K091736 · Daniels Corporation Pty, Inc. · MMK · Oct 30, 2009 · General Hospital
Device Facts
Record ID
K091736
Device Name
SHARPSMART DISPOSABLE SHARPS CONTAINER, MODEL S2
Applicant
Daniels Corporation Pty, Inc.
Product Code
MMK · General Hospital
Decision Date
Oct 30, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
Sharpsmart™ S2 disposable sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps. This device is intended for "Over-the-Counter" use.
Device Story
Sharpsmart S2 is a 1-quart, two-piece, injection-molded polypropylene copolymer container; used for safe disposal of hazardous medical sharps. Device features closure mechanism allowing multiple lid re-openings while ensuring security between uses; includes optional bracket for vertical/horizontal surface mounting. Used in patient rooms, medication carts, ORs, and emergency vehicles. Healthcare professionals or small quantity waste generators insert sharps vertically through top flaps; flaps lock for final disposal. Translucent design with fill line and overfill warning allows visual monitoring of capacity. Device provides puncture resistance, leak-proof sides/bottom, and stability to prevent toppling. Benefits include safe containment of hazardous sharps, reducing risk of accidental needle sticks during disposal.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified via standardized testing: UN Recommendations on the Transport of Dangerous Goods (15th Ed. 6.1.5.3) and AS/NZS 4031:1992. Passed tests for puncture resistance, leak resistance, impact, handle integrity, drop, stacking, and vibration.
Technological Characteristics
Polypropylene copolymer construction; injection-molded two-piece design. 1-quart capacity (1.6 qt total volume, 1.16 qt fill capacity). Dimensions: 7.5"h x 4.5"w x 5.5"d. Features top-loading flap mechanism for sharps insertion and locking closure. Non-sterile. Standalone or bracket-mounted. Complies with AS/NZS 4031:1992 and UN Transport of Dangerous Goods standards.
Indications for Use
Indicated for safe disposal of hazardous sharps in healthcare facilities, nursing stations, medication carts, labs, dental offices, ERs, surgical rooms, treatment rooms, emergency vehicles, and veterinary offices. Intended for over-the-counter use by healthcare professionals and small quantity waste generators.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
K964387 — POINT OF USE II SHARPS-A-GATOR · Graphic Controls Corp. · Mar 27, 1997
K130281 — OAK RIDGE PRODUCTS SHARP CONTAINER · Oak Ridge Products, LLC · Jun 21, 2013
K072667 — SHARPS AWAY DISPOSABLE CONTAINER · Solutions, Inc. · Oct 10, 2007
K190240 — Tiger Sharps Containers · International Marketing Specialists, Inc. · May 31, 2019
K960525 — BEMIS ONE QUART PHLEBOTOMY CONTAINER (PART NUMBER 100) · Bemis Mfg. Co. · Mar 28, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
KC091736
510(k)
### Daniels Sharpsmart
Page 3
## 510(k) Summary
Manufacturer Name:
The Daniels Corporation PTY, LTD
Company Address:
Representative
135 S. LaSalle St. Ste 2850 Chicago, IL 60603
Andrea Arredondo Daniels Sharpsmart, Inc. 2133-126 Upton Dr., #436 Virginia Beach, VA 23454 (540) 818-0783 (757) 299-8363 (fax) aarredondo@danielsinternational.com
OCT 3 0 2009
Summary Preparation Date: Sept 29, 2009
DEVICE NAME
| Trade/Proprietary Name: | Sharpsmart™ S2 disposable sharps container |
|-------------------------|------------------------------------------------------------------|
| Common Name: | Disposable sharps disposal container |
| Classification Name: | Accessory to hypodermic single lumen needle;<br>Sharps Container |
# PREDICATE DEVICE IDENTIFICATION:
21CFR880.5570
Sharps Container
Product Code MMK
Device Class: II
## LEGALLY MARKETED EQUIVALENT DEVICE:
| Company | Product | 510(k)# |
|----------------|----------------------------------|---------|
| Solutions, Inc | Sharps Away Disposable Container | K072667 |
{1}------------------------------------------------
#### 510(k)
### DESCRIPTION OF DEVICE:
The Sharpsmart™ S2 disposable sharps container is intended to be used in patient rooms, medication rooms, operating rooms, physicians' offices or any other patient care area requiring the use of a sharps container. It is a two-piece injection-molded container made from a polypropylene copolymer. The container is designed with a closure mechanism that allows multiple lid re-openings, while ensuring secure between uses. Both handle and lid can be secured out of the way for safe, unobstructed disposal of sharps. The S2 container is capable of standalone use. A Unique bracket design is offered as an additional measure to secure mounting on vertical or horizontal surfaces to prevent toppling. The bracket is not required for the use of the container. It only offers a secondary option for the user for ease of use.
The containers are closable, puncture resistant, leak proof on the sides and bottom and stable. The label, containing an overfill warning and an easily visible fill line, is adhered to the container at the factory. The containers are translucent and are red or yellow. The label is white and red with the biohazard warning.
Sharpsmart™ S2 disposable sharps containers are available in a 1 quart size. Each product has two parts, which assemble together to form a unit. Assembly is easily visualized. They can be nested together to reduce storage space.
The container is designed and certified to comply with penetration requirements, as represented though product testing for Test methods: AS/NZS 4031:1992 and Test Method: roprosence and the Transport of Dangerous Goods. 15th revised Ed. 6.1.5.3The Sharpsmart™ S2 disposable sharps containers meet or exceed OSHA recommendations for sharps containers.
#### INTENDED USE:
Sharpsmart™ S2 disposable sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps. This device is intended for "Over-the-Counter" use.
S2 model is a 1 quart size. The container has a volume of 1.6quarts quarts, fill capacity of 1.16 quarts and an empty weight of 0.95 lb.
S2 model Color - Red with translucent lid or yellow with translucent lid.
S2 model outer dimensions of 7.50 "h x 4.50"w x 5.50 "d. (190mm x 115mm x 140mm ).
{2}------------------------------------------------
# PREDICATE PRODUCT COMPARISON TABLE:
| Manufacturer | Daniels Sharpsmart, Inc | Solutions, Inc |
|---------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Sharpsmart Disposable<br>Sharps Container | Sharps Away Disposable<br>Container |
| K Number | K091736 | K072667 |
| Indication for Use | Sharpsmart Disposable<br>Containers are intended to<br>be used for the safe disposal<br>of hazardous sharps | Sharps Away Disposable<br>Containers are intended to<br>be used for the safe disposal<br>of hazardous sharps |
| Target population | Healthcare professional | Healthcare professional |
| Where used | Healthcare facilities | Healthcare facilities |
| Material | Polypropylene | Polypropylene |
| Sharp Access | Sharps inserted through the<br>top in a vertical position<br>with sharp side down<br>through a hole formed with<br>flaps | Sharps inserted through the<br>top in a vertical position<br>with sharp side down<br>through a hole formed with<br>flaps |
| Sharps closure | Flaps are closed and locked<br>in place for removal | Flaps are closed and locked<br>in place for removal |
| Impact Resistance | Yes | Yes |
| Puncture Resistance | Yes | Yes |
| Leak Resistance | Yes | Yes |
| Single Use | Yes | Yes |
| Non-sterile | Yes | Yes |
## SUBSTANTIAL EQUIVALENCE DISCUSSION OF SIMILARITIES AND DIFFERENCES:
The Sharpsmart S2 Disposable Sharps Containers are similar to the Solution, Inc Sharps Away Disposable Containers in:
Non-Clinical test comparisons are viewed through: Intended Use, Materials and design.
Clinical tests are discussed in the Performance testing section.
Daniels submits the following information to demonstrate that the Sharpsmart S2 Disposable Sharps Container shares indications, design principles, materials and properties with the following predicate devices, which have been determined by FDA to be white the 2010 wais alent to pre-amendment devices: Sharps Away Disposable Containers (K072667) from Solutions, Inc.
Intended Use Comparison
{3}------------------------------------------------
The indications for use of the Sharpsmart S2 Disposable Sharps Container are not new indications in that they are the same as or are included in those for the predicate devices. Sharpsmart and the predicate devices are containers intended to be used for the disposal of contaminated medical sharps in health care facilities.
### Design and Materials Comparison
The design and functional characteristics of the Sharpsmart and the predicate devices are similar. They are constructed from polymeric materials. The Disposable Sharps Containers and Sharps Away containers are intended to be disposed once used. All of the devices conform to national or international standards for puncture resistance, impact resistance and leakage. They have means to prevent contact between the user and the contents, and are designed with features to easily and safely determine if they are full. They do not have features to bend, break or shear needles.
### Performance Testing Comparison
Test methods: UN Recommendations on the Transport of Dangerous Goods. 15th revised Ed. 6.1.5.3; AS/NZS 4031:1992
| Puncture Resistance -Passed | Handle Testing - Passed |
|-----------------------------|-------------------------|
| Leak Resistance - Passed | Drop Test - Passed |
| Impact Testing - Passed | Stacking Test - Passed |
Vibration - Passed
#### Conclusion:
The Sharpsmart S2 Disposable Sharps Containers introduces no new questions concerning the safety or effectiveness of the Sharpsmart Disposable Sharps Containers and is thus substantially equivalent to the predicate device.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Daniels Corporation PTY, Limited C/O Ms. Andrea Arredondo Compliance Manger Daniels Sharpsmart, Incorporated 2133-126 Upton Drive, #436 Virginia Beach, Virginia 23454
OCT 3 0 2009
Re: K091736
Trade/Device Name: Sharpsmart™ S2 Disposable Sharps Contianer Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK Dated: October 2, 2009 Received: October 2, 2009
Dear Ms. Arredondo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2- Ms. Arredondo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRH Offices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Clinton D. Van
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use:
510(k) Number (if known): K091736 -
Device Name: Sharpsmart™ S2 disposable sharps container
Model: S2
Indications For Use: Sharpsmart™ S2 disposable sharps container are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps. This device is intended for "Over-the-Counter" use.
Physical Attributes: S2 model is a 1 quart size. The container has a volume of 1.6quarts quarts, fill capacity of 1.16 quarts and an empty weight of 0.95 lb.
S2 model Color - Red with translucent lid or yellow with translucent lid.
S2 model outer dimensions of 7.50 "h x 4.50"w x 5.50 "d. (190mm x 115mm x 140mm ).
Prescription Use __
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ X
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K 0 917 3 6
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.