Sterile Disposable Syringe, Sterile Disposable Syringe with needle, Sterile Disposable Syringe with safety needle
K232950 · Perfect Medical Industry (Vn) Co., Ltd. · FMF · Jun 12, 2024 · General Hospital
Device Facts
Record ID
K232950
Device Name
Sterile Disposable Syringe, Sterile Disposable Syringe with needle, Sterile Disposable Syringe with safety needle
Applicant
Perfect Medical Industry (Vn) Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
Jun 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sterile Disposable Syringe is intended to be used to withdraw solutions into patient's body. Sterile Disposable Syringe with needle are intended to be used with a luer slip or luer lock syringe to withdraw solutions and insert solutions into patient's body. Sterile Disposable Syringe with safety needle is intended to be used with a luer slip or luer lock syringe to withdraw solutions into patient's body. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
Device Story
Manual piston-style syringe system consisting of barrel, plunger, and piston; available in luer slip or luer lock configurations. Used in hospital settings by clinicians to withdraw and inject fluids. Models include standard syringes, syringes with attached needles, and syringes with safety needles. Safety needle variants feature a manually activated shield to cover the needle post-use. Device operation is entirely manual; no electronic or automated components. Benefits include precise fluid delivery and reduced risk of sharps injury for healthcare providers.
Clinical Evidence
Bench testing only. Compliance with ISO 10993 (biocompatibility), ISO 7886-1 (syringes), ISO 7864 (needles), ISO 80369-7 (connectors), and ISO 23908 (sharps injury protection) verified performance. Testing included sterility, seal integrity (ASTM F88, F1929), and mechanical strength.
Technological Characteristics
Materials: Polypropylene (barrel, plunger, cap), Polyisoprene (gasket), Stainless Steel 304 (needle). Manual operation. EO sterilization. Connectivity: Luer slip/lock per ISO 80369-7. Dimensions/specs comply with ISO 7886-1, ISO 7864, ISO 9626, and ISO 6009.
Indications for Use
Indicated for use in patients requiring aspiration and injection of fluids. Sterile Disposable Syringe with safety needle includes a manual safety shield to minimize accidental needlestick risk after use.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 12, 2024
Perfect Medical Industry (VN) Co., Ltd. % Thomas Huang Lead Consultant, Quality and Regulatory Affairs Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin. Texas 78746
# Re: K232950
Trade/Device Name: Sterile Disposable Syringe, Sterile Disposable Syringe with Needle, Sterile Disposable Syringe with Safety Needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, FMF, FMI, MEG Dated: May 9, 2024 Received: May 9, 2024
# Dear Thomas Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Shruti N. Mistry -S
Shruti Mistry Assistant Director Division of Drug Delivery and General Hospital Devices, and Human Factors Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232950
#### Device Name
Sterile Disposable Syringe, Sterile Disposable Syringe with Needle, Sterile Disposable Syringe with Safety Needle
#### Indications for Use (Describe)
Sterile Disposable Syringe is intended to be used to withdraw solutions into patient's body. Sterile Disposable Syringe with needle are intended to be used with a luer slip or luer lock syringe to withdraw solutions and insert solutions into patient's body. Sterile Disposable Syringe with safety needle is intended to be used with a luer slip or luer lock syringe to withdraw solutions into patient's body. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# 510(k) Submitter
[K232950]
| Device Submitter: | Perfect Medical Industry (VN) Co., Ltd.<br>Block D7/I, No. 1B Road, Vinh Loc Industrial Zone, Binh Hung Hoa B<br>Ward, Binh Tan District,<br>Ho Chi Minh City, 700000 Vietnam<br>Telephone: +0084.915.521.775<br>Fax: +0084.283.765.3135<br>2017@pwuhui.com.tw |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Thomas Huang<br>Lead Consultant, Quality and Regulatory Affairs<br>Emergo Global Consulting, LLC<br>2500 Bee Cave Road, Building 1, Suite 300<br>Austin, Texas 78746 USA<br>Telephone: (512) 327-9997<br>Fax: (512) 327-9998<br>LST.AUS.ProjectManagement@ul.com |
| Date Prepared | May 03, 2024 |
### l Device
| Trade Name of Device: | Sterile Disposable<br>Syringe, Sterile<br>Disposable Syringe<br>with Needle, Sterile<br>Disposable Syringe<br>with Safety Needle |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number: | 21 CFR 880.5860,<br>21 CFR 880.5570 |
| Classification Name: | Piston syringe |
| Product Code: | FMF, FMI, MEG |
| Regulation Name: | Piston Syringe |
| Common name | Syringes with Needle |
| Regulatory Class | II |
| Review Panel | Injection |
{4}------------------------------------------------
#### II Predicate Devices
| 510k Number | K211329 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name of Device: | Sterile Disposable<br>Syringe with Safety<br>Needle, Sterile<br>Disposable Syringe<br>with Needle, Sterile<br>Disposable Syringe,<br>Sterile Disposable<br>Safety Needle, Sterile<br>Disposable Needle |
| Regulation Number: | 21 CFR 880.5860,<br>880.5570 |
| Regulation Name: | Piston Syringe |
| Regulatory Class | II |
| Product Code: | MEG, FMF, FMI |
#### III Device Descriptions
The Sterile Disposable Syringe consists of barrel, plunger and piston. The proposed device is available in luer slip or luer lock connector types which are intended to be connected with a hypodermic needle. The Sterile Disposable Syringe with Needle is available in 18-19, 21-26 gauge and 3/8 – 2 inch lengths. The proposed device is also available with fixed needle 25-26 gauge and 1/2 – 1 inch length. The Sterile Disposable Syringe with Safety Needle is available in 18-19, 21-26 gauge and 3/8 – 2 inch lengths. The safety cap will be manually activated to cover the needle immediately after use to minimize risk of accidental needle sticks. All subject device are operated manually.
{5}------------------------------------------------
# IV Indications for use
V Technological Characteristics Comparison
Sterile Disposable Syringe are intended to be used to withdraw solutions and insert solutions into patient's body. Sterile Disposable Syringe with needle are intended to be used with a luer slip or luer lock syringe to withdraw solutions and insert solutions into patient's body. Sterile Disposable Syringe with safety needle are intended to be used with a luer slip or luer lock syringe to withdraw solutions and insert solutions into patient's body. After withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
| Attribute | Subject Device | Predicate Device, K211329 | Comparison |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Syringe Type: | FMF, FMI, MEG | FMF, FMI, MEG | Same |
| Indications for Use: | 1. Sterile Disposable Syringe are intended to be used to withdraw solutions and insert solutions into patient's body.<br>2. Sterile Disposable Syringe with needle are intended to be used with a luer slip or luer lock syringe to withdraw solutions and insert solutions into patient's body.<br>3. Sterile Disposable Syringe with safety needle are intended to be used with a luer slip or luer lock syringe to withdraw solutions and insert | 1. The Sterile Disposable Syringe is a sterile luer slip or luer lock syringe which is intended to be used with a hypodermic needle for the aspiration and injection of fluids for medical purpose.<br>2. The Sterile Disposable Syringe with Needle is intended for use in the aspiration and injection of fluids for medical purpose.<br>3. The Sterile Disposable Syringe with Safety Needle is intended for use in the aspiration | Same |
# VI-1: Comparison of Piston Syringe
{6}------------------------------------------------
| solutions into<br>patient's body. After<br>withdrawal of the<br>needle from the<br>body, the attached<br>needle safety shield<br>can be manually<br>activated to cover<br>the needle<br>immediately after<br>use to minimize risk<br>of accidental<br>needlestick. | and injection of<br>fluids for medical<br>purpose. After<br>withdrawal of the<br>needle from the<br>body, the attached<br>needle safety shield<br>can be manually<br>activated to cover<br>the needle<br>immediately after<br>use to minimize risk<br>of accidental needle<br>sticks.<br>4. The Sterile<br>Disposable Needle<br>is intended to be<br>used with a luer slip<br>or luer lock syringe<br>for aspiration and<br>injection of fluids for<br>medical purpose.<br>5. The Sterile<br>Disposable Safety<br>Needle is intended<br>to be used with a<br>luer slip or luer lock<br>syringe for<br>aspiration and<br>injection of fluids for<br>medical purpose.<br>After withdrawal of<br>the needle from the<br>body, the attached<br>needle safety shield<br>can be manually<br>activated to cover<br>the needle<br>immediately after<br>use to minimize risk |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{7}------------------------------------------------
| | | of accidental needle | |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| | | sticks. | |
| Principle of<br>Operation | For manual use only | For manual use only | Same |
| Specific Drug Use | N/A | N/A | Same |
| Length | Complied with ISO<br>7886-1:2017 | Complied with ISO<br>7886-1:2017 | Same |
| Diameter | Complied with ISO<br>7886-1:2017 | Complied with ISO<br>7886-1:2017 | Same |
| Tip Type | Luer lock, luer slip,<br>fixed | Luer lock, luer slip | Different, see<br>analysis below |
| Volume | 1mL, 3mL, 5mL,<br>10mL, 20mL, 30mL,<br>50mL | 1ml, 2ml, 3ml, 5ml,<br>10ml, 20ml, 30mL,<br>50ml, 60ml | Different, see<br>discuss below |
| Needle Length | 3/8", 1/2", 5/8", 3/4",<br>1", 1-1/4", 1-1/2", 2" | 1/2", 5/8", 3/4", 1",<br>1-1/4", 1-1/2" | Different, see<br>discuss below |
| Needle Gauge | 18G, 19G, 21G, 22G,<br>23G, 24G, 25G, 26G | 23G | Different, see<br>discuss below |
| Needle Tip<br>Configuration | Bevel tip<br>Complied with ISO<br>7864:2016 | Complied with ISO<br>7864:2016 | Same, see analysis<br>below |
| Nozzle Type | Luer lock, luer slip | Luer lock, luer slip | Same |
| Barrel Marking Spec | Graduated scale:<br>1cc: 0.01 mL; 3cc:0.1<br>mL; 5cc, 10cc:0.2<br>mL;<br>20cc, 30cc,<br>50cc:1mL.<br>Length of scale<br>(Min):<br>1cc: 57mm, 3cc:<br>27mm, 5cc: 36mm,<br>10cc: 44mm, 20cc:<br>52mm, 30cc:67mm,<br>50cc: 75mm<br>Complied with ISO<br>7886-1:2017 | Complied with ISO<br>7886-1:2017 | Same, see analysis<br>below |
| Gradations Legibility | Complied with ISO<br>7886-1:2017 | Complied with ISO<br>7886-1:2017 | Same |
| Needle Cover<br>Dimensions | ID : min 5.9mm; OD:<br>min 7.2mm;<br>Length:<br>42mm~70mm<br>Complied with ISO<br>7864:2016 | Complied with ISO<br>7864:2016 | Same, see analysis below |
| Needle Cover Color | unpigmented<br>Complied with ISO<br>7864:2016 | Complied with ISO<br>7864:2016 | Same, see analysis below |
| Lubricant<br>Composition | Silicone oil | Silicone oil | Same |
| Lubricant<br>Amount/cm2 | Complied with ISO<br>7886-1:2017 | Complied with ISO<br>7886-1:2017 | Same |
| Barrel Transparency | Transparent<br>Complied with ISO<br>7886-1:2017 | Complied with ISO<br>7886-1:2017 | Same, see analysis below |
| Delivery Accuracy | 1cc and 3cc: ±5% of<br>expelled volume<br>when it filled equal<br>to or greater than<br>half nominal<br>capacity;<br>±(1.5% of volume +<br>2% of expelled<br>volume when it<br>filled less than half<br>nominal capacity<br>5cc, 10cc, 20cc, 30cc,<br>50cc: ±4% of<br>expelled volume<br>when it filled equal<br>to or greater than<br>half nominal<br>capacity;<br>±(1.5% of volume +<br>1% of expelled<br>volume when it<br>filled less than half<br>nominal capacity | Complied with ISO<br>7886-1:2017 | Same, see analysis below |
| | Complied with ISO<br>7886-1:2017 | | |
| Reuse Durability | Single use | Single use | Same |
| Needle Cover<br>Strength | 0.4-1.8kgf<br>Complied with ISO<br>7864:2016 | Complied with ISO<br>7864:2016 | Same, see analysis<br>below |
| Hub/Needle Bond<br>Strength | Min. 22-69N<br>Complied with ISO<br>7864:2016 | Complied with ISO<br>7864:2016 | Same, see analysis<br>below |
| Biocompatibility | Cytotoxicity,<br>Irritation,<br>Sensitization,<br>Systemic Toxicity,<br>Hemolysis, Pyrogen | Cytotoxicity,<br>Irritation,<br>Sensitization,<br>Systemic Toxicity,<br>Hemolysis, Pyrogen | Same |
| Materials | Gasket:<br>Polyisoprene<br>Rubber; Barrel:<br>Polypropylene (PP);<br>Plunger:<br>Polypropylene (PP);<br>Needle: Stainless<br>Steel 304;<br>Needle cap:<br>Polypropylene (PP) | Gasket:<br>Polyisoprene<br>Rubber; Barrel:<br>Polypropylene (PP);<br>Plunger:<br>Polypropylene (PP);<br>Needle: Stainless<br>Steel SUS 304;<br>Needle cap:<br>Polypropylene (PP) | Same |
| Labeling | Complied with 21<br>CFR part 801 | Complied with 21<br>CFR part 801 | Same |
| Sterilization method | EO Sterilized | EO Sterilized | Same |
| Environment of use | Hospital | Hospital | Same |
| Shelf life | 3 years.<br>Complied with<br>ASTM F1980-07 | 5 years.<br>Complied with<br>ASTM F1980-07 | Different, see<br>analysis below |
| Transport<br>conditions | Complied with ISTA<br>2A:2011 | Not public | Different, see<br>analysis |
{8}------------------------------------------------
{9}------------------------------------------------
# Difference Analysis:
Tip Type: The tip configuration of the proposed device is similar to the predicate device. However, this difference is in a fixed connector. The difference in technological characteristics does not raise new questions of safety and effectiveness. Connectivity to the needle is
{10}------------------------------------------------
assessed with performance data.
Volume: The Syringe volume for the subject device is different from the predicate device. However, this difference is in dimension. Different volume devices will be selected by the physician per the patient's condition. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device. Differences are assessed via performance data and conformance to the same ISO standards as the predicate device.
Needle Length: The needle length for the subject device is different from the predicate device. This difference is in dimension. Different length devices will be selected by the physician per the patient's condition. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device. Differences are assessed via performance data and conformance to the same ISO standards as the predicate device.
Needle Gauge: The needle size for the subject device is different from the predicate device. This difference is in dimension. Different size devices will be selected by the physician per the patient's condition. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device. Differences are assessed via performance data and conformance to the same ISO standards as the predicate device.
Needle Tip Configuration: The needle tip for the subject device is different in needle point geometry from the predicate device. This difference will not raise new questions of safety and effectiveness of the subject device. Differences are assessed via performance data and conformance to the same ISO standards as the predicate device.
Barrel Marking Spec: The Syringe Barrel Marking Spec for the subject device is different from the predicate device. However, different Marking Specs will be dependent on the Syringe volume. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device. Differences are assessed via performance data and conformance to the same ISO standards as the predicate device.
Needle Cover Dimensions: The Needle Cover Dimensions for the subject device is different from the predicate device. However, this difference is in dimension. Different Needle Cover Dimensions will be dependent on the Needle Length. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device. Differences are assessed via performance data and conformance to the same ISO standards as the predicate device.
Needle Cover Color: The Needle Cover Color for the subject device is different from the predicate device. However, this difference is in Color. Different colors will be dependent on either pigmented or unpigmented material. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device. Differences are assessed via performance data and conformance to the same ISO standards as the predicate device.
Barrel transparency: The Barrel transparency for the subject device is different from the predicate device. However, this difference is in color. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device. Differences are assessed via performance data and conformance to the same ISO standards as the predicate device.
{11}------------------------------------------------
Delivery accuracy: The Delivery accuracy for the subject device is different from the predicate device. However, this accuracy is dependent on the Syringe volume. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device. Differences are assessed via performance data and conformance to the same ISO standards as the predicate device.
Needle cover strength: The needle cover strength for the subject device is different from the predicate device. However, differences are assessed via performance data and conformance to the same ISO standards as the predicate device. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device.
Hub/needle bond strength: The hub/needle bond strength for the subject device is different from the predicate device. However, differences are assessed via performance data and conformance to the same ISO standards as the predicate device. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device.
Shelf life: The Shelf life for subject device is different from the predicate device. However, differences are assessed via performance data and conformance to the same ISO standards as the predicate device. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device.
Transport conditions: The Transport conditions for the subject device is different from the predicate device. However, differences are assessed via performance data and conformance to the same ISO standards as the predicate device. Therefore, this difference will not raise new questions of safety and effectiveness of the subject device.
#### VI Summary of Non-clinical Testing (Bench)
Nonclinical tests were conducted to verify that the proposed device met all design specifications to establish Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- . ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in Vitro Cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and skin sensitization
- ISO 10993-23:20212021 Biological evaluation of medical devices- Part 23: Test for irritation
- ISO 10993-11:2017 Biological evaluation of medical devices- Part 11: Tests for systemic toxicity
- ISO 10993-4:2017 Biological evaluation of medical devices--Part 4: Selection of Tests for Interactions with Blood
- ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
- ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for ●
{12}------------------------------------------------
Flexible Packaqinq by Visual Inspection
- . ASTM F88/F88M-15, Standard Test Method for Seal Strength of Flexible Barrier Materials. (Sterility)
- . ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
- ISO 7864:2016 Sterile hypodermic needles for single use - Requirements and test methods
- ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices ●
- ISO 6009:2016 Hypodermic needles for single use Colour coding for identification ●
- ISO 80369-7:2016 Small-bore connectors for liguids and gases in healthcare applications ● — Part 7: Connectors for intravascular or hypodermic applications
- ISO 7886-1:2017 Sterile hypodermic syringes for single use- Part 1: Syringes for manual . use
- ISO 10993-7:2008 Biological Evaluation of Medical Device-Part 7: Ethylene Oxide ● Sterilization Residuals
- . ISO 23908:2011 Sharps injury protection - Requirements and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
- USP<85> Bacterial Endotoxins Test
- USP<151> Pyrogen Test
- USP<788> Particulate Matter in Injections ●
- Guidance on the Content of Premarket Notification [510(K)] Submissions for Hypodermic ● Single Lumen Needles
- Guidance on the Content of Premarket Notification [510(K)] Submissions for Piston Svringes
- Medical Devices with Sharps Injury Prevention Features - Guidance for Industry and FDA Staff
#### VII Conclusion
The subject device has the same indications for use as the predicate device. Any minor differences in the technological characteristics of the subject device when compared to the predicate device have been successfully evaluated through appropriate safety and performance testing which demonstrates that the subject device, when compared to the predicate device does not raise any new questions of safety and effectiveness. Therefore, the subject device has been determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.