K070936 · Welmed, Inc. · FMF · Jul 16, 2007 · General Hospital
Device Facts
Record ID
K070936
Device Name
WELMED HYPODERMIC SYRINGE
Applicant
Welmed, Inc.
Product Code
FMF · General Hospital
Decision Date
Jul 16, 2007
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Welmed, Inc. piston syringes is to inject fluids into or withdraw fluids from the body.
Device Story
Welmed Hypodermic Syringe is a manual piston-style device consisting of a calibrated hollow barrel and movable plunger. It features a male Luer Lock connector at the barrel end for attachment to hypodermic needles or other female Luer-compatible devices. Operated manually by a clinician or user to inject or withdraw fluids. Device design and operational principles are standard for manual syringes. Safety and performance are established through conformance to ISO 7886-1:1993.
Clinical Evidence
Bench testing only. Conformance to ISO 7886-1:1993 (Sterile hypodermic syringes for single use) was used to demonstrate safety and performance.
Technological Characteristics
Piston syringe; manual operation; male Luer Lock connector; calibrated barrel; plunger mechanism. Conforms to ISO 7886-1:1993. Materials are typical for general-purpose syringes.
Indications for Use
Indicated for the injection of fluids into or withdrawal of fluids from the body. No specific patient population, age, or gender restrictions are stated.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
K113091 — Syringes, With or without Needles · Jiangyin Caina Technology Co., Ltd. · Oct 28, 2011
K024052 — MERIT MEDICAL 1-ML SYRINGE · Merit Medical Systems, Inc. · Feb 21, 2003
K232950 — Sterile Disposable Syringe, Sterile Disposable Syringe with needle, Sterile Disposable Syringe with safety needle · Perfect Medical Industry (Vn) Co., Ltd. · Jun 12, 2024
K213811 — Sterile Hypodermic Syringes for Single Use · Berpu Medical Technology Co., Ltd. · May 12, 2022
K182279 — Merit Syringe · Merit Medical Systems, Inc. · Nov 6, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 16 2007
## Section 5: 510(k) Summary
| 1. Preparation Date: | March 30, 2007 |
|----------------------|-------------------------------------------------------|
| 2. Submitted by: | Welmed Inc.<br>691 Lake Street<br>Grayslake, IL 60030 |
## Contact Person/Prepared by:
Darren Reeves Management Representative Phone (866) 393-4954 Fax (866) 393-4954 E-mail: dpdist@bww.com
#### 3. Device Identification:
| Trade Name: | Welmed Hypodermic Syringe (various sizes) |
|----------------------|----------------------------------------------------|
| Common Name: | Syringes, Hypodermic |
| Classification Name: | Piston Syringe (21 CFR 880.5860; Product Code FMF) |
#### 4. Predicate Device: Merit Medical Syringe (K024052)
#### 5. Device Description:
The Welmed, Inc. Syringe is a device intended for medical purposes, consisting of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male Luer Lock connector (nozzle) for attaching the female Luer connector (hub) of a hypodermic single lumen needle, or for attaching other devices with a female Luer.
#### 6. Intended Use:
The intended use of the Welmed, Inc. piston syringe is to inject fluids into or withdraw fluids from the body.
#### 7. Statement to conform to ISO 7886-1; 1993
Welmed Inc. has established that its family of syringes corm to the FDArecognized consensus standard, ISO 7886-1:1993, Sterile hypodermic syringes for single use - Part 1: Syringes for manual use. Data supporting conformance with the standard is available from Welmed
8. Conclusion - Based on conformance with the recommended standard, Welmed syringes are safe and effective for the intended use.
{1}------------------------------------------------
### 9. Similarities/ Differences of the proposed device when compared to the predicate:
#### 9.1.1 Intended Use
The Merit Medical syringes are intended to inject fluids into, or withdraw fluids from, the body. As such, the Intended Uses of the predicate and Welmed syringes are equivalent.
#### 9.1.2 Materials
Materials used in the manufacture of Welmed syringes are typically used in the manufacture of general-purpose syringes, including the predicate device,
#### 9.1.3 Design
The design of the Welmed syringe is typical for syringes, including that of the predicate.
#### 9.1.4 Operational Principles
The Welmed syringe is manually operated by advancing and withdrawing the plunger in the barrel. The operating principles are identical for all manual syringes, including the predicate.
#### 9.1.5 Technology
The same fundamental technology is used in the design of the Welmed syringes as is employed in the design of all manual syringes, including the predicate.
#### 9.1.6 Safety and Performance
Welmed has provided a statement that its syringes conform to the requirements of IS0 7886-1 :I 993, an FDA- recognized consensus standard. This statement and the data that has been collected to support conformance has been used to demonstrate safety and performance in lieu of demonstrating substantial equivalence with the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circle around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 16 2007
Welmed, Incorporated C/O Mr. Darren Reeves Management Representative DP Distribution and Consulting, LLC 15637 Fox Cove Circle Moseley, Virginia 23120
Re: K070936
Trade/Device Name: Welmed Hypodermic Syringes (1,3,5,10,20,30 and 60 ml) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: May 23, 2007 Received: May 29, 2007
Dear Mr. Reeves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Reeves
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a sequence of alphanumeric characters. The characters appear to be handwritten and include the letter 'K' followed by the numbers '070436'. The characters are written in a simple, clear style, making them easily readable. The image seems to be a close-up of the characters.
# Food and Drug Administratio
ER FOR DEVICES AND RADIOLOGIC. FDA Home Page | CDRH Home Page | Search | CDRH A-Z Index | Co
# Indications for Use
510(k) Number (if known): __K070936___________________________________________________________________________________________________________________________________________
Device Name: Welmed Hypodermic Syringes_
Indications for Use:
The intended use of the Welmed, Inc. piston syringes is to inject fluids into or withdraw fluids from the body.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
;
12 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
re
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _ K4 7φ936
Page 1 of 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.