Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd.
Product Code
JKA · Clinical Chemistry
Decision Date
Mar 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Intended Use
The Blood Collection Needle (with/without Needle Holder) is intended to be used with a vacuum blood collection tube for the collection of venous blood.
Device Story
Blood collection needle used in routine venipuncture; forms channel between patient vein and evacuated blood collection tube. Device consists of stainless steel needle tube, hub, rubber cap, and optional needle holder. Available in gauges 25G-18G and lengths 3/4"-1 1/2". Operated by trained healthcare professionals in clinical settings. Needle inserted into vein; vacuum tube attached to holder end to facilitate blood withdrawal. Single-use, sterile, disposable device. Benefits include standardized, safe venous blood collection for diagnostic testing.
Clinical Evidence
No clinical data was submitted. Substantial equivalence is supported by non-clinical bench testing, including performance testing (ISO 9626, ISO 7864, ISO 80369-7/20, ISO 6009), biocompatibility (ISO 10993 series), and sterilization/packaging integrity validation.
Technological Characteristics
Materials: Stainless steel (needle), PP (bottom sheath/holder), PE (upper sheath), ABS (hub), Polydimethysiloxane (lubricant). Sensing/Actuation: Mechanical venipuncture. Energy: None. Connectivity: None. Sterilization: Ethylene Oxide (EO). Standards: ISO 9626, ISO 7864, ISO 80369-7/20, ISO 10993-1, ISO 11135, ASTM F1929, ASTM F1980, ASTM D3078, ASTM F88/F88M.
Indications for Use
Indicated for the collection of venous blood in patients requiring blood sampling via vacuum blood collection tubes. For use by trained healthcare professionals only.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 24, 2023
Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd. % Charles Mack Principal Engineer Irc 2950 E Lindrick Drive Chandler, Arizona 85249
Re: K223323
Trade/Device Name: Blood Collection Needle (with/without Needle Holder) Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA, FMI Dated: February 21, 2023 Received: February 22, 2023
Dear Charles Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K223323
#### Device Name
Blood Collection Needle (with/without Needle Holder)
Indications for Use (Describe)
The Blood Collection Needle (with/without needle holder) is intended to be used with a vacuum blood collection tube for the collection of venous blood.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains a logo and text. The logo is on the left and consists of the letters 'i', 'w', and 'm' in red and blue. To the right of the logo is the text "安徽宏宇五洲医疗器械股份有限公司" in Chinese, followed by "ANHUI HONGYU WUZHOU MEDICAL MANUFACTURER CO., LTD" in English. The text likely represents the name of a company.
# K223323 - 510(k) SUMMARY
| Preparation Date: | March 24, 2023 |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer's Name and Address: | Anhui Hongyu Wuzhou Medical<br>Manufacturer Co., Ltd.<br>No. 2 Guanyin Road, Economic<br>Development Zone, Taihu, Anqing City,<br>Anhui Province, China 246400<br>Tel: +86- 556 5129666 |
| Corresponding Official: | Charles Mack |
| Telephone Number: | 931-625-4938 |
| Email Address: | charliemack@irc-us.com |
| Trade Name: | Blood Collection Needle<br>(with/without Needle Holder) |
| Common Name: | Blood Collection Tubes, Vials,<br>Systems, Serum Separators |
| Regulation Name: | Blood specimen collection device |
| Regulation Number: | 21CFR862.1675 |
| Primary Product Code:<br>Secondary Product Code: | JKA<br>FMI |
| Device Class: | Class II |
| Predicate Device:<br>510K Number:<br>Product Name: | K200027<br>Blood Collection Needle with/without<br>Holder |
| Regulation Number(s):<br>Product Code: | 21CFR862.1675<br>JKA |
{4}------------------------------------------------
# Device Description:
The subject devices are blood collection devices used in routine venipuncture procedures, which form a channel between the patient's vein and the evacuated blood collection tube intended to collect blood.
The Blood Collection Needle (with/without Needle Holder) is intended for single use only and consists of a bottom sheath or needle holder, needle tube, rubber cap, upper sheath, and hub. They are available in different configurations.
The product is to be used by appropriately trained healthcare professionals only in accordance with these instructions.
It is manufactured from tubular stainless steel sharpened at both ends and is attached to the hub:
- -The subject devices are available with a pre-attached needle holder or without a needle holder; the hub is threaded on one side to connect with the needle holder, which is used to guide the needle into an evacuated blood collection tube. This needle's end is shorter and fitted with a protective rubber cap and a needle holder.
- The opposite end of the needle is longer for withdrawing blood and is fitted with a needle sheath.
The subject devices are available in different specifications of needle gauge (25G-18G) and length (3/4"-1 1/2").
The proposed devices are only packaged as sterile, disposable, and single-patient use.
The devices are packaged for sterile, single-use, and single-patient use only.
#### Indications for Use
The Blood Collection Needle (with/without Needle Holder) is intended to be used with a vacuum blood collection tube for the collection of venous blood.
{5}------------------------------------------------
| Feature | Subject Device | Predicate Device | Remark |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| FDA510(K) | Pending | K200027 | N/A |
| Device Name | Blood Collection Needle (with/without<br>Needle Holder) | Blood Collection Needle with/without<br>Holder;<br>Safety Blood Collection Needle<br>with/without Holder;<br>Luer Access Device-holder with<br>Preattached Multiple Sample Adapter | Note 1 |
| Product code | JKA | JKA | Identical |
| Subsequent<br>product code: | FMI | FMI | Identical |
| Indication<br>for<br>Use | The Blood Collection Needle<br>(with/without Needle Holder) is intended<br>to be used with a vacuum blood<br>collection tube for the collection of<br>venous blood. | The Blood Collection Needle<br>with/without holder is intended to be<br>used with vacuum blood collection tube<br>for the collection of venous blood.<br>The Safety Blood Collection Needle<br>with/without holder is intended to be<br>used with vacuum blood collection tube<br>for the collection of venous blood. The<br>safety shield is intended to aid in the<br>protection against accidental needle<br>stick injury.<br>The Luer access device- holder with<br>preattached multiple sample adapter is<br>a sterile, non-invasive device used to<br>connect devices with male or female<br>luer connectors to blood collection tubes<br>for the collection of blood. | Note 1 |
| Needle gauge | 25G, 24G, 23G, 22G, 21G, 20G, 19G,<br>18G | 27G, 25G, 23G, 22G, 21G, 20G, 18G | Note 2 |
| Needle Length | 3/4", 1", 1 1/8", 1 1/4",1 2/5", 1 1/2" | 1", 1 1/4", 1 1/2" | Note 3 |
| Needle wall<br>thickness | Regular Wall and Thin Wall | Regular Wall and Thin Wall | Identical |
{6}------------------------------------------------
| Feature | Subject Device | Predicate Device | Remark |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Configuration<br>and material | Bottom Sheath (PP)<br>Upper Sheath (PE)<br>Non- patient Needle Tube (Stainless<br>Steel)<br>Patient Needle Tube (Stainless Steel)<br>Rubber Sleeve<br>Needle Hub (ABS)<br>Lubricant (Polydimethysiloxane)<br>Holder (PP) | Non-patient Needle Cap(PP)<br>Patient Needle Cap (PP)<br>Non- patient Needle Tube (Stainless<br>Steel)<br>Patient Needle Tube (Stainless Steel)<br>Rubber Sleeve (Case Gather<br>Isoprene Rubber)<br>Needle Hub (PP or MABS)<br>Lubricant (Polydimethysiloxane)<br>Holder (PP) | Note 4 |
| Biocompatibility | Complies with ISO10993-1<br>In Vitro Cytotoxicity<br>Skin Sensitization<br>Intracutaneous Reactivity<br>Acute Systemic Toxicity | Complies with ISO10993-1<br>In Vitro Cytotoxicity<br>Skin Sensitization<br>Intracutaneous Reactivity<br>Acute Systemic Toxicity | Identical |
| | Hemocompatibility | Hemocompatibility | |
| Performance | Complies with<br>ISO 7864<br>ISO 9626<br>ISO 11135:2014<br>ISO 80369-7<br>ISO 80369-20<br>USP788 | Complies with<br>ISO 7864<br>ISO 9626<br>ISO 11135:2014<br>ISO 80369-7<br>ISO 80369-20<br>USP788 | Identical |
| Label/labeling | Conform with 21 CFR Part 801 | Conform with 21 CFR Part 801 | Identical |
| Sterilization | EO | EO | Identical |
| Sterile | Yes | Yes | Identical |
| SAL | 10-6 | 10-6 | Identical |
| For Single use | Yes | Yes | Identical |
| Single Patient<br>Use | Yes | Yes | Identical |
#### Note 1: Predicate Device and IFU statement
The predicate device included three devices; We only chose one "Blood Collection Needle with/without Needle Holder" as the predicate device;
The subject device's Indication for Use is precisely the same as the selected predicate device "Blood Collection Needle with/without Needle Holder" in K200027; it doesn't raise new questions on the safety and effectiveness of the subject device.
#### Note 2: Needle Gauge
The needle gauge of the subject devices differs from the predicate device, but they still conform to the same applicable performance standards of ISO 9626. Therefore, the difference does not raise new questions about the safety and effectiveness of the proposed device.
{7}------------------------------------------------
#### Note 3: Needle Length
The needle length of the subject devices is available in more sizes than the predicate device, but they conform to the same applicable performance standards. Therefore, the difference does not raise new questions about the safety and effectiveness of the proposed device.
#### Note 4: Material
Although the materials of these components of subject devices are different from predicate devices, they conform to the same ISO10993-1 biocompatibility standards;
{8}------------------------------------------------
#### Performance Testing
To establish substantial equivalence to the identified predicate devices, we performed the tests noted below on the subject devices. The testing results proved that the device complies with the applicable standards requirement and is substantially equivalent to the predicate devices.
#### Non-Clinical Performance Testing
Testing was performed to evaluate the functional performance and safety of the subject device with the following standards:
#### Performance:
- ISO 9626 Second edition 2016-08-01 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- ISO 7864 Fourth edition 2016-08-01 Sterile hypodermic needles for single use --Requirements and test methods
- ISO 80369-7 First edition 2016-10-15 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
- -ISO 80369-20 First edition 2015-05-15, small-bore connectors for liquids and gases in healthcare applications part 20: standard test methods. (General I (QS/RM))
- ISO 6009 Fourth edition 2016-08-01 Hypodermic needles for single use Colour coding for identification
{9}------------------------------------------------
### Biocompatibility
The subject device is classified as an external communication device → Circulating Blood -> Limited Contact Duration (≤24h).
- -ISO 10993-5: 2009 In Vitro Cytotoxicity
- -ISO 10993-10: 2010 Skin Sensitization
- ISO 10993-10: 2010 Intracutaneous Reactivity -
- ISO 10993-11: 2017 Acute Systemic Toxicity -
- ISO 10993-4: 2017 & ASTM F756-17: Hemolytic Properties -
- ISO 10993-4: 2017 Complement activity -
- ISO 10993-4:2017 Coagulation PPT Test -
- ISO10993-11: 2017 USP 43-NF38 <151> Pyrogen Test -
### Sterilization Validation:
- । ISO11135-1: Sterilization of health care products - ethylene oxide - part 1: requirements for the development, validation, and routine control of a sterilization process for medical devices.
- ISO11737-1: Sterilization of medical devices-Microbiological methods-Part -1: Determination of the population of microorganisms on the product.
- ISO11737-2: Sterilization of medical devices -- Microbiological methods ---Part 2: Tests of sterility performed in the validation of a sterilization process.
- ISO 10993-7: Biological evaluation of medical devices Part 7: Test of -Ethylene Oxide Residuals.
- AAMI / ANSI ST72: Bacterial endotoxins Test methods, routine monitoring, and alternatives to batch testing.
## Package and Shelf Life:
### Applicable Standards:
- AAMI/ANSI/ISO 11137-1:2006/(R) 2010 Sterilization of health care products -- Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices
- AAMI/ANSI/ISO 11737-2:2009 Sterilization of medical devices --Microbiological methods - Part 2: Tests of sterility performed in the definition, validation, and maintenance of a sterilization process
- AAMI/ANSI/ISO 11607-1:2006/(R) 2010 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems, and packaging systems, 3ed.
{10}------------------------------------------------
- ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous -Medical Packaging by Dye Penetration
- ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier -Systems for Medical Devices
- ASTM D3078-02 (2021), Standard Test Method for Determination of Leaks in Flexible Packaging by Bubble Emission. (Sterility)
- DIN58953-6: 2016 Sterilization Sterile supply Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilized.
- ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible -Barrier Materials
- USP 41 <1207> Package Integrity Evaluation Sterile Products; <1207.1> -Package Integrity Testing in the Product Life Cycle - Test Method Selection and Validation; <1207.2> Package Integrity Leak Test Technologies.
# The package integrity test noted below was conducted for the sterilization package of the subject devices by accelerated aging testing followed by simulated shipping distribution testing:
- Accelerated Aging Test l
- Simulated shipping distribution testing -
- Visual inspection -
- Performance Inspection (Chemical performance and Physical performance) -
- Sterile Test -
- Vacuum Leak Test -
- Dye penetration test -
- Agar Contact-Attack Test -
- Tensile Seal Strength Test -
- -Container Closure Integrity Test
# Clinical Data
No clinical data was submitted in this submission.
{11}------------------------------------------------
#### Conclusions:
Based on the verification test results, the subject devices conform to the exact standards requirements, such as performance and biocompatibility as the predicate devices. The subject device uses the same fundamental scientific technology, the same indications for use, sterilization methods, and the same shelf life and packaging. The different technological characteristics do not raise new or different questions about the safety and effectiveness of the proposed device. END
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.