BLOOD COLLECTION NEEDLE, MODEL 20G 21G 22G

K073127 · Shandong Weigao Group Medical Polymer Co., Ltd. · FMI · Dec 20, 2007 · General Hospital

Device Facts

Record IDK073127
Device NameBLOOD COLLECTION NEEDLE, MODEL 20G 21G 22G
ApplicantShandong Weigao Group Medical Polymer Co., Ltd.
Product CodeFMI · General Hospital
Decision DateDec 20, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2

Indications for Use

Blood Collection Needles is designed for use in the daily blood collection routine when delegated by a qualified practitioner.

Device Story

Sterile, disposable blood collection needle; consists of puncturing needle for vein access, bottle needle for collection tube insertion, needle holder, protective caps, and latex cover. Used by healthcare professionals for routine blood collection. Device facilitates venous blood sampling; output is collected blood sample. Benefits include safe, standardized blood collection.

Clinical Evidence

No clinical data. Bench testing only. Performance evaluated against ISO 7864:1993 and ISO 9626:1991/AMD: 2001. Biocompatibility testing conducted per ISO 10993-1:2003(E) including cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, and haemo-compatibility.

Technological Characteristics

Sterile, disposable hypodermic single lumen needle. Available in 20G, 21G, 22G. Materials comply with ISO 10993 biocompatibility requirements. Performance standards: ISO 7864:1993 (needles) and ISO 9626:1991/AMD: 2001 (tubing).

Indications for Use

Indicated for use in routine blood collection procedures when delegated by a qualified practitioner.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 9807.92. The Assigned 510(k) Number is: ### 1. Applicant Device Information Trade/Proprietary Name: Blood Collection Needle Classification Information: DEC 2 0 2007 ### (1) Classification Name: Needle, Hypodermic, Single Lumen (2) Regulation Number: 880.5570 (3) Product Code: FMI (4) Class: II (5) Review Panel: General Hospital ### 2. Submitter Information #### Manufacturer Name: ShanDong WeiGao Group Medical Polymer Products Co., LTD No.312, Shichang Road Weihai, Shandong, China, 264209 #### Contact Person of the Submission: Ms. Diana. Hong Mr. Eric. Chen Suite 8D, Zhongxin Zhongshan Mansion, No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, China 20030 Phone: +86-21-64264467 x 152 Fax: +86-21-64264468 x 809 Email: Diana.hong@mid-link.net Eric.chen@mid-link.net ### 3. Predicate Device ### K number: K061483 Trade Name: VACUETTE® VISIO PLUS Blood Collection Needles Common Name: Blood Collection Needles Shanghai Midlink Business Consulting Co.,Ltd {1}------------------------------------------------ Kop3107; 062 Classification Name: Needle, Hypodermic, Single Lumen Product Code: FMI ### 4. Device Description Blood collection needle is a sterile and disposable medical device. It is consist of a puncturing needle which is punctured into vein, a bottle needle which is insert to the blood collection bottle to collect blood, a needle holder which connects the puncturing needle and bottle needle, two needle caps which protect puncturing needle and bottle needle and a latex cover to protect bottle needle. The performances of the applicant device of blood collection needle comply with ISO 7864:1993 and ISO9626:1991/AMD: 2001. The applicant device of blood collection needle is available in 20G, 21G and 22G. ### 5. Substantially Equivalence Determination #### Comparison Analysis: The applicant device has same classification information, same indications and intended use, similar product design, same performance effectiveness, performance safety as the predicate device. #### Conclusion: The applicant device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the applicant device is determined as safe and effectiveness. ### 6. Effectiveness and Safety Considerations #### Effectiveness: All the variant models of the applicant device are evaluated regarding the performance. #### Safety Considerations: With accordance with the Table 1 Initial Evaluation Tests for Consideration and Table 2 Supplementary Evaluation Tests for Consideration in ISO 10993-1:2003(E), Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing, the necessary tests for Biocompatibility Testing includes: Cytotoxicity, Sensitization or Intracutaneous Reactivity, Systemic Toxicity (Acute), Haemo-compatibility. Conclusion: The all conducted Biological Evaluation Tests are in compliance with the standards of ISO {2}------------------------------------------------ K(0.310) 33 10993, "Biological Evaluation of Medical Devices". The compatibility of all the possible skin-contact component material in the finished product meets the requirement of Biocompatibility The applicant device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the applicant device is determined as safe and effectiveness. Shanghai Midlink Business Consulting Co.,Ltd {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features the department's name in a circular arrangement around a stylized eagle symbol. The eagle is depicted with three curved lines representing its body and wings, giving it a modern and abstract appearance. The text reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 2 0 2007 ShanDong WeiGao Group Medical Polymer Products Company, Limited C/O Ms. Diana Hong General Manager Shanghai Mid-Link Business Consulting Company, Limited Suite 8D, No. 19, Lane 999 Zhongshan No. 2 Road Shanghai CHINA 200030 Re: K073127 Trade/Device Name: Blood Collection Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: October 31, 2007 Received: November 6, 2007 Dear Ms. Hong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Ms. Hong Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Susan Runne Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Korgu 121 ## Indications for Use 510(k) Number:_ Device Name: _ Blood Collection Needle Indications for Use: Blood Collection Needles is designed for use in the daily blood collection routine when delegated by a qualified practitioner. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page Anthony D. Mnt (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...