K190240 · International Marketing Specialists, Inc. · MMK · May 31, 2019 · General Hospital
Device Facts
Record ID
K190240
Device Name
Tiger Sharps Containers
Applicant
International Marketing Specialists, Inc.
Product Code
MMK · General Hospital
Decision Date
May 31, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
Tiger Sharps Containers are intended to provide a receptacle for used, contaminated medical sharps and act as an enclosure during transport to ultimate disposal. The Containers are single-use, disposable, non-sterile containers intended to be used for health-care purposes for safe disposal of hazardous sharps such as hypodermic needles, syringes, lancets and blood needles. The target population is for qualified personnel in health care facilities in which medical sharps may be used. All device models are not for use in areas with unsupervised patient access. All device models only be used with appropriate mounting accessories.
Device Story
Tiger Sharps Containers are injection-molded polypropylene receptacles for used, contaminated medical sharps; designed for single-use disposal. Containers feature puncture-resistant, leak-resistant, and impact-resistant construction; translucent lids allow visual fill-level determination. Models range from 1 Quart to 8 Gallons with various lid configurations (hinged, sliding, counterbalanced, rotary) and integrated needle unwinder ports. Used in healthcare facilities by qualified personnel; mounted via wall brackets, stabilizers, or free-standing. Device acts as a secure enclosure during transport to disposal. Benefits include safe containment of hazardous sharps, reducing risk of accidental injury to healthcare staff.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM F2132-01(2008)e1 (puncture resistance) and ISO 23907-2012 (drop, handle strength, stability). Transport safety verified via 49 CFR 178.606 (stacking), 49 CFR 178.608 (vibration), and 49 CFR 178.603 (drop test).
Indicated for safe disposal of hazardous sharps (hypodermic needles, syringes, lancets, blood needles) by qualified personnel in healthcare facilities. Not for use in areas with unsupervised patient access.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 31, 2019
International Marketing Specialists Inc. % Charles Mack Principal Engineer IRC 2950 E Lindrick Drive Chandler, Arizona 85249
Re: K190240
Trade/Device Name: Tiger Sharps Containers Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: MMK Dated: April 21, 2019 Received: April 29, 2019
Dear Charles Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For David Krause, PhD Acting Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K190240
Device Name Tiger Sharps Containers
Indications for Use (Describe)
Tiger Sharps Containers are intended to provide a receptacle for used, contaminated medical sharps and act as an enclosure during transport to ultimate disposal.
The Containers are single-use, disposable, non-sterile containers intended to be used for health-care purposes for safe disposal of hazardous sharps such as hypodermic needles, syringes, lancets and blood needles. The target population is for qualified personnel in health care facilities in which medical sharps may be used. All device models are not for use in areas with unsupervised patient access.
All device models only be used with appropriate mounting accessories.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| Model | Weight (empty) | Capacity (total) | Capacity (full line) | Dimensions of finished goods (in) (L x W x H) | Colors | Acceptable sites of use | Model | Lid<br>configuration | Dimensions of Sharps disposal<br>aperture (inches) | Locking<br>mechanism | Needle<br>unwinder | Requirements<br>for<br>mounting |
|------------|----------------|------------------|----------------------|-----------------------------------------------|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------|-------------------------------------------------------------------------------|----------------------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| 1 Quart | 109 grams | 28 Oz | 23.5 Oz | 3.91x3.91x6.37 | Red base and transparent lid | The target population is for qualified personnel in health care facilities and other facilities in which medical sharps may be used. All the containers are intended to be used in areas where there is no unsupervised patient access. | 1 Quart | Dual opening<br>Hinge | Opening 1: 1.70 x 1.18;<br>Opening 2: 0.7 diameter;<br>Opening 3: 0.83 x 0.31 | Hinged closure | located in the lid<br>and above the<br>containment<br>area. The<br>unwinder has a<br>round entry port<br>for the needle to | a. Non-Locking Wire<br>Wall Bracket<br>(Covidien)<br>b. Stabilizer/holder with<br>Double face adhesive<br>tape mounting<br>c. Free standing |
| 5 Quart | 450 grams | 140 Oz | 119 Oz | 10.60 x4.70x10.64 | | | 5 Quart | Counter<br>Balance | 8.67 x 2.36 x 2.27 | Hinged closure | pass through,<br>allowing it to be<br>fully enclosed<br>within the<br>container. | a. Locking wall bracket<br>(Covidien)<br>b. Wall enclosure<br>c. Free standing |
| 2 Gallon | 453 grams | 7.8 Quarts | 6 Quarts | 10.84x6.86x10.36 | | | 2 Gallon | Sliding | 2.40 x 5.04 | Sliding | | a. Locking wall bracket<br>(Covidien)<br>b. Plastic wall bracket<br>c. Stainless steel or<br>plastic stabilizer/holder<br>d. Free standing |
| 2 Gallon B | 453 grams | 7.8 Quarts | 6 Quarts | 10.22x7.02x13.30 | | | 2 Gallon B | Star Hinge | 2.83 Diameter | Sliding | | a. Locking wall bracket<br>(Covidien)<br>b. Plastic wall bracket<br>c. Stainless steel or<br>plastic stabilizer/holder<br>d. Free standing |
| 15 liter | 625 grams | 15 liter | 12 liter | 9.50x14.77x12.03 | | | 15 liter | Sliding | 9.98 x 3.43 | Sliding | | a. Stainless steel or plastic<br>stabilizer/holder<br>b. Free standing |
| 3 Gallon | 559 grams | 2.9 Gal | 2.45 Gal | 14.86x7.25x14.00 | | | 3 Gallon | Counter<br>Balance | 9.37 x 3.34 | Counterbalanced<br>door | | a. Locking wire bracket<br>(Bemis)<br>b. Plastic wall bracket<br>c. Plastic<br>stabilizer/holder<br>d. Free standing |
| 8 Gallon | 1172 grams | 7.8 Gal | 6.5 Gal | 10.92x15.61x17.27 | | | 8 Gallon | Rotary/Hinge | 3.94 x 1.70 | Rotary door and<br>hinge closure | | a. Stainless steel or plastic<br>stabilizer/holder<br>b. Free standing |
{4}------------------------------------------------
{5}------------------------------------------------
#### K190240
# 510(k) Summary
Date of Preparation: April 21, 2019
# I. Submitter Information:
| Submitter Name: | International Marketing Specialists Inc. |
|-----------------|--------------------------------------------|
| Address: | 1278 Highway 461, Somerset, Kentucky 42503 |
| Contact Person: | Mr. Rod Calderon, Manager |
### US Agent and Correspondent
Mr. Charles Mack Principal Engineer IRC 2950 E Lindrick Drive, Chandler, Arizona 85249 USA Tel: 931-625-4938 Email: charliemack@irc-us.com
# II. Device
| Trade Name: | Tiger Sharps Container |
|--------------------|--------------------------------|
| Common Name: | container, sharps |
| Regulation Number: | 21 CFR§880.5570 |
| Regulation Name: | Hypodermic single lumen needle |
| Regulatory Class: | II |
| Product Code: | MMK |
# III. Predicate Device Information
| Manufacturer | Predicate Device | 510(k) Number | Submitted Device |
|------------------------------|------------------------------|---------------|-------------------------|
| Medline Sharps<br>Containers | Medline Sharps<br>Containers | K143693 | Tiger Sharps Containers |
{6}------------------------------------------------
### IV. Device Description:
Tiger Sharps Containers constructed of injection molded polypropylene plastic. They are designed for a single-use (disposable) by healthcare professionals. No part of the container is intended to come in contact with patients. The containers are designed to be puncture resistant, leak resistant on the sides and bottom, impact resistant, closable and stable.
Lids and closures are uncolored translucent material allowing for a visual determination of fill level. The base is made from a high strength material to support the capacity of the container. The recommended fill level is engraved onto the plastic and corresponds to the level line on the product identification label.
| Parts & Accessories | Material | Material Specification | Patient<br>Contact<br>(Direct<br>/Indirect)? |
|---------------------|--------------------------|-----------------------------------------|----------------------------------------------|
| Lid | Polypropylene | Formolene® Polypropylene<br>Homopolymer | No |
| Base | Polypropylene | Formolene® Polypropylene<br>Homopolymer | No |
| Color power | Hififast Scarlet<br>HF4Y | PPR007BTS | No |
#### V. Intended Use / Indications for Use
Tiger Sharps Containers are intended to provide a receptacle for used, contaminated medical sharps and act as an enclosure during transport to ultimate disposal.
The Containers are single-use, disposable, non-sterile containers intended to be used for health-care purposes for safe disposal of hazardous sharps such as hypodermic needles, syringes, lancets, and blood needles. The target population is for qualified personnel in health care facilities and other facilities in which medical sharps may be used. All device models are not for use in areas with unsupervised patient access.
All device models only are used with appropriate mounting accessories.
{7}------------------------------------------------
| Model | Weight<br>(empty) | Capacity<br>(total) | Capacity<br>(full<br>line) | Dimensions of<br>finished goods (in)<br>(L x W x H) | Colors | Acceptable sites of use |
|------------|-------------------|---------------------|----------------------------|-----------------------------------------------------|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 Quart | 109<br>grams | 28 Oz | 23.5 Oz | 3.91x3.91x6.37 | Red base<br>and<br>transparent<br>lid | The target population is for<br>qualified personnel in<br>health care facilities and<br>other facilities in which<br>medical sharps may be<br>used. All the containers are<br>intended to be used in<br>areas where there is no<br>unsupervised patient<br>access. |
| 5 Quart | 450<br>grams | 140 Oz | 119 Oz | 10.60 x4.70x10.64 | | |
| 2Gallon | 453<br>grams | 7.8<br>Quarts | 6 Quarts | 10.84x6.86x10.36 | | |
| 2 Gallon B | 453<br>grams | 7.8<br>Quarts | 6 Quarts | 10.22x7.02x13.30 | | |
| 15 liter | 625<br>grams | 15 liter | 12 liter | 9.50x14.77x12.03 | | |
| 3 Gallon | 559<br>grams | 2.9 Gal | 2.45 Gal | 14.86x7.25x14.00 | | |
| 8 Gallon | 1172<br>grams | 7.8 Gal | 6.5 Gal | 10.92x15.61x17.27 | | |
| Model | Lid<br>configuration | Dimensions of<br>Sharps disposal<br>aperture (inches) | Locking<br>mechanism | Needle<br>unwinder | Requirements for<br>mounting |
|-------------|-----------------------|-------------------------------------------------------------------------------------|----------------------|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| 1 Quart | Dual opening<br>Hinge | Opening 1: 1.70 x<br>1.18; Opening 2:<br>0.7 diameter;<br>Opening 3: 0.83 x<br>0.31 | Hinged closure | located in<br>the lid and<br>above the<br>containment<br>area. The<br>unwinder<br>has a round | a. Non-Locking Wire<br>Wall Bracket<br>(Covidien)<br>b. Stabilizer/holder with<br>Double face adhesive<br>tape mounting<br>c. Freestanding |
| 5 Quart | Counter<br>Balance | 8.67 x 2.36 x 2.27 | Hinged closure | entry port<br>for the<br>needle to<br>pass | a. Locking wall bracket<br>(Covidien)<br>b. Wall enclosure<br>c. Freestanding |
| 2<br>Gallon | Sliding | 2.40 x 5.04 | Sliding | through,<br>allowing it<br>to be fully<br>enclosed<br>within the<br>container. | a. Locking wall bracket<br>(Covidien)<br>b. Plastic wall bracket<br>c. Stainless steel or plastic<br>stabilizer/holder<br>d. Freestanding |
{8}------------------------------------------------
| 2<br>Gallon<br>B | Star Hinge | 2.83 Diameter | Sliding | a. Locking wall bracket<br>(Covidien)<br>b. Plastic wall bracket<br>c. Stainless steel or plastic<br>stabilizer/holder<br>d. Freestanding |
|------------------|--------------------|--------------------|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| 15 liter | Sliding | $9.98 \times 3.43$ | Sliding | a. Stainless steel or plastic<br>stabilizer/holder<br>b. Freestanding |
| 3<br>Gallon | Counter<br>Balance | $9.37 \times 3.34$ | Counterbalanced<br>door | a. Locking wire bracket<br>(Bemis)<br>b. Plastic wall bracket<br>c. Plastic stabilizer/holder<br>d. Free standing |
| 8<br>Gallon | Rotary/Hinge | $3.94 \times 1.70$ | Rotary door and<br>hinge closure | a. Stainless steel or plastic<br>stabilizer/holder<br>b. Freestanding |
{9}------------------------------------------------
| Element of comparison | Subject Device | Predicate Device |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | International Marketing Specialists Inc. | Medline Industries, Inc. |
| FDA510(K) Number | K190240 | K143693 |
| Device Name | Tiger Sharps Containers | Medline Sharps Containers |
| Intended Use | Tiger Sharps Containers are intended to provide a receptacle for used, contaminated medical sharps and act as an enclosure during transport to ultimate disposal. | Medline Sharps Containers are intended to provide a receptacle for used, contaminated medical sharps and act as an enclosure during transport to ultimate disposal. |
| Product Code | MMK | MMK |
| Regulation Number | 21CFR880.5570 | 21CFR880.5570 |
| Class | II | II |
| Capacities | 1 Quart / 5 Quart / 2 Gallon / 2 Gallon<br>B / 15 Liter<br>3 Gallon/ 8 Gallon | 1G / 2G / 8G / 10G / 12G / 18G |
| Prescribed | OTC | OTC |
| Weight Range (grams) | 109 - 1172 | 303 - 2388 |
| No. of Piece | 2-3 | 2-3 |
| Material | Polypropylene | Polypropylene |
| Color | Base: Red<br>Lid: Transparent | Red |
| Clarity | Opaque/translucent | Opaque/translucent |
| Non-sterile | Yes | Yes |
| Method of Manufacture | Injection Molded | Injection Molded |
| Performance testing (puncture, impact, drop, stability, integrity etc.) | Pass | Pass |
| Standards | ISO 23907, ASTM F 2132 | ISO 23907, ASTM F 2132 |
| Disposable or Re-usable? | Disposable | Disposable |
# Comparison of Technological Characteristics with the Predicate Device
{10}------------------------------------------------
#### VI. Non Clinical Performance Testing:
Performance testing was provided in support of the substantial equivalence determination and to validate and verify that the Tiger Sharps Container met all of the requirements of related international standards, including biocompatibility, sterility, and product specifications. Results of these tests demonstrate compliance with the requirements of the consensus standards noted below.
### Performance Testing
The following performance tests were conducted for the subject device: Puncture: per ASTM< F2132-01(2008)e1 = Pass Resistance to Damage/Leaking after Drop: per ISO 23907-2012 = Pass Handle Strength: per ISO 23907-2012 = Pass Container Stability: per ISO 23907-2012 = Pass Stacking: Per 49 CFR 178.606 = Pass Vibration: Per 49 CFR 178.608 = Pass Drop test: Per 49 CFR 178.603 = Pass
# Sterility Information
The Tiger Sharps container is non-sterile; therefore no sterility testing was performed.
# Clinical Testing:
Not Applicable.
#### VIII. Conclusions:
The non-clinical data demonstrate that the Tiger Sharps Container is as safe, as effective, and performs as well as the predicate device, Medline Sharps Containers , cleared under K143693.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.