K222271 · Promisemed Hangzhou Meditech Co., Ltd. · FMI · Feb 15, 2023 · General Hospital
Device Facts
Record ID
K222271
Device Name
Verisafe Safety sterile needles
Applicant
Promisemed Hangzhou Meditech Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Feb 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
It is intended for use in the aspiration and injection of fluids for medical purposes. When the safety shield is manually activated, it will cover the needle immediately after use to minimize risk of accidental needle stick.
Device Story
Verisafe Safety sterile needles are single-use, sterile hypodermic needles with an integrated manual safety shield; designed for use with syringes. The device consists of a stainless steel needle, polypropylene hub, and safety shield. Operation involves manual activation of the shield post-injection to cover the needle point, reducing accidental needle stick risk. Used in clinical settings by healthcare professionals for fluid aspiration and injection. The device provides a physical barrier to sharps injuries, enhancing safety during disposal. Bench testing confirms performance against ISO standards for hypodermic needles and sharps injury protection.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via testing against ISO 7864 (hypodermic needles), ISO 9626 (tubing), ISO 80369-7 (connectors), and ISO 23908 (sharps injury protection). Biocompatibility confirmed via ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis).
Technological Characteristics
Materials: Stainless steel (06Cr19Ni10) needle, polypropylene (PP) hub and cap. Dimensions: 18G-30G gauges, various lengths. Connectivity: None. Sterilization: Ethylene Oxide (EO) per EN ISO 11135. Standards: ISO 7864, ISO 9626, ISO 80369-7, ISO 23908, ISO 6009 (color coding).
Indications for Use
Indicated for use in the aspiration and injection of fluids for medical purposes in patients requiring hypodermic needle access.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 15, 2023
Promisemed Hangzhou Meditech Co., Ltd. Zearou Yang Regulatory Affairs Manager No. 1388 Cangxing Street. Cangqian Community Yuhang District Hangzhou City, Zhejiang 311121 China
Re: K222271
Trade/Device Name: Verisafe Safety sterile needles Regulation Number: 21 CFR 880.5570 Regulation Name: Needle, hypodermic, single lumen Regulatory Class: Class II Product Code: FMI Dated: December 14, 2022 Received: December 14, 2022
Dear Zearou Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan M.
Stevens
-S3
CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22271
Device Name
Verisafe Safety Sterile Needles
#### Indications for Use (Describe)
It is intended for use in the aspiration of fluids for medical purposes. When the safety shield is manually activated, it will cover the needle immediately after use to minimize risk of accidental needle stick.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K22271 510k Summary
#### 1. Date Prepared
February 14th, 2023
## 2. Submitter's Information
#### Name of Sponsor:
Promisemed Hangzhou Meditech Co., Ltd.
#### Address:
No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121
Zhejiang, China
Contact Name: Zearou Yang
#### Telephone No.:
+86 571 88772985
Fax No.:
+86 571 88772985
Email Address:
zearou.yang@promisemed.ca
### 3. Trade Name, Common Name, Classification
Trade/Product Name: Verisafe Safety sterile needles
Common Name: Safety sterile needles
Classification name: Needle, hypodermic, single lumen
Regulation Number: 880.5570
Device Class: Class II
Product Code: FMI
### 4. Identification of Predicate Device
K123684: Sol-Care Safety Needle
### 5. Description of the Device
The Verisafe Safety sterile needles are sterile, single use, standard hypodermic needle with a shield to enclose the needle after use.
The Verisafe Safety sterile needles are for use with syringes.
The device mainly consists of hub, needle cap and safety shield. It is packaged in a
{5}------------------------------------------------
sealed sterility barrier. This is a single-use device and delivered sterilization process is validated according to EN ISO 11135. Sterilization process undergoes routine control.
Sterilized by ETO, shelf life 5 years.
#### 6. Indication
It is intended for use in the aspiration and injection of fluids for medical purposes. When the safety shield is manually activated, it will cover the needle immediately after use to minimize risk of accidental needle stick.
#### 7. Similarities and Differences of the Proposed Devices to the Predicate Devices
Promisemed Safety sterile needles are substantially equivalent to the predicate device, Sol-Care Safety Needle, K123684 in that these devices have same intended use and technological characteristics. The basic technological and operating principles are the same for both devices. Both the subject and predicate devices are disposable, sterile, single patient use devices. The differences between the subject device and predicate device do not affect the basic design principle, usage of the subject device. A detailed comparison to the predicate is provided in Table 1.
| Items | Subject Device | Predicate Device<br>(K123684) | Comments |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Trade Name | Verisafe Safety sterile<br>needles | Sol-Care Safety Needle; | |
| Manufacturer | Promisemed Hangzhou<br>Meditech Co., Ltd | Sol-Millennium. Medical.<br>Inc. | |
| Device Class | Class II | Class II | Similar |
| Product Code | FMI | FMI | Similar |
| Regulation number | 880.5570 | 880.5570 | Similar |
| Regulation Name | Needle, hypodermic,<br>single lumen | Needle, hypodermic,<br>single lumen | Similar |
| Intended Use/<br>Indications for Use | It is intended for use in the<br>aspiration and injection of<br>fluids for medical purposes.<br>When the safety shield is<br>manually activated, it will<br>cover the needle<br>immediately after use to<br>minimize risk of accidental<br>needle stick. | It is intended for use in the<br>aspiration and injection of<br>fluids for medical purposes.<br>When the safety shield is<br>manually activated, it will<br>cover the needle<br>immediately after use to<br>minimize risk of accidental<br>needle stick. | Similar |
| Operating<br>Principle | When the safety shield is<br>manually activated, it will<br>cover the needle immediately<br>after use to minimize risk of<br>accidental needle stick. | The Sol-Care Safety Needle has<br>a shield that covers the needle<br>point after use. In the<br>activated position, the needle<br>shield guards against<br>accidental needle stick during<br>normal handling and disposal<br>of the used needle/syringe<br>combination. | Similar |
| Needle Gauge | 18G, 19G, 20G, 21G, 22G, 23G,<br>24G, 25G, 26G, 27G, 28G, 29G,<br>30G | 18G, 19G, 20G, 21G, 22G, 23G,<br>24G, 25G, 26G, 27G, 28G, 29G,<br>30G | Similar |
{6}------------------------------------------------
| Needle Length | 1/2in, 3/4in, 5/8in, 1in, 1 1/4in,<br>1 1/2in, 2 in | 3/8 in to 1 1/2 in | Different, Subject<br>devices have more<br>needle lengths than<br>predicate device. |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Needle hub<br>color | Per ISO 6009:<br>Pink/18G, Cream/19G,<br>Yellow/20G, Light green/21G,<br>Black/22G, Light blue/23G,<br>Medium purple /24G,<br>Orange /25G, Brown /26G,<br>Medium grey /27G,<br>Blue - green /28G, Red /29G,<br>Yellow /30G. | Per ISO 6009:<br>Pink/18G, Cream/19G,<br>Yellow/20G, Light green/21G,<br>Black/22G, Light blue/23G,<br>Medium purple /24G,<br>Orange /25G, Brown /26G,<br>Medium grey /27G,<br>Blue - green /28G, Red /29G,<br>Yellow /30G. | Each color is in<br>accordance with ISO<br>6009. |
| Reuse durability | Single Use | Single Use | Similar |
| Biocompatibility | Complied with ISO10993 series<br>standards, and the following<br>tests are performed<br>- Cytotoxicity: No cytotoxicity<br>- Intracutaneous Reactivity<br>No evidence of irritation<br>- Skin Sensitization: No<br>evidence of sensitization<br>- Acute Systemic Toxicity: No<br>systemic toxicity<br>- Pyrogen: Non pyrogenic<br>- Platelet Count: No effect<br>- Hemolysis: No effect | Per ISO 10993-1 | Similar |
| Materials | - Needle: Stainless Steel<br>(06Cr19Ni10)<br>- Hub: Polypropylene (PP)<br>- Need Cap: Polypropylene<br>(PP) | - Needle: Stainless Steel<br>(06Cr19Ni10)<br>- Hub: Polypropylene (PP)<br>- Need Cap: Polypropylene<br>(PP) | Similar |
| Performance | Complied with ISO 7864, ISO<br>9626, ISO 80369-7 | Complied with ISO 7864, ISO<br>9626, ISO 80369-7 | Similar |
| Sterilization<br>method | EO Sterilization | EO Sterilization | Similar |
#### Performance Testing Summary 8.
The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device.
These tests and standard are provided in Table 2 and Table 3.
### • Table 2 Performances
| No. | Test item | Reference recognized standards |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | 1.1 Appearance | EN ISO 7864:2016, Sterile hypodermic syringes for single use -<br>Requirements and test methods |
| | 1.2 Cleanliness | |
| 2. | Color code | EN ISO 6009:2016, Hypodermic needles for single use - Colour<br>coding for identification |
| 3. | 3.1 Dimensions | EN ISO 7864:2016, Sterile hypodermic syringes for single use -<br>Requirements and test methods |
| | | EN ISO 9626:2016, Stainless steel needle tubing for the<br>manufacture of medical devices — Requirements and test methods |
| | 3.2 Length of the<br>needle | EN ISO 7864:2016, Sterile hypodermic syringes for single use –<br>Requirements and test methods |
| 4. | Lubricant | EN ISO 7864:2016, Sterile hypodermic syringes for single use -<br>Requirements and test methods |
| 5. | Needle point | EN ISO 7864:2016, Sterile hypodermic syringes for single use -<br>Requirements and test methods |
| 6. | Penetration force | EN ISO 7864:2016, Sterile hypodermic syringes for single use -<br>Requirements and test methods |
| 7. | Bond between<br>needle and hub | EN ISO 7864:2016, Sterile hypodermic syringes for single use -<br>Requirements and test methods |
| 8. | Patency of lumen | EN ISO 7864:2016, Sterile hypodermic syringes for single use –<br>Requirements and test methods |
| 9. | Needle cap | / |
| 10. | 10.1 Stiffness | EN ISO 9626:2016, Stainless steel needle tubing for the<br>manufacture of medical devices - Requirements and test methods |
| | 10.2 Resistance to<br>breakage | EN ISO 9626:2016, Stainless steel needle tubing for the<br>manufacture of medical devices - Requirements and test methods |
| | 10.3 Resistance to<br>corrosion | EN ISO 9626:2016, Stainless steel needle tubing for the<br>manufacture of medical devices — Requirements and test methods |
| 11. | 11.1<br>Safety feature<br>(Activate force) | EN ISO 23908:2013, Sharps injury protection- Requirements and<br>test methods — Sharps protection features for single-use<br>hypodermic needles, introducers for catheters and needles used for<br>blood sampling |
| | 11.2<br>Safety feature<br>(Reuse<br>Prevention) | EN ISO 23908:2013, Sharps injury protection- Requirements and<br>test methods — Sharps protection features for single-use<br>hypodermic needles, introducers for catheters and needles used for<br>blood sampling |
| | 11.3<br>Safety feature<br>(Safety<br>Performance) | EN ISO 23908:2013, Sharps injury protection- Requirements and<br>test methods — Sharps protection features for single-use<br>hypodermic needles, introducers for catheters and needles used for<br>blood sampling |
| 12. | 12.1 Dimension | ISO 80369-7:2021, Small-bore connectors for liquids and gases in<br>healthcare applications - Part 7: Connectors for intravascular or<br>hypodermic applications |
| 12.2 Positive<br>pressure liquid<br>leakage | ISO 80369-7:2021, Small-bore connectors for liquids and gases in<br>healthcare applications - Part 7: Connectors for intravascular or<br>hypodermic applications | |
| 12.3 Sub<br>atmospheric<br>pressure air leakage | ISO 80369-7:2021, Small-bore connectors for liquids and gases in<br>healthcare applications - Part 7: Connectors for intravascular or<br>hypodermic applications | |
| 12.4 Stress cracking | ISO 80369-7:2021, Small-bore connectors for liquids and gases in<br>healthcare applications - Part 7: Connectors for intravascular or<br>hypodermic applications | |
| 12.5 Resistance to<br>separation from<br>axial load | ISO 80369-7:2021, Small-bore connectors for liquids and gases in<br>healthcare applications - Part 7: Connectors for intravascular or<br>hypodermic applications | |
| 12.6 Resistance to<br>separation from<br>unscrewing | ISO 80369-7:2021, Small-bore connectors for liquids and gases in<br>healthcare applications - Part 7: Connectors for intravascular or<br>hypodermic applications | |
| 12.7 Resistance to<br>overriding | ISO 80369-7:2021, Small-bore connectors for liquids and gases in<br>healthcare applications - Part 7: Connectors for intravascular or<br>hypodermic applications | |
| 13. | Particulate<br>contamination | USP 39 $<788>$ Particulate matter in injections |
| 14. | Acidity and<br>alkalinity | EN ISO 7864:2016, Sterile hypodermic syringes for single use –<br>Requirements and test methods |
| 15. | Extractable metals | EN ISO 7864:2016, Sterile hypodermic syringes for single use –<br>Requirements and test methods |
| 16. | Package<br>appearance | ASTM F1886/F1886M – 16 Standard Test Method for Determining<br>Integrity of Seals for Flexible Packaging by Visual Inspection |
| 17. | Seal strength | EN 868-5:2018 Packaging for terminally sterilized medical devices<br>Part 5: Sealable pouches and reels of porous materials and plastic<br>film construction - Requirements and test methods |
| 18. | Dye penetration | ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in<br>Porous Medical Packaging by Dye Penetration |
| 19. | Sterility | USP 39 $<71>$ Sterility tests |
| 20. | Bacterial endotoxin | USP 40 $<85>$ Bacterial endotoxins test |
{7}------------------------------------------------
{8}------------------------------------------------
# ・ Table 3 Biocompatibility
| No. | Test item | Reference recognized standards |
|-----|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| 1. | In Vitro Cytotoxicity - MTT Method,<br>MEM with 10% FBS extract | ISO 10993-5:2009 – Biological Evaluation of<br>Medical Devices -- Part 5: Tests for in Vitro<br>Cytotoxicity |
| 2. | Skin Sensitization – Guinea Pig<br>Maximization, 0.9% sodium chloride<br>extract and sesame oil extract | ISO 10993-10:2021, Biological Evaluation of<br>Medical Devices -- Part 10: Tests for Irritation and<br>Skin Sensitization |
{9}------------------------------------------------
| 3. | Intradermal reaction test,<br>0.9% sodium chloride extract and<br>sesame oil extract | ISO 10993-23:2021, Biological evaluation of<br>medical devices - Part 23: Tests for irritation |
|----|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 4. | Acute systemic toxicity<br>0.9% sodium chloride extract and<br>sesame oil extract | ISO 10993-11:2017, Biological evaluation of<br>medical devices - Part 11: Tests for systemic<br>toxicity- Acute systemic toxicity and pyrogen test |
| 5. | Pyrogen study in rabbits<br>0.9% sodium chloride injection | ISO 10993-11:2017, Biological evaluation of<br>medical devices - Part 11: Tests for systemic<br>toxicity- Acute systemic toxicity and pyrogen test<br>USP43-NF37<151> Pyrogen Test |
| 6. | In Vitro Hemolytic Properties Test | ASTM F756-17 Standard Practice for Assessment<br>of Materials |
| 7. | Ethylene Oxide residues EO ECH Test | ISO 10993-7:2008, Biological evaluation of<br>medical devices — Part 7: Ethylene oxide<br>sterilization residuals |
#### Conclusion 9.
The comparison of intended use, technological characteristics, and performance data of the subject device has demonstrated substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.