Merit Syringe
K182279 · Merit Medical Systems, Inc. · FMF · Nov 6, 2018 · General Hospital
Device Facts
| Record ID | K182279 |
| Device Name | Merit Syringe |
| Applicant | Merit Medical Systems, Inc. |
| Product Code | FMF · General Hospital |
| Decision Date | Nov 6, 2018 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 880.5860 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Merit Syringe is used to inject fluids into, or withdraw fluids from, the body.
Device Story
Merit Syringe is a manual, three-piece piston syringe consisting of a clear hollow barrel, a plunger, and an O-ring seal. It features an ISO 594-2 compliant fixed male luer connector for compatibility with female luer hubs. Operated manually by advancing or withdrawing the plunger within the barrel to inject or withdraw fluids. Used in clinical settings by healthcare professionals. Benefits include precise fluid delivery or aspiration. The device utilizes a fluorocarbon-based non-silicone lubricant.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against ISO 7886-1 (manual syringes) and ISO 594-2 (luer fittings). Biocompatibility testing conducted per ISO 10993-1 for externally communicating, indirect blood-contacting devices (<24 hours), including cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, and hemolysis.
Technological Characteristics
Three-piece piston syringe. Barrel: cyclo-olefin polymer. Plunger: glass-filled nylon. O-ring: silicone. Lubricant: fluorocarbon-based non-silicone. Features ISO 594-2 compliant male luer lock. Sterilization: Ethylene oxide (ISO 11135). Compliant with ISO 7886-1 for graduation accuracy.
Indications for Use
Indicated for injection of fluids into or withdrawal of fluids from the body for patients requiring such procedures.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
- Merit 20 mL Syringe (K152783)
Related Devices
- K171362 — Merit Syringe · Merit Medical Systems, Inc. · Aug 7, 2017
- K173601 — Merit Syringe · Merit Medical Systems, Inc. · Jan 31, 2018
- K182216 — Merit Syringe · Merit Medical Systems, Inc. · Nov 6, 2018
- K163084 — Merit Coronary Control Syringe · Merit Medical Systems, Inc. · Jun 15, 2017
- K223678 — Sterile Syringes for Single Use · Shanghai Kindly Enterprise Development Group Co., Ltd. · Feb 9, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 6, 2018
Merit Medical Systems, Inc. John Skousen Senior Regulatory Affairs Specialist 1600W Merit Parkway South Jordan, Utah 84095
Re: K182279
Trade/Device Name: Merit Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: August 13, 2018 Received: August 22, 2018
Dear John Skousen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, nına Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182279
Device Name Merit Syringe
Indications for Use (Describe)
The Merit Syringe is used to inject fluids into, or withdraw fluids from, the body.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Commercial and/or Industrial OFF-Highway Diesel Engine Oil</label> |
| <label><input type="checkbox"/> In Use Fleet Automotive Vehicle OFF-Highway Diesel Engine Oil</label> |
|× | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K182279 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date Prepared:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 316-3724<br>(801) 826-4112<br>Dr. John Skousen<br>August 21, 2018<br>1721504 |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Merit Syringe<br>Piston Syringe<br>Syringe, Piston<br>2<br>FMF<br>880.5860<br>General Hospital |
| Predicate<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Premarket Notification:<br>Manufacturer:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Merit 20 mL Syringe<br>Piston Syringe<br>Syringe, Piston<br>2<br>K152783<br>Merit Medical Systems, Inc.<br>FMF<br>880.5860<br>General Hospital |
| Device<br>Description | The Merit Syringe contains a calibrated hollow barrel into which is<br>inserted a closely fitted, plunger, and O-ring seal. The barrel contains<br>an ISO 594-2 compliant fixed male luer connector, which is<br>compatible with ISO 594-2 compliant female luer hubs. The syringe<br>uses a fluorocarbon based non-silicone lubricant. | |
| Indications<br>for Use | The Merit Syringe is used to inject fluids into, or withdraw fluids from,<br>the body. | the body. |
| Comparison<br>to Predicate<br>Device | | The proposed Merit Syringe incorporates the same Intended Use with<br>similar materials, design, and principle of operation as the predicate<br>device. Differences include the nominal capacity; plunger material;<br>plunger seal design and material; and lubricant. The safety and<br>effectiveness of the new materials were verified via biocompatibility |
{4}------------------------------------------------
testing. Performance testing of the subject device was completed against FDA recognized consensus standards ISO 7886-1 and ISO 594-2. These differences between the subject device and predicate device do not raise different questions of safety and effectiveness.
| Comparison<br>to Predicate<br>Device | Attribute | Subject Device<br>- Merit Syringe | Predicate<br>Device - Merit<br>20 mL Syringe | Comments |
|---------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| (Continued) | Design | Standard three-<br>piece piston<br>syringe<br>constructed with<br>a clear hollow<br>barrel into which<br>is inserted a<br>closely fitting<br>plunger and O-<br>ring seal. Fitting<br>offered with male<br>luer lock<br>connector. | Standard three-<br>piece piston<br>syringe<br>constructed with<br>a clear hollow<br>barrel into which<br>is inserted a<br>closely fitting<br>plunger and<br>plunger tip seal.<br>Fitting offered<br>with male luer<br>lock connector. | Different<br>plunger<br>and<br>plunger<br>seal design |
| | Material | The barrel is<br>constructed from<br>clear cyclo-olefin<br>polymer; the<br>plunger from<br>glass-filled nylon;<br>the O-ring is<br>made of silicone;<br>fluorocarbon<br>based non-<br>silicone lubricant. | The barrel is<br>constructed from<br>clear cyclo-olefin<br>polymer; the<br>plunger from<br>polycarbonate or<br>ABS material;<br>the plunger tip<br>seal is made of<br>silicone; the<br>lubricant is<br>silicone. | Same<br>Barrel<br>Material<br>Different<br>plunger<br>material<br>Different tip<br>seal design<br>Different<br>lubricant |
| | Principle of<br>Operation | Manually<br>operated by<br>advancing and<br>withdrawing the<br>plunger within<br>the barrel. | Manually<br>operated by<br>advancing and<br>withdrawing the<br>plunger within<br>the barrel. | Same |
| | Operational<br>Volume | 0.25 mL<br>operational<br>volume. | 20 mL<br>operational<br>volume. | Different |
| Graduation | Printed with accurate<br>graduation lines<br>that are<br>compliant with<br>ISO 7886-1. | Printed with accurate<br>graduation lines<br>that are<br>compliant with<br>ISO 7886-1. | Different<br>graduations<br>but both<br>follow<br>same<br>standard | |
| Indications<br>for Use /<br>Intended<br>Use | The Merit<br>Syringe is used<br>to inject fluids<br>into, or withdraw<br>fluids from, the<br>body. | The Merit<br>Medical 20 mL<br>Syringe is used<br>to inject fluids<br>into, or withdraw<br>fluids from, the<br>body. | Same | |
{5}------------------------------------------------
FDA quidance and recognized consensus standards have been established for Piston Syringes under FDA Product Code FMF and 21 CFR 880.5860. A battery of tests was performed based on the Performance requirements of the below recognized consensus standards and guidance, as well as biocompatibility, sterilization, and packaging Data standards and guidance. ISO 7886-1: 2017, Sterile hypodermic syringes for single use – . Part 1: Syringes for manual use ISO 594-2:1998, Conical fittings with 6% (Luer) taper for ● syringes, needles and certain other medical equipment - Part 2: Lock fittings ISO 11135:2014, Sterilization of health care products – Ethylene ● oxide - Requirements for the development, validation, and routine control of a sterilization process for medical devices ● ISO 10993-7:2008, Biological evaluation of medical devices -Part 7: Ethylene oxide sterilization residuals . AAMI TIR 28:2009. Product adoption and process equivalency for ethylene oxide sterilization . ISO 11607-1:2006, Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems [Including: Amendment 1 (2014)] ASTM D4169-16: 2016, Standard Practice for Performance ● Testing of Shipping Containers and Systems . ASTM F1980-16: Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices . ISO 2233:2000, Packaging - Complete, filled transport packages and unit loads - Conditioning for testing
- ISO 10993-1:2009, Biological evaluation of medical devices -● Part 1: Evaluation and Testing within a risk management process
- ISO 10993-4:2017, Biological evaluation of medical devices -● Part 4: Selection of tests for interactions with blood
{6}------------------------------------------------
| Performance<br>Data<br>(Continued) | ISO 10993-5:2009, Biological evaluation of medical devices –<br>Part 5: Tests for <i>in vitro</i> cytotoxicity ISO 10993-10:2010, Biological evaluation of medical devices –<br>Part 10: Tests for irritation and skin sensitization ISO 10993-11:2006, Biological evaluation of medical devices –<br>Part 11: Tests for systemic toxicity ASTM F756-17:2017, Standard Practice for Assessment of<br>Hemolytic Properties of Materials AAMI/ANSI ST72:2011/(R) 2016, Bacterial Endotoxins – Test<br>methods, routine monitoring, and alternatives to batch testing United States Pharmacopeia 40, National Formulary 35, <151><br>Pyrogen Test (2017) FDA Guidance, <i>Guidance on the Content of Premarket<br/> Notification [510(k)] Submissions for Piston Syringes</i> , April 1993 FDA Guidance, <i>Use of International Standard ISO 10993-1,<br/> Biological Evaluation of Medical Devices Part 1: Evaluation and<br/> Testing within a risk management process</i> , June 2016 <b>Performance Bench Testing</b><br>The Merit Syringe complies with the FDA recognized consensus<br>standards ISO 7886-1 and ISO 594-2, as outlined within this<br>submission. Results of the testing demonstrate that the subject<br>device met the acceptance criteria sufficient for its intended use.<br>Testing included the following from these standards:<br><br><b>Biocompatibility testing</b><br>A biocompatibility evaluation for the Merit Syringe was conducted in<br>accordance with the FDA Guidance Document “Use of International<br>Standard ISO-10993, ‘Biological Evaluation of Medical Devices Part<br>1: Evaluation and Testing within a Risk Management Process,’” June<br>16, 2016, and International Standard ISO 10993-1 “Biological<br>Evaluation of Medical Devices - Part 1: Evaluation and Testing<br>Within a Risk Management Process,” as recognized by FDA. The<br>Merit Syringe is an externally communicating device with indirect<br>blood contacting for a duration of less than 24 hours. Testing included<br>the following:<br>Cytotoxicity Sensitization Irritation Acute Systemic Toxicity Pyrogenicity Hemolysis |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of<br>Substantial<br>Equivalence | Differences between the technological characteristics of the subject<br>device as compared to the predicate do not raise different questions<br>of safety and effectiveness. The performance of the device is<br>supported by non-clinical testing and risk management activities. The<br>Merit Syringe is Substantially Equivalent (SE) to the Merit 20 mL<br>Syringe, cleared under K152783. |