The powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
AMADEX Nitrile Powder Free Examination Gloves are single-use, non-sterile, disposable gloves; available in blue, black, and orange. Device functions as a physical barrier to prevent cross-contamination between patient and examiner. Used in clinical settings by healthcare professionals. No electronic components, software, or algorithms involved.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ASTM D6319-19, ASTM D5151-19, ASTM D6124-06, and ISO 10993-10 (irritation and sensitization) confirmed performance metrics meet or exceed standards.
Technological Characteristics
Material: Nitrile. Form factor: Disposable, powder-free, finger-textured examination glove. Standards: ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ISO 10993-10. Non-sterile. No energy source or connectivity.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
K070096 — POWDER FREE NITRILE EXAMINATION GLOVES, BLUE, WHITE, PINK, BLACK AND GRAY COLOR · Gx Corporation Sdn Bhd · Mar 14, 2007
K222813 — Nitrile Examination Gloves · Ecoglove Medical Manufacturing Company Limited · Mar 16, 2023
K090464 — BLACK NITRILE EXAMINATION GLOVES, POWDER FREE, NON STERILE · Pt Mahakarya Inti Buana · Apr 23, 2009
K220232 — Nitrile Powder Free Examination Gloves, Non-Sterile - Blue, Black, Orange · Megine Industries Sdn Bhd · Feb 20, 2022
K102230 — NUGARD NITILE GREY COLOR · Terang Nusa Sdn Bhd · Sep 10, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
June 16, 2021
Megine Industries Sdn Bhd % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K211310
Trade/Device Name: AMADEX - Nitrile Powder Free Examination Gloves Non-Sterile (Blue, Black, Orange) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA Dated: June 11, 2021 Received: June 14, 2021
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211310
### Device Name
AMADEX - Nitrile Powder Free Examination Gloves Non-Sterile, (Blue, Black, Orange)
Indications for Use (Describe)
The powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# MEGINE INDUSTRIES SDN BHD
# 510 (K) SUMMARY ----------
This summary of 510(K) is being submitted in Accordance with requirements of SMDA 1990 and 21 CFR807.92.
Collicit conceller and en artifies
# The assigned 510 (K) number is: K211310
# I. APPLICANT INFORMATION:
| 510 (K) Owner's Name | MEGINE INDUSTRIES SDN BHD |
|----------------------|----------------------------------------------------------------------------------------------------------|
| Address | B-05-02, The Vertical Business Suites, 8, Jalan Kerinchi,<br>Bangsar South, 59200 Kuala Lumpur, Malaysia |
| Phone | +60328566180 |
| Fax | +60328566180 |
| E-mail | benjamin.ng@meginegroup.com.my |
| Contact Person | Benjamin Ng Soong Jing |
| Designation | Managing Director |
| Contact Number | +60328566180 |
| Contact Email | benjamin.ng@rneginegroup.com.my |
| Date Submission | 28th April 2021 |
AT REAL ME
# 2. DEVICE IDENTIFICATION:
글
| Name of the Device | AMADEX-Nit1ile Powder Free Examination Gloves Non-<br>Sterile, (Blue, Black, Orange) |
|--------------------------------------|--------------------------------------------------------------------------------------|
| Product Proprietary or<br>Trade Name | AMADEX |
| Common or Usual Name | Examination Gloves |
| Classification Name | Patient Examination Glove |
| Device Classification | Class - I |
| Product Code | LZA |
| Regulation Number | 21 CFR 880.6250 |
| Review Panel | General Hospital |
{4}------------------------------------------------
#### 3. PREDICATE DEVICE INFORMATION:
| Predicate Device | Nitrite Examination Powder Free Glove, Black, Nitrile<br>Examination Powder Free Glove, Orange. |
|------------------|-------------------------------------------------------------------------------------------------|
| 510 (K) Number | K172867 |
| Regulatory Class | I |
| Product Code | LZA |
#### 4. DEVICE DESCRIPTION:
AMADEX - Nitrile Powder Free Examination Gloves Non-Sterile, (Blue, Black, Orange) are Class I patient examination gloves bearing the product code Nitril e - LZA (2 ICFR880.6250). AMADEX - Nitrile Powder Free Examnation Gloves Non-Sterile,(Blue, Black, Orange) meet all the requirements of ASTM standard D 6319-10 and FDA 21 CFR 880.6250. The AMADEX - Nitrile Powder Free Examination Gloves Non-Sterile, (Blue, Black, Orange) is sing le-use device to prevent contamination between patient and examiner.
# 5. INDICATION OF USE:
The powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner' s hand or finger to prevent contamination between patient and examiner.
| Characteristics | Standards | Device Performance | | Remark s |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| | | Predicate | Subject | - |
| 5 10 (K) number | | K172867 | (Kxxxxxx) | - |
| Name of Device | | Nitrile<br>Examination<br>Powder Free<br>Glove, Blac k,<br>Nitrile<br>Examination<br>Powder Free<br>Glove, Orange. | AMADEX-<br>Nitrile Powder<br>Free<br>Examination<br>Gloves Non-<br>Sterile, (Blue,<br>Black, Orange) | - |
| Dimension s | ASTM D 6319-10<br>Length:<br>Range - Length: Min -240<br>mm<br>(All Size s ) | Res ult:<br>Length h:<br>Min 240mm<br>By Sizes: | Result:<br>Length: Min<br>240mm<br>By Sizes: | Sam e |
| | Width:<br>Range value -<br>XS : Length : 220mm,<br>Width: 70 ± 10 mm<br>S : Length : 220mm,<br>Width: 80 ± 10mm<br>M : Length : 230mm,<br>Width: 95 ± 10 mm<br>L : Length : 230mm,<br>Width: 110 ± 10mm<br>XL : Length : 230mm,<br>Width: 120 ± 10mm | XS-<br>Width: 76 ± 3<br>mm<br>S -<br>Width: 84 ± 3<br>mm<br>M-<br>Width: 94 ± 3<br>mm<br>L -<br>Width: 105 ± 3<br>mm<br>XL -<br>Width: 113 ± 3<br>mm | XS -<br>Width: 76 ± 3<br>mm<br>S-<br>Width: 84± 3<br>mm<br>M-<br>Width: 94 ± 3<br>mm<br>L -<br>Width: 105 ± 3<br>mm<br>XL -<br>Width: 113 ± 3<br>mm | |
| Physical<br>Properties | ASTM D 6319-10<br>Tensile, Elongation -<br>Before Aging<br>Range:<br>Tensile Strength min: > 14<br>Mpa<br>Ultimate Elongation Min:<br>500%<br>After Aging<br>Range:<br>Tensile Strength min: > 14<br>Mpa<br>Ultimate Elongation Min:<br>400% | Before Aging<br>Result: 14 Mpa<br>500%<br>After Aging<br>Result: 14 Mpa,<br>400% | Before Aging<br>Result: 14 Mpa<br>500%<br>After Aging<br>Result: 14 Mpa<br>400% | Same |
| Thickness | ASTM D 6319-10<br>Range:<br>Thickness (mm)<br>Single wall ( <i>All Sizes</i> )<br>Finger:<br>Typical Value: (0.10 -<br>0.12)<br>Palm:<br>Typical Value: (0.07 -<br>0.08) | Result:<br>Finger: 0.10 ±<br>0.02<br>Palm: 0.07 ±<br>0.02 | Result:<br>Finger: 0.10 ±<br>0.02<br>Palm: 0.07 ±<br>0.02 | Same |
| Powder Free | ASTM S 6319-10 | Result: | Result: | Same |
| | Range:<br>< 2 mg /glove | XS: 0.6 mg /<br>glove<br>S: 0.8 mg /<br>glove<br>M: 1.2 mg/glove<br>L: 0.6 mg /glove<br>XL: 1.0 mg /<br>glove | XS: 0.6 mg /<br>glove<br>S: 0.8 mg /<br>glove<br>M: 1.2 mg/glove<br>L: 0.6 mg /glove<br>XL: 1.0 mg /<br>glove | Same |
| Biocompatibility | Primary Skin<br>Irritation - ISO 10993-<br>10:2010 (E) | Under the<br>condition on<br>of study not<br>an irritant | Under the<br>condition of<br>study not<br>an irritant | Same |
| | Dermal Sensitization Assay<br>(Skin Maximization) - ISO<br>10993-10:2010 (E) | Under the<br>condition of<br>study not an<br>irritant | Under the<br>condition of<br>study not an<br>irritant | Same |
#### TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE: 6.
{5}------------------------------------------------
{6}------------------------------------------------
# MEG/NEINDUSTRIES SDN BHD
| Characteristics | Standards | Device Performance | | Remark s |
|--------------------------------|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | | Predicate | Subject | |
| Intended use | - | A patient<br>examination<br>glove is a<br>disposable glove<br>intended for<br>medical purpose<br>that is worn on<br>the examiner' s<br>hand or finger to<br>prevent<br>contamination<br>between patient<br>and exam in er. | A powder-free<br>patient<br>examination glove<br>is a disposable<br>device intended<br>for medical<br>purposes that is<br>worn on the<br>examiner's hand<br>or finger to<br>prevent<br>contamination<br>between patient<br>and examiner | - |
| Water Tight<br>(1000 ml) | ASTM D 5151-06 | AQL 2.5<br>PASSES | AQL 2.5<br>PASSES | Same |
| Material | ASTM D 6910-10 | Nitrile | Nitrile | Same |
| Colour | - | Black,<br>Orange | Black,<br>Orange | Same |
| | | | Blue | Different |
| Available Sizes<br>with colour | | Black (XS, S, M,<br>L, XL) | Black (XS, S, M,<br>L, XL) | Same |
{7}------------------------------------------------
| | | O range<br>(XS,S,M,L,XL) | O range<br>(XS,S,M,L,XL) | |
|--------------|----------------------------------------------|--------------------------------------------------------|--------------------------------------------------------|-----------|
| | | | Blu e<br>(XS,S,M,L,XL) | Different |
| Texture | - | Finger Texture | Finger Texture | Same |
| Sizes | ASTM D 6319-10 | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large | Same |
| Single Use | Medical Glove Guidance<br>Manual - Labelling | Single Use | Single Use | Same |
| Manufacturer | | GMP Medicare<br>Sdn Bhd,<br>Malaysia | Megine<br>Industries Sdn<br>Bhd, Malaysia | - |
- 7. SUMMARY OF NON-CLINCIAL TESTING SUMMARY:
The standards used for Megine Industries Sdn Bhd glove production are Based on the listed standards. All testing meets requirements for Physical and Dimensions testing conducted on gloves.
- · ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
- · ASTM D5151-19 Standard Test Method for Detection of Hole s in Medical Gloves
- ASTM D6124 -06 Standard Test Method for Residual Powder on Medical Gloves
- • ASTM D3578-05,2015 Standard Specification for Rubber Examination Gloves
- ISO 10993-10: 2010, Biological evaluation of medical de vices Part 10: Test for irritation and Skin Sensitization
- ISO 10993-10: 2010, Biologicalevaluation of medical devices Part 10: Test for irritation and Skin Sensitization
- · ISO 1522 3-1:2016, Medical Devices Symbols to be used with medical device labels, labelling ad information to be supplied - Part I: General Requirements.
There are no special labeling claims and there is no claim as hypoallergic on the labels.
- a) Performance Data -
| Test Method | Purpose | Acceptance Criteria | Result |
|-------------|---------|---------------------|--------|
|-------------|---------|---------------------|--------|
{8}------------------------------------------------
# MEG/NE INDUSTRIES SON BHD
| MEG/NE INDUSTRIES SON BHD | | | |
|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|----------------------------------------|---------------------------|
| ASTM D 6319- 10<br>(Reapproved 2015)<br>Standard specification<br>for Nitrile<br>Examination Gloves<br>for Medical<br>Application | To determine the<br>length of the gloves | Min 230 mm | >230 mm (Passes) |
| ASTM D 6319-10<br>(Reapproved 2015)<br>Standard specification<br>for Nitrile<br>Examination Gloves<br>for Medical<br>Application | To determine the<br>width of the gloves | $95 \pm 10$ mm | $98 \pm 2$ mm (Passes) |
| ASTM D 6319-10<br>(Reapproved 2015) | To determine the<br>thickness of the gloves | Palm 0.05 mm min<br>Finger 0.05 mm min | Palm: 0.09 mm<br>(Passes) |
| Standard specification<br>for Nitrile<br>Examination Gloves<br>for Medical<br>Application | | | Finger: 0.12 mm<br>(Passes) |
|----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| ASTM D 6319-10<br>(Reapproved 2015)<br>Standard specification<br>for Nitrile<br>Examination Gloves<br>for Medical<br>Application | To determine the<br>physical properties<br>Tensile strength | Before Ageing:<br>Tensile strength<br>14Mpa min | 24.5 Mpa (Passes) |
| ASTM D 5151-06<br>(Reapproved 2015)<br>Standard Test method<br>for Detection of Holes<br>in Medical Gloves | To determine the holes<br>in the gloves | AQL 2.5 | AQL 2.5 (Passes) |
| ASTM D 6124-06<br>(Standard Test Method<br>for Residual Powder on<br>medical Gloves) | To determine the<br>amount of residual<br>powder and non-<br>powder solids found<br>on gloves | Less than 2 mg per<br>glove | 0.60 mg / glove<br>(Passes) |
| ASTM D 3578-05<br>(Reapproved 2015)<br>Standard Specification<br>for Rubber<br>Examination Gloves | To determine the<br>specifications for<br>rubber examination<br>gloves | Dimension, AQL: 4.0<br>Length: min 230 mm<br>Force at Break, AQL:<br>4.0, Tensile and<br>Modul us | Meet the requirements<br>of ASTM 3578 until 5<br>years (Passes) |
- b) Bio-Compatibility Data
| Test Method | Purpose | Acceptance Criteria | Re sult |
|-------------|---------|---------------------|---------|
| | | | |
{9}------------------------------------------------
## MEG/NE INDUSTRIES SON BHD
| ISO 1099 3-10<br>Biological Evaluation<br>of Medic al Device<br>Test for Irritation and<br>Skin Sensitization.<br>Test done for<br>Irritation. | To deter mine the<br>potential irritation of<br>the effect likely to<br>arise from a single<br>exposure of test<br>material on the intact<br>skin of rabbits. | Under the condition of<br>study not an irritant | Under the condition of<br>study not an irritant |
|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|-------------------------------------------------|
| ISO 10993-10<br>Biological Evaluation<br>of Medical De vice<br>Test for Irritation and<br>Skin Sensitization.<br>Test done for<br>sensitization. | To evaluate the<br>potential of a test<br>material to cause a<br>delayed<br>hypersensitivity<br>reaction (Type IV)<br>following exposure of<br>the skin of guinea pigs. | Under the condition of<br>study not an irritant | Under the condition of<br>study not an irritant |
- 8. Discussion of Clinical tests performed: Not Applicable - Clinical Data is not needed for Gloves or for most devices cleared by the 510 (K) process.
- 9. Conclusion:
The conclusion drawn from the nonclinical test demonstrate that the subject device in 510 (K) submission Kl 72867. AMADEX Nitrile Powder Free Examination Gloves Non-Sterile, (Blue, Black, and Orange) is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.