K072667 · Solutions, Inc. · MMK · Oct 10, 2007 · General Hospital
Device Facts
Record ID
K072667
Device Name
SHARPS AWAY DISPOSABLE CONTAINER
Applicant
Solutions, Inc.
Product Code
MMK · General Hospital
Decision Date
Oct 10, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Sharps Away Disposable Containers and accessories are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps.
Device Story
Sharps Away Disposable Containers are injection-molded polypropylene receptacles designed for safe disposal of hazardous sharps. Containers feature puncture-resistant, leak-proof construction with integrated fill lines and overfill warnings. Available in various sizes (0.5 quart to 6 gallon) and colors (red/yellow), with optional wall brackets or countertop holders. Used in clinical settings by healthcare professionals to collect used needles and syringes. Sharps are inserted vertically through top flaps that lock for secure disposal. Device provides stable, single-use containment to prevent accidental needle sticks and exposure to biohazardous waste.
Clinical Evidence
Bench testing only. Performance testing included puncture resistance (per 49 CFR 178.609(h)(2), ASTM D7P1152/ASTM F2132-01), leak resistance, impact testing, and handle strength testing. All tests passed.
Indicated for the safe disposal of hazardous sharps in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, and veterinarian offices.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
K072667
## Executive Summary
307 1 0 2007
Company Name: Company Address Solutions Inc. 3600 Chamberlain Lane, Ste. 104 Louisville, KY 40241
Telephone: Fax:
(502) 228-2901 (502) 228-2931
Victor Anderson Contact Person:
Summary Preparation Date: July 27, 2007
Trade Name: Sharps Away Disposable Containers
Common Name: Sharps Container
Classification Name: Container, Sharps
Predicate Device Identification: CFR21 880.5570 Sharps Container Product Code: MMK Device Class: II
#### Legally Marketed Equivalent Device:
| <b>Company</b> | <b>Product</b> | <b>510(k) #</b> |
|------------------|-----------------------------------------|-----------------|
| Becton Dickinson | B-D Guardian Nestable Sharps Collectors | K943575 |
### Description:
Sharps Away Disposable Containers are injection molded polypropylene plastic. The containers are closable, puncture resistant, leak proof on the sides and bottom and stable. The label, containing an overfill warning and an easily visible fill line, is adhered to the container at the factory. The containers are translucent and are red and yellows. The label is white with the biohazard warning fluorescent orange,
Sharps Away Disposable Containers are available in a variety of sizes. Each product has two parts, which assemble together to form a unit. Assembly is easily visualized. They can be nested together to reduce storage space.
Handles are present on the 2-quart, 5-quart, 2.5-gallon and the 6-gallon containers. They are strong, easy to grasp and permits control of the container with one hand. Wall brackets are available for the 2-quart, 5 quart and the 2.5-gallon size. A counter top holder is available for the 1-quart size.
{1}------------------------------------------------
The Sharps Away Disposable Containers meet or exceeds OSHA recommendations for sharps containers.
### Intended Use:
Sharps Away Disposable Containers are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps.
| Manufacturer | Solutions Inc | Becton Dickinson |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Sharps Away Disposable Container | B-D Guardian Nestable Sharps |
| K Number | | K943575 |
| Indication for<br>Use | Sharps Away Disposable Containers are<br>intended to be used for the safe disposal<br>of hazardous sharps | B-D sharps containers are intended to<br>be used for the safe disposal of<br>hazardous sharps |
| Target<br>population | Healthcare professional | Healthcare professional |
| Where used | Healthcare facilities | Healthcare facilities |
| Material | polypropylene | polypropylene |
| Sharps access | Sharps inserted through the top in a<br>vertical position with sharp side down<br>through the hole formed with flaps<br>through which the sharp is inserted | Sharps inserted through the top in a<br>vertical position with sharp side down<br>through the hole formed with flaps<br>through which the sharp is inserted |
| Sharps closure | Flaps are closed and locked in place for<br>removal | Flaps are closed and locked in place<br>for removal |
| Impact<br>Resistance | Yes | Yes |
| Puncture<br>Resistance | Yes | Yes |
| Leak resistance: | Yes | Yes |
| Single Use: | Yes | Yes |
| Non-sterile: | Yes | Yes |
### Predicate Product Comparison Table:
Substantial Equivalence Discussion of Similarities and Differences:
The Sharps Away Disposable Containers are similar to the B-D Guardian Nestable Sharps containers in:
- . Intended Use
- . Target Population
- . Materials
- Design
- . Performance testing
Sharps Away Disposable Sharps Container introduces no new questions concerning the safety or effectiveness of the Sharps Away disposable container and is thus substantially equivalent to the predicate device.
{2}------------------------------------------------
K072667
### Performance Testing
Puncture Resistance - 49 CFR; Section 178.609(h)(2); ASTM 07P1152/ ASTM F2132-01 - Passed
Leak Resistance Test -Passed
Impact Test-Passed
Handle Test - Passed
{3}------------------------------------------------
Image /page/3/Picture/12 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized caduceus symbol. The caduceus is represented by a staff with a snake winding around it, symbolizing medicine and health. The seal is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### OCT 1 0 2007
Solutions, Incorporated Mr. Jeff D. Rongero Responsible Third Party Official Underwriters Laboratories, Incorporated 12 Laboratory Drive Research Triangle, North Carolina 27709
Re: K072667
Trade/Device Name: Sharps Away Disposable Container Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK Dated: October 4, 2007 Received: October 9, 2007
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Rongero
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Chiu-Ling, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): _K O7 2 ம்மு 2 முடி 2 ப
Device Name: Sharps Away Disposable Container
Indications for Use:
Sharps Away Disposable Containers and accessories are intended to be used in healthcare facilities, including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, emergency vehicles, veterinarian office and other small quantity waste generators for the safe disposal of hazardous sharps.
There are fourteen (14) containers and four (4) holders included in this submission. The model number, size, color and syringe size specification of each container and holder are as follows: (See attached sheet.)
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shule A. Murphy MD
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page of
510(k) Number: -
K072667
{6}------------------------------------------------
# K072667
## Indications for Use
## Product specifications
| Item # | Size | Color | Syringe Size |
|--------|---------------------------|--------|--------------|
| 1901 | 1 Quart | Red | .5ml - 20ml |
| 1902 | 1 Quart | Yellow | .5ml - 20ml |
| 1903 | 2 Quart | Red | .5ml - 60ml |
| 1904 | 2 Quart | Yellow | .5ml - 60ml |
| 1905 | 5 Quart | Red | .5ml - 60ml |
| 1906 | 5 Quart | Yellow | .5ml - 60ml |
| 1907 | 2.5 Gal w/Horizontal Lid | Red | .5ml - 60ml |
| 1908 | 2.5 Gal w/Horizontal Lid | Yellow | .5ml - 60ml |
| 1909 | 2.5 Gal w/Vertical Lid | Red | .5ml - 60ml |
| 1910 | 2.5 Gal w/Vertical Lid | Yellow | .5ml - 60ml |
| 1911 | 6 Gal | Red | .5ml - 60ml |
| 1912 | 6 Gal | Yellow | .5ml - 60ml |
| 1915 | .5 Quart | Red | Needles only |
| 1916 | .5 Quart | Yellow | Needles only |
| 1951 | 2 Quart Wall Bracket | Beige | N/A |
| 1952 | 5 Quart Wall Bracket | Beige | N/A |
| 1953 | 2.5 Gal Wall Bracket | Beige | N/A |
| 1954 | 1 Quart Countertop Holder | Blue | N/A |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.