The 5 FR DL Piper PICC Catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and prolonged exposure to intraluminal solutions containing up to 70% ethanol. The maximum recommended infusion rate is 5mL/sec for power injection of contrast media.
Device Story
5 FR Dual Lumen Piper PICC; peripherally inserted central catheter; provides peripheral access to central venous system for IV therapy, blood sampling, and power injection of contrast media. Device features reverse-tapered polyurethane shaft; dual-lumen; trimmable; radiopaque distal tip. Clinician inserts via Seldinger or modified-Seldinger technique; tip positioned in lower 1/3 of Superior Vena Cava. Compatible with intraluminal solutions containing up to 70% ethanol. Max infusion rate 5 mL/sec. Provides securement via suture wings. Benefits include reliable venous access for prolonged therapy and contrast delivery. Used in clinical settings by trained healthcare providers.
Clinical Evidence
Bench testing only. No clinical trials performed. Verification and validation testing conducted per 21 CFR 820.30, including dimensional analysis, radiopacity, leak testing, pump flow, tensile strength, fatigue, power injection, and biocompatibility (ISO 10993). Thrombogenicity evaluated in vivo against reference device (Bard 5F Dual Lumen PowerPICC) per ISO 10993-4, showing equivalent or better performance.
Technological Characteristics
5 French dual-lumen catheter; material: Quadrasil (siliconized aromatic polycarbonate polyurethane); reverse-tapered design; 55cm length; radiopaque distal tip. Standards: ISO 10555-1/3, ISO 594-1/2, ASTM F640-79. Sterilization: Ethylene Oxide (SAL 10^-6). Biocompatibility: ISO 10993-1. Power injection rated to 300 psi at 5 mL/sec.
Indications for Use
Indicated for patients requiring short or long-term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and prolonged exposure to intraluminal solutions containing up to 70% ethanol.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
K070002 — BOSTON SCIENTIFIC PERIPHERALLY INSERTED CENTRAL CATHETER · Boston Scientific · Aug 17, 2007
K071361 — R4 VASCULAR INC. PEGASUS CT PICC, MODEL 500005-001,002, 004, 005, 500004-001, 002,004,005 · R4 Vascular, Inc. · Aug 20, 2007
K080604 — PRESSURE INJFECTABLE PICC · Arrow International, Inc. · Jun 27, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
September 25, 2018
Piper Access LLC Jay Muse CEO 3981 South 700 East Suite 15 Salt Lake City, Utah 84107
Re: K181904
Trade/Device Name: 5 FR Dual Lumen Piper PICC Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: July 13, 2018 Received: July 16, 2018
Dear Jay Muse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181904
Device Name 5 FR DL Piper PICC
#### Indications for Use (Describe)
The 5 FR DL Piper PICC Catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and prolonged exposure to intraluminal solutions containing up to 70% ethanol. The maximum recommended infusion rate is 5mL/sec for power injection of contrast media.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Piper Access. The logo features a stylized, curved line in a light orange color above the company name. The text "PIPERACCESS" is written in a simple, sans-serif font, with "PIPER" in a darker gray and "ACCESS" in a lighter gray, and a trademark symbol to the right.
## SECTION 2
# 510(k) Summary
# K181904
## Applicant
| Name: | Piper Access, LLC |
|----------|----------------------------------------------------------|
| Address: | 3981 South 700 East Suite 15<br>Salt Lake City, UT 84107 |
| Phone: | 801-210-2886 |
| Contact: | Shanna Ryan, Regulatory Affairs Project Manager |
| Email: | shanna.ryan@gmail.com |
September 13, 2018
Date Prepared:
## Subject Device
| Trade Name: | 5 FR DL Piper PICC |
|----------------------|---------------------------------------------------------------------|
| Manufacturer: | Piper Access |
| Common Name: | Catheter, Intravascular, Therapeutic, Long-term Greater than 30days |
| Classification Name: | Percutaneous, Implanted, Long-term Intravascular Catheter |
| Class: | II |
| Product Code: | LJS |
| Regulation: | 21 CFR §880.5970 |
| Panel: | General Hospital |
### Primary Predicate
| Premarket Notification: | K151985 (clearance date June 14, 2016) |
|-------------------------|---------------------------------------------------------------------|
| Trade Name: | PowerPICC® EtOH Catheter |
| Manufacturer: | Bard Access Systems, Inc. |
| Common Name: | Catheter, Intravascular, Therapeutic, Long-term Greater than 30days |
| Classification Name: | Percutaneous, Implanted, Long-term Intravascular Catheter |
| Class: | II |
| Product Code: | LJS |
| Regulation: | 21 CFR §880.5970 |
| Panel: | General Hospital |
### Reference Device
| Premarket Notification: | K051672 (clearance date Nov 23, 2005) |
|-------------------------|---------------------------------------------------------------------|
| Trade Name: | 5 Fr DL PowerPICC® Catheter |
| Manufacturer: | Bard Access Systems, Inc. |
| Common Name: | Catheter, Intravascular, Therapeutic, Long-term Greater than 30days |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Piper Access. The logo features the text "PIPERACCESS" in a sans-serif font, with "PIPER" in gray and "ACCESS" in orange. Above the text is a stylized graphic resembling a bird in flight, with the head and neck in orange and the body in gray. The logo is clean and modern, suggesting a focus on accessibility and innovation.
VOLUME 2
SECTION 2
| Classification Name: | Percutaneous, Implanted, Long-term Intravascular Catheter |
|----------------------|-----------------------------------------------------------|
| Class: | II |
| Product Code: | LJS |
| Regulation: | 21 CFR §880.5970 |
| Panel: | General Hospital |
#### Device Description
The 5 French (FR) Dual Lumen (DL) Piper PICC is a single use peripherally inserted central catheter made from a specially-formulated medical grade polyurethane material.
The Piper PICC is designed to perform infusion, intravenous therapy, blood sampling and power injection of contrast media, and is compatible with intraluminal solutions containing up to 70% ethanol. The maximum recommended infusion rate is 5 mL/sec.
The Piper PICC is a 5 French catheter with two 18-gauge lumens and will be provided with 55cm of effective length. The catheter has a reverse-tapered design. The catheter is inserted peripherally by the clinician and is trimmable to fit different patient sizes. It is inserted with a Seldinger or modified-Seldinger technique with compatibility to guidewires up to 0.019" in outer diameter. The catheter distal tip is positioned in the lower 1/3 of the Superior Vena Cava (SVC). The effective length of the catheter including the distal end is radiopaque, which allows for catheter tip visualization.
The Piper PICC catheter includes an extruded polyurethane catheter shaft molded to an injection molded polyurethane hub with extruded extension legs molded to luer-lock fittings, which provide attachments for IV administration. The junction has suture wings to allow for securement to the patient. Clamps are attached to both extension legs on the catheter.
The distal end of the catheter shaft is a dual lumen symmetrical D-shape design that does not differ in material from the remainder of the shaft. With the exception of the reverse-tapered section of the shaft, the distal end is also dimensionally identical to the remainder of the shaft.
The proximal end of the catheter consists of two power injectable extension legs, which each have a luer lock style connection depicting gage size, thumb clamp, ID tag. The catheter has an average priming volume is 0.61 mL
The Piper PICC is packaged in a catheter only kit and provided sterile to the end user. There are no additional components in the kit.
#### Intended Use
The Piper PICC Catheter is intended for short- or long-term peripheral access to the central venous system for intravenous therapy and blood sampling.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Piper Access. The logo features a stylized, curved line in two colors: the top portion is a light orange, and the bottom portion is light gray. Below the line is the text "PIPERACCESS™", with "PIPER" in light gray and "ACCESS" in light orange. The trademark symbol is also in light orange.
#### SECTION 2
### Technological Characteristics and Comparison to the Predicate Device
Both the subject device and predicate are intended for short- or long-term peripheral access to the central venous system for intravenous therapy and blood sampling. The subject and predicate devices are based on the following same technological elements: Intended use, mechanism of action, placement technique and insertion site, duration of use/dwell time, distal tip placement. design features, dimensions, power injection, sterility, packaging, sterilization, and biocompatibility.
The differences between the subject device and cited predicate include:
- Materials: The Piper PICC is comprised of Quadrasil, a siliconized aromatic polycarbonate ● polyurethane which differs from the material used in the predicate. The material was evaluated using the same test standards as the predicate devices. Performance testing demonstrated that the material meets the mechanical testing and biocompatibility requirements and demonstrated resistance to intraluminal exposure to 70% ethanol.
The following table shows a comparison of the technological characteristics between the Piper PICC and the cited predicate in sufficient detail to provide an understanding of the basis for determining substantial equivalence.
| Feature | PREDICATE<br>BARD ACCESS<br>PowerPICC® EtOH Catheter | SUBJECT DEVICE<br>5 FR Dual Lumen<br>PIPER PICC | | | | | | | | |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|--|--|--|--|--|--|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The PowerPICC EtOH Catheter is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy including prolonged exposure to intraluminal solutions containing up to 70% ethanol, power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion, or therapy, use a 4 French or larger catheter. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used. Catheter Size Maximum Flow Rate 4F Single Lumen 5 ml/sec 5F Dual Lumen 5 ml/sec 5F Triple Lumen 4 ml/sec | | | | | | | | | The 5 FR DL Piper PICC Catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and prolonged exposure to intraluminal solutions containing up to 70% ethanol. The maximum recommended infusion rate is 5mL/sec for power injection of contrast media. |
| Placement Technique | Seldinger, Modified-Seldinger | Same | | | | | | | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Piper Access. The logo features a stylized, curved line above the words "PIPER ACCESS." The word "PIPER" is in gray, and the word "ACCESS" is in orange.
VOLUME 2
#### SECTION 2
| Feature | PREDICATE<br>BARD ACCESS<br>PowerPICC® EtOH Catheter | SUBJECT DEVICE<br>5 FR Dual Lumen<br>PIPER PICC |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Duration of Use | Short (<30 days) or long-term (≥30 days) | Same |
| Insertion Site | Peripheral | Same |
| Placement of Distal Tip | Lower 1/3 of the Superior Vena Cava | Same |
| ID | 18 Gauge | Same |
| OD | 5 French | Same |
| Length | 55 cm (Usable) | Same |
| Number of Lumens | Dual | Same |
| Catheter configuration | Two power injection lumens | Same |
| Internal Lumen Configuration | Double D | Same |
| Catheter Tip Configuration | Trimmable | Same |
| Maximum Power Injection | 300 psi | Same |
| Max Power Injection Flow Rate | 5/mL sec | Same |
| Ethanol Resistance | Yes | Same |
| Catheter Design | Reverse Taper | Same |
| Catheter Materials-<br>Catheter Shaft, Extension<br>Legs, Molded Junction | Aromatic polycarbonate polyurethane | Siliconized aromatic<br>polycarbonate polyurethane<br>(Quadrasil) |
| Catheter Materials - Luers | Quadraplast polyurethane | Same |
| Printing | Fr size, depth markings, gauge size | Same |
| Sterility | Provided sterile (SAL 10⁻⁶) / EO | Same |
| Biocompatibility | To ISO 10993-1 | Same |
| Non-clinical performance testing | To 1SO - 10555-1 and 1SO - 10555-<br>3; ISO 594-1 and 594-2; FDA<br>Guidance, ASTM F640-79 | Same |
#### Performance Testing Summary
Verification and validation tests have been performed in accordance with Design Controls as per 21 CFR §820.30. The following guidance and standards used to determine test methods and acceptance criteria:
- Guidance on Premarket Notification [510(k)] Submission for Short- Term and Long-Term ● Intravascular Catheters, March 16, 1995
- Design Control Guidance for Medical Device Manufacturers, March 11, 1997 ●
- ISO 10555-1: 2013, Sterile, single-use intravascular catheters, Part 1: General requirements ●
- ISO 10555-3: 2013, Intravascular catheters--Sterile and single-use catheters, Part 3: Central ● venous catheters
- . ISO 594-1: 1986, Conical fittings with 6% luer taper for syringes, needles and certain other medical equipment - Part 1: General Requirements
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Piper Access. The logo features a stylized, curved line in orange and gray above the words "PIPERACCESS" in gray. The word "ACCESS" has a trademark symbol next to it.
- ISO 594-2: 1998, Conical fittings with 6% luer taper for syringes, needles and certain other medical equipment - Part 2: Lock Fittings
- ASTM F640-79 (reapproved 2000): 2012, Standard Test Methods for Radiopacity of ● Plastics for Medical Use
Sterilization and Packaging Guidance and standards:
- ISO 11135:2014, Medical Devices Validation and Routine Control of Ethylene Oxide ● Sterilization
- ISO 11607-1: 2006, Packaging for terminally sterilized medical devices Part 1: ● Requirements for materials, sterile barrier systems and packaging systems
- ISO 11607-1 AMD 1: 2014, Packaging for terminally sterilized medical devices Part 1: ● Requirements for materials, sterile barrier systems and packaging systems
- ISO 10993-7: 2008. Biological Evaluation of Medical Devices Part 7: Ethylene Oxide Sterilization Residuals
- ANSI/AAMI ST72:2011, Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing
- FDA Guidance for Industry Pyrogen and Endotoxins Testing: Questions and Answers, ● 2010
The following tests were performed:
- Dimensional Analysis ●
- Ink Permanence
- Radiopacity ●
- Leak Testing ●
- Extension Leg Leak w/ Clamp ●
- Pump Flow
- Priming Volume ●
- Gravity Flow
- Kink Diameter
- Catheter Collapse
- Tip Stability
- Suture Wing Integrity
- Assembly Tensile
- Shaft Tensile
- Catheter Elongation
- Catheter Modulus ●
- Catheter Fatigue ●
- Power Injection ●
- Assembly Burst ●
- Extension Leg Burst
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Piper Access. The logo features a stylized, curved line above the words "PIPERACCESS". The line starts with an orange color and fades to gray. The word "PIPER" is in gray, while "ACCESS" is in orange.
### SECTION 2
- ISO Luer Gauging ●
- ISO Luer Testing
- Sterilization ●
- Packaging Validation
- Residuals. EO and ECH ●
- Pyrogenicity Bacterial Endotoxin Test (LAL) ●
Biocompatibility evaluations were completed per ISO 10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing for externally communicating, blood contacting, permanent devices and FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. The following were evaluated:
- Cytotoxicity ●
- Sensitization
- Irritation/Intracutaneous Reactivity
- Acute Systemic Toxicity ●
- Material Mediated Pyrogenicity
- Hemolysis (indirect/direct)
- Complement Activation ●
- Partial Thromboplastin Time
- Sheep Thrombogenicity
- Implantation 13 weeks ●
- Chemical Characterization
- Subacute/ Sub-chronic Toxicity ●
- Genotoxicity
- Chronic Toxicity
- Carcinogenicity ●
To meet the requirements in ISO10993-4, a reference device, the Bard 5F Dual Lumen PowerPICC was used for in vivo thrombogenicity evaluations as the legally marketed comparative device (LMCD). The Bard 5F Dual Lumen PowerPICC was selected as the Bard PowerPICC EtOH Catheter was not available on the market and could not be obtained for use as the LMCD. The Bard 5F Dual Lumen PowerPICC provides a point of comparison and is one of the most commonly used PICC products throughout the world, and has proven acceptable clinical performance with over a decade of use. Further justifying the acceptability of the 5F Dual Lumen is the fact that it is the device to which the cited predicate device claimed substantial equivalence. During evaluations reported within this 510k submission, it has been shown that the subject device performs equivalent to, or better than, the reference device with respect to thrombogenic potential.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Piper Access. The logo features a stylized, curved line above the words "PIPER ACCESS". The word "PIPER" is in a light gray color, while the word "ACCESS" is in a light orange color. The curved line above the words is also in light gray and orange.
SECTION 2
The 5 FR DL Piper PICC met all of the predetermined acceptance criteria derived from the standards and guidance listed above. Risk management of the subject device was conducted in accordance with ISO 14971:2007, Medical Devices – Risk Management for Medical Devices.
### Conclusions
The subject device, 5 FR DL Piper PICC, has the same intended use as the cited predicate device. Testing performed on the subject device demonstrates the device meets the requirements and is substantially equivalent to the predicate.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.