprietary Name: Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim
Device Name: Nitrile Patient Examination gloves.
Device Classification Name: Non-Powdered Patient Examination gloves (21 CFR 880.6250). Device Class: Class I.
Product Code: LZA, LZC, OPJ, and QDO
Page 1 of 9
{6}
ECO Medi Glove Sdn. Bhd. (815262-D)
Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2 : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my
TEL +60-5-806 2316
FAX +60-5-895 7668
SST NO.: A11-1808-21015730
## 4.0 Identification of The Legally Marketed Device:
Predicate Device: K171339, EMG Blue Nitrile Examination Glove Powder Free with Chemotherapy Drugs.
## 5.0 Device Description
The subject device in this 510(k) Notification is EMG Blue Powder Free Nitrile Examination Gloves with Chemotherapy Drugs and Fentanyl Test Claim. The subject device is a patient examination glove made from nitrile latex compound, blue color, powder-free, and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-19, the Standard Specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility and permeability to chemotherapy drugs and fentanyl test claims.
The Blue Nitrile Medical Examination Gloves, Powder Free, is a disposable device intended for medical purposes that are worn on the examiner's hand or finger to prevent contamination between patient and examiner (product Code LZA) and are used with chemotherapy drugs (Product code LZC and OPJ) and fentanyl test (Product code QDO).
Page 2 of 9
{7}
ECO Medi Glove Sdn. Bhd. (815262-D)
Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my
TEL +60-5-806 2316
FAX +60-5-895 7668
SST NO.: A11-1808-21015730
## 6.0 Intended Use/Indications for Use
Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim is a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use.
In addition, these gloves were tested for use with Chemotherapy drugs and Fentanyl Test claim in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs and Fentanyl Test claim.
| | Chemotherapy Drug and Fentanyl Test Claim Concentration | Minimum Breakthrough detection time in Minutes, ug/cm²/minute |
| --- | --- | --- |
| 1 | Carboplatin (10mg/ml or 10,000ppm) | >240 minutes |
| 2 | Carmustine (3.3mg/ml or 3000ppm), | 24.0 minutes |
| 3 | Cisplatin (1.0mg/ml or 1,000ppm), | >240 minutes |
| 4 | Cyclophosphamide (20mg/ml or 20,000ppm), | >240 minutes |
| 5 | Dacarbazine (DTIC) (10.0mg/ml) | >240 minutes |
| 6 | Doxorubicin Hydrochloride (2.0mg/ml or 2000ppm), | >240 minutes |
| 7 | Etoposide (20mg/ml or 20,000ppm), | >240 minutes |
| 8 | Fluorouracil (50mg/ml or 50,000), | >240 minutes |
| 9 | Ifosfamide (50mg/ml) | >240 minutes |
| 10 | Methorexate (25mg/ml or 25,000ppm), | >240 minutes |
| 11 | Mitomycin (0.5mg/ml or 500ppm) | >240 minutes |
| 12 | Mitoxantrone (2mg/ml or 2,000ppm), | >240 minutes |
| 13 | Paclitaxel (6mg/ml or 6,000ppm), | >240 minutes |
| 14 | Thiotepa (10mg/ml or 10,000ppm), | 56.9 minutes |
| 15 | Vincristine Sulfate (1mg/ml), | >240 minutes |
| 16 | Fentanyl Citrate Injection 100mg/2ml | >240 minutes |
The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time.
Carmustine (BCNU) (3.3mg/ml)- Minimum Breakthrough detection time 24.0 ug/cm²/minute.
Thiotepa (10ug/ml) - Minimum Breakthrough detection time 56.9 ug/cm²/minute.
WARNING: Not for use with Carmustine and ThioTepa
Page 3 of 9
{8}
ECO Medi Glove Sdn. Bhd. (815262-D)
Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2 : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my
TEL +60-5-806 2316
FAX +60-5-895 7668
SST NO.: A11-1808-21015730
## 7.0 Specification for the subject Nitrile gloves:
## 7.1 Dimension and Thickness of Gloves
| Dimension | Size XXS | Size XS | Size S | Size M | Size L | Size XL | Size XXL | Size XXXL |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Overall Length (mm) | 230min | 230min | 230min | 230min | 230min | 230min | 230min | 230min |
| Width (± 10mm) | 65 | 75 | 85 | 95 | 105 | 115 | 125 | 135 |
| Thickness at Palm (mm) | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min |
| Thickness at Finger Tip (mm) | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min | 0.05min |
## 7.2 Gloves Physical Properties and Holes
| Measurement | Before Ageing | After Aging at 70°C for 168 hrs @ 100°C for 22 hrs |
| --- | --- | --- |
| Tensile Strength (MPa) | 14min | 14 Min |
| Ultimate Elongation (%) | 500min | 400min |
| Pin-hole Level | AQL 2.5
Inspection Level G-1 | AQL 2.5
Inspection Level G-1 |
Gloves meet all the specifications listed in ASTM D 6319-19
Page 4 of 9
{9}
ECO Medi Glove Sdn. Bhd. (815262-D)
Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my
TEL +60-5-806 2316
FAX +60-5-895 7668
SST NO.: A11-1808-21015730
# Technological Characteristics Comparison of the Proposed and Predicate Devices
| Characteristics | Acceptance Criteria | Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim K233405 | Predicate Device EMG Blue Nitrile Examination Glove Powder-Free with Chemotherapy Drugs K171339 | Assessment Similarities and Differences |
| --- | --- | --- | --- | --- |
| Product Code | | LZA, LZC, OPJ, and QDO | LZA, LZC | Similar |
| Intended use/Indications for Use | | A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use. | A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner’s hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use. | same |
| Material use | Nitrile compound | Nitrile compound | Nitrile compound | Same |
| Color | | Blue | Blue | Same |
| Sterility | | Non sterile | Non sterile | Same |
| Surface | Finger Textured | Finger Textured | Finger Textured | Same |
| Cuffing Beading | Rolled Beading | Rolled Beading | Rolled Beading | Same |
| Design | Ambidextrous | Ambidextrous | Ambidextrous | Same |
| Dimensions | Overall Length (mm)
Min 230mm
Width (± 10mm)
Size XXS = 65mm
Size XS = 75mm
Size S = 85mm
Size M = 95mm
Size L = 105mm
Size XL = 115mm
Size XXL=125mm
Size XXXL=135mm
Thickness at
Palm (mm)
Min; 0.05 mm
Thickness at
Finger Tip (mm)
Min 0.05 mm | Overall Length (mm)
Min 230mm
Width (± 10mm)
Size XXS = 65mm
Size XS = 75mm
Size S = 85mm
Size M = 95mm
Size L = 105mm
Size XL = 115mm
Size XXL=125mm
Size XXXL=135mm
Thickness at
Palm (mm)
Min; 0.05 mm
Thickness at
Finger Tip (mm)
Min 0.05 mm | Overall Length (mm)
Min 230 mm
Width (± 5mm)
Size S = 85mm
Size M = 95mm
Size L = 105mm
Size XL = 115mm
Thickness at Palm (mm)
Min; 0.05 mm
Thickness at Finger Tip (mm)
Min 0.05 mm | Similar additional sizes were added to the subject device |
Page 5 of 9
{10}
ECO Medi Glove Sdn. Bhd. (815262-D)
Plant 1 : Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2 : Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my
TEL +60-5-806 2316
FAX +60-5-895 7668
SST NO.: A11-1808-21015730
| Characteristics | Acceptance Criteria | Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim, | Predicate Device EMG Blue Nitrile Examination Glove Powder-Free with Chemotherapy Drugs K171339 | Assessment Similarities and Differences |
| --- | --- | --- | --- | --- |
| Physical properties | **Before Ageing**
Tensile Strength (MPa) = 14min
Ultimate Elongation (%) = 500min
**After Aging at 70°C for 168 hrs @ 100°C for 22 hrs**
Tensile Strength (MPa) = 14min
Ultimate Elongation (%) = 400min | Meets ASTM D6319-19 | Meets ASTM D6319-19 | same |
| Freedom from pinholes | AQL 2.5
Inspection Level G-1 | Meets ASTM D5151-19 | Meets ASTM D5151-19 | same |
| Residual Powder | ≤ 2.0 mg/pc | Meets ASTM D6124-06 | Meets ASTM D6124-06 | same |
| Biocompatibility | ISO 10993-23: 2021 Biological evaluation of medical devices- Part 23: Test for irritation | Under the conditions of this study, the test article was a non-irritant. | Under the conditions of this study, the test article was a non-irritant. | same |
| | ISO 10993-10: 2021 Biological Evaluation on Medical Device – Part 10: Test for Skin Sensitization | Under the conditions of this study, the test article was a non-sensitizer. | Under the conditions of this study, the test article was a non-sensitizer. | same |
| | ISO 10993-11 Biological evaluation on medical device Part 11 – Test for systemic toxicity | Under the conditions of this study, not induce any acute systemic toxicity | Under the conditions of this study, not induce any acute systemic toxicity | Same |
Page 6 of 9
{11}
ECO Medi Glove Sdn. Bhd. (815262-D)
Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my
TEL +60-5-806 2316
FAX +60-5-895 7668
SST NO.: A11-1808-21015730
| Characteristics | Acceptance Criteria | Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim, | Predicate Device EMG Blue Nitrile Examination Glove Powder-Free with Chemotherapy Drugs K171339 | Assessment Similarities and Differences |
| --- | --- | --- | --- | --- |
| Resistance against Chemotherapy Drugs | Standards Practice for Assessment of resistance of Medical Glove to Permeation by Chemotherapy drugs ASM D6978-05 | 1) Carboplatin (10mg/ml or 10,000ppm) >240 minutes
2) Carmustine (3.3mg/ml or 3000ppm), 24.0 minutes
3) Cisplatin (1.0mg/ml or 1,000ppm), >240 minutes
4) Cyclophosphamide (20mg/ml or 20,000ppm), >240 minutes
5) Dacarbazine (DTIC) (10.0mg/ml) >240 minutes
6) Doxorubicin Hydrochloride (2.0mg/ml or 2000ppm), >240 minutes
7) Etoposide (20mg/ml or 20,000ppm), >240 minutes
8) Fluorouracil (50mg/ml or 50,000), >240 minutes
9) Ifosfamide (50mg/ml) >240 minutes
10) Methorexate (25mg/ml or 25,000ppm), >240 minutes
11) Mitomicin (0.5mg/ml or 500ppm) >240 minutes
12) Mitoxantrone (2mg/ml or 2,000ppm), >240 minutes
13) Paclitaxel (6mg/ml or 6,000ppm), >240 minutes
14) Thiotepa (10mg/ml or 10,000ppm), 56.9 minutes
15) Vincristine Sulfate (1mg/ml), >240 minutes
16) Fentanyl Citrate Injection (100mg/2ml) >240 minutes
Warning: Not for use with Carmustine or Thio Tepa | 1) Carmustine (BCNU) 3.3mg/ml 24.0 minutes
2) Cisplatin 1.0mg/mL, >240 minutes
3) Cyclophosphamide (Cytoxan) (20mg/ml) >240 minutes
4) Dacarbazine (10mg/ml) >240 minutes
5) Doxorubicin HCI (2.0mg/ml) >240 minutes
6) Etoposide (20.0mg/ml >240 minutes
7) Fluorouracil (50.0mg/ml) >240 minutes
8) Methotrexate (25mg/ml) >240 minutes
9) Paclitaxel (6.0mg/ml) >240 minutes
10) ThioTepa (10.0mg/ml) 56.9 minutes
11) Vincristine Sulfate 1.0mg/ml, >240 minutes
12) Ifosfamide (50mg/ml)
13) Mitoxantrone (2mg/ml or 2,000ppm),
14) Carboplatin (10mg/ml or 10,000ppm)
15) Vincristine Sulfate (1mg/ml),
Warning: Not for use with Carmustine or Thio Tepa | Similar. Additional Fentanyl test is conducted for Device sample |
Page 7 of 9
{12}
ECO Medi Glove Sdn. Bhd. (815262-D)
Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my
TEL +60-5-806 2316
FAX +60-5-895 7668
SST NO.: A11-1808-21015730
# 8.0 Summary of Non-Clinical Performance test data
| Characteristics | Test Standard | Acceptance Criteria | Test Result | Remark |
| --- | --- | --- | --- | --- |
| Freedom from Pin holes | ASTM D5151 -19
(Re-approved 2011) | AQL 2.5
Inspection Level G-1 | Meets ASTM D5151-19 | Per Lot Size 500,000pcs |
| Dimensions | ASTM D6319 -19 | Length
Min 230mm | Meets ASTM D6319-19 | Lot Size 500,000pcs |
| | ASTM D6319 -19 | Width
Size XXS = 65mm±10
Size XS = 75mm±10
Size S = 85mm±10
Size M = 95mm±10
Size L = 105mm±10
Size XL = 115mm±10
SizeXXL =125mm±10
Size XXXL = 135mm±10 | ISO 2859-1 / S2/AQL 4.0 | |
| | ASTM D6319 -19 | Thickness at Palm (mm)
Min; 0.05 mm
Thickness at Finger Tip (mm)
Min 0.05 mm | ISO 2859-1 / S2/AQL 4.0 | |
| Physical properties | ASTM D6319 -19
ASTM D412-16(2021) | Before Ageing
Tensile Strength (MPa) = 14min
Ultimate Elongation (%) = 500min | Meets ASTM D6319-19 | |
| | ASTM D6319 -19 and
ASTM D412-06(2021) | After Aging at 70°C for 168 hrs @ 100°C for 22 hrs
Tensile Strength (MPa) = 14min
Ultimate Elongation (%) = 400min | | |
| Powder-free residue | ASTM D6124-06 | ≤ 2.0 mg/pc | Meets ASTM D6124-06 (2022) | Once every PO |
| Biocompatibility | ISO 10993-23: 2021
Biological evaluation of medical devices- Part 23: Test for irritation | The Device shall be non-irritant. | Under the conditions of this study, the test article was a non-irritant | As per Device with same color |
| | ISO 10993-10: 2021
Biological Evaluation on Medical Device – Part 10: Test for Skin Sensitization | The device shall be non-sensitizer. | Under the conditions of this study, the test article was a non-sensitizer. | As per Device with same color |
| | ISO 10993-11
Biological evaluation on medical device Part 11 – Test for systemic toxicity | The device shall not induce any acute systemic toxicity | Under the conditions of this study, not induce any acute systemic toxicity | As per Device with same color |
{13}
ECO Medi Glove Sdn. Bhd. (815262-D)
Plant 1: Lot 32586, No. 118, Jalan Logam 7, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
Plant 2: Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, 34600 Taiping. Perak Darul Ridzuan. MALAYSIA
E-mail: sales@ecomediglove.com.my Website: www.ecomediglove.com.my
TEL +60-5-806 2316
FAX +60-5-895 7668
SST NO.: A11-1808-21015730
Summary of Clinical Testing: Not applicable.
## 9.0 Conclusion
The Conclusion drawn from the non-Clinical test demonstrates that the subject device in 510(K) Submission, Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl test claim is as safe, as effective, and performs as well as or better than the legally marketed Predicate device cleared under K171339.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.