K103587 · Weihai Hongyu Medical Devices Co., Ltd. · FMI · Apr 6, 2011 · General Hospital
Device Facts
Record ID
K103587
Device Name
DISPOSABLE BLLOD COLLECTION NEEDLE
Applicant
Weihai Hongyu Medical Devices Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Apr 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
Disposable Blood Collection Needles is designed for use in the daily blood collection routine when delegated by a qualified practitioner.
Device Story
Disposable blood collection needle; sterile, single-use device. Components: puncturing needle for vein access, bottle needle for blood collection tube insertion, needle holder, protective needle caps, and latex cover for bottle needle. Used in clinical settings for routine blood draws. Operated by qualified practitioners or delegated personnel. Device facilitates venous blood collection into vacuum tubes. Benefits include standardized, sterile blood sampling.
Clinical Evidence
No clinical data. Bench testing only. Device validated against ISO 7864, ISO 9626, and ISO 10993 series (biological evaluation, blood interaction, cytotoxicity, irritation, sensitization, and systemic toxicity).
Technological Characteristics
Stainless steel needle tubing (ISO 9626:1991). Sterile, single-use. Available in 18G, 20G, 21G, 22G, 23G sizes. Includes latex cover. Complies with ISO 7864:1993 and ISO 10993 series for biocompatibility.
Indications for Use
Indicated for routine blood collection procedures in patients when performed by or delegated by a qualified practitioner.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
APR - 6 2011
ﯿﺴﮧ
510(k) Report for Disposable Blood Collection Needle- Section 111 510(k) Summary
# Section III. 510(k) Summary
### Disposable Blood Collection Needle
Weihai Hongyu Medical Devices Co., Ltd.
(As required by 21 CFR 807.92)
#### 1. Date Prepared: Nov.18, 2010
### 2. Sponsor Information
Weihai Hongyu Medical Devices Co., Ltd. No.86, Shenzhen Road, Weihai City, China. 264205 Contract Person: YanPing. Li Phone: +86-631-3636943 Fax: +86-631-3636910 Email: whlyp1979@vip.sina.com
### 3. Submission Correspondent
Ms. Diana Hong Mr. Tarzan Wang Shanghai Mid-Link Business Consulting Co., Ltd P.O. BOX 237-023 Shanghai, 200237, China Tel: +86-21-64264467 Fax:+86-21-64264468 Email: Diana.hong(@mid-link.net
{1}------------------------------------------------
# 4. Proposed Device Information
Device Common or Usual name: Disposable Blood Collection Needle; Device Trade or Proprietary Name: Blood Collection Needle Classification Name: Needle, Hypodermic, Single Lumen; Regulatory Class: II Regulation number: 21CFR 880.5570; Product Code: FMI; Panel: General Hospital Model: 18G, 20G, 21G, 22G, 23G
#### 5. Predicate Device
Blood Collection Needle (K073127)
## 6. Device Description
Blood collection needle is a sterile and disposable medical device. It is consist of a puncturing needle which is punctured into vein, a bottle needle which is insert to the blood collection bottle to collect blood, a needle holder which connects the puncturing needle and bottle needle, two needle caps which protect puncturing needle and bottle needle and a latex cover to protect bottle needle.
#### 7. Intended use
Disposable Blood Collection Needles is designed for use in the daily blood collection routine when delegated by a qualified practitioner.
#### 8. Substantial Equivalence
The Disposable Blood Collection Needles shares the similar indications for use, design features, functional features, same safety compliance. Therefore the proposed device is substantially equivalent (SE) to the predicate devices.
# 9. Testing
{2}------------------------------------------------
The Disposable Blood Collection Needle is designed, tested and will be manufactured in accordance with both mandatory and voluntary standards, including:
ISO 7864:1993 Sterile hypodermic needles for single use;
ISO 9626:1991 Stainless steel needle tubing for the manufacture of medical devices/ Amendment:2001
ISO 10993, Biological Evaluation of Medical Devices; Including:
ISO 10993-1:2003, Biological Evaluation of Medical Devices - Part 1: Guidance on selection of tests:
ISO 10993-4:2002, Biological Evaluation of Medical Devices – Part 4: Selection of test for interactions with Blood
ISO 10993-5:1999, Biological Evaluation of Medical Devices - Part 5: Test for in vitro cytotoxicity;
ISO 10993-10:2002, Biological Evaluation of Medical Devices - Part 10: Tests for irritation and delayed-type hypersensitivity;
ISO 10993-11:2006, Biological Evaluation of Medical Devices - Part 11: Tests for systemic toxicity;
ISO 10993-12:2002, Biological Evaluation of Medical Devices – Part 12: Sample preparation and reference materials
Non-Clinical Conclusion:
Laboratory testing was conducted to validate and verify that the proposed device, Disposable Blood Collection Needle met all design specifications and was substantially equivalent to the predicate device. No Clinical Information is required
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66~G609 Silver Spring, MD 20993-0002
Weihai Hongyu Medical Devices Company, Limited C/O Ms. Diana Hong Shanghai Mid-Link Business Consulting Company, Limited P.O. Box 237-023 Shanghai CHINA 264200
APR - 6 2011
Re: K103587
Trade/Device Name: Disposable Blood Collection Needle 18G, 20G, 21G, 23G, 23G Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: March 18, 2011 Received: March 21, 2011
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any l'ederal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.ida.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jams I Desotud
for 7
Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
# Section II. Indications for Use Statement
510(k) Number: Device Name:
Disposable Blood Collection Needle 18G, 20G, 21G, 22G, 23G
Indications for Use:
Disposable Blood Collection Needles is designed for use in the daily blood collection routine when delegated by a qualified practitioner.
Prescription Use ____X AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ald C. Chiga 4/6/11
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control and Dental Devices
510(k) Number: K10358
Page 1 of 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.