K123684 · Sol-Millennium Medical · FMI · May 8, 2013 · General Hospital
Device Facts
Record ID
K123684
Device Name
SOL-CARE SAFETY NEEDLE
Applicant
Sol-Millennium Medical
Product Code
FMI · General Hospital
Decision Date
May 8, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sol-Care Safety Hypodermic Needle is used in conjunction with a standard syringe. This device is used for aspiration and administration of medication. The Sol-Care Safety Hypodermic Needle safety mechanism covers the needle after use. In the activated position, the needle cover guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination.
Device Story
Sol-Care Safety Needle is a sterile, single-use hypodermic needle (18-30 gauge; 3/8"-1-1/2" length) featuring an integrated safety shield. Device attaches to standard syringes for fluid aspiration or medication administration. After use, user manually activates shield to cover needle tip, reducing risk of accidental needle sticks during disposal. Used in clinical or home settings. Safety mechanism provides physical barrier between user and contaminated needle point. Benefits include enhanced sharps injury protection for clinicians and patients.
Clinical Evidence
No clinical data submitted. Safety and effectiveness supported by non-clinical bench testing, including simulated use studies, accelerated aging, and compliance with ISO standards (ISO 7864, ISO 594-2, ISO 9626, ISO 14971, ISO 11135, ISO 11607, ISO 10993-1, and ISO 23908).
Technological Characteristics
Sterile, single-use hypodermic needle with integrated safety shield. Materials: stainless steel tubing (ISO 9626), Luer taper (ISO 594-2). Sterilization: Ethylene Oxide (ISO 11135). Dimensions: 18-30 gauge, 3/8"-1-1/2" length. Mechanical safety feature per ISO 23908.
Indications for Use
Indicated for aspiration and administration of medication using a standard syringe. Intended for use by healthcare professionals or patients for general purpose injection and fluid aspiration below the skin surface, incorporating a safety mechanism to prevent accidental needle sticks during handling and disposal.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
K123684
510(k) Summary (As required by 21 CFR 807.92(a))
# MAY 0 8 2013
Summary of Safety and Effectiveness for the Safety Needle
Date Prepared: January 31, 2013
Submitter Information A.
> Sol-Millennium Medical, Inc. 5415 Sugarloaf Parkway. Suite 2203 Lawrenceville, GA 30043
Phone Number: Fax Number:
949-433-3058 760-295-4381
Trade Name:
Sol-Care Safety Needle
Sol-Care Safety Needle
### B. Device Information
Trade/Proprietary Name:
Common name of device:
Classification Name:
Product Code:
Regulatory Class:
Classification Number:
Reason for 510(k):
Needle, Hypodermic, Single Lumen
80 EMI
ll
Hypodermic Needle
880.5570
Significant modification of device previously found equivalent (Expanded intended use to include the addition of a safety feature)
- j Predicate Device: BD Eclipse Hypodermic Needle Predicate 510(k) #: K010188 Predicate product code: FMI
16-2
{1}------------------------------------------------
#### D. Device Description
The Sol-Care Safety Needle is a sterile, single use, standard hypodermic needle with a shield to enclose the needle after use. The device is available in 18 to 30 gauge and in lengths from 3/8" to 1-1/2". In addition, the needle tip is available in a regular bevel.
The Sol-Care Safety Needle has a shield that covers the needle point after use. In the activated position, the needle shield guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination.
The Sol-Care Safety Needles are sterilized by Ethylene Oxide Gas and supplied sterile in blister pack. One hundred blister packs are packaged in a chipboard box. Each Blister pack and chipboard box is labeled with the contents and the appropriate information per the FDA's Quality System Regulation and Labeling requirements.
- E. Statement of Indications for Use
The Sol-Care Hypodermic Needle is used for general purpose injection and aspiration of fluids from vials, ampules and parts of the body below the surface of the skin.
The Sol-Care Hypodermic Needle contains a mechanism that covers the needle point after use. In the activated position, the needle cover guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination.
#### E. Comparison of Required Technological Characteristics:
Information was submitted to demonstrate that there are no significant differences in technological characteristics between the Sol-Care Safety Needle and the BD Eclipse Needles. This included:
- 1. Label Review
:,
- a. Both devices were similar in design, function and intended use.
- b. Both devices were labeled as EO Sterilized, non pyrogenic,
non toxic, single use, latex free and with a five year shelf life.
- 2. Intended Use Comparison
- a. The intended use for both devices is the same.
- 3. Material Comparison
- a. Both devices are fabricated from the same materials.
16-3
{2}------------------------------------------------
- 4. Physical, mechanical and biological specifications
- a. A comparison of physical, mechanical and biological specifications showed that the Sol-Care Safety Needles are substantially equivalent to the predicate device.
- G. Summary and Conclusion of Nonclinical and Clinical Tests:
The Sol-Care Safety Needles met the appropriate requirements contained in the following standards:
- 1. ISO 7864:1993, Sterile Hypodermic Needles for Single Use;
- 2. ISO 594-2:1998, Conical Fittings with a 6% (Luer) Taper for Syringes, Needles, and Certain Other Medical Equipment;
- 3. ISO.9626:1991, Stainless Steel Needle Tubing for Manufacture of Medical Devices;
- 4. ISO 14971:2009, Medical Devices, Application of Risk Analysis
- ISO 11135:2007, Medical Apparatus Epoxy Ethane Sterilization న. Confirmation and Routine Control
- 11607-1,-1:2006, Packaging for terminally sterilized medical devices 6.
- ISO 10993-1:2006, Biological evaluation of medical devices Part 1: 7. Evaluation and testing
- 8. ISO 23908:2007, Sharps injury protection Requirements and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling"
FDA Guidance on the content of premarket notification ರಿ. [510(k)] submissions for hypodermic single lumen needles, April, 1993.
- 10. Guidance for Industry and FDA Staff Medical Devices with Sharps Injury Prevention Features. Document Issued on: August 9, 2005
- H. Discussion of Clinical Tests:
None submitted
i.y
{3}------------------------------------------------
Conclusions Demonstrating Safety, Effectiveness and Performance:
I.
The device has been tested and found to meet all product specifications and requirements. Accelerated aging was used to verify the performance of the product over the life of the device.
Instructions for Use detail how to use the devices and the conditions of use. Product labeling clearly shows that the device is for single patient use only.
A Simulated Use Study was conducted with the Sol-Care Safety Necdle. The conclusion of the study showed that the Sol-Care Safety-Needle is safe and effective for its intended use.
11-5
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized representation of an eagle or bird-like figure with three curved lines forming its body and wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 8, 2013
Mr. Jim Barley Director of Regulatory Affairs/Quality Assurance Sol-Millennium Medical 5415 Sugarloaf Pkwy Suite 2203 Lawrenceville, GA 30043
Re: K123684
Trade/Device Name: Sol-Care Safety Needle Regulation Number: 21 CFR 880,5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: January 31, 2013 Received: February 21, 2013
Dear Mr. Barley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Barley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/5/Picture/8 description: The image contains a handwritten inscription that appears to say "for with our wits". The text is stylized and somewhat difficult to read due to the cursive handwriting and overlapping strokes. The inscription is surrounded by some geometric shapes.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications For Use
K125684 510(k) Number (if known):
Device Name: Sol-Care Safety Needle
Indications For Use:
The Sol-Care Safety Hypodermic Needle is used in conjunction with a standard syringe. This device is used for aspiration and administration of medication.
The Sol-Care Safety Hypodermic Needle safety mechanism covers the needle after use. In the activated position, the needle cover guards against accidental needle stick during normal handling and disposal of the used needle/syringe combination.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Richard C. Chapman 2013.05.07 12:31:53 -04'00' Sivision Sign-Off) ·vision of Anesthesiology, General Hospital iection Control, Dental Devices Page 1 of ﻒ ﺇ K12360 510(k) Number:
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.