EXELINT SecureTouch Safety Hypodermic Needle

K152183 · Exelint International, Co. · FMI · Nov 2, 2015 · General Hospital

Device Facts

Record IDK152183
Device NameEXELINT SecureTouch Safety Hypodermic Needle
ApplicantExelint International, Co.
Product CodeFMI · General Hospital
Decision DateNov 2, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2
AttributesTherapeutic

Intended Use

EXELINT SecureTouch Safety Hypodermic Needle is intended for use with syringes for general purpose aspiration and medication administration. It has an antistick mechanism to cover the needle after use.

Device Story

Sterile, single-use, luer lock compatible hypodermic needle; features integrated protection shield to enclose needle post-use. Available in 18-30 gauge, 10mm-50mm lengths. Operated manually by clinicians to aspirate or administer medication; safety mechanism activated post-injection to guard against accidental needle sticks during handling and disposal. Benefits include reduced risk of sharps injury to healthcare personnel.

Clinical Evidence

No clinical data; bench testing only. Performance verified via ISO 7864, ISO 9626, and ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis). Package integrity validated per ASTM F88 and F1140. Mechanical testing included activation/detachment force, bond strength, and patency.

Technological Characteristics

Materials: Stainless steel cannula, polypropylene hub, shield, and cap. Dimensions: 18-30G, 10-50mm length. Energy: Manual. Connectivity: None. Sterilization: Ethylene Oxide (EO). Standards: ISO 7864, ISO 9626, ISO 10993, ASTM F756, ASTM F88, ASTM F1140, USP <85>, USP <151>.

Indications for Use

Indicated for use with syringes for general purpose aspiration and medication administration in patients requiring hypodermic injections.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with a wing-like shape above them. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 2, 2015 EXELINT International Co. Mr. Navid Hamid Senior Regulatory Affairs 2500 Santa Fe Avenue Redondo Beach, California 90278 Re: K152183 Trade/Device Name: EXELINT SecureTouch Safety Hypodermic Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: August 4, 2015 Received: August 5, 2015 Dear Mr. Hamid: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152183 Device Name EXELINT SecureTouch Safety Hypodermic Needle Indications for Use (Describe) EXELINT SecureTouch Safety Hypodernic Needle is intended for use with syringes for general purpose aspiration and medication administration. It has an antistick mechanism to cover the needle after use. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary - 1. Date of Preparation: October 15, 2015 - 2. Sponsor Identification ### EXELINT International, Co. 2500 Santa Fe Avenue Redondo Beach, CA 90278 Establishment Registration Number: 1035907 Contact Person: Navid Hamid Position: Manager Tel: 310-649-0707 Fax: 310-649-1168 Email: navid(@exelmed.com - 3. Identification of Proposed Device Trade Name: EXELINT SecureTouch Safety Hypodermic Needle Common Name: Safety Hypodermic Needle Regulatory Information: Classification Name: Hypodermic single lumen needle Classification: 2 Product Code: FMI Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital Intended Use Statement: EXELINT SecureTouch Safety Hypodermic Needle is intended for use with syringes for general purpose aspiration and medication administration. It has an antistick mechanism to cover the needle after use. #### Device Description: EXELINT SecureTouch Safety Hypodermic Needle is a sterile, single use, standard luer lock compatible hypodermic needle with a protection shield to enclose the needle after use The device is available in 18 to 30 gauge in length from 10mm to 50mm. The purpose of this submission is to enable Exelint International, Co. to market a line of safety hypodermic needle. The proposed device is available in EO sterilized sealed in a sterility maintenance package. - 4. Identification of Predicate Device(s) 510(k) Number: K123684 Product Name: Sol-Care Safety Needle - న్. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - ISO 7864 Third Edition 1993-05-15, Sterile Hypodermic Needles For Single Use. A - A ISO 9626 First Edition 1991-09-01, Stainless Steel Needle Tubing For The Manufacture Of Medical Devices [Including: Amendment 1 (2001)] {4}------------------------------------------------ - > ISO 10993-5:2009/(R) 2014, Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity. - > ISO 10993-7:2008(R) 2012, Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals. - > ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization. - > ISO 10993-11:2006/(R) 2010, Biological Evaluation Of Medical Devices -- Part 11: Tests For Systemic Toxicity. - > ASTM F756-13, Standard Practice for Assessment of Hemolytic Properties of Materials. - > ASTM F88/F88M-09, Standard Test Method for Seal Strength of Flexible Barrier Materials. - A ASTM F1140/F1140M-13, Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages. - > USP <85> Bacterial Endotoxins Test. - A USP <151> Pyrogen Test. The performance tests performed on the proposed device are as follows: - Test for Tolerance on Length i. - ii. Test for Lubricant - Test for Bond between Hub and Needle Tube iii. - Test for Patency Lumen iv. - V. Dimension Test - vi. Stiffness Test - vii. Test for Resistance of Tubing to Breakage - viii. Test for Resistance to Corrosion - ix. Testing of Force to Activate the Safety Feature - Testing of Force to Detach the Safety Feature X. - 6. Clinical Test Conclusion No clinical study is included in this submission. - 7. Substantially Equivalent (SE) Comparison | Item | Proposed Devices | Predicate Devices | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Code | FMI | FMI | | Regulation<br>Number | 21 CFR 880.5570 | 21 CFR 880.5570 | | Class | 2 | 2 | | Intended Use | EXELINT SecureTouch Safety<br>Hypodermic Needle is intended<br>for use with syringes for<br>general purpose aspiration and<br>medication administration. It<br>has an antistick mechanism to<br>cover the needle after use. The<br>safety mechanism covers the<br>needle after use. In the<br>activated position, the needle<br>cover guards against accidental<br>needle stick during normal<br>handling and disposal of the<br>used needle/syringe<br>combination. | The Sol-Care Safety Hypodermic<br>Needle is used in conjunction with a<br>standard syringe. This device is used<br>for aspiration and administration of<br>medication. The Sol-Care Safety<br>Hypodermic Needle safety<br>mechanism covers the needle after<br>use. In the activated position, the<br>needle cover guards against<br>accidental needle stick during normal<br>handling and disposal of the used<br>needle/syringe combination. | Table 1 Comparison of Technology Characteristics {5}------------------------------------------------ | Configuration | Hub, Needle Cannula, Needle<br>Cap, Protection Shield | Hub, Needle Cannula, Needle Cap,<br>Protection Shield | |--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------| | Force to<br>activate the<br>safety feature | ≤4N | 30G~28G<4N;<br>27G~21G<5.5N;<br>20G~18G<7.5N | | Force to detach<br>the safety<br>feature | ≥30N | ≥30N | | Operation<br>Mode | Manual | Manual | | Single Use | Yes | Yes | | Needle Gauge | 18G, 19G, 20G, 21G, 22G,<br>23G, 24G, 25G, 26G, 27G,<br>28G, 29G, 30G | 18G, 19G, 20G, 21G, 22G, 23G, 24G,<br>25G, 26G, 27G, 28G, 29G, 30G | | Performance<br>specification | Comply with ISO 7864, ISO<br>9626 | Comply with ISO 7864, ISO 9626 | | Material | Needle cannula: Stainless Steel | Needle cannula: Stainless Steel | | | Needle hub and protection<br>shield: Polypropylene | Needle hub and protection shield:<br>Polypropylene | | | Needle cap: Polypropylene | Needle cap: Polypropylene | | Biocompatibilit<br>y | Conforms to ISO 10993<br>Tests: Cytotoxicity, Skin<br>Sensitization, Intracutaneous<br>Irritation, Systemic Toxicity, In<br>Vitro Hemolysis, Pyrogen | Conforms to ISO 10993 | #### Substantially Equivalent (SE) Conclusion 8. Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...