Nitrile Powder Free Blue Patient Examination Gloves, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate
K223415 · Mah Sing Healthcare Sdn. Bhd. · LZA · Jan 26, 2023 · General Hospital
Device Facts
Record ID
K223415
Device Name
Nitrile Powder Free Blue Patient Examination Gloves, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate
Applicant
Mah Sing Healthcare Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Jan 26, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 (2019), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable, non-sterile, ambidextrous nitrile examination glove; blue color; powder-free. Designed to prevent contamination between patient and examiner. Tested for permeation resistance against various chemotherapy drugs and fentanyl citrate per ASTM D6978-05. Used in clinical settings by healthcare professionals. Provides a physical barrier; protects the wearer from exposure to hazardous drugs and patient contaminants. Does not include software or electronic components.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated via ASTM D6124-06 (residual powder), ASTM D5151-19 (freedom from holes), ASTM D6319-19 (dimensions/physical properties), and ASTM D6978-05 (chemotherapy drug permeation). Biocompatibility confirmed via ISO 10993-10 (irritation/sensitization) and ISO 10993-11 (acute systemic toxicity).
Technological Characteristics
Material: Acrylonitrile-butadiene copolymer (nitrile). Form: Ambidextrous, powder-free, blue, five sizes (XS-XL). Standards: ASTM D6319-19, ASTM D6978-05, ASTM D5151-19, ASTM D6124-06. Biocompatibility: ISO 10993-10, ISO 10993-11. Non-sterile. No software or electronic components.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs and fentanyl citrate.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Blue Colored, Powder Free Nitrile Examination Gloves, Non-sterile, and Tested for Use With Chemotherapy Drugs and Fentanyl Citrate (K192954)
Related Devices
K212532 — Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs) · Crdlight Optoelectronic Technology Co., Ltd. · Nov 11, 2021
K214110 — Nitrile Powder Free Blue Patient Examination Gloves Non-Sterile, Tested For Use With Chemotherapy Drugs · Mah Sing Healthcare Sdn. Bhd. · Feb 19, 2022
K213408 — BPG ChemoPlus Nitrile Patient Examination Gloves, Powder Free · Meditech Gloves Sdn Bhd · Apr 17, 2022
K233405 — Blue Powder Free Nitrile Examination Glove with Chemotherapy Drugs and Fentanyl Test Claim · Eco Medi Glove Sdn. Bhd. · Dec 28, 2023
K181857 — Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Low Dermatitis Potential. Tested for Use with Chemotherapy Drugs and Fentanyl Citrate · Kossan International Sdn Bhd · Nov 20, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below.
February 2, 2023 Mah Sing Healthcare Sdn Bhd Ivan Tan Chee Wei Senior QA Manager Wisma Mah Sing, Penthouse Suite 1, 163 Jalan Sungai Besi Kuala Lumpur, Kuala Lumpur 57100 Malaysia
Re: K223415
Trade/Device Name: Nitrile Powder Free Blue Patient Examination Gloves, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ, QDO
Dear Ivan Tan Chee Wei:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated 01/27/23. Specifically, FDA is updating this SE Letter to correct the contact name to include the two middle names as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Bifeng Qian, OHT4: Office of Surgical and Infection Control Devices, at: 301-796-2261 or bifeng.gian@fda.hhs.gov.
Sincerely,
Bifeng Qian -S
Bifeng Qian, MD, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
January 26, 2023
Mah Sing Healthcare Sdn Bhd Ivan Wei Senior QA Manager Wisma Mah Sing, Penthouse Suite 1, 163 Jalan Sungai Besi Kuala Lumpur, Kuala Lumpur 57100 Malaysia
## Re: K223415
Trade/Device Name: Nitrile Powder Free Blue Patient Examination Gloves, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QDO Dated: November 1, 2022 Received: November 10, 2022
Dear Ivan Wei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bifeng Qian -S
Bifeng Qian, M.D. Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223415
#### Device Name
Nitrile Powder Free Blue Patient Examination Gloves, Non-Sterile, Tested For Use With Chemotherapy Drugs and Fentanyl Citrate
#### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
These gloves were tested for use with Chemotherapy Drugs as per ASTM D6978-05 (2019), Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
Test Chemotherapy Drugs are as follows:
| Test Chemotherapy Drugs Concentration | Minimum Breakthrough Detection Time in minutes |
|---------------------------------------|------------------------------------------------|
| *Carmustine 3.3mg/ml | 14.1 |
| Cisplatin1 mg/ml | >240 |
| Cyclophosphamide 20 mg/ml | >240 |
| Dacarbazine 10 mg/ml | >240 |
| Doxorubicin, HCl2 mg/ml | >240 |
| Etoposide20 mg/ml | >240 |
| Fluorouracil 50 mg/ml | >240 |
| Methotrexate 25 mg/ml | >240 |
| Mitomycin 0.5 mg/ml | >240 |
| Oxaliplatin 5 mg/ml | >240 |
| Paclitaxel 6 mg/ml | >240 |
| *Thiotepa 10 mg/ml | 57.2 |
| Vincristine 1 mg/ml | >240 |
| Test Fentanyl Citrate are as follows: | |
Test Chemotherapy Drugs Concentration Fentanyl Citrate Injection 100mcg/2ml
Minimum Breakthrough Detection Time in minutes >240
Warning: Do not use with Carmustine and Thiotepa. Please note that the following drugs have low permeation times: (1)Carmustine - 14.1 minutes (2)Thiotepa - 57.2 minutes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
× Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for MahSing Healthcare. The word "MahSing" is in a large, bold, red font. Below that, the word "HEALTHCARE" is in a smaller, black, sans-serif font. The logo is simple and clean, with a focus on the company name.
## Premarket Notification [510(k)] No: K223415
| 510 (K) SUMMARY | | |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| 1.0 Device Name | Nitrile Powder Free Blue Patient Examination Gloves, Non-Sterile,<br>Tested For Use With Chemotherapy Drugs and Fentanyl Citrate. | |
| 2.0 Submitter name /<br>Contact details | Mah Sing Healthcare Sdn. Bhd<br>Wisma Mah Sing, Penthouse Suite 1,<br>163 Jalan Sungai Besi,<br>Kuala Lumpur<br>57100<br>MALAYSIA | |
| | Contact Person Details:<br>Ivan Tan Chee Wei (Mr)<br>E-mail: ivan.tan@mshealthcare.com<br>Tel: +60-3-3396 2288, Extn: 2213<br>Fax: +60-3-3396 2299 | |
| 3.0 Summary<br>Preparation Date | January 18, 2023 | |
| 4.0 Device Name &<br>Classification | Trade Name: Nitrile Powder Free Blue Patient Examination Gloves<br>Non-sterile, Tested for use with Chemotherapy Drugs<br>and Fentanyl Citrate. | |
| | Common Name: Nitrile Powder Free Patient Examination Glove | |
| | Classification Name: Patient Examination Gloves Specialty<br>Polymer Patient Examination Gloves | |
| | Device Classification: I | |
| | Regulation Number: 21 CFR 880.6250 | |
| | Panel: General Hospital | |
| | Product Code: LZC, LZA, OPJ, QDO | |
| 5.0 Identification of<br>The Legally Marketed<br>Device | Predicate Device Name: Blue Colored, Powder Free Nitrile Examination<br>Gloves, Non-sterile, and Tested for Use<br>With Chemotherapy Drugs and Fentanyl Citrate | |
| | Predicate 510(K) Number: K192954 | |
| | Manufacture's Name: Comfort Rubber Gloves Industries Sdn Bhd. | |
| DrivingInnovation. Protecting Lives. | MAH SING HEALTHCARE SDN BHD<br>Lot 6478, Lorong Sungai Puloh / KU6,<br>Kawasan Industri Sungai Puloh,<br>42100 Klang, Selangor.<br>MALAYSIA | T 603 3396 2288<br>F 603 3396 2299<br>mshealthcare.com |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for "Mah Sing Healthcare". The word "MahSing" is in red, with the word "Healthcare" in black underneath. The font is sans-serif and bolded.
| 6.0 Description of Device | Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested<br>For Use With Chemotherapy Drugs and Fentanyl Citrate meets all the<br>requirements of ASTM standards D6319-19, D6978-05 (2019) and FDA<br>21 CFR 880.6250. | | | |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|--------------------------------------------------------|--|
| | Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested<br>For Use With Chemotherapy Drugs are Class I patient examination gloves<br>bear the product code Nitrile - LZA, LZC, OPJ and QDO (21CFR880.6250).<br>The gloves are made from acrylonitrile-butadiene copolymer dispersion.<br>Inner surface of gloves undergoes surface treatment process to produce a<br>smooth surface that assists the user in donning the gloves without using<br>any lubricant such as powder on the glove surface. These gloves are blue<br>in color and are powder free. The gloves are ambidextrous i.e., can be worn<br>on right hand or left hand, single use disposable devices that come in five<br>sizes (XS, S, M, L and XL). The physical properties of glove, i.e., tensile<br>strength meet ASTM D 6319-19. | | | |
| | | | | |
| 7.0 Indications for Use | A patient examination glove is a disposable device intended for medical<br>purpose that is worn on the examiner's hand or finger to prevent<br>contamination between patient and examiner.<br>These gloves were tested for use with Chemotherapy Drugs as per<br>ASTM D6978-05 (2019), Standard Practice for Assessment of Medical<br>Gloves to Permeation by Chemotherapy Drugs: | | | |
| | Test Chemotherapy Drugs are as follows: | | | |
| | Test Chemotherapy | Concentration | Minimum Breakthrough | |
| | Drugs | | Detection Time in minutes | |
| | *Carmustine | 3.3mg/ml | 14.1 | |
| | Cisplatin | 1 mg/ml | >240 | |
| | Cyclophosphamide | 20 mg/ml | >240 | |
| | Dacarbazine | 10 mg/ml | >240 | |
| | Doxorubicin, HCl | 2 mg/ml | >240 | |
| | Etoposide | 20 mg/ml | >240 | |
| | Fluorouracil | 50 mg/ml | >240 | |
| | Methotrexate | 25 mg/ml | >240 | |
| | Mitomycin | 0.5 mg/ml | >240 | |
| | Oxaliplatin | 5 mg/ml | >240 | |
| | Paclitaxel | 6 mg/ml | >240 | |
| | *Thiotepa | 10 mg/ml | 57.2 | |
| | Vincristine | 1 mg/ml | >240 | |
| | Test Fentanyl Citrate are as follows:<br>Test Chemotherapy Drugs | Concentration | Minimum Breakthrough | |
| | | | Detection Time in minutes | |
| | Fentanyl Citrate Injection | 100mcg/2ml | >240 | |
| | Warning: Do not use with Carmustine and Thiotepa. | | | |
| | *Note:<br>Please note that the following drugs have low permeation times:<br>(1) Carmustine - 14.1 minutes<br>(2) Thiotepa - 57.2 minutes. | | | |
| ovation. Protecting Lives. | MAH SING HEALTHCARE SDN BHD<br>Lot 6478, Lorong Sungai Puloh / KU6,<br>Kawasan Industri Sungai Puloh,<br>42100 Klang, Selangor.<br>MALAYSIA | | T 603 3396 2288<br>F 603 3396 2299<br>mshealthcare.com | |
DrivingInn
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for "Mah Sing Healthcare". The words "MahSing" are in a large, bold, red font. Below that, the word "HEALTHCARE" is in a smaller, black, bold font. The logo is simple and professional.
### 8.0 Summary of the Technological Characteristic of the Device
Nitrile Powder Free Blue Patient Examination Gloves Non-sterile, Tested For Use with Chemotherapy Drugs and Fentanyl Citrate meets all the requirements of ASTM standards D6319-19, D6978-05 (2019) and FDA 21 CFR 880.6250.
| 1<br>able | 1 |
|-----------|---|
|-----------|---|
| Characteristics and<br>Parameters | Standard | Proposed Device | Predicate device | Comparison<br>Analysis |
|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| 510(k) Number | - | K223415 | K192954 | Different |
| Manufacturer | - | Mah Sing Healthcare Sdn.<br>Bhd. | Comfort Rubber Gloves<br>Industries Sdn Bhd | Different |
| Name of device | - | Nitrile Powder Free Blue<br>Patient Examination<br>Gloves, Non-Sterile,<br>Tested for use with<br>chemotherapy drugs and<br>fentanyl citrate | Blue Colored, Powder<br>Free Nitrile Examination<br>Gloves, Non-Sterile, and<br>Tested for use with<br>chemotherapy drugs and<br>fentanyl citrate | Similar |
| Device Classification<br>Name/Regulation Number | Patient Examination<br>Glove,<br>21 CFR Part 880.6250 | Patient Examination<br>Glove,<br>21 CFR Part 880.6250 | Patient Examination<br>Glove,<br>21 CFR Part 880.6250 | Same |
| Product Code | - | LZA, LZC, OPJ, QDO | LZA, LZC, QDO | Same |
| Classification | - | Class 1 | Class 1 | Same |
| Raw Rubber Material | ASTM D 6319-19 | Nitrile | Nitrile | Same |
| Color | - | Blue | Blue | Same |
| Characteristics and<br>Parameters | Standard | Proposed Device | Predicate device | Comparison<br>Analysis |
| Chemotherapy Drug<br>Permeation Test | ASTM D6897-05 | | | |
| Test Chemotherapy<br>Drugs | Concentration | Minimum Breakthrough Detection Time (min) | | |
| *Carmustine | 3.3mg/ml | 14.1 | 18.2 | Similar<br>Below 240 minutes<br>permeation times |
| Cisplatin | 1 mg/ml | >240 | >240 | Similar |
| Cyclophosphamide | 20 mg/ml | >240 | >240 | Similar |
| Dacarbazine | 10 mg/ml | >240 | >240 | Similar |
| Doxorubicin, HCl | 2 mg/ml | >240 | >240 | Similar |
| Etoposide | 20 mg/ml | >240 | >240 | Similar |
| Fluorouracil | 50 mg/ml | >240 | >240 | Similar |
| Methotrexate | 25 mg/ml | >240 | Not tested | Optional, Subject<br>device perform<br>additional<br>Chemotherapy drug<br>test |
| Mitomycin | 0.5 mg/ml | >240 | Not tested | Optional, Subject<br>device perform<br>additional<br>Chemotherapy drug<br>test |
| Oxaliplatin | 5 mg/ml | >240 | Not tested | Optional, Subject<br>device perform<br>additional<br>Chemotherapy drug<br>test |
| Paclitaxel | 6 mg/ml | >240 | >240 | Similar |
| *Thiotepa | 10 mg/ml | 57.2 | 57.3 | Similar<br>Below 240 minutes<br>permeation times |
| Vincristine | 1 mg/ml | >240 | >240 | Similar |
| Fentanyl Citrate Injection | 100 mcg/2ml | >240 | >240 | Similar |
| Characteristics and<br>Parameters | Standard | Proposed Device…
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.